Abstract

Background

Radiostereometric Analysis (RSA) is used to measure fixation of joint prosthesis. This study compared radiation dose and image quality of a digital radiography (DR) RSA system and a computed radiography (CR) RSA system in a clinical setting.

Methods

RSA recordings of 24 hips and shoulders were analyzed. We compared two systems: (1) Arcoma T0 with ST-VI image plates and Profect CR-IR 363 reader to (2) AdoraRSA with CXDI-70C wireless DR detectors in a clinical uniplanar RSA set-up with a ± 20 degrees tube angulation and 35 cm × 43 cm detectors. Effective dose was calculated using dedicated software. Image quality was evaluated using calibration errors as calculated by the RSA software.

Results

The mean dose for hips was 0.14 (SD 0.04) mSv in the CR system and 0.05 (SD 0.02) mSv in the DR system. The mean dose for shoulders was 0.16 (SD 0.07) mSv in the CR system and 0.09 (SD 0.03) mSv in the DR system. Radiation dose was 64% (p < 0.001) and 43% (p = 0.03) lower in the DR system compared with the CR system for hip and shoulder RSA, respectively. Image quality was better for the DR system with 60–80% less calibration errors compared to the CR system.

Conclusion

Owing to highly efficient detectors and added filtration at the x-ray tubes, the DR system considerably reduced radiation dose compared with the CR system without compromising image quality.
Based on the findings in this study, we recommend replacing CR RSA systems with DR RSA systems.

Registration

Patients were selected from clinical studies performed on the two systems and approved by the local ethics committee [20060165, M-20100112, M-20070082, M-20110224, and 20070258] and registered with ClinicalTrials.gov [NCT00408096, NCT01289834, NCT00913679, NCT02311179, and NCT00679120].
Anhänge
Nur mit Berechtigung zugänglich
Literatur
7.
Zurück zum Zitat Teeuwisse W BR, Geleijns J (1998) Stralenbelasting bij orthopedische radiologie n.d. Gamma:3. Teeuwisse W BR, Geleijns J (1998) Stralenbelasting bij orthopedische radiologie n.d. Gamma:3.
10.
Zurück zum Zitat Tapiovaara M, Siiskonen T (2008) A PC-based Monte Carlo program for calculating patient doses in medical x-ray examinations. Tapiovaara M, Siiskonen T (2008) A PC-based Monte Carlo program for calculating patient doses in medical x-ray examinations.
Metadaten
Titel
Radiostereometric analysis: comparison of radiation dose and precision in digital and computed radiography
verfasst von
Peter Bo Jørgensen
Nicolai Krag-Nielsen
Lars Lindgren
Rikke Margrethe Staghøj Mørup
Bart Kaptein
Maiken Stilling
Publikationsdatum
23.11.2022
Verlag
Springer Berlin Heidelberg
Erschienen in
Archives of Orthopaedic and Trauma Surgery / Ausgabe 9/2023
Print ISSN: 0936-8051
Elektronische ISSN: 1434-3916
DOI
https://doi.org/10.1007/s00402-022-04674-0

Weitere Artikel der Ausgabe 9/2023

Archives of Orthopaedic and Trauma Surgery 9/2023 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Knie-TEP: Kein Vorteil durch antibiotikahaltigen Knochenzement

29.05.2024 Periprothetische Infektionen Nachrichten

Zur Zementierung einer Knie-TEP wird in Deutschland zu über 98% Knochenzement verwendet, der mit einem Antibiotikum beladen ist. Ob er wirklich besser ist als Zement ohne Antibiotikum, kann laut Registerdaten bezweifelt werden.

Häusliche Gewalt in der orthopädischen Notaufnahme oft nicht erkannt

28.05.2024 Häusliche Gewalt Nachrichten

In der Notaufnahme wird die Chance, Opfer von häuslicher Gewalt zu identifizieren, von Orthopäden und Orthopädinnen offenbar zu wenig genutzt. Darauf deuten die Ergebnisse einer Fragebogenstudie an der Sahlgrenska-Universität in Schweden hin.

Fehlerkultur in der Medizin – Offenheit zählt!

28.05.2024 Fehlerkultur Podcast

Darüber reden und aus Fehlern lernen, sollte das Motto in der Medizin lauten. Und zwar nicht nur im Sinne der Patientensicherheit. Eine negative Fehlerkultur kann auch die Behandelnden ernsthaft krank machen, warnt Prof. Dr. Reinhard Strametz. Ein Plädoyer und ein Leitfaden für den offenen Umgang mit kritischen Ereignissen in Medizin und Pflege.

Mehr Frauen im OP – weniger postoperative Komplikationen

21.05.2024 Allgemeine Chirurgie Nachrichten

Ein Frauenanteil von mindestens einem Drittel im ärztlichen Op.-Team war in einer großen retrospektiven Studie aus Kanada mit einer signifikanten Reduktion der postoperativen Morbidität assoziiert.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.