The effect of different posterior inclinations of tibial component on tibiofemoral contact pressures after unicompartmental knee arthroplasty

verfasst von: Bo Yuan, Zhongjun Mo, Kuan Zhang, Xu Zhu, Songhua Yan, Jizhou Zeng

Erschienen in: Journal of Orthopaedic Surgery and Research | Ausgabe 1/2023

Abstract

Background

Different posterior inclinations of tibial component after unicompartmental knee arthroplasty (UKA) may lead to different biomechanical characteristics of the knee joint. This finite element study was designed to investigate the tibiofemoral contact pressures after UKA with different posterior inclinations of tibial component.

Methods

Finite element model of a healthy knee joint was constructed, and mobile-bearing (MB) UKA models with 5 different posterior inclinations (3°, 5°, 7°, 9° and 11°) of tibial components were simulated. The maximum contact pressures of tibial plateau cartilage in the lateral compartment and polyethylene insert in the medial compartment were calculated based on the ground reaction force and the angle of the knee flexion obtained by 3D motion capture system.

Results

The loading ratio of medial and lateral compartments during standing stance (medial 54.49%, lateral 45.51%) and tibial anterior displacement (134 N, 3.89 mm) of healthy knee was basically consistent with previous experimental data. The maximum contact pressures of the medial meniscus and lateral tibial plateau cartilage of the healthy knee during standing stance were 2.14 MPa and 1.57 MPa, respectively. At the static standing phase, the maximum contact pressures of the polyethylene insert decreased from 17.90 to 17.29 Mpa, and the maximum contact pressures of the tibial plateau cartilage in the lateral compartment increased from 0.81 to 0.92 Mpa following an increase in the posterior inclination of the tibial component. At the first peak of ground reaction force, the maximum contact pressures of polyethylene insert increased from 22.37 to 25.16 MPa, and the maximum contact pressures of tibial plateau cartilage in the lateral compartment increased from 3.03 to 3.33 MPa, with the increase in the posterior inclination of the tibial component. At the second peak of ground reaction force, the maximum contact pressures of polyethylene insert decreased from 2.34 to 2.22 MPa with the increase in posterior inclination of tibial component.

Conclusion

The preoperative and postoperative finite element models of MB UKA were well established. The results showed that the maximum contact pressures of the polyethylene insert did not change significantly with the increase in the posterior inclination of the tibial prosthesis, while the maximum contact pressures of the tibial plateau cartilage of the lateral compartment increased when the posterior inclination of the tibial prosthesis was > 7°. Our results also show that the maximum contact pressures were greater with an excessive inclination angle (11°) of the tibial component, and the pressures of the tibial plateau cartilage in the lateral compartment were more concentrated on the posterior area. This study, therefore, proposes that excessive osteotomy should be avoided.
Hinweise
Co-first author: Bo Yuan, Zhongjun Mo.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
UKA
Unicompartmental knee arthroplasty
MB UKA
Mobile-bearing unicompartmental knee arthroplasty
KOA
Knee osteoarthritis
TKA
Total knee arthroplasty
CoCrMo ally
Cobalt–chromium–molybdenum ally
UHMWPE
Ultra-high molecular weight polyethylene

Introduction

Knee osteoarthritis (KOA) is a common and frequently occurring disease in the elderly [1], and most KOA patients suffer anterior medial osteoarthritis. At present, surgical treatment methods for anterior medial osteoarthritis are mainly divided into two categories: total knee arthroplasty (TKA) and unicompartmental knee arthroplasty (UKA). Compared with TKA, UKA is characterized by less trauma and quicker recovery periods [2]. As a kind of UKA surgery, MB UKA’s unique prosthesis design shares close similarity to the natural knee joint and can better restore the knee joint function. However, clinical outcomes have shown that the survival rate of MB UKA is significantly lower than that of TKA [35], mainly due to aseptic loosening and progressive development of lateral compartment osteoarthritis [6].
Contact stress is one of the important factors affecting the survival rate of MB UKA. The alteration of the alignment of the prosthesis, especially the posterior inclination of tibial prosthesis, leads to the change of contact pressures of the knee, which may cause aseptic loosening and concomitant osteoarthritis of MB UKA in the lateral compartment [7]. Excessive posterior inclination of tibial prosthesis may increase the probability of prosthesis loosening, the risk of anterior cruciate ligament injury and surgical revision [8].
Although different posterior inclinations of tibial prosthesis have different effects on knee function, scholars and clinicians still have greatly differing opinions on what is the best osteotomy angle should be. Weber et al. [9] suggested that a 4–8° posterior inclination of tibial prosthesis could help reduce the wearing of polyethylene insert. Aleto and Weber et al. [10, 11] believed that the posterior inclination of tibial prosthesis should be kept within the range of 0°–7° after surgery. Simpson et al. [12] found that the mean von Mises stress of the anteromedial tibia region did not change significantly with the increase in inclination of MB UKA. However, Small et al. [13] believed that the strain increased significantly in the posteromedial region of the proximal tibia when the tilt angle of tibial prosthesis increased from 5° to 10°. Therefore, it is of great significance to study the influence of posterior angle of tibial prosthesis on biomechanics of knee joint after MB UKA.
In this study, a finite element model of healthy knee joint was developed, and a MB UKA surgical simulation conducted to establish the influence of the maximum contact pressures on the polyethylene insert and the lateral tibial plateau cartilage surface after subjecting the knee to different posterior inclinations of tibial prosthesis. These pressures were calculated based on the joint force and motion characteristics of the gait cycle.

Methods

Modeling of healthy and MB UKA knee joint

A healthy female (48 years old, 165 cm height, 65 kg weight) with no history of knee degeneration or trauma as confirmed by X-ray and magnetic resonance imaging (MRI) was recruited in the study. Following a signed informed consent for imaging, 256-slice spiral CT (Philips, Brilliance iCT) was performed for full length CT scan of the left lower limb, ranging from hip- to ankle joint. The scanning layer thickness was 1 mm; A 3.0 T magnetic resonance scanner (Combined image UI770) was used to perform sagittal MR scan at the left knee joint with a thickness of 1 mm.
Mimics 20.0 (Materialise Ltd., Leuven, Belgium) was used to extract 3D models of bone, cartilage, meniscus, anterior and posterior cruciate ligaments, medial and lateral collateral ligaments from CT and MR images. Rapidform 2006 reverse engineering software (INUS Technology, Inc., Seoul, South Korea) was used to materialize the 3D model of each structure. Abaqus 6.14–2 finite element analysis software (Dassault Systemes SimuliaCoip., Providence, RI, USA) was imported for assembly and registration, and the 3D solid model of healthy knee joint was developed.
3D point cloud data of femoral prosthesis (S), tibial prosthesis (A) and polyethylene insert were obtained using 3D scanner (ARTEC, EVA), and the Rapidform software was used for reverse 3D reconstruction to obtain the geometric model of the prosthesis. On the basis of the above established healthy knee joint, according to the Oxford III generation UKA standard surgical technique, a 7 mm thickness osteotomy was performed with Boolean operation with the 0° inclination in the coronal plane and the 7° posterior inclination in the sagittal plane. On the basis of this model, the tibial prosthesis was rotated to establish 3°, 5°, 9° and 11° posterior inclination models (Fig. 1).

Material property assignment and mesh

All models were divided by ten-node modified tetrahedral elements. There were 153,727 units of healthy knee joint model. A total of five MB UKA finite element models with different posterior inclinations of tibial prosthesis were established, and each model had about 300,000 units. Cartilage and meniscus were defined as isotropic linear elastic materials [14]. The ligament was defined as an incompressible transversely isotropic hyperelastic material [14], and the Neo Hookean constitutive model was applied. Its constitutive equation is:
$$\Psi = C_{1} \times (I_{1} - 3)$$
where C1 is the initial shear modulus and I1 is the first modified invariant of Cauchy Green strain tensor. C1 values used for the anterior and posterior cruciate ligaments, the medial and lateral collateral ligaments were 6.06, 6.43, 5.83 and 6.06 MPa, respectively. The material parameters of other structures are shown in Table 1.
Table 1
Material properties in the finite element models
 
Young’s modulus (MPa)
Poisson’s ratio
Bone
17,000
0.3
Cartilage
15
0.46
Meniscus
27.5
0.33
CoCrMo ally
195,000
0.3
UHMWPE
685
0.4
CoCrMo ally Cobalt–chromium–molybdenum ally; UHMWPE Ultra-high molecular weight polyethylene

Load and boundary conditions

Healthy knee joint finite element model using frictionless and limited slip surface-to-surface contact relations of the six (contact on location for: room between the medial and lateral cartilage of the femur and tibia, the femoral cartilage and meniscus between surface under the surface, tibial cartilage and meniscus) was made and the calculation made based on penalty function algorithm. MB UKA model had five contacts to set; the medial compartment surface contact which was using a frictionless contact, and the lateral compartment which was using coulomb friction contact, with a friction coefficient of 0.04.
Model verification under axial load: The degree of freedom in flexion and extension direction of femur was fixed, whereas the degree of freedom in other directions of femur was not fixed. While the lower surface of tibia and fibula remained completely fixed, the cartilage and bone as well as ligament and bone were connected in a binding form, and the front and back corners of the meniscus were bound to the tibial plateau. The reference point of femur was defined at the midpoint of the medial chamber and medial epicondyles of femur. An axial load was applied to this reference point (the direction of the load was down the mechanical axis, and the load size was 1000N), and the load distribution of the medial and lateral compartment was calculated and verified by comparison with the literature results [15].
Tibialis anterior drawer under the force model validation: Tibial reference point was defined at the center of the ankle, tibia and fibula coupling constraint main limit degree of freedom completely fixed in the direction of flexion and extension, femoral and tibial reference points put on forward force (the size of the force of 134 N), and the displacement of the tibia compared with literature results [16].
Static standing load model: Model verification was done under constrained coaxial load. A load of 325 N was applied along the mechanical axis of the tibia at the reference point of the femur, with a buckling angle of 0°. The load ratio of the medial and lateral compartment was then calculated.
Model under gait cycle load: The femur reference point was coupled with the proximal femur, and the model verification done under constrained coaxial load. A force load along the mechanical axis of the tibia and a sagittal angle (flexion and extension) were applied at the femur reference point. The force load and flexion angle were provided by the joint forces and range of motion of the subjects at different points in the gait cycle [17] (Fig. 2). The knee joint force and the angle of flexion and extension at the first peak of ground reaction force and the second peak of ground reaction force in the gait cycle were selected as the loading boundary conditions to calculate the load distribution of the medial and lateral compartments of the knee joint, and the maximum contact pressure between the polyethylene insert and the tibial cartilage of the lateral compartment.

Results

Verification of the finite element models of healthy and MB UKA knee joint

Verification under axial load: In the finite element model of healthy knee joint, the load distribution proportion of the medial compartment accounted for 54.49%, whereas the load distribution proportion of the lateral compartment accounted for 45.51% of the total load, which was consistent with the literature [15] (Fig. 3).
The verification under tibialis anterior drawer force: Under the action of 134 N forward and backward drawer force, tibia moved forward 3.89 mm, which was also consistent with literature results [16] (Fig. 4).
Comparison of MB UKA and healthy knee finite element model: In the MB UKA model, tilting the of tibial prosthesis 3°, 5°, 7°, 9° and 11° resulted in bearing of 45%, 44.73%, 44.87%, 44.94% and 45.07% of the total load by the medial compartment, respectively. Similarly, the lateral compartment bore 55%, 55.27%, 55.13%, 55.06% and 54.93% of the total load, respectively. The medial and lateral compartments of the healthy knee joint model bore 54.92% and 45.08% of the total load, respectively. The maximum contact pressures of the medial meniscus and lateral tibial plateau cartilage of the healthy knee were 2.14 MPa and 1.57 MPa, respectively. The load distribution proportions of the medial and lateral compartments of the knee after MB UKA were similar to the literature findings [18] (Fig. 5).

The maximum contact pressures of polyethylene insert

At the static standing phase, the maximum contact pressures of polyethylene insert in MB UKA model with 3°, 5°, 7°, 9° and 11° inclinations were 17.90 MPa, 17.62 MPa, 16.78 MPa, 17.41 MPa and 17.29 MPa, respectively. At the first peak of ground reaction force, the maximum contact pressures of polyethylene insert in MB UKA models with 3°, 5°, 7°, 9° and 11° inclinations were 22.37 MPa, 23.90 MPa, 25.10 MPa, 25.31 MPa and 25.16 MPa, respectively. At the second peak of ground reaction force, the maximum contact pressures of polyethylene insert in MB UKA models with 3°, 5°, 7°, 9° and 11° inclinations were 19.13 MPa, 18.82 MPa, 18.34 MPa, 19.18 MPa and 19.21 MPa, respectively (Fig. 6).

The maximum contact pressures of tibial plateau cartilage in the lateral compartment

At the static standing phase, the maximum contact pressures of tibial plateau cartilage in the lateral compartment were 0.81 MPa, 0.81 MPa, 0.80 MPa, 0.77 MPa and 0.92 MPa, respectively. At the first peak of ground reaction force, the maximum contact pressures of tibial plateau cartilage in the lateral compartment were 3.03 MPa, 3.02 MPa, 3.02 MPa, 3.13 MPa and 3.33 MPa, respectively. At the second peak of ground reaction force, the maximum contact pressures of tibial plateau cartilage in the lateral compartment were 2.34 MPa, 2.36 MPa, 2.34 MPa, 2.01 MPa and 2.22 MPa, respectively (Fig. 7).

Discussion

Normally, there is a complex biomechanical effect of the posterior inclination on mobile-bearing UKA. The serious problems that arise from malpositioning of the mobile-bearing UKA are progressive ostearthritis, rapid wear on the PE insert and its dislocation, and failure of the ACL [19]. However, to the best of our knowledge, the biomechanical effects of posterior inclination of the tibial prosthesis in mobile-bearing UKA on the knee joint have not been fully elucidated and/or defined through an acceptable research approach. Our results showed that the maximum contact pressures of the polyethylene insert did not change significantly with the increase in the posterior inclination of the tibial prosthesis, while the maximum contact pressures of the tibial plateau cartilage of the lateral compartment increased when the posterior inclination of the tibial prosthesis was > 7°.
In this study, a combination of MRI and CT scans was used to maximize the reconstruction of bone, cartilage, meniscus, and meniscus structures. Due to the complex structure of the model, all the structures adopted a tetrahedral mesh. Within the academic circles, there is still lack of highly recognized process for verification of finite element model. Oberkampf et al. [20], however, advocated four main levels of model verification, namely organizational level, single structure level, multi-structure level and overall level. In the selection of inspection structures, considering that the focus of this study was on load distribution and contact state of the whole knee joint in the gait cycle, the cartilage, meniscus and ligament did not have too much influence on the overall experimental study of the knee joint.
Comparing to the healthy knee model, the MB UKA model was more difficult to verify. Tuncer and Innocenti et al. [21, 22] established the UKA model based on the imaging data obtained from cadaver specimen and carried out matching in vitro tests simultaneously to verify the model. Netter et al. [23] analyzed and compared the data obtained from UKA wear tester and the wear rates obtained from UKA model calculation by using a knee simulation to obtain verification results. Due to the limitation of funds and materials in this study, our finite element model adopted an indirect verification method, and compared the load proportion before and after MB UKA. It further compared the maximum contact stress profile of tibial plateau cartilage in the lateral compartment and polyethylene insert with other studies to verify the validity of the model.
In our study, the proportion of internal and lateral compartment load after MB UKA was consistent with the literature results [18]. Innocenti et al. [18] found that the lateral compartment load of all UKA models was greater than the medial one by studying tibial component alignment from varus 6° to valgus 6°, and similar results were obtained in this study. We believe that the main reason for this result is that the elastic modulus of the prosthesis is much larger than that of the cartilage and meniscus, leading to the redistribution of loads in the knee joint, with a focus on the unreplaced side. Excessive elastic modulus of tibial prosthesis will not only lead to the redistribution of medial and lateral loads, but also change the stress distribution of bone around tibial prosthesis due to the stress shielding effect, which may lead to pain and increase the incidence of aseptic loosening [21].
The greater the maximum contact pressures of the tibial cartilage in the contralateral compartment, the higher the incidence of progressive osteoarthritis in this compartment [2427]. Studies have shown that the posterior inclination of the tibial prosthesis for MB UKA should not exceed 7°. The tibial inclination of the healthy knee in this study was 7°. Therefore, the 7° inclination of the tibial prosthesis was used as a reference and the original inclination was assumed to be retained. The posterior inclination of tibial prosthetics in normal knee joints has been proven to have a wide range, with an average of 9° [28]. In order to restore the biomechanical environment of the knee as much as possible, the optimal posterior inclination of tibial prosthesis in the sagittal plane of UKA has always been the objective pursued by surgeons. However, the optimal inclination has not been determined clinically and biomechanically. In general, it is recommended that the surgeon restores the patient’s initial tibial posterior inclination after UKA. Our results showed that the maximum contact pressures and pressure concentrations of the tibial plateau cartilage in the lateral compartment were greater when the posterior inclination of the tibial component was excssive (11°).However, these were lesser than the those of a healthy knee joint.

Conclusions

In this study, finite element models of healthy knee joints including 3D ligament structures were established and verified. Likewise, MB UKA finite element models bearing 5 different posterior inclinations of tibial prostheses (3°, 5°, 7°, 9° and 11°) were established. The results showed that the maximum contact pressures of the polyethylene insert did not change significantly with the increase in the posterior inclination of the tibial prosthesis, while the maximum contact pressures of the tibial plateau cartilage of the lateral compartment increased when the posterior inclination of the tibial prosthesis was > 7°. The stress clouds show that the contact pressures of the polyethylene insert did not change significantly with the increase in the posterior inclination of the tibial component, while the pressure concentrations on the tibial plateau cartilage in the lateral compartment gradually occurred, mainly in the posterior side. The preoperative and postoperative finite element models of MB UKA established in this study reasonably analyzed the contact stress of knee joint, providing a reference method for subsequent MB UKA research. We therefore proposed that excessive osteotomy should be avoided.

Acknowledgements

Not applicable.

Declarations

The study was approved by the Ethics Committee of Capital Medical University.
Not applicable.

Competing interests

The authors declare that they have no competing interests.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Siddiq MAB, Clegg D, Jansen TL, et al. Emerging and new treatment options for knee osteoarthritis. Curr Rheumatol Rev. 2022;18(1):20–32.CrossRefPubMed Siddiq MAB, Clegg D, Jansen TL, et al. Emerging and new treatment options for knee osteoarthritis. Curr Rheumatol Rev. 2022;18(1):20–32.CrossRefPubMed
2.
Zurück zum Zitat Crawford DA, Berend KR, Thienpont E. Unicompartmental knee arthroplasty: US and global perspectives. Orthop Clin N Am. 2020;51(2):147–59.CrossRef Crawford DA, Berend KR, Thienpont E. Unicompartmental knee arthroplasty: US and global perspectives. Orthop Clin N Am. 2020;51(2):147–59.CrossRef
3.
Zurück zum Zitat Sava MP, Hara H, Alexandra L, et al. Verasense sensor-assisted total knee arthroplasty showed no difference in range of motion, reoperation rate or functional outcomes when compared to manually balanced total knee arthroplasty: a systematic review. Knee Surg Sports Traumatol Arthrosc. 2023;31:1851–8.CrossRefPubMedPubMedCentral Sava MP, Hara H, Alexandra L, et al. Verasense sensor-assisted total knee arthroplasty showed no difference in range of motion, reoperation rate or functional outcomes when compared to manually balanced total knee arthroplasty: a systematic review. Knee Surg Sports Traumatol Arthrosc. 2023;31:1851–8.CrossRefPubMedPubMedCentral
4.
Zurück zum Zitat Röttinger T, Lisitano L, Wiedl A, et al. The anterior impingement after mobile-bearing unicomparimental knee arthroplasty—a neglected problem. A clinical report of 14 cases. Arthroplast Today. 2022;17:94–100.CrossRefPubMedPubMedCentral Röttinger T, Lisitano L, Wiedl A, et al. The anterior impingement after mobile-bearing unicomparimental knee arthroplasty—a neglected problem. A clinical report of 14 cases. Arthroplast Today. 2022;17:94–100.CrossRefPubMedPubMedCentral
5.
Zurück zum Zitat Wu LP, Mayr HO, Zhang X, et al. Knee scores of patients with non-lateral compartmental knee osteoarthritis undergoing mobile, fixed-bearing unicompartmental knee and total knee arthroplasties: a randomized controlled trial. Orthop Surg. 2022;14(1):73–87.CrossRefPubMed Wu LP, Mayr HO, Zhang X, et al. Knee scores of patients with non-lateral compartmental knee osteoarthritis undergoing mobile, fixed-bearing unicompartmental knee and total knee arthroplasties: a randomized controlled trial. Orthop Surg. 2022;14(1):73–87.CrossRefPubMed
6.
Zurück zum Zitat Migliorini F, Maffulli N, Cuozzo F, et al. Mobile bearing versus fixed bearing for unicompartmental arthroplasty in monocompartmental osteoarthritis of the knee: a meta-analysis. J Clin Med. 2022;11(10):2837.CrossRefPubMedPubMedCentral Migliorini F, Maffulli N, Cuozzo F, et al. Mobile bearing versus fixed bearing for unicompartmental arthroplasty in monocompartmental osteoarthritis of the knee: a meta-analysis. J Clin Med. 2022;11(10):2837.CrossRefPubMedPubMedCentral
7.
Zurück zum Zitat Chatellard R, Sauleau V, Colmar M, et al. Medial unicompartmental knee arthroplasty: Does tibial component position influence clinical outcomes and arthroplasty survival? Orthop Traumatol Surg Res. 2013;99(4 Suppl):S219–25.CrossRefPubMed Chatellard R, Sauleau V, Colmar M, et al. Medial unicompartmental knee arthroplasty: Does tibial component position influence clinical outcomes and arthroplasty survival? Orthop Traumatol Surg Res. 2013;99(4 Suppl):S219–25.CrossRefPubMed
8.
Zurück zum Zitat Hernigou P, Deschamps G. Posterior slope of the tibial implant and the outcome of unicompartmental knee arthroplasty. J Bone Jt Surg Am. 2004;86(3):506–11.CrossRef Hernigou P, Deschamps G. Posterior slope of the tibial implant and the outcome of unicompartmental knee arthroplasty. J Bone Jt Surg Am. 2004;86(3):506–11.CrossRef
9.
Zurück zum Zitat Meng Q, Jin Z, Wilcox R, et al. Computational investigation of the time-dependent contact behaviour of the human tibiofemoral joint under body weight. Proc Inst Mech Eng H. 2014;228(11):1193–207.CrossRefPubMedPubMedCentral Meng Q, Jin Z, Wilcox R, et al. Computational investigation of the time-dependent contact behaviour of the human tibiofemoral joint under body weight. Proc Inst Mech Eng H. 2014;228(11):1193–207.CrossRefPubMedPubMedCentral
10.
Zurück zum Zitat Aleto TJ, Berend ME, Ritter MA, et al. Early failure of unicompartmental knee arthroplasty leading to revision. J Arthroplast. 2008;23(2):159–63.CrossRef Aleto TJ, Berend ME, Ritter MA, et al. Early failure of unicompartmental knee arthroplasty leading to revision. J Arthroplast. 2008;23(2):159–63.CrossRef
11.
Zurück zum Zitat Weber P, Schroder C, Schmidutz F, et al. Increase of tibial slope reduces backside wear in medial mobile bearing unicompartmental knee arthroplasty. Clin Biomech. 2013;28(8):904–9.CrossRef Weber P, Schroder C, Schmidutz F, et al. Increase of tibial slope reduces backside wear in medial mobile bearing unicompartmental knee arthroplasty. Clin Biomech. 2013;28(8):904–9.CrossRef
12.
Zurück zum Zitat Weber P, Woiczinski M, Steinbrück A, et al. Increase in the tibial slope in unicondylar knee replacement: analysis of the effect on the kinematics and ligaments in a weight-bearing finite element model. BioMed Res Int 2018;2018. Weber P, Woiczinski M, Steinbrück A, et al. Increase in the tibial slope in unicondylar knee replacement: analysis of the effect on the kinematics and ligaments in a weight-bearing finite element model. BioMed Res Int 2018;2018.
13.
Zurück zum Zitat Small SR, Berend ME, Rogge RD, et al. Tibial loading after UKA: evaluation of tibial slope, resection depth, medial shift and component rotation. J Arthroplasty. 2013;28(9 Suppl):179–83.CrossRefPubMed Small SR, Berend ME, Rogge RD, et al. Tibial loading after UKA: evaluation of tibial slope, resection depth, medial shift and component rotation. J Arthroplasty. 2013;28(9 Suppl):179–83.CrossRefPubMed
14.
Zurück zum Zitat Wen PF, Guo WS, Gao FQ, et al. Effects of lower limb alignment and tibial component inclination on the biomechanics of lateral compartment in unicompartmental knee arthroplasty. Chin Med J. 2017;130(21):2563–8.CrossRefPubMedPubMedCentral Wen PF, Guo WS, Gao FQ, et al. Effects of lower limb alignment and tibial component inclination on the biomechanics of lateral compartment in unicompartmental knee arthroplasty. Chin Med J. 2017;130(21):2563–8.CrossRefPubMedPubMedCentral
15.
Zurück zum Zitat Pe AE, Martínez MA, Calvo B, et al. A finite element simulation of the effect of graft stiffness and graft tensioning in ACL reconstruction. Clin Biomech. 2005;20(6):636–44.CrossRef Pe AE, Martínez MA, Calvo B, et al. A finite element simulation of the effect of graft stiffness and graft tensioning in ACL reconstruction. Clin Biomech. 2005;20(6):636–44.CrossRef
16.
Zurück zum Zitat Mary TG, Eric KW, Savio LW, et al. Distribution of in situ forces in the anterior cruciate ligament in response to rotatory loads. J Orthop Res. 2004;22(1):85–9.CrossRef Mary TG, Eric KW, Savio LW, et al. Distribution of in situ forces in the anterior cruciate ligament in response to rotatory loads. J Orthop Res. 2004;22(1):85–9.CrossRef
17.
18.
Zurück zum Zitat Innocenti B, Pianigiani S, Ramundo G, et al. Biomechanical effects of different varus and valgus alignments in medial unicompartmental knee arthroplasty. J Arthroplast. 2016;31(12):2685–91.CrossRef Innocenti B, Pianigiani S, Ramundo G, et al. Biomechanical effects of different varus and valgus alignments in medial unicompartmental knee arthroplasty. J Arthroplast. 2016;31(12):2685–91.CrossRef
19.
Zurück zum Zitat Tay ML, Matthews BG, Monk AP, et al. Disease progression, aseptic loosening and bearing dislocations are the main revision indications after lateral unicompartmental knee arthroplasty: a systematic review. J ISAKOS. 2022. Tay ML, Matthews BG, Monk AP, et al. Disease progression, aseptic loosening and bearing dislocations are the main revision indications after lateral unicompartmental knee arthroplasty: a systematic review. J ISAKOS. 2022.
20.
Zurück zum Zitat Henninger HB, Reese SP, Anderson AE, et al. Validation of computational models in biomechanics. Proc Inst Mech Eng Part H J Eng Med. 2010;224(7):801–12.CrossRef Henninger HB, Reese SP, Anderson AE, et al. Validation of computational models in biomechanics. Proc Inst Mech Eng Part H J Eng Med. 2010;224(7):801–12.CrossRef
21.
Zurück zum Zitat Innocenti B, Bilgen ÖF, Labey L, et al. Load sharing and ligament strains in balanced, overstuffed and understuffed UKA A validated finite element analysis. J Arthroplast. 2014;29(7):1491–8.CrossRef Innocenti B, Bilgen ÖF, Labey L, et al. Load sharing and ligament strains in balanced, overstuffed and understuffed UKA A validated finite element analysis. J Arthroplast. 2014;29(7):1491–8.CrossRef
22.
Zurück zum Zitat Tuncer M, Cobb JP, Hansen UN, et al. Validation of multiple subject-specific finite element models of unicompartmental knee replacement. Med Eng Phys. 2013;35(10):1457–64.CrossRef Tuncer M, Cobb JP, Hansen UN, et al. Validation of multiple subject-specific finite element models of unicompartmental knee replacement. Med Eng Phys. 2013;35(10):1457–64.CrossRef
23.
Zurück zum Zitat Netter J, Hermida J, Kester M, et al. Effect of polyethylene crosslinking and bearing design on wear of unicompartmental arthroplasty. Orthop Proc. 2018;95-B(S15). Netter J, Hermida J, Kester M, et al. Effect of polyethylene crosslinking and bearing design on wear of unicompartmental arthroplasty. Orthop Proc. 2018;95-B(S15).
24.
Zurück zum Zitat Setton LA, Mow VC, Howell DS. Mechanical behavior of articular cartilage in shear is altered by transection of the anterior cruciate ligament. J Orthop Res. 1995;13(4):473–82.CrossRefPubMed Setton LA, Mow VC, Howell DS. Mechanical behavior of articular cartilage in shear is altered by transection of the anterior cruciate ligament. J Orthop Res. 1995;13(4):473–82.CrossRefPubMed
25.
Zurück zum Zitat Pena E, Calvo B, Martinez MA, et al. Why lateral meniscectomy is more dangerous than medial meniscectomy. A finite element study. J Orthop Res. 2006;24(5):1001–10.CrossRefPubMed Pena E, Calvo B, Martinez MA, et al. Why lateral meniscectomy is more dangerous than medial meniscectomy. A finite element study. J Orthop Res. 2006;24(5):1001–10.CrossRefPubMed
26.
Zurück zum Zitat Segal NA, Anderson DD, Iyer KS, et al. Baseline articular contact stress levels predict incident symptomatic knee osteoarthritis development in the MOST cohort. J Orthop Res. 2009;27(12):1562–8.CrossRefPubMedPubMedCentral Segal NA, Anderson DD, Iyer KS, et al. Baseline articular contact stress levels predict incident symptomatic knee osteoarthritis development in the MOST cohort. J Orthop Res. 2009;27(12):1562–8.CrossRefPubMedPubMedCentral
27.
Zurück zum Zitat Kwon OR, Kang KT, Son J, et al. Importance of joint line preservation in unicompartmental knee arthroplasty: finite element analysis. J Orthop Res. 2017;35(2):347–52.CrossRefPubMed Kwon OR, Kang KT, Son J, et al. Importance of joint line preservation in unicompartmental knee arthroplasty: finite element analysis. J Orthop Res. 2017;35(2):347–52.CrossRefPubMed
28.
Zurück zum Zitat Utzschneider S, Goettinger M, Weber P, et al. Development and validation of a new method for the radiologic measurement of the tibial slope. Knee Surg Sports Traumatol Arthrosc. 2011;19(10):1643–8.CrossRefPubMed Utzschneider S, Goettinger M, Weber P, et al. Development and validation of a new method for the radiologic measurement of the tibial slope. Knee Surg Sports Traumatol Arthrosc. 2011;19(10):1643–8.CrossRefPubMed
Metadaten
Titel
The effect of different posterior inclinations of tibial component on tibiofemoral contact pressures after unicompartmental knee arthroplasty
verfasst von
Bo Yuan
Zhongjun Mo
Kuan Zhang
Xu Zhu
Songhua Yan
Jizhou Zeng
Publikationsdatum
01.12.2023
Verlag
BioMed Central
Erschienen in
Journal of Orthopaedic Surgery and Research / Ausgabe 1/2023
Elektronische ISSN: 1749-799X
DOI
https://doi.org/10.1186/s13018-023-04222-5

Weitere Artikel der Ausgabe 1/2023

Journal of Orthopaedic Surgery and Research 1/2023 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Knie-TEP: Kein Vorteil durch antibiotikahaltigen Knochenzement

29.05.2024 Periprothetische Infektionen Nachrichten

Zur Zementierung einer Knie-TEP wird in Deutschland zu über 98% Knochenzement verwendet, der mit einem Antibiotikum beladen ist. Ob er wirklich besser ist als Zement ohne Antibiotikum, kann laut Registerdaten bezweifelt werden.

Häusliche Gewalt in der orthopädischen Notaufnahme oft nicht erkannt

28.05.2024 Häusliche Gewalt Nachrichten

In der Notaufnahme wird die Chance, Opfer von häuslicher Gewalt zu identifizieren, von Orthopäden und Orthopädinnen offenbar zu wenig genutzt. Darauf deuten die Ergebnisse einer Fragebogenstudie an der Sahlgrenska-Universität in Schweden hin.

Fehlerkultur in der Medizin – Offenheit zählt!

28.05.2024 Fehlerkultur Podcast

Darüber reden und aus Fehlern lernen, sollte das Motto in der Medizin lauten. Und zwar nicht nur im Sinne der Patientensicherheit. Eine negative Fehlerkultur kann auch die Behandelnden ernsthaft krank machen, warnt Prof. Dr. Reinhard Strametz. Ein Plädoyer und ein Leitfaden für den offenen Umgang mit kritischen Ereignissen in Medizin und Pflege.

Mehr Frauen im OP – weniger postoperative Komplikationen

21.05.2024 Allgemeine Chirurgie Nachrichten

Ein Frauenanteil von mindestens einem Drittel im ärztlichen Op.-Team war in einer großen retrospektiven Studie aus Kanada mit einer signifikanten Reduktion der postoperativen Morbidität assoziiert.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.