Dear editor,
Orotracheal intubation is the most common method for establishing artificial airways and plays a significant role in rescuing critically ill patients [1]. However, orotracheal intubation is associated with several complications, including iatrogenic mucocutaneous damage, catheter dislodgement, and tooth loosening [1, 2]. These complications may result in prolonged hospitalisation, increased infection opportunities, and burden the healthcare system and society [2, 3]. Therefore, proper endotracheal tube fixation is essential for preventing complications in patients undergoing orotracheal intubation. Herein, we report the successful treatment of a 79-year-old patient with lip pressure injury using a novel self-developed mask endotracheal tube fixation device (utility model patent of China, ZL 2022 2 1745523.5; Fig. 1A). The fixation device comprises a triangular headband, base, connecting bridge, fixation nut, and bite block (Fig. 1B). The base comprises a mask frame and silicone pad. The connecting bridge is printed using a photosensitive resin. The fixation nut is adapted from a traditional endotracheal tube fixator.
×
Anzeige
A 79-year-old Chinese woman was admitted to an intensive care unit with dyspnoea. She had a history of diabetes mellitus, hypertension, and cerebral infarction. Using a traditional endotracheal tube fixator, orotracheal intubation connected to a ventilator for assisted breathing, and oral care was provided every 6 h. Six days later, three pressure injuries were observed on her lips and the surrounding regions (Fig. 1C). Pathological examination of tissue samples revealed epidermal fractures, focal pseudoepithelioma-like hyperplasia, dermal vascular expansion, acute and chronic inflammatory cell infiltration in the epidermis and dermis, parakeratosis of the shed squamous epithelial cell mass, bacterial colonies in the inflammatory exudative, and necrotic tissue (Fig. 1D). Additionally, tissue-print cytology indicated inflammatory exudation, necrosis, and large amounts of red blood cells (Fig. 1E). Sustained pressure above a threshold can cause prolonged ischaemia and lead to tissue necrosis [4]. The pressure wounds were likely initially colonised by bacterial flora and that a bacterial imbalance resulted in a bacterial infection [3]. Bacterial infections impede wound healing, especially in diabetic patients [5]. To prevent further lip pressure injury and considering the patient’s history of diabetes mellitus, we replaced the traditional endotracheal tube fixator with a self-developed fixation device. After explaining the procedure to the patient’s husband, we placed the self-developed fixation device. On the third day after device replacement, the ruptured regions surrounding the right upper lip was smaller than those earlier (Fig. 1F). Moreover, tissue-print cytology showed decreased inflammatory exudation and necrosis (Fig. 1G). On the fifth day after device replacement, the upper edge of the lip peak and the vermilion border of the right lower lip were completely healed, and the ruptured regions surrounding the right upper lip was significantly smaller (Fig. 1H). Tissue-print cytology indicated a significant reduction in inflammatory exudation and necrosis, without any bacterial colonies or repaired epithelial cells (Fig. 1I). The endotracheal tube did not dislodge, and no new pressure injury occurred while the self-developed fixation device was in place.
To the best of our knowledge, this is the first report of a patient with lip pressure injury treated with this novel endotracheal tube fixation device combined with histopathological analyses. Compared with the traditional endotracheal tube fixator, our new device has two advantages. First, it avoids the perioral tissues, improves the labial blood circulation, and promotes the repair of pressure injury. Second, it increases the contact area on the face, thus reducing the pressure per unit area and preventing pressure injuries. Our report introduces a novel endotracheal tube fixation device that is safe and useful for nursing patients undergoing orotracheal intubation.
Acknowledgements
Not applicable.
Declarations
Ethics approval and consent to participate
The study was approved by the Institutional Review Board of Nanjing Lishui People's Hospital. The patient provided their written informed consent to participate in this study.
Anzeige
Consent for publication
The authors declare to have informed consent of the patient to publish the image.
Competing interests
The authors have disclosed that they do not have any conflicts of interest.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Der optimale Ansatz für die Blutdruckkontrolle bei Patientinnen und Patienten mit akutem Schlaganfall ist noch nicht gefunden. Ob sich eine frühzeitige Therapie der Hypertonie noch während des Transports in die Klinik lohnt, hat jetzt eine Studie aus China untersucht.
Laut einer Studie aus den USA und Kanada scheint es bei der Reanimation von Kindern außerhalb einer Klinik keinen Unterschied für das Überleben zu machen, ob die Wiederbelebungsmaßnahmen während des Transports in die Klinik stattfinden oder vor Ort ausgeführt werden. Jedoch gibt es dabei einige Einschränkungen und eine wichtige Ausnahme.
Eine ältere Frau trinkt regelmäßig Sennesblättertee gegen ihre Verstopfung. Der scheint plötzlich gut zu wirken. Auf Durchfall und Erbrechen folgt allerdings eine Hyponatriämie. Nach deren Korrektur kommt es plötzlich zu progredienten Kognitions- und Verhaltensstörungen.
In der Notaufnahme wird die Chance, Opfer von häuslicher Gewalt zu identifizieren, von Orthopäden und Orthopädinnen offenbar zu wenig genutzt. Darauf deuten die Ergebnisse einer Fragebogenstudie an der Sahlgrenska-Universität in Schweden hin.
Update AINS
Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.