Erschienen in: BMC Ophthalmology 1/2024

Open Access 01.12.2024 | Research

The effect of corneal power on the accuracy of 14 IOL power formulas

verfasst von: Jialin Xu, Lu Zhang, Er Mo, Kaiyi Zhu, Yitong Zhu, Ke Feng, Zunting Wu, Yangran Zheng, Fang Huang, Xianhui Gong, Jin Li

Erschienen in: BMC Ophthalmology | Ausgabe 1/2024

Abstract

Background

This study evaluates the impact of corneal power on the accuracy of 14 newer intraocular lens (IOL) calculation formulas in cataract surgery. The aim is to assess how these formulas perform across different corneal curvature ranges, thereby guiding more precise IOL selection.

Methods

In this retrospective case series, 336 eyes from 336 patients who underwent cataract surgery were studied. The cohort was divided into three groups according to preoperative corneal power. Key metrics analyzed included mean prediction error (PE), standard deviation of PE (SD), mean absolute prediction error (MAE), median absolute error (MedAE), and the percentage of eyes with PE within ± 0.25 D, 0.50 D, ± 0.75 D, ± 1.00 D and ± 2.00 D.

Results

In the flat K group (Km < 43 D), VRF-G, Emmetropia Verifying Optical Version 2.0 (EVO2.0), Kane, and Hoffer QST demonstrated lower SDs (± 0.373D, ± 0.379D, ± 0.380D, ± 0.418D, respectively) compared to the VRF formula (all P < 0.05). EVO2.0 and K6 showed significantly different SDs compared to Barrett Universal II (BUII) (all P < 0.02). In the medium K group (43 D ≤ Km < 46 D), VRF-G, BUII, Karmona, K6, EVO2.0, Kane, and Pearl-DGS recorded lower MAEs (0.307D to 0.320D) than Olsen (OLCR) and Castrop (all P < 0.03), with RBF3.0 having the second lowest MAE (0.309D), significantly lower than VRF and Olsen (OLCR) (all P < 0.05). In the steep K group (Km ≥ 46D), RBF3.0, K6, and Kane achieved significantly lower MAEs (0.279D, 0.290D, 0.291D, respectively) than Castrop (all P < 0.001).

Conclusions

The study highlights the varying accuracy of newer IOL formulas based on corneal power. VRF-G, EVO2.0, Kane, K6, and Hoffer QST are highly accurate for flat corneas, while VRF-G, RBF3.0, BUII, Karmona, K6, EVO2.0, Kane, and Pearl-DGS are recommended for medium K corneas. In steep corneas, RBF3.0, K6, and Kane show superior performance.
Hinweise
Jialin Xu and Lu Zhang co-first authors who contributed equally to this work.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Introduction

In cataract surgery, achieving precise refractive outcomes is crucial and largely depends on accurate ocular biometric measurements. One of the key factors in this process is the selection of appropriate intraocular lens (IOL) power calculation formulas. Despite advancements in technology and methodology, prediction errors (PE) continue to pose challenges, particularly in cases with unique ocular characteristics like axial length (AL), anterior chamber depth (ACD, measured from corneal epithelium to lens), corneal power, and lens thickness (LT) [1].
The development of newer IOL power calculation formulas such as the Barrett Universal II(hereafter BUII) [2], Castrop [3], Emmetropia Verifying Optical Version 2.0(EVO2.0), Hoffer QST [4], Kane [5], Karmona [6], Cook K6(K6), Naeser2 [7], Olsen C [8, 9], Pearl­DGS [10], Radial Basis Function Version 3.0 (RBF3.0), T2 [11], VRF [12] and VRF-G [13] marks significant progress in this field [1316]. These formulas have shown enhanced accuracy, especially in eyes with atypical AL, [1719]ACD, [19] and other specific ocular parameters [20].
While AL and ACD are often highlighted in IOL power calculation, the role of corneal power in influencing the accuracy of these formulas is equally critical. Studies with comprehensive datasets have highlighted that variations in corneal power can significantly affect the precision of IOL power calculation formulas [2125]. Traditional comparisons primarily focused on the performance of third- and fourth-generation formulas in eyes with atypical corneal power, revealing limitations in formulas like Haigis [26], Hoffer Q [27], and SRK/T [28] in cases of steep corneal power [22, 24, 25]. However, evaluations of the newer-generation formulas, such as RBF and Olsen C, suggest substantial improvements in accuracy [21, 22].
This study aims to examine the influence of corneal power on the prediction accuracy of a range of IOL calculation formulas. By comparing newer IOL power calculation formulas (BUII, Castrop, EVO2.0, Hoffer QST, Kane, Karmona, K6, Naeser2, Olsen (OLCR), Pearl-DGS, RBF3.0, T2, VRF, and VRF-G), we intend to provide insights for selecting the most suitable IOL formula based on individual corneal characteristics. This approach will enhance our understanding of the relationship between corneal power and IOL formula accuracy, aiding in more precise postoperative refractive outcomes.

Materials and methods

Patients and measurements

This retrospective case series study was conducted from January 2019 to December 2021 at the Eye Hospital of Wenzhou Medical University. The patient cohort consisted of individuals who underwent uncomplicated cataract surgery via phacoemulsification. The surgeries were performed by two experienced cataract surgeons, LJ and HF, with each patient receiving the same intraocular lens (IOL) model (SN6CWS, Alcon, Fort Worth, TX, USA). Selection criteria for study participants were aligned with the IOL power calculation guidelines proposed by Hoffer et al. in 2020 [29]. In cases where patients underwent sequential bilateral cataract surgery, the right eye was preferentially included in the study. Inclusion criteria were a postoperative corrected distance visual acuity of at least 20/40. Exclusion criteria encompassed patients with a history of eye disease, prior ocular surgery, invalid biometry, intraoperative or postoperative complications, or lack of postoperative manifest refraction data.
Preoperative ocular parameters were measured using the Lenstar LS900 (Haag-Streit AG, Koeniz, Switzerland Biometry: v2.5.2, IOL: v4.2.1), covering axial length (AL), anterior chamber depth (ACD), flat and steep keratometry readings (K1 and K2), central corneal thickness (CCT), horizontal corneal diameter (CD), and lens thickness (LT). The average of the keratometry reading (Km) was calculated from the flat and steep keratometry readings. Postoperative manifest refraction with a constant distance of 6 m from the phoropter to the optotype screen was assessed between one and three months after surgery to ensure stabilization of refractive outcomes.

IOL power calculation

The study utilized a range of formulas for spherical equivalent prediction, including BUII, Castrop, EVO2.0, Hoffer QST, Kane, Karmona, K6, Naeser2, Olsen (OLCR), Pearl-DGS, RBF3.0, T2, VRF, and VRF-G. Several of these formulas (Hoffer QST, Kane, RBF3.0, VRF-G) also incorporated gender in their calculations [30].
Refractive prediction error (PE) was determined by comparing the spherical equivalent of the postoperative manifest refraction with each formula’s predicted spherical equivalent using the IOL power actually implanted. The standard deviation (SD) of PE’s error was calculated, with positive and negative PE values indicating hyperopic and myopic shifts, respectively. The mean refractive prediction error (ME), mean absolute error (MAE), median absolute error (MedAE), and the percentages of eyes with PE within ± 0.25 D, ± 0.5 D, ± 0.75 D, ± 1.00 D and ± 2.00 D were also computed.All formulas were individually optimized by their respective authors to achieve a mean PE of zero.

Statistical analysis

The data were analyzed with the SPSS software (version 25.0, IBM Corp.) and R Project for Statistical Computing (https://​www.​r-project.​org). The Kolmogorov-Smirnov test assessed data normality. The PE of Castrop, Karmona, and Olsen (OLCR) showed normal distribution, while nonparametric Wilcoxon tests were applied to the PE of all formulas. Heteroscedastic method [31]was used to evaluate SD, MAE, MedAE, and the proportion of eyes within different diopter ranges. The Holm-Bonferroni correction was applied for multiple comparisons to determine adjusted P-values. A P-value of less than 0.05 was considered statistically significant.

Results

This study included 336 eyes from 336 patients, with an average participant age of 70.29 years (range 33–87 years), predominantly women (66.1%, n = 222) and right eyes (64.3%, n = 216). Pre-surgical biometric ocular parameters are detailed in Table 1. Based on mean keratometry (Km), patients were categorized into three groups: flat K (Km < 43 D), medium K (43 D ≤ Km < 46 D), and steep K (Km ≥ 46 D). Both the flat and steep K groups represented around 20% of the cohort each.
Table 1
Preoperative patient biometric ocular parameters
Parameter
Mean ± SD/ Median (IQR)
Range
Axial length, mm*
23.33 (0.7)
21.21–31.98
Anterior chamber depth, mm
3.01 ± 0.44
1.69–4.28
Flat keratometry (K1), D
44.14 ± 1.71
39.77–49.36
Steep keratometry (K2), D
44.85 ± 1.71
40.65–50.76
Mean of keratometry (Km), D
44.49 ± 1.68
40.21–50.06
Corneal central thickness, µm
535.09 ± 34.13
439–649
Horizontal corneal diameter, mm
11.54 ± 0.46
9.89–13.07
Lens thickness, mm
4.47 ± 0.49
2.77–5.88
Axial length distribution, n (%)
  
AL<22.0 mm
29
8.63%
22.0 mm ≤ AL<26.0 mm
289
86.01%
AL ≥ 26.0 mm
18
5.36%
Keratometry subgroups, n (%)
  
Km<43.0D (Flat)
65
19.35%
43.0D ≤ Km<46.0D (Medium)
205
61.01%
Km ≥ 46.0D (Steep)
66
19.64%
ACD, as measured from the corneal epithelium to the lens
*Data with a non-normal distribution was shown as the median and interquartile range (IQR)
SD, standard deviation; D, diopter; IOL, intraocular lens

Formula accuracy in all patients

Table 2 summarizes the outcomes for 14 IOL formulas. It details the optimized constants, PE, SD, MAE, MedAE, and the percentage of eyes within specific PE ranges. The ME for all formulas was not significantly different from zero (P > 0.05), indicating overall accurate predictions. Formulas with the highest accuracy included K6 (SD ± 0.399D), EVO2.0 (SD ± 0.403D), VRF-G (SD ± 0.403D), Kane (SD ± 0.404D), and RBF3.0 (SD ± 0.404D). The Olsen (OLCR) formula showed the largest SD (± 0.459D), yet no statistical difference was observed in the SDs across all formulas (P > 0.05). EVO2.0, Pearl-DGS, RBF3.0, and BUII outperformed in achieving a PE within ± 0.25D, with over 53% of eyes falling in this category. In contrast, Olsen (OLCR), Castrop, VRF, and T2 had less than 50% of eyes reaching a PE within ± 0.25D.
Table 2
Refractive outcomes and optimized constants obtained by each formula in all eyes
Formula
Optimized Constants
PE
SD
MAE
MedAE
Eyes within PE (%)
PE ≤ 0.25 D
PE ≤ 0.50 D
PE ≤ 0.75 D
PE ≤ 1.00 D
PE ≤ 2.00 D
Barrett Universal II
1.940
0.000
0.425
0.319
0.237
53.27
78.87
91.37
97.92
100.00
Castrop
0.420
0.150
0.003
0.442
0.348
0.281
45.24
75.30
89.58
97.02
100.00
EVO 2.0
119.068
0.000
0.403
0.306
0.236
54.17
78.87
94.64
98.21
100.00
Hoffer QST
5.620
0.000
0.416
0.318
0.255
50.30
79.46
92.56
97.62
100.00
Cooke K6
119.250
0.000
0.399
0.305
0.234
52.38
80.06
94.35
97.92
100.00
Kane
119.043
0.000
0.404
0.307
0.238
52.38
78.87
93.15
97.62
100.00
Karmona
119.430
0.005
0.415
0.325
0.250
51.79
78.57
93.45
97.92
100.00
Naeser 2
1.439
0.940
0.004
0.447
0.342
0.259
50.30
76.79
90.77
96.43
100.00
Olsen (OLCR)
4.94
0.000
0.459
0.364
0.302
43.75
74.11
89.88
97.02
100.00
Pearl-DGS
119.299
0.000
0.408
0.313
0.243
53.57
77.98
93.15
98.21
100.00
RBF 3.0
119.031
0.000
0.404
0.306
0.233
53.27
80.95
93.15
98.51
100.00
T2
119.034
0.000
0.424
0.327
0.266
49.70
77.38
92.26
96.73
100.00
VRF
5.598
0.001
0.445
0.342
0.273
48.21
75.00
90.77
97.62
100.00
VRF-G
119.088
0.002
0.403
0.305
0.238
52.38
81.55
94.64
97.62
100.00
PE = mean prediction error. SD = standard deviation of the error. MedAE = median absolute error. MAE = mean absolute error. D = diopter

Formula accuracy according to corneal power

Table 3; Figs. 1 and 2 present the performance of each IOL formula across different corneal power subgroups.
Table 3
Predictive outcomes of IOL calculation formulas according to corneal power
 
BU II
Castrop
EVO 2.0
Hoffer QST
Cook K6
Kane
Karmona
Naeser 2
Olsen (OLCR)
Pearl-DGS
RBF 3.0
T2
VRF
VRF-G
Km<43.0D(n = 65)
PE
0.037
-0.048
0.014
-0.030
-0.031
-0.054
-0.137
-0.138
-0.139
-0.061
-0.037
0.020
-0.018
-0.047
SD
0.469
0.400
0.379 a,b
0.418 a
0.383 b
0.380 a
0.436
0.453
0.407
0.382
0.418
0.443
0.468
0.373a
MAE
0.340
0.318
0.291
0.324
0.293
0.299
0.369
0.368
0.355
0.301
0.326
0.341
0.365
0.294
MedAE
0.243
0.291
0.265
0.280
0.242
0.293
0.299
0.266
0.292
0.239
0.281
0.299
0.303
0.229
Percentage of Eyes within Diopter Range Indicated
± 0.25D
50.77
41.54
47.69
47.69
52.31
46.15
44.62
47.69
44.62
52.31
47.69
47.69
43.08
52.31
±0.50D
78.46
81.54
84.62
76.92
83.08
86.15
75.38
70.77
75.38
80.00
81.54
73.85
70.77
81.54
± 0.75D
90.77
93.85
96.92
93.85
96.92
95.38
95.38
87.69
92.31
96.92
93.85
89.23
92.31
96.92
± 1.00D
95.38
96.92
98.46
98.46
96.92
98.46
95.38
96.92
98.46
98.46
98.46
95.38
98.46
98.46
43.0D ≤ Km<46D(n = 205)
PE
-0.010
0.005
0.007
0.016
0.007
0.010
0.016
0.003
0.007
-0.006
-0.003
0.008
0.015
0.004
SD
0.409 a,c
0.450
0.417 c
0.426
0.408 c,e
0.415 c
0.406
0.441
0.470
0.414 c,e
0.405 a, c,d
0.429
0.448
0.413
MAE
0.311 c,e
0.356
0.315 c,e
0.324
0.314 c,e
0.315 c,e
0.313 c,e
0.338
0.368
0.320 c,e
0.309 a,c
0.331
0.342
0.307 c,e
MedAE
0.233
0.276
0.233c
0.269
0.234
0.227
0.249
0.252
0.308
0.248
0.239
0.267
0.267
0.235
Percentage of Eyes within Diopter Range Indicated
± 0.25D
55.61
46.34
56.10
49.76
51.71
53.17
52.20
50.73
43.41
53.17
52.68
49.27
49.27
52.68
±0.50D
79.51
72.68
75.61
79.02
77.56
76.59
78.54
78.05
74.15
76.59
80.00
76.10
75.12
80.98
± 0.75D
93.17
89.27
93.17
91.22
93.66
92.68
93.66
91.22
89.27
92.68
93.66
92.68
89.76
94.63
± 1.00D
98.54
97.56
98.05
96.59
98.05
97.07
98.54
96.59
96.10
98.05
98.05
96.59
97.07
97.07
Km ≥ 46D(n = 66)
PE
-0.008
0.049
-0.036
-0.021
0.013
0.022
0.109
0.145
0.116
0.078
0.048
-0.044
-0.025
0.043
SD
0.428
0.448
0.375
0.377
0.380
0.384
0.381
0.417
0.436
0.399
0.376
0.386
0.409
0.398
MAE
0.327
0.353
0.294
0.292
0.290e
0.291 e
0.316
0.331
0.361
0.302
0.279 e
0.301
0.317
0.310
MedAE
0.278
0.28
0.234
0.231
0.221
0.228
0.215
0.248
0.292
0.194
0.185
0.236
0.26
0.257
Percentage of Eyes within Diopter Range Indicated
± 0.25D
48.48
45.45
54.55
54.55
54.55
56.06
57.58
51.52
43.94
56.06
60.61
53.03
50
51.52
±0.50D
77.27
77.27
83.33
83.33
84.85
78.79
81.82
78.79
72.73
80.30
83.33
84.85
78.79
83.33
± 0.75D
86.36
86.36
96.97
95.45
93.94
92.42
90.91
92.42
89.39
90.91
90.91
93.94
92.42
92.42
± 1.00D
98.48
95.45
98.48
100.00
98.48
98.48
98.48
95.45
98.48
100.00
100.00
98.48
98.48
98.48
a, Heteroscedastic test, significantly lower than VRF, P < 0.05
b, Heteroscedastic test, significantly lower than BUII, P < 0.015
c, Heteroscedastic test, significantly lower than Olsen (OLCR), P < 0.015
d, Heteroscedastic test, significantly lower than Naeser, P < 0.025
e, Heteroscedastic test, significantly lower than Castrop, P < 0.025
In the group with corneal power less than 43 D (n = 65), VRF-G, EVO2.0, Kane, and K6 showed SDs of ± 0.370D, ± 0.379D, ± 0.380D, and ± 0.383D, respectively, and MAEs of 0.294D, 0.291D, 0.299D, and 0.294D. The proportion of eyes achieving a PE within ± 0.50D ranged from 81.54 to 86.15%. The BUII formula had an SD of ± 0.469D, higher compared to EVO2.0 and K6 (all P < 0.001). The VRF formula had higher SD of ± 0.468D than EVO2.0, Kane, Hoffer QST and VRF-G formulas (all P < 0.05).
In the subgroup with corneal power between 43 D and 46 D (n = 205), the RBF3.0, Karmona, K6, BUII, VRF-G, Pearl-DGS, Kane, and EVO2.0 formulas had SDs ranging from ± 0.405D to ± 0.417D and MAEs from 0.307D to 0.320D. The percentage of eyes within ± 0.50D PE for VRF-G and RBF3.0 was above 80.0%. Olsen (OLCR) and Castrop showed higher SDs of ± 0.470D and ± 0.450D, and MAEs of 0.368D and 0.356D, respectively. A statistical analysis of the SDs and MAEs revealed significant differences among all formulas in Table 3.
In the group with corneal power of 46 D or higher (n = 66), EVO2.0, RBF3.0, Hoffer QST, K6, and Kane had SDs between ± 0.375D and ± 0.384D. The RBF3.0, K6 and Kane also achieved the lowest MAEs (0.279D, 0.290D, 0.291D, respectively) and which were significantly lower than the Castrop (all P < 0.001). Olsen (OLCR) and Castrop recorded higher SDs of ± 0.436D and ± 0.448D, and higher MAEs of 0.361D and 0.353D, with the lowest percentages of eyes achieving PE within ± 0.25D and ± 0.50D.

Discussion

This study conducted a comprehensive assessment of 14 newer IOL calculation formulas, with specific emphasis on their accuracy in predicting outcomes for different corneal curvatures. Our findings provide valuable insights into the nuanced performance of these formulas in relation to corneal power, offering valuable guidance for their appropriate application in specific corneal profiles.
Our analysis revealed that the VRF-G, EVO2.0, Kane, and K6 formulas demonstrate exceptional accuracy in eyes with flat corneal power. These formulas seem to effectively compensate for the unique optical characteristics presented by flat corneas. The VRF-G formula, which incorporates elements of theoretical optics, regression analysis, and ray tracing [13], contributing to its high precision in this group. This aligns with the accuracy levels reported in earlier studies [16, 17, 32]. Additionally, the K6 formula, which has not been as extensively studied as others, exhibited notable performance in eyes with short and long ALs, displaying results comparable to those of the Kane and EVO2.0 formulas [10, 17, 18]. Kane and EVO2.0 also performed well and have good stability in flat cornea group. This finding was consistent with many previous studies [16, 19, 33, 34].
The middle range of corneal curvature presented a different challenge, with formulas such as RBF3.0, Karmona, K6, BUII, VRF-G, Pearl-DGS, Kane, and EVO2.0 showing commendable accuracy. Interestingly, formulas like Pearl-DGS [10], despite being not better than the other new formulas in several studies [17, 34, 35], showcased promising results, comparable to established formulas like Kane and EVO2.0 in the medium K group. The Karmona formula was designed and programmed in Shiny-RStudio version 1.1.423 (R Foundation, Boston, USA) by David Carmona González [6] and reported better results (SD = ± 0.30D) than ours (SD = ± 0.415D). This variance could be attributed to our inability to obtain the mean keratometry of posterior surface to substitute into the calculations. To the best of our knowledge, few studies investigated the accuracy of Karmona formula in different range of corneal power and in our results, the Karmona formula showed good accuracy in the medium K group.
For steep corneal powers, our findings suggested a superior performance from formulas like RBF3.0, K6, and Kane. The EVO2.0 and Hoffer QST also performed well. The RBF3.0 formula was found to have good results at different corneal curvatures in our previous studies in long eyes [33]. The K6 formula showed its effectiveness in both flat and steep corneas, although its efficacy in long eyes with abnormal corneal power was somewhat less pronounced [33]. The Kane formula maintained its accuracy across eyes with abnormal corneal power, in both normal and long ALs, as supported by the results of this and previous studies [33].
Some limitations have been identified in this study. Firstly, we refrained from comparing classic formulas due to prior studies [1, 23] evaluating the efficacy of conventional formulas across various ranges of corneal curvature. Furthermore, the exclusion of more extreme ocular parameters, particularly in eyes with atypical ALs, might limit the applicability of our findings to a broader patient population. Also, additional ocular biology measurements should be obtained, such as the mean keratometry of posterior surface and total keratometry, which may have had a impact on the postoperative refraction and needs more investigation.
In summary, our study offers valuable insights into the performance of various newer-generation IOL calculation formulas across different corneal curvature groups. The VRF-G, EVO2.0, Kane, K6 showed good accuracy in flat K eyes, while the RBF3.0, K6 and Kane performed better in eyes with steep corneal power.

Acknowledgements

We thank Oleksiy V. Voytsekhivskyy MD, Jack Kane MD, Tun Kuan Yeo MD, David Cooke MD and Guillaume Debellemanière MD for their invaluable contributions to the optimization and data analysis of newly developed formulas. Their expertise and insights, shared through personal communication, have been instrumental in the advancement of this work.

Declarations

All included patients provided informed written content. The Declaration of Helsinki tenets were followed in this study. Ethical approval was obtained from the Institution Review Board of the Eye Hospital of Wenzhou Medical University (Ethics approval number: 2022-042-K-27).
Not applicable.

Competing interests

The authors declare no competing interests.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Melles RB, Holladay JT, Chang WJ. Accuracy of intraocular lens calculation formulas. Ophthalmol. 2018;125:169–78.CrossRef Melles RB, Holladay JT, Chang WJ. Accuracy of intraocular lens calculation formulas. Ophthalmol. 2018;125:169–78.CrossRef
2.
Zurück zum Zitat Barrett GD. An improved universal theoretical formula for intraocular lens power prediction. J Cataract Refract Surg. 1993;19:713–20.CrossRefPubMed Barrett GD. An improved universal theoretical formula for intraocular lens power prediction. J Cataract Refract Surg. 1993;19:713–20.CrossRefPubMed
3.
Zurück zum Zitat Langenbucher A, Szentmáry N, Cayless A, Weisensee J, Fabian E, Wendelstein J, et al. Considerations on the Castrop formula for calculation of intraocular lens power. PLoS ONE. 2021;16:e0252102.CrossRefPubMedPubMedCentral Langenbucher A, Szentmáry N, Cayless A, Weisensee J, Fabian E, Wendelstein J, et al. Considerations on the Castrop formula for calculation of intraocular lens power. PLoS ONE. 2021;16:e0252102.CrossRefPubMedPubMedCentral
4.
Zurück zum Zitat Taroni L, Hoffer KJ, Pellegrini M, Lupardi E, Savini G. Comparison of the new Hoffer QST with 4 modern accurate formulas. J Cataract Refract Surg. 2023;49:378–84.CrossRefPubMed Taroni L, Hoffer KJ, Pellegrini M, Lupardi E, Savini G. Comparison of the new Hoffer QST with 4 modern accurate formulas. J Cataract Refract Surg. 2023;49:378–84.CrossRefPubMed
5.
Zurück zum Zitat Connell BJ, Kane JX. Comparison of the Kane formula with existing formulas for intraocular lens power selection. BMJ Open Ophthalmol. 2019;4:e000251.CrossRefPubMedPubMedCentral Connell BJ, Kane JX. Comparison of the Kane formula with existing formulas for intraocular lens power selection. BMJ Open Ophthalmol. 2019;4:e000251.CrossRefPubMedPubMedCentral
6.
Zurück zum Zitat Carmona González D, Palomino Bautista C. Accuracy of a new intraocular lens power calculation method based on artificial intelligence. Eye. 2021;35:517–22.CrossRefPubMed Carmona González D, Palomino Bautista C. Accuracy of a new intraocular lens power calculation method based on artificial intelligence. Eye. 2021;35:517–22.CrossRefPubMed
7.
Zurück zum Zitat Næser K, Savini G. Accuracy of thick-lens intraocular lens power calculation based on cutting-card or calculated data for lens architecture. J Cataract Refract Surg. 2019;45:1422–9.CrossRefPubMed Næser K, Savini G. Accuracy of thick-lens intraocular lens power calculation based on cutting-card or calculated data for lens architecture. J Cataract Refract Surg. 2019;45:1422–9.CrossRefPubMed
8.
Zurück zum Zitat Preussner P-R, Wahl J, Lahdo H, Dick B, Findl O. Ray tracing for intraocular lens calculation. J Cataract Refract Surg. 2002;28:1412–9.CrossRefPubMed Preussner P-R, Wahl J, Lahdo H, Dick B, Findl O. Ray tracing for intraocular lens calculation. J Cataract Refract Surg. 2002;28:1412–9.CrossRefPubMed
9.
Zurück zum Zitat Olsen T, Hoffmann P. C constant: new concept for ray tracing-assisted intraocular lens power calculation. J Cataract Refract Surg. 2014;40:764–73.CrossRefPubMed Olsen T, Hoffmann P. C constant: new concept for ray tracing-assisted intraocular lens power calculation. J Cataract Refract Surg. 2014;40:764–73.CrossRefPubMed
10.
Zurück zum Zitat Debellemanière G, Dubois M, Gauvin M, Wallerstein A, Brenner LF, Rampat R, et al. The PEARL-DGS formula: the development of an open-source machine learning-based thick IOL calculation formula. Am J Ophthalmol. 2021;232:58–69.CrossRefPubMed Debellemanière G, Dubois M, Gauvin M, Wallerstein A, Brenner LF, Rampat R, et al. The PEARL-DGS formula: the development of an open-source machine learning-based thick IOL calculation formula. Am J Ophthalmol. 2021;232:58–69.CrossRefPubMed
11.
Zurück zum Zitat Sheard RM, Smith GT, Cooke DL. Improving the prediction accuracy of the SRK/T formula: the T2 formula. J Cataract Refract Surg. 2010;36:1829–34.CrossRefPubMed Sheard RM, Smith GT, Cooke DL. Improving the prediction accuracy of the SRK/T formula: the T2 formula. J Cataract Refract Surg. 2010;36:1829–34.CrossRefPubMed
12.
Zurück zum Zitat Voytsekhivskyy OV. Development and clinical accuracy of a new intraocular lens power formula (VRF) compared to other formulas. Am J Ophthalmol. 2018;185:56–67.CrossRefPubMed Voytsekhivskyy OV. Development and clinical accuracy of a new intraocular lens power formula (VRF) compared to other formulas. Am J Ophthalmol. 2018;185:56–67.CrossRefPubMed
13.
Zurück zum Zitat Hipólito-Fernandes D, Elisa Luís M, Gil P, Maduro V, Feijão J, Yeo TK, et al. VRF-G, a new intraocular lens power calculation formula: a 13-formulas comparison study. Clin Ophthalmol Auckl NZ. 2020;14:4395–402.CrossRef Hipólito-Fernandes D, Elisa Luís M, Gil P, Maduro V, Feijão J, Yeo TK, et al. VRF-G, a new intraocular lens power calculation formula: a 13-formulas comparison study. Clin Ophthalmol Auckl NZ. 2020;14:4395–402.CrossRef
14.
Zurück zum Zitat Darcy K, Gunn D, Tavassoli S, Sparrow J, Kane JX. Assessment of the accuracy of new and updated intraocular lens power calculation formulas in 10 930 eyes from the UK National Health Service. J Cataract Refract Surg. 2020;46:2–7.PubMed Darcy K, Gunn D, Tavassoli S, Sparrow J, Kane JX. Assessment of the accuracy of new and updated intraocular lens power calculation formulas in 10 930 eyes from the UK National Health Service. J Cataract Refract Surg. 2020;46:2–7.PubMed
15.
Zurück zum Zitat Cheng H, Kane JX, Liu L, Li J, Cheng B, Wu M. Refractive predictability using the IOLMaster 700 and Artificial Intelligence-based IOL power formulas compared to standard formulas. J Refract Surg (Thorofare NJ 1995). 2020;36:466–72.CrossRef Cheng H, Kane JX, Liu L, Li J, Cheng B, Wu M. Refractive predictability using the IOLMaster 700 and Artificial Intelligence-based IOL power formulas compared to standard formulas. J Refract Surg (Thorofare NJ 1995). 2020;36:466–72.CrossRef
16.
Zurück zum Zitat Voytsekhivskyy OV, Hoffer KJ, Tutchenko L, Cooke DL, Savini G. Thorofare NJ. Accuracy of 24 IOL power calculation methods. J Refract Surg (1995). 2023;39:249–256. Voytsekhivskyy OV, Hoffer KJ, Tutchenko L, Cooke DL, Savini G. Thorofare NJ. Accuracy of 24 IOL power calculation methods. J Refract Surg (1995). 2023;39:249–256.
17.
Zurück zum Zitat Voytsekhivskyy OV, Hoffer KJ, Savini G, Tutchenko LP, Hipólito-Fernandes D. Clinical accuracy of 18 IOL power formulas in 241 short eyes. Curr Eye Res. 2021;46:1832–43.CrossRefPubMed Voytsekhivskyy OV, Hoffer KJ, Savini G, Tutchenko LP, Hipólito-Fernandes D. Clinical accuracy of 18 IOL power formulas in 241 short eyes. Curr Eye Res. 2021;46:1832–43.CrossRefPubMed
18.
Zurück zum Zitat Shammas HJ, Taroni L, Pellegrini M, Shammas MC, Jivrajka RV. Accuracy of newer intraocular lens power formulas in short and long eyes using sum-of-segments biometry. J Cataract Refract Surg. 2022;48:1113–20.CrossRefPubMedPubMedCentral Shammas HJ, Taroni L, Pellegrini M, Shammas MC, Jivrajka RV. Accuracy of newer intraocular lens power formulas in short and long eyes using sum-of-segments biometry. J Cataract Refract Surg. 2022;48:1113–20.CrossRefPubMedPubMedCentral
19.
Zurück zum Zitat Mo E, Lin L, Wang J, Huo Q, Yang Q, Liu E, et al. Clinical accuracy of six intraocular lens power calculation formulas in elongated eyes, according to anterior chamber depth. Am J Ophthalmol. 2022;233:153–62.CrossRefPubMed Mo E, Lin L, Wang J, Huo Q, Yang Q, Liu E, et al. Clinical accuracy of six intraocular lens power calculation formulas in elongated eyes, according to anterior chamber depth. Am J Ophthalmol. 2022;233:153–62.CrossRefPubMed
20.
Zurück zum Zitat Hipólito-Fernandes D, Luís ME, Serras-Pereira R, Gil P, Maduro V, Feijão J, et al. Anterior chamber depth, lens thickness and intraocular lens calculation formula accuracy: nine formulas comparison. Br J Ophthalmol. 2020;106:349–55.CrossRefPubMed Hipólito-Fernandes D, Luís ME, Serras-Pereira R, Gil P, Maduro V, Feijão J, et al. Anterior chamber depth, lens thickness and intraocular lens calculation formula accuracy: nine formulas comparison. Br J Ophthalmol. 2020;106:349–55.CrossRefPubMed
21.
Zurück zum Zitat Chang P-J, Zhang F, Wang J-J, Zhao Y-E. Accuracy of constant C for ray tracing: assisted intraocular lens power calculation in normal ocular axial eyes. Ophthalmic Res. 2021;64:85–90.CrossRefPubMed Chang P-J, Zhang F, Wang J-J, Zhao Y-E. Accuracy of constant C for ray tracing: assisted intraocular lens power calculation in normal ocular axial eyes. Ophthalmic Res. 2021;64:85–90.CrossRefPubMed
22.
Zurück zum Zitat Reitblat O, Levy A, Kleinmann G, Lerman TT, Assia EI. Intraocular lens power calculation for eyes with high and low average keratometry readings: comparison between various formulas. J Cataract Refract Surg. 2017;43:1149–56.CrossRefPubMed Reitblat O, Levy A, Kleinmann G, Lerman TT, Assia EI. Intraocular lens power calculation for eyes with high and low average keratometry readings: comparison between various formulas. J Cataract Refract Surg. 2017;43:1149–56.CrossRefPubMed
23.
Zurück zum Zitat Röggla V, Langenbucher A, Leydolt C, Schartmüller D, Schwarzenbacher L, Abela-Formanek C, et al. Accuracy of common IOL power formulas in 611 eyes based on axial length and corneal power ranges. Br J Ophthalmol. 2021;105:1661–5.CrossRefPubMed Röggla V, Langenbucher A, Leydolt C, Schartmüller D, Schwarzenbacher L, Abela-Formanek C, et al. Accuracy of common IOL power formulas in 611 eyes based on axial length and corneal power ranges. Br J Ophthalmol. 2021;105:1661–5.CrossRefPubMed
24.
Zurück zum Zitat Zhang Z, Miao Y, Fang X, Luo Q, Wang Y. Accuracy of the Haigis and SRK/T formulas in eyes longer than 29.0 mm and the influence of central corneal keratometry reading. Curr Eye Res. 2018;43:1316–21.CrossRefPubMed Zhang Z, Miao Y, Fang X, Luo Q, Wang Y. Accuracy of the Haigis and SRK/T formulas in eyes longer than 29.0 mm and the influence of central corneal keratometry reading. Curr Eye Res. 2018;43:1316–21.CrossRefPubMed
25.
Zurück zum Zitat Zhang C, Dai G, Pazo EE, Xu L, Wu X, Zhang H, et al. Accuracy of intraocular lens calculation formulas in cataract patients with steep corneal curvature. PLoS ONE. 2020;15:e0241630.CrossRefPubMedPubMedCentral Zhang C, Dai G, Pazo EE, Xu L, Wu X, Zhang H, et al. Accuracy of intraocular lens calculation formulas in cataract patients with steep corneal curvature. PLoS ONE. 2020;15:e0241630.CrossRefPubMedPubMedCentral
26.
Zurück zum Zitat Haigis W, Lege B, Miller N, Schneider B. Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis. Graefes Arch Clin Exp Ophthalmol. 2000;238:765–73.CrossRefPubMed Haigis W, Lege B, Miller N, Schneider B. Comparison of immersion ultrasound biometry and partial coherence interferometry for intraocular lens calculation according to Haigis. Graefes Arch Clin Exp Ophthalmol. 2000;238:765–73.CrossRefPubMed
27.
Zurück zum Zitat Hoffer KJ. The Hoffer Q formula: a comparison of theoretic and regression formulas. J Cataract Refract Surg. 1993;19:700–712; errata 1994: 20:677 and 2007;33(1):2–3. Hoffer KJ. The Hoffer Q formula: a comparison of theoretic and regression formulas. J Cataract Refract Surg. 1993;19:700–712; errata 1994: 20:677 and 2007;33(1):2–3.
28.
Zurück zum Zitat Retzlaff JA, Sanders DR, Kraff MC. Development of the SRK/T intraocular lens implant power calculation formula. J Cataract Refract Surg. 1990;16:333–340; erratum: 1990;16(4):528. Retzlaff JA, Sanders DR, Kraff MC. Development of the SRK/T intraocular lens implant power calculation formula. J Cataract Refract Surg. 1990;16:333–340; erratum: 1990;16(4):528.
29.
Zurück zum Zitat Hoffer KJ, Savini G. Update on intraocular lens power calculation study protocols: the better way to design and report clinical trials. Ophthalmol. 2020;:S0161-6420(20)30638-2. Hoffer KJ, Savini G. Update on intraocular lens power calculation study protocols: the better way to design and report clinical trials. Ophthalmol. 2020;:S0161-6420(20)30638-2.
30.
Zurück zum Zitat Hoffer KJ, Savini G. Effect of gender and race on ocular biometry. Int Ophthalmol Clin. 2017;57:137–42.CrossRefPubMed Hoffer KJ, Savini G. Effect of gender and race on ocular biometry. Int Ophthalmol Clin. 2017;57:137–42.CrossRefPubMed
31.
Zurück zum Zitat Holladay JT, Wilcox RR, Koch DD, Wang L. Review and recommendations for univariate statistical analysis of spherical equivalent prediction error for IOL power calculations. J Cataract Refract Surg. 2021;47:65–77.CrossRefPubMed Holladay JT, Wilcox RR, Koch DD, Wang L. Review and recommendations for univariate statistical analysis of spherical equivalent prediction error for IOL power calculations. J Cataract Refract Surg. 2021;47:65–77.CrossRefPubMed
32.
Zurück zum Zitat Voytsekhivskyy O, Tutchenko L. Comparison of the prediction accuracy of 13 formulas in long eyes. Graefes Arch Clin Exp Ophthalmol. 2023;261:2575–83.CrossRefPubMed Voytsekhivskyy O, Tutchenko L. Comparison of the prediction accuracy of 13 formulas in long eyes. Graefes Arch Clin Exp Ophthalmol. 2023;261:2575–83.CrossRefPubMed
34.
Zurück zum Zitat Savini G, Di Maita M, Hoffer KJ, Næser K, Schiano-Lomoriello D, Vagge A, et al. Comparison of 13 formulas for IOL power calculation with measurements from partial coherence interferometry. Br J Ophthalmol. 2021;105:484–9.CrossRefPubMed Savini G, Di Maita M, Hoffer KJ, Næser K, Schiano-Lomoriello D, Vagge A, et al. Comparison of 13 formulas for IOL power calculation with measurements from partial coherence interferometry. Br J Ophthalmol. 2021;105:484–9.CrossRefPubMed
35.
Zurück zum Zitat Rocha-de-Lossada C, Colmenero-Reina E, Flikier D, Castro-Alonso F-J, Rodriguez-Raton A, García-Madrona J-L, et al. Intraocular lens power calculation formula accuracy: comparison of 12 formulas for a trifocal hydrophilic intraocular lens. Eur J Ophthalmol. 2021;31:2981–8.CrossRefPubMed Rocha-de-Lossada C, Colmenero-Reina E, Flikier D, Castro-Alonso F-J, Rodriguez-Raton A, García-Madrona J-L, et al. Intraocular lens power calculation formula accuracy: comparison of 12 formulas for a trifocal hydrophilic intraocular lens. Eur J Ophthalmol. 2021;31:2981–8.CrossRefPubMed
Metadaten
Titel
The effect of corneal power on the accuracy of 14 IOL power formulas
verfasst von
Jialin Xu
Lu Zhang
Er Mo
Kaiyi Zhu
Yitong Zhu
Ke Feng
Zunting Wu
Yangran Zheng
Fang Huang
Xianhui Gong
Jin Li
Publikationsdatum
01.12.2024
Verlag
BioMed Central
Erschienen in
BMC Ophthalmology / Ausgabe 1/2024
Elektronische ISSN: 1471-2415
DOI
https://doi.org/10.1186/s12886-024-03395-9

Weitere Artikel der Ausgabe 1/2024

BMC Ophthalmology 1/2024 Zur Ausgabe

Neu im Fachgebiet Augenheilkunde

Metastase in der periokulären Region

Metastasen Leitthema

Orbitale und periokuläre metastatische Tumoren galten früher als sehr selten. Aber mit der ständigen Aktualisierung von Medikamenten und Nachweismethoden für die Krebsbehandlung werden neue Chemotherapien und Strahlenbehandlungen eingesetzt. Die …

Staging und Systemtherapie bei okulären und periokulären Metastasen

Metastasen Leitthema

Metastasen bösartiger Erkrankungen sind die häufigsten Tumoren, die im Auge diagnostiziert werden. Sie treten bei ungefähr 5–10 % der Patienten mit soliden Tumoren im Verlauf der Erkrankung auf. Besonders häufig sind diese beim Mammakarzinom und …

CME: Wundheilung nach Trabekulektomie

Trabekulektomie CME-Artikel

Wird ein Glaukom chirurgisch behandelt, ist die anschließende Wundheilung von entscheidender Bedeutung. In diesem CME-Kurs lernen Sie, welche Pathomechanismen der Vernarbung zugrunde liegen, wie perioperativ therapiert und Operationsversagen frühzeitig erkannt werden kann.

„standard operating procedures“ (SOP) – Vorschlag zum therapeutischen Management bei periokulären sowie intraokulären Metastasen

Metastasen Leitthema

Peri- sowie intraokuläre Metastasen sind insgesamt gesehen selten und meist Zeichen einer fortgeschrittenen primären Tumorerkrankung. Die Therapie ist daher zumeist palliativ und selten kurativ. Zudem ist die Therapiefindung sehr individuell. Die …

Update Augenheilkunde

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.