Silica coated high performance oxide ceramics promote greater ossification than titanium implants: an in vivo study
verfasst von:
Filippo Migliorini, Hanno Schenker, Marcel Betsch, Nicola Maffulli, Markus Tingart, Frank Hildebrand, Sophie Lecouturier, Björn Rath, Jörg Eschweiler
This in vitro study investigated the osseointegration and implant integration of high performance oxide ceramics (HPOC) compared to titanium implants in rabbits.
Methods
Histomorphometry was conducted around the distal, proximal, medial, and lateral aspects of the HPOC to quantify the amount of mature and immature ossification within the bone interface. Histomorphometry was conducted by a trained musculoskeletal pathologist. The region of interest (ROI) represented the percentage of surrounding area of the implant. The percentage of ROI covered by osteoid implant contact (OIC) and mature bone implant contact (BIC) were assessed. The surrounding presence of bone resorption, necrosis, and/or inflammation were quantitatively investigated.
Results
All 34 rabbits survived the 6- and 12-week experimental period. All HPOC implants remained in situ. The mean weight difference from baseline was + 647.7 mg (P < 0.0001). The overall OIC of the ceramic group was greater at 6 weeks compared to the titanium implants (P = 0.003). The other endpoints of interest were similar between the two implants at all follow-up points. No difference was found in BIC at 6- and 12-weeks follow-up. No bone necrosis, resorption, or inflammation were observed.
Conclusion
HPOC implants demonstrated a greater osteoid implant contact at 6 weeks compared to the titanium implants, with no difference found at 12 weeks. The percentage of bone implant contact of HPOC implants was similar to that promoted by titanium implants.
Hinweise
Filippo Migliorini and Hanno Schenker have contributed equally to the final version of the manuscript and share the first author
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
RA-CPSC
Risedronate calcium phosphate silicate cements
SBA15
Mesoporous silica
MCP
Monocalcium phosphate
PCR
Polymerase chain reaction
PTHrP
Parathyroid hormone-related protein
APN
Adiponectin
HA
Hydroxyapatite
PBS
Phosphate-buffered saline
TNT
TiO2 nanotube
TNT-HA
Hydroxyapatite-TiO2 nanotube
CPC
Calcium phosphate ceramic
PCL
Poly(ε-caprolactone)
TCP
Tricalcium phosphate 70%
SE
Standard error
CI
95% Confidence interval
PE-CVD
Plasma-enhanced chemical vapor deposition
SiOx
Silicon suboxide
HPOC
High performance oxide ceramics
PMMA
Polymethyl methacrylate
Introduction
Alloy implants are commonly used in musculoskeletal medicine [1]. Bony osteointegration is pivotal to ensure implant survivorship. Osseointegration is a foreign body reaction to shield off the implant from the tissues [2]. Osteointegration is the capability of the implant to fuse with the surrounding bone, interacting and integrating with it [3, 4]. The rate of aseptic loosening and related consequences (stress-shielding, persistent pain, inflammation) are still a concern in musculoskeletal medicine following implantation of foreign materials [5, 6]. Given their optimal ossification, titanium and its alloys are commonly used [7‐9]. High performance oxide ceramics (HPOC) has attracted much interest [10, 11]. HPOC have several advantages: high hardness and wear resistance, light weight, outstanding resistance to creep and compressive stress, and avoid imaging artefacts [12‐14]. However, the biologically inactivity of ceramics, and consequently limited potential for integration, may limit their application. To overcome this limitation to clinical application, HPOC biologically functionalised with a silicate coating alumina AL2O3 have been developed [15, 16]. HPOC implants have been recently introduced in musculoskeletal medicine, and their application has been successfully supported by recently published clinical trials [17‐24]. HPOC have been developed and investigated at our institution during the past years. The strength of the bonding between the functionalised ceramic and surrounding layer was evaluated by the application of polymethyl methacrylate (PMMA) to the surface, as per manufacturer instructions [25]. The contact guidance, adhesions, surrounding mesenchymal stromal cells osteogenic differentiation, and their cytotoxic potential has also been evaluated in previous studies conducted at our institution [25, 26]. An experimental study in rabbits was conducted to characterise osseointegration and implant integration of HPOC. The bone implant interface is responsible for the primary and secondary stability of the implant, and it is suggested to be the weakest domain in the bone implant system, where most failures occur [27, 28]. We hypothesised that HPOC implants promote ossification and transplant integration. To validate this hypothesis, osteointegration of HPOC were compared to that provided by commercial titanium implants. Histomorphometry was conducted around the distal, proximal, medial, and lateral aspect of the HPOC implants to quantify the amount of mature and immature ossification within the bone implant interface.
Materials and method
Type of implants
For the experimental group, HPOC were manufactured and functionalized at the RWTH University Aachen, Germany, Department of Dental Materials Science and Biomaterial Research in a previously published fashion [25, 26, 29, 30]. Briefly, standard 5.5 × 8 mm Al2O3 ceramic based cylinders were used. To facilitate the coupling of stable organosilane monolayers on the monolithic Al2O3 ceramic based cylinders, plasma-enhanced chemical vapor deposition (PE-CVD) was performed [31]. Silicon suboxide (SiOx) was deposited on the polished and cleaned Al2O3 ceramic based cylinders to activate the ceramics [32]. Functionalised ceramic cylinders were then air-dried, cured at 80° for 45 min, and stored in liquid nitrogen until use. For the control group, Sandblasted titanium Ti-6Al-4 V 5.5 × 8 mm implants (Zimmer Biomet GmbH, Neu-Ulm, Germany) were used.
Anzeige
Surgical procedure
The present investigation was conducted following the Animal Welfare Act of the Federal Republic of Germany in 2017, and approved by the Federal Office for Nature, Environment and Consumer Protection (Landesamt für Natur, Umwelt und Verbraucherschutz, LANU) of North Rhine-Westphalia, Federal Republic of Germany (Approval ID: 84–02.04.2016.A434). 34 New Zealand adult female white rabbits (weight > 3 kg) were used. Rabbits were allocated into four groups (Fig. 1).
×
0.1 ml/mg/kg bodyweight Medetomidin (Domitor) combined with a 0.2 ml Ketamin (Narketan) in a subcutaneous injection were used to induct the anaesthesia. The surgical site was shaved, disinfected with iodine and ethanol, and draped in a sterile fashion. 10 mg/kg bodyweight Enrofloxacin subcutaneous injection were injected before the incision. The skin incision was performed over the right lateral femoral condyle. After dissection through fascia and muscles, the condyle was exposed, and the lateral collateral ligament identified. With a 5.5 mm trephine, a unicortical drill hole was performed under irrigation to avoid thermal necrosis, making sure that the lateral collateral ligament was sparred. Once the bony cylinder was extracted, either a titanium or HPCO cylinder was inserted in a press fit fashion, without damaging the knee capsule. Tissues were closed in layers in a standard fashion. Finally, the skin was stapled and sprayed with chelated silver. For the first three days after surgery, 4 mg/kg bodyweight Carprofen were administered every 24 h. Six or 12 weeks postoperatively, the rabbits were euthanised using 2 ml/kg bodyweight Natriumpentobarbital (160 mg Natriumpentobarbital/ml). The femoral condyles were harvested and examined.
Sample preparation
The femoral condyles were harvested en bloc. Fixation was performed over 12 days with 4% paraformaldehyde followed by an alcohol series with ethanol of 50–100% and xylol. The specimens were embedded in Technovit® 9100 (Kulzer GmbH, Hanau, Germany). Using a diamond band saw Exakt 300CL (EXAKT Technologies Inc, Oklahoma City, US), 60–70 µm coplanar cuts of the specimens were performed. Grinding of titanium cylinders was conducted using sandpaper, while HPCO implants were ground with diamond paper. All specimens were stained with haematoxylin eosin, trichrome, and toluidine. Histomorphometry was conducted by a professional pathologist with OLYMPUS digital microscope DSX-1000 and stream desktop- Software (Olympus Hamburg, Germany).
Histomorphometry
At microscopy, specimens were divided as follows: lateral (K1), distal (K2), medial (K3), and proximal (K4). The surrounding area between the bone-implant interface (red zones) was the region of interest (ROI, Fig. 2).
×
Anzeige
The outcome of interest was to investigate the potential of osteointegration of HPCO in comparison to the standard titanium. Hence, the percentage immature, and unmineralized bone matrix (osteoid implant contact, OIC) within the ROI was assessed. The percentage of mature, mineralized bone (bone implant contact, BIC) within the ROI was also quantified. The surrounding presence of bone resorption, necrosis, and/or inflammation were quantitatively evaluated and classified as follows: 0 (none), 1 (minimal), 2 (low), 3 (moderate), 4 (severe).
Statistical analysis
The IBM SPSS (version 25) was used for statistical analyses. Mean and standard deviation were used for descriptive statistics. The mean difference effect measure was adopted for continuous variables, with standard error (SE) and 95% confidence interval (CI). Values of t-test < 0.05 were considered statistically significant.
Results
Animal data
All 34 rabbits survived the 6- or 12-week experimental period. Four wounds dehiscence were further stapled without any signs of wound infection. No rabbit died during the experimental period. At euthanasia, no clinical signs of inflammation or adverse tissue reactions were observed. All implants remained in situ. At baseline, rabbits had a mean weight of 3254.2 ± 199.6 mg. At last follow-up, rabbits had a mean weight of 3901.9 ± 275.0 mg. The mean weight difference from baseline was + 647.7 mg (P < 0.0001).
Result syntheses
The overall OIC of the ceramic group was greater at 6 weeks compared to the titanium implants (P = 0.0001). The other endpoints of interest were similar between the two implants at both follow-up times (P ≥ 0.05). No difference was found in BIC at 6- and 12-weeks. The results of the quantitative analyses are shown in detail in Table 1. Figures 3 and 4 show an overview at 1 × magnification of the implants at six and 12 weeks, respectively. Figures 5, 6, 7 and 8 show respectively an overview at 42 × , 140 × , 300 × , 400 × magnification of the implants at 12 weeks.
Table 1
Comparison of HPOC versus titanium at 6- and 12-weeks follow-up (MD: mean difference)
Endpoint
6 weeks
12 weeks
Ceramic
Titanium
MD
P
Ceramic
Titanium
MD
P
Lateral
BIC (%)
3.0 ± 2.6
3.1 ± 3.9
− 0.1
0.5
2.8 ± 3.5
3.9 ± 3.7
− 1.0
0.3
OIC (%)
0.9 ± 1.7
0.0 ± 0.3
0.9
0.07
1.6 ± 2.1
0.4 ± 0.1
1.1
0.09
Distal
BIC (%)
13.2 ± 6.6
11.5 ± 2.1
1.7
0.2
13.9 ± 7.6
13.6 ± 7.7
0.2
0.5
OIC (%)
5.8 ± 2.5
1.6 ± 3.5
4.2
0.003
5.6 ± 4.6
5.0 ± 3.6
0.6
0.4
Medial
BIC (%)
5.6 ± 3.7
5.3 ± 3.1
0.2
0.4
8.2 ± 4.5
7.1 ± 3.5
1.1
0.3
OIC (%)
2.2 ± 1.5
− 0.3 ± 1.9
2.5
0.003
2.8 ± 2.3
1.4 ± 1.4
1.3
0.1
Proximal
BIC (%)
12.0 ± 6.2
10.8 ± 8.7
1.2
0.4
12.6 ± 6.2
11.6 ± 3.9
1.0
0.4
OIC (%)
5.0 ± 0.0
5.0 ± 0.9
0.0
1.0
5.0 ± 0.0
4.3 ± 2.4
0.7
0.1
Overall
BIC (%)
0.3 ± 0.1
0.3 ± 0.1
0.0
0.4
0.4 ± 0.1
0.3 ± 0.7
0.0
0.2
OIC (%)
13.9 ± 3.8
6.3 ± 1.7
7.6
0.0001
14.9 ± 7.7
11.1 ± 6.1
3.8
0.2
BIC (%)
0.6 ± 0.2
0.6 ± 0.6
0.0
0.5
0.3 ± 0.1
0.4 ± 0.4
0.0
1.0
There is evidence of greater osteoid ossification in the ceramics group compared to the titanium cohort only at 6 weeks in the distal (P = 0.003), medial (P = 0.003) sides and at the overall level analysis (P = 0.0001)
No other difference between the two implants has been evidenced in the other endpoints at both 6 or 12 weeks
×
×
×
×
×
×
Discussion
The present study investigated the integration potential of HPOC implants, and then compared in vivo to titanium implants at 6 and 12 weeks. HPOC implants promote similar osteointegration to commercial titanium implants in rabbits. HPOC implants demonstrated greater osteoid implant contact at 6 weeks compared to the titanium implants, with no difference at 12 weeks. The percentage of bone implant contact of HPOC implants was similar to the titanium implants. No bone necrosis, resorption, nor inflammation was observed.
Bone is a dynamic tissue which undergoes continuous modifications and adaptations. For end stage degenerative or traumatic bony ailments, permanent or temporary implants are commonly used to restore bony function and quality of life. In the case of permanent implants (e.g. arthroplasty) the integration of the implant to the surrounding bone is crucial to ensure its longevity.
Recently, implants coating has attracted much interest. Titanium alloys may be responsible of hypersensitivity reactions, which may compromise implant longevity [33‐38]. When low-grade infection and other mechanical problems have been excluded, symptoms such as pruritus, pain, effusion, erythema, hypersensitivity reactions should be taken into consideration [39]. Ions released by corrosion of metallic wear debris may impair ossification and metal particles can be found in the soft tissues surrounding the implant [40]. Particles and ions may become clinically relevant for sensitive patients. According to the 2016 Australian Arthroplasty Register, approximately 2% of revision TKAs are attributed to metal-related pathology [41]. In selected patients with hypersensitivity, non-metallic implant solution as possible savage revision is limited with unpredictable results [42].
Current researches to develop alternative to the metal alloys are ongoing. In this context, ceramics are attracting growing interest and broad researches. Implant coating aims to optimize the biological properties of the surface to improve ossification. Ceramics are biologically inert, and consequently do not promote implant ossification. However, ceramics have high hardness and wear resistance, light weight, outstanding resistance to creep and compressive stress, avoiding imaging artefacts [12‐14]. The biological inactivity of ceramics also prevents to systemic reactions to ceramic wear debris and aseptic osteolysis. Therefore, following the development of novel coating methodologies, ceramic implants have triggered renewed interest [15, 16, 43, 44].
Anzeige
Several coating process to functionalise the surface of ceramics have been developed with promising pre-clinical and clinical results [45‐50]. Previous studies conducted in rabbit models used heterogeneous coating methodologies, follow-up, analyses, and implantation sites (Table 2). The current evidence on the clinical applications of ceramics is still limited, and translational studies are required to clarify the best coating process.
Table 2
Previous studies conducted on rabbit models investigating ceramic coating methodologies
SBA15 and SiO2. The surface was chemically modified with an organic modification of silica walls with alkoxysilane, n-octyltriethoxysilane and functionalized by soaking the mesoporous in a solution of PTHrP in PBS
2 weeks
–
–
No inflammation. Increase in staining for PCNA, Runx2 antibody, osteopontin, and/or VEGF in rabbits in both Group II and IV
TNT were immersed in supersaturated Ca(OH)2 solution, Ca(NO3)2·4H2O (0.2 M) and (NH4)2HPO4 (0.2 M) solutions to create a coating of HA. TNT-HA was subsequently functionalized with Aln by immersion in Aln solution (physical absorption)
12 weeks
no dislocation and inflammation occurred. Increase in BV/TV and Tb.Th in Group IV and III than Group I and II. Increase in
Increase in interfacial strength in Group IV than other groups and in Group III than Group I and II
Increase in osteoid tissue in Group IV and III than other Groups
Increase in BV/TV, Tb.Th, Tb.N, connectivity density in Group III than Group I. Reduction of Tb.Sp, SMI and total porosity in Group III than Group I and Group II
Two implants (2-mm diameter, 10-mm depth) into each femur
Ti–6Al–4 V implants (ø10 × 2 mm) coated by means of plasma-spray technique with HA or CaSiO3 or zinc-modified calcium silicate (Ca2ZnSi2O7) at two different Zn contents
4, 8, 12 weeks
All parameters ↑ from 4 to 8 weeks. ↑ highest values of all parameters in Group V than all other Groups
No new bone in Group I. Small bone in Group II and III stating from 4 weeks. Strong new bone formation in Group IV and V and improved osteointegration
A composite material of HA/TCP granules with or without Sr substitution
12 weeks
–
–
Greater mean trabecular area in Group I in comparison to other groups. Increase of NFkB, OPG, OC, BMP 2/4, COL-1α and IL-1 in Group III in comparison to Group IV
The current evidence on the clinical application of HPOC is limited. Alumina ceramic for talar replacement has been developed over the past 20 years as an alternative to ankle arthrodesis [59]. In 2015, Tanighuki et al. [17] published their results on 55 total talar replacements using customised HPOC, with improvement in range of motion and patient reported outcome measures [17]. Moreover, all patients returned to their activities of daily living with no complication [17]. Several studies investigated HPOC implants for total knee arthroplasty. Most studies used an alumina [18‐21] or a mixed alumina/zirconia [22‐24] femoral component and metal tibial component.
Few studies investigated the clinical outcomes of a fully alumina [60] or mixed alumina/zirconia [24] total knee implant. Clinical trials indicated that HPOC implants for total knee arthroplasty performed as well as conventional metallic components in improving patient reported outcome measures (PROMs) or inducing complications. In a recent systematic review investigating the efficacy and safety of HPOC implants for total knee arthroplasty in 14 clinical studies, wear and breakage of the prosthesis only occurred in three knees after following up for 18.1 years and in 3 patients after a minimum follow-up of 5 years, respectively [61]. The overall estimate of the rate of revision was 0.03% [61]. Despite the limited use of ceramic knee implants worldwide, and considering the advanced tribological features of the ceramic components, these components demonstrated a promising role in knee implant surgery.
This study has several limitations. The assessors (pathologist and surgeon) were not blinded to the nature of the implant, potentially increasing the risk of detection and performance biases and overestimating the results. A formal mechanical test to prove implant stability (e.g. pull-out torque test) was not conducted. However, previous reports demonstrated a definite association between pull-out torque test and histomorphometry [62‐65]. The absence of physiological load on the prostheses also impairs the validity of the present conclusion, and clinical studies should be performed. Rabbits, being reproducible, low cost, and easy to handle, are commonly used as animal model. However, between species differences exist, and must be considered in clinical translation. Current evidence could benefit of longer follow-up investigation. The results of the present study should encourage further translational studies to verify the ossification potential of ceramics in a clinical setting. Biomaterial research underlined that is possible to design and develop new bone implants with geometry, and chemical characteristics able to improve bone in growth and prevent infection. Two time points (6 and 12 weeks) were used to evaluate osteointegration. In the current literature, heterogeneous time points for analysis of implant osteointegration are reported. Though bone can be already observed at one week after implantation, proper bone remodelling starts between 6 and 12 weeks with primary and secondary osteons forming compact bone [66, 67]. Though implant roughness is a key parameter for the in vivo performance of a bone integrating insert, this parameter was not analysed in the present investigation, and further studies are required to overcome this limitation. Future studies should improve antibiofilm and antibacterial strategies to the novel materials with complementarity between knowledge and expertise in biology, chemistry, engineering and physics, associating biological models with material sciences and other technologies.
Anzeige
Conclusions
HPOC implants demonstrated a greater osteoid implant contact at 6 weeks compared to the titanium implants, and no difference at 12 weeks. The percentage of osteoid production and bone implant contact of HPOC implants was similar to that observed in titanium implants. No bone necrosis, resorption, or inflammation were observed.
Acknowledgements
None
Declarations
Ethics approval and consent to participate
This study was conducted according to the Animal Welfare Act of the Federal Republic of Germany. This study was approved by the Federal Office for Nature, Environment and Consumer Protection (Landesamt für Natur, Umwelt und Verbraucherschutz, LANU) of North Rhine-Westphalia, Federal Republic of Germany (Approval ID: 84–02.04.2016.A434).
Consent for publication
Not applicable.
Anzeige
Competing interests
Professor Maffulli is the Editor in Chief of the Journal of Orthopaedic Surgery and Research.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Zur Zementierung einer Knie-TEP wird in Deutschland zu über 98% Knochenzement verwendet, der mit einem Antibiotikum beladen ist. Ob er wirklich besser ist als Zement ohne Antibiotikum, kann laut Registerdaten bezweifelt werden.
In der Notaufnahme wird die Chance, Opfer von häuslicher Gewalt zu identifizieren, von Orthopäden und Orthopädinnen offenbar zu wenig genutzt. Darauf deuten die Ergebnisse einer Fragebogenstudie an der Sahlgrenska-Universität in Schweden hin.
Darüber reden und aus Fehlern lernen, sollte das Motto in der Medizin lauten. Und zwar nicht nur im Sinne der Patientensicherheit. Eine negative Fehlerkultur kann auch die Behandelnden ernsthaft krank machen, warnt Prof. Dr. Reinhard Strametz. Ein Plädoyer und ein Leitfaden für den offenen Umgang mit kritischen Ereignissen in Medizin und Pflege.
Ein Frauenanteil von mindestens einem Drittel im ärztlichen Op.-Team war in einer großen retrospektiven Studie aus Kanada mit einer signifikanten Reduktion der postoperativen Morbidität assoziiert.
Update Orthopädie und Unfallchirurgie
Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.