Positional transfer accuracy of titanium base implant abutment provided by two different scan body designs: an invitro study

verfasst von: Rania E. Ramadan, Mahmoud Khamis Abdel Razek, Faten S. Mohamed, Rania A. Fahmy, Mervat E. Abd-Ellah

Erschienen in: BMC Oral Health | Ausgabe 1/2023

Abstract

Background

The variabilities in design and material of scan bodies have a major role in the positional transfer accuracy of implants. The purpose of this invitro study was to compare the 3D transfer accuracy (trueness and precision) of titanium base (TB) abutment position provided by 2 different scan bodies: one-piece scan body (SB) in comparison to two-piece healing abutment and scan peg (HA-SP).

Methods

A maxillary model with a dummy implant in the 2nd premolar (Proactive Tapered Implant; Neoss) was 3D printed and TB (Ti Neolink Mono; Neoss) was tightened on the implant and scanned by using a laboratory scanner (inEos X5; Dentsply Sirona) (reference scan). An SB (Elos Medtech) and an HA-SP (Neoss) were subsequently connected to the implant and were scanned 10 times each by using the same scanner (test scans). All the scans were exported as STL files and imported into CAD software where the TBs were formed. Test scans were superimposed on reference scans for transfer accuracy analysis using 3D metrology software (GOM Inspect; GOM GmbH) in terms of angular deviation in vertical and horizontal directions, linear deviation in each XYZ axis of TBs and total linear deviation in all axes. Statistical analysis was done using independent sample t test. When Levene’s test for equality of variances was significant, Welch’s t-test was used. (P value < 0.05)

Results

Significant differences were found amongst the tested groups in both angular and linear deviation in terms of trueness with less deviation values for the SB group (P < 0.001). For the precision, significant differences were found amongst the tested groups in angular deviation in vertical direction with less deviation value for the SB group compared to HA-SP group (P < 0.001). However, no significant difference was found between the tested groups regarding the angular deviation in horizontal direction (P = 1.000). Moreover, significant differences were found amongst the tested groups in linear deviations with less linear deviations in XYZ axes for SB compared to HA-SP group (P = 0.020, < 0.001, = 0.010 respectively).

Conclusions

SB showed less angular and linear deviation values in the 3D positional transfer of TB than HA-SP indicating higher degree of accuracy of SB.
Hinweise

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
3D
Three-dimensional
HA-SP
Healing abutment-scan peg
NIST
National Institute of Standards and Technology
PEEK
PolyEtherEtherKetone
PTB
Physikalisch-Technische Bundesanstalt
SB
Scan body
STL
Standard Tessellation Language
TB
Titanium base

Background

Transfer accuracy of implant position is considered a crucial step in implantology as it is a fundamental requirement in the long-term effectiveness of implant-supported restorations [13]. Because an implant’s mobility is 10 times lower than that of a natural tooth, implant prosthodontics requires extremely high accuracy in transferring the 3D implant position intraorally to a model virtually. Inaccurate implant position’s transfer may eventually result in biological, mechanical, and functional complications [47].
Scan bodies are described by Mizumoto and Yilmaz [2] as “complex implant position transfer devices.“ Scan bodies may be attached to an implant analog extraorally on a stone cast after conventional impression and scanned by a laboratory scanner (indirect digital impression) or may be attached to an implant intraorally and scanned by an intraoral scanner (direct digital impression) [811]. The direct digital impression offers certain benefits over the indirect approach including the direct transfer of intraoral data to the laboratory, and the elimination of potential errors associated to dimensional changes of the conventional impression and stone cast [2, 12].
Scan bodies from different manufacturers come in varying designs, sizes, implant connection types and are made of different materials. They diverge in several ways. They may be made of titanium, PEEK, or a combination [2]. Also, they may be designed as one or two pieces, used single or several times, screw-retained or snap-on (friction grip), radiolucent or radiopaque, having a sandblasted or coated surface, tall or short, with a wide or narrow width, and with a simplified or complicated design [13]. These differences can play a role in the transfer accuracy of implant position [57].
Commonly, most scan bodies are one-piece of cylindrical or conical design with little or no taper and their geometry usually does not simulate the natural tooth’s emergence profile. All of them consist of an upper scan region, a middle body, and a base that is connected to the implant [2]. The part that is scanned by the intraoral scanner is the scan region, however the base should be adequately seated on the implant connection. Thus, the design of scan base depends on the type of implant connection and dimension of platform. In addition, the base material may have an impact on the fit when screwing on the implant, whereas the surface characteristics of the material used for the scan region may affect how many spots the intraoral scanner can detect digitally [2, 3, 14].
However, other scan bodies are 2-pieces like HA-SP (Neoss) that is formed of a contoured HA and a scannable body called SP. The SP is maintained by friction between an indentation on the SP and a vertical depression in the contoured HA, thus acting as anti-rotational feature [1518]. HA is made entirely of PEEK, and SP is made of medical grade acrylic-based polymer. It does not need to be sprayed with nonreflective powder required for metals during scanning and contrary to titanium scan bodies, and the scan is captured at the HA’s level [1518]. Furthermore, the soft tissues can be properly contoured for appropriate emergence profile, which makes it easier for the soft tissues to accept the definitive implant-supported restoration. The implant and peri-implant tissues are scanned together, and the CAD software includes the contoured HA, allowing its duplication on the definitive restoration. Consequently, reducing the dental chairside time and improving the satisfaction of patients and clinicians [19].
Evaluating the 3D transfer accuracy of TB abutment position is always a very important objective as it corresponds to the accuracy of implant position. However, it is challenging to conduct it in a clinical study due to the lack of a reference model since a patient’s jaw cannot be analyzed using high-precision laboratory scanners [4, 20]. In addition, in vitro studies led to a coordinate-based evaluation [21]. Therefore, the current invitro study was conducted.
Based on International Organization for Standardization 5725-1 [22], accuracy includes trueness in addition to precision. Trueness is the proximity to the actual position of a reference object, while precision is the proximity of same object’s position to one another which is attained through repetitive measurements.
The purpose of this invitro study was to compare the 3D transfer accuracy of TB abutment position provided by 2 different designs of scan bodies: one-piece SB in comparison to two-piece HA-SP. (Fig. 1) The null hypothesis was that no difference would be found in the 3D transfer accuracy (trueness and precision) of TB position using the mentioned scan bodies.

Materials and methods

Sample size calculation

Based on previous studies [5, 16] a sample size of 10 scans for each group was calculated with a software program (G*Power v3.1.9.2; Heinrich Heine University Düsseldorf) [23] supposing a 95% level of confidence and 80% power. Following the preliminary findings, it was determined that the sample size permitted the identification of high statistic significant differences; hence, it was decided that no additional scans were required.

Fabrication of 3D printed implant model

A maxillary resin model with an implant site at the right 2nd premolar was designed on CAD design software (Dental DB 3.0 Galway; Exocad GmbH) using the model creator module and the Neoss implant library as shown in Fig. 2. Then, the model was additively manufactured using 3D printer (Dent2 3D printer; Mogassam Co LLC). (Fig. 3A) A dummy implant (4.0 × 11 mm) (Proactive Tapered Implant; Neoss) was cemented in the corresponding site using cyanoacrylate adhesive (Super Bonder; Henkel Loctite Corp) so that the vertical mark on the screwdriver was positioned mid-buccally during implant insertion for correct further orientation of TB abutment and scan bodies [19]. (Fig. 3B)

Data acquisition

A hexed TB abutment (Ti Neolink Mono; Neoss) was screwed on the implant by a manual torque wrench to 32 Ncm using the gold prosthetic screw so that the flat surface was positioned buccally. (Fig. 4A) The model with TB was scanned by using a high precision laboratory scanner (inEos X5; Dentsply Sirona) [24, 25] and was exported as an STL file (STL-1) that was further used to get the reference scan. (Fig. 4B)
Then, the one-piece SB (Elos Medtech) was tightened on the implant to 10 Ncm so that the beveled surface was placed buccally, and the model was scanned by the same scanner 10 consecutive times without taking the model out of the scanning tray in between scans (STL-2). (Fig. 4C, D)
Following that, the SB was removed, and the premolar HA (Healing Abutment-Scan Peg; Neoss) was inserted so that the vertical groove on the HA was placed buccally and tightened to 10 Ncm. Then, the SP was secured on the HA by means of a vertical indentation on the SP and a vertical groove in the HA, thus acting as an anti-rotational means and enabling proper seating of the SP in the HA. The model with HA-SP was scanned 10 consecutive times with the same scanner without taking the model out of the scanning tray in between scans (STL-3). (Fig. 4E, F)
All the scans were made on the same day of printing the model [26]. Each scan was checked for imperfections, especially on the investigated scan bodies’ surfaces. A scan was accepted when no significant flaws or voids were found [27].

Reference scan

The reference scan was obtained through the following steps:
Step 1: An STL file of the scanned model with SB (STL-2) was imported to the CAD design software. (Fig. 5A)
Step 2: The SB corresponding to hexed TB used was selected from Neoss library (Neolink mono 3.5–5.5 IO) on the CAD software and superimposed using the “best-fit matching” tool. (Fig. 5B)
Step 3: The TB was formed, trimmed from the model, and exported solely as an STL-4. (Fig. 5C)
Step 4: An STL file of scanned model with TB (STL-1) was imported to the software (Fig. 5D) together with STL-4 which was subsequently aligned through fixing points on the 2 meshes (Fig. 5E) to get a well-defined structure of hexed TB. This STL was considered as the reference STL. (Fig. 5F).

SB test scans

An STL file of scanned model with SB (STL-2) was imported to the CAD software and superimposed using the “best-fit matching” tool with the SB in the library to form the virtual position of the TB (Fig. 6).

HA-SP test scans

AnSTL file of scanned model with HA-SP (STL-3) was imported to the CAD software. The scan marker used was selected from the library (Scan peg premolar Neolink) on the software and superimposed using the “best-fit matching” tool to form the virtual position of the TB. (Fig. 7)
The accuracy of CAD software “best-fit matching” for each scan was checked by the section view that could show how much the scan marker in CAD library corresponds with the used scan bodies and the color map that was displayed where the blue color means the zero deviation [2, 13]. For each test scan in each group, this alignment process was carried out once. After being aligned, the model was exported as a single STL file in the form of a model with TB in position. For further accuracy analysis, a total of 10 aligned models were produced for each group having the virtual position of the corresponding TBs.

Accuracy analysis

Positional transfer accuracy was analyzed with 3D metrology software (GOM Inspect; GOM GmbH). The measurements were conducted by a single operator (RER). The measurements were done 3 times and the average was obtained. There was no need for intra observer reliability because every step made was based on computer-assisted algorithms with reproducible outcomes [28].
Regarding trueness analysis, angular and linear deviations between reference and test scans were measured [22]. Each test scan was compared to the reference scan as SB test scan having the position of TB formed by the SB was compared to the reference scan having the original position of TB. Similarly, HA-SP test scan having the position of TB formed by HA-SP was compared to the reference scan having the original position of TB.
For precision analysis, the variance of angular and linear deviations within each test group was measured [22]. In other words, the repeatability of SB or HA-SP to form the same position of TB every time.
The software’s “pre-alignment” tool was used to initially align the received STLs. The “local best-fit” tool was then used to select all teeth other than the TB site for further alignment. (Fig. 8A, B) A coordinate system was constructed and applied during the whole analysis to calculate the angular and linear deviation of the TBs position. The XYZ axes were oriented on each scan originating from a selected point on the occlusobuccal surface of the TB so that the buccopalatal axis was the X axis, the mesiodistal axis was Y axis and the occlusocervical axis was Z axis. These coordinates were used to make sure that the TBs were always set to the same axes for subsequent comparisons. (Fig. 9A, B)

Measurement of angular deviations

Angular deviations were measured in a vertical direction and a horizontal direction by measuring the angle between predetermined lines traced on the test TB compared to the reference TB. The surface of the scans was presented as series of flat polygons or triangles virtually to allow for appropriate point selection [2]. Points 1 and 2 were constructed to draw line 1 on occlusocervical direction of the reference TB and points 1 and 3 were constructed to draw line 2 on mesiodistal direction of the reference TB. (Fig. 10A) Points 4 and 5 were constructed to draw line 3 on occlusocervical direction of the test TB and points 4 and 6 were constructed to draw a line 4 on mesiodistal direction of the test TB. (Fig. 10B) Then, the angles between line 1 on reference scan and line 3 on the test scan were calculated resembling angular deviation in vertical direction. Also, the angles between line 2 on reference scan and line 4 on the test scan were calculated resembling angular deviation in horizontal direction. (Fig. 10C)

Measurement of linear deviations in XYZ axes

Regarding the linear deviations, one fitting cylinder was constructed for each TB on the reference and the test scans. This fitting cylinder was constructed using the “Gaussian best-fit” method by selecting all cervical surface of the TB to guarantee repeatable cylinder’s center points and remove human-created mistakes affecting positional transfer accuracy measurements [28]. (Fig. 11A, B) A 2-point distance (point7-8) was constructed between both fitting cylinders’ center points corresponding to linear deviations in XYZ axes on the reference and the test scans to be compared. (Fig. 11C) The 3D linear deviation value was calculated according to the following formula: 3D linear deviation = \((\sqrt{{X}^{2}+{Y}^{2} +{Z}^{2}}\) ) [9, 17, 29, 30].
Low deviation value indicated a high degree of 3D matching of aligned files, which was translated to high trueness and precision.

Statistical analysis

Statistical analysis was done by using a statistical software program (IBM SPSS Statistics, v25; IBM Corp). Kolmogorov-Smirnov test of normality showed no significant difference in the distribution of the variables; hence parametric statistics was used. Comparisons were done between two studied independent normally distributed variables using independent sample t test. When Levene’s test for equality of variances was significant, Welch’s t-test was applied. Statistical significance was tested at P value < 0.05.

Results

Regarding the trueness, significant differences were found amongst the tested groups in vertical and horizontal angular deviation, linear deviation in each axis and 3D linear deviation in XYZ axes. SB group showed statistically significant less deviation values compared to HA-SP group (P < 0.001) as shown in Table 1. Thus, SB group had higher trueness values than HA-SP group.
Table 1
Comparison of trueness between SB group and HA-SP group
 
Group
Test of significance
P value
SB
(n = 10)
HA-SP
(n = 10)
Angular deviation in vertical direction (°)
 
- Min – Max
- Mean ± SD
- Standard error of mean
- 95% CI of the mean
- 25th Percentile − 75th Percentile
0.350–0.410
0.379 ± 0.023
0.007
0.363–0.395
0.360–0.400
1.590–1.780
1.676 ± 0.073
0.023
1.624–1.728
1.600–1.750
t(W)(df=10.734) = 53.482
P < 0.001
Angular deviation in horizontal direction (°)
 
- Min – Max
- Mean ± SD
- Standard error of mean
- 95% CI of the mean
25th Percentile − 75th Percentile
0.220–0.430
0.288 ± 0.071
0.023
0.237–0.339
0.220–0.350
0.500–0.650
0.571 ± 0.044
0.014
0.539–0.603
0.540–0.600
t(df=18) = 10.658
P < 0.001
Linear deviation in X axis (μm)
 
- Min – Max
- Mean ± SD
- Standard error of mean
- 95% CI of the mean
- 25th Percentile − 75th Percentile
0.001–0.005
0.002 ± 0.001
0.000
0.001–0.003
0.001–0.003
0.153–0.167
0.157 ± 0.005
0.002
0.154–0.161
0.154–0.160
t(W)(df=10.507) = 94.452
P < 0.001
Linear deviation in Y axis (μm)
 
- Min – Max
- Mean ± SD
- Standard error of mean
- 95% CI of the mean
- 25th Percentile − 75th Percentile
0.001–0.005
0.002 ± 0.001
0.000
0.001–0.003
0.001–0.003
0.064–0.067
0.065 ± 0.001
0.000
0.064–0.066
0.064–0.066
t(df=18) = 108.393
P < 0.001
Linear deviation in Z axis (μm)
 
- Min – Max
- Mean ± SD
- Standard error of mean
- 95% CI of the mean
- 25th Percentile − 75th Percentile
0.053–0.055
0.054 ± 0.001
0.000
0.053–0.055
0.053–0.055
0.063–0.065
0.064 ± 0.001
0.000
0.063–0.065
0.063–0.065
t(df=18) = 26.678
P < 0.001
3D linear deviation (μm)
 
- Min – Max
- Mean ± SD
- Standard error of mean
- 95% CI of the mean
- 25th Percentile − 75th Percentile
0.053–0.055
0.054 ± 0.001
0.000
0.053–0.055
0.053–0.055
0.178–0.190
0.182 ± 0.004
0.001
0.179–0.185
0.179–0.184
t(W)((df=9.742) = 89.841
P < 0.001
Min – Max Minimum – Maximum; SD Standard Deviation; CI Confidence interval; t Independent samples t test;
W Welch test; df degree of freedom
For the precision, significant differences were found amongst the tested groups in angular deviation in vertical direction. SB group showed statistically significant less angular deviation in vertical direction compared to HA-SP group (P < 0.001). However, no significant difference was found between the tested groups regarding the angular deviation in horizontal direction (P = 1.000). Moreover, significant differences were found amongst the tested groups in linear deviations in each axis as well as 3D linear deviation in XYZ axes. SB group showed statistically significant less linear deviation values in each axis and 3D linear deviation value compared to HA-SP group (P = 0.020, < 0.001, = 0.010, = 0.009 respectively) as shown in Table 2. Thus, SB group had higher precision values than HA-SP group except for the angular deviation in horizontal direction that was the same precision.
Table 2
Comparison of precision between SB group and HA-SP group
 
Group
Test of significance
P value
SB
(n = 10)
HA-SP
(n = 10)
Angular deviation in vertical direction (°)
 
- Min – Max
- Mean ± SD
- Standard error of mean
- 95% CI of the mean
- 25th Percentile − 75th Percentile
0.110–0.360
0.248 ± 0.087
0.028
0.186–0.310
0.180–0.300
0.400–0.860
0.641 ± 0.163
0.052
0.524–0.758
0.530–0.800
t(df=18) = 6.711
P < 0.001
Angular deviation in horizontal direction (°)
 
- Min – Max
- Mean ± SD
- Standard error of mean
- 95% CI of the mean
- 25th Percentile − 75th Percentile
0.290–0.790
0.447 ± 0.178
0.056
0.320–0.574
0.320–0.580
0.020–0.710
0.447 ± 0.247
0.078
0.270–0.624
0.270–0.690
t(df=18) = 0.000
P = 1.000
Linear deviation in X axis (μm)
 
- Min – Max
- Mean ± SD
- Standard error of mean
- 95% CI of the mean
- 25th Percentile − 75th Percentile
0.024–0.046
0.038 ± 0.007
0.002
0.032–0.043
0.037–0.042
0.024–0.101
0.065 ± 0.031
0.010
0.043–0.087
0.030–0.093
t(W)(df=10.055) = 2.775
P = 0.020
Linear deviation in Y axis (μm)
 
- Min – Max
- Mean ± SD
- Standard error of mean
- 95% CI of the mean
- 25th Percentile − 75th Percentile
0.000-0.018
0.010 ± 0.006
0.002
0.006–0.014
0.005–0.014
0.017–0.060
0.035 ± 0.015
0.005
0.024–0.046
0.021–0.049
t(W)(df=11.566) = 4.813
P < 0.001
Linear deviation in Z axis (μm)
 
- Min – Max
- Mean ± SD
- Standard error of mean
- 95% CI of the mean
- 25th Percentile − 75th Percentile
0.022–0.042
0.032 ± 0.006
0.002
0.028–0.036
0.029–0.034
0.019–0.102
0.063 ± 0.031
0.010
0.041–0.085
0.032–0.091
t(W)(df=9.588) = 3.196
P = 0.010
3D linear deviation (μm)
 
- Min – Max
- Mean ± SD
- Standard error of mean
- 95% CI of the mean
- 25th Percentile − 75th Percentile
0.033–0.061
0.051 ± 0.008
0.002
0.045–0.056
0.049–0.055
0.037–0.154
0.098 ± 0.045
0.014
0.066–0.130
0.047–0.133
t(W)(df=9.588) = 3.278
P = 0.009
Min – Max Minimum – Maximum; SD Standard Deviation; CI Confidence interval; t Independent samples t test;
W Welch test; df degree of freedom

Discussion

The positional transfer accuracy (trueness and precision) of TB abutment provided by one-piece SB was different from that provided by two-piece HA-SP in terms of angular and linear deviations of TB position and accordingly the null hypothesis was rejected. The current study analyzed the positional transfer accuracy of TB provided by two scan body designs only as these were the only designs compatible to the used Neoss implant.
A laboratory scanner with 2.1 μm accuracy according to DIN EN ISO 12836.2015 was employed in the current study to standardize the digitizing process by reducing human errors that could occur by intraoral scanners, since a laboratory scanner automatically rotates the scanned object into various locations to provide the ideal illumination for the best data acquisition [31]. This scanner utilized to create the reference and test STL files has also been recommended in previous studies on deviation analyses [24, 25]. All the scans were made on the same day of printing the model to overcome the dimensional changes that may occur to the printed model over time, thus obtaining accurate superimposition of the scans [26].
Regarding the accuracy analysis, trueness and precision were analyzed based on ISO 5725-1 [22]. The ISO technique was chosen as the standard methodology, which the authors of the current study believe will be useful for further comparison of these findings with other studies [32].
The used 3D metrology software to make superimpositions of different scans had been widely used for evaluation of accuracy [1517, 28]. This software program was classified as class1 (lowest measurement deviations) by the PTB and the NIST [33]. Although the best-fit alignment is often employed for analyses of accuracy, it has little shortcomings; the software looks for the perfect superimposition of two surface scans with the minimal deviation between all surface points [16]. The distance between two corresponding points may as a result be underestimated [34]. Thus, the current study utilized a local rather than an overall best-fit alignment except for the site of TB. Following the local best-fit algorithm, selected points on the TBs were used for the comparisons to reduce the underestimate [16].
Previous studies [15, 16] investigated the intraoral scan accuracy of HA-SP by measuring distance and angular deviations. However, these studies did not investigate which SB was more accurate in transfer of 3D implant position. Thus, it was evaluated in the current study.
The present study measured the angular deviation of TBs position using standardized points of measurement that were selected on occlusocervical and mesiodistal directions of TBs to reduce the possibility of random measurement errors in 3D analysis. These points were selected after virtual presentation of the scan surface as series of flat polygons or triangles as mentioned in a systematic review by Mizumoto and Yilmaz [2]. Additionally, the study measured the linear deviation values of TBs in XYZ axes by forming fitting cylinders on the cervical area of TBs similar to a previous study by Atalay et al. [17] who created fitting circles 3 mm below and parallel to the upper plane of HA-SP to evaluate the impact of implant location and operator on the scan accuracy provided by HA-SP. This technique may be advantageous as all available information can be used for comparisons and it did not depend on certain points that may not be visible as reported by Atalay et al. [17].
In the current study, SB group showed lower 3D deviation value than HA-SP group in terms of both trueness and precision. However, Yilmaz et al., [16] performed an in vitro study on intraoral scan accuracy using SB and HA-SP using points selected on certain landmarks and reported that SB had lower linear deviation than HA-SP in terms of trueness in one point only which was middle point on buccal coronal slope. Additionally, no significant difference between SB and HA-SP were reported in their study in terms of precision. This difference in the findings may be attributed to dissimilarities in the methodology between the studies.
The positional transfer accuracy of TB abutment provided by one-piece SB was higher than that provided by two-piece HA-SP and this may be related to several factors. The variability in the design characteristics between the 2 mentioned scan bodies may have led to the difference in accuracy. HA-SP was formed of 2 pieces and the seating of the 2 pieces may be influenced by the manufacturing tolerances, that could have an effect on the accuracy as reported in previous in vitro studies [6, 35]. Furthermore, Donmez et al., [18] reported that the absence of a flat surface in HA-SP might have caused the higher deviation values upon investigation of the congruence between the meshes of HA-SP obtained by four different intraoral scanners used in their study and the corresponding library file. This may explain the higher deviation values of HA-SP in comparison to SB in the present study. Additionally, the difference in the implant connection’s material of each scan body was another contributing factor affecting the positional transfer accuracy of implant as reported by Schmidt et al., [5] that may contribute to the final prosthetic misfit. The one-piece SB was made of PEEK and its implant connection was in titanium. However, the HA was made entirely of PEEK, its implant connection was in PEEK and its SP was made of medical grade acrylic-based polymer. Furthermore, the difference in the installation method of each scan body was another factor influencing the transfer accuracy as one-piece SB was tightened directly on implants, but the SP in the HA-SP system was fastened to HA through autorotational frictional means without screw [1518].

Limitations and recommendations

The limitations of this in vitro study included that the model utilized in the current study did not accurately represent the intraoral circumstances; teeth, soft tissues, and the existence of saliva or blood [27, 36]. The use of a laboratory scanner is another limitation, thus future studies should be conducted utilizing an industrial high-accuracy scanner [37]. The used CAD software “best-fit matching” tool in the current study for generating the reference and test scans is considered as a limitation due to the lack of data in the literature about its margin of error, therefore further use of other CAD software programs should be done. In addition, the reference scan was produced by aligning a TB from the CAD library to the original scanned model and this may also introduce some degree of error in the reference scan. Furthermore, angular and linear deviations may influence the proximal and occlusal contacts of definitive implant-supported restorations [15, 16]. Therefore, clinical performance and positional accuracy of definitive crowns regarding proximal contacts with the neighboring teeth and occlusal contacts with the opposing teeth should be further evaluated using both scan bodies through clinical studies as well as the effect of HA-SP on the contouring of the soft tissues.

Conclusions

Within the limitations of this in vitro study, the following conclusions were drawn:
1.
One-piece SB showed less angular and linear deviation in the 3D transfer of TB position than two-piece HA-SP indicating a higher degree of accuracy of SB.
 
2.
Different designs, materials and installation methods of the scan bodies could affect the 3D positional transfer accuracy of TB.
 

Acknowledgements

The authors are grateful to Dr. Akram F. Neena, Lecturer, Department of Prosthodontics, Faculty of Dentistry, Alexandria University, Alexandria, Egypt for his valuable support during the conduction of this study.

Declarations

Not applicable.
Not applicable.

Competing interests

The authors declare no competing interests.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Flügge T, van der Meer WJ, Gonzalez BG, Vach K, Wismeijer D, Wang P. The accuracy of different dental impression techniques for implant-supported dental prostheses: a systematic review and meta-analysis. Clin Oral Implants Res. 2018;29:374–92.CrossRefPubMed Flügge T, van der Meer WJ, Gonzalez BG, Vach K, Wismeijer D, Wang P. The accuracy of different dental impression techniques for implant-supported dental prostheses: a systematic review and meta-analysis. Clin Oral Implants Res. 2018;29:374–92.CrossRefPubMed
2.
Zurück zum Zitat Mizumoto RM, Yilmaz B. Intraoral scan bodies in implant dentistry: a systematic review. J Prosthet Dent. 2018;120:343–52.CrossRefPubMed Mizumoto RM, Yilmaz B. Intraoral scan bodies in implant dentistry: a systematic review. J Prosthet Dent. 2018;120:343–52.CrossRefPubMed
3.
Zurück zum Zitat Rutkunas V, Geciauskaite A, Jegelevicius D, Vaitiekunas M. Accuracy of digital implant impressions with intraoral scanners. A systematic review. Eur J Oral Implantol. 2017;10:101–20.PubMed Rutkunas V, Geciauskaite A, Jegelevicius D, Vaitiekunas M. Accuracy of digital implant impressions with intraoral scanners. A systematic review. Eur J Oral Implantol. 2017;10:101–20.PubMed
4.
Zurück zum Zitat Schmidt A, Wöstmann B, Schlenz MA. Accuracy of digital implant impressions in clinical studies: a systematic review. Clin Oral Implants Res. 2022;33:573–85.CrossRefPubMed Schmidt A, Wöstmann B, Schlenz MA. Accuracy of digital implant impressions in clinical studies: a systematic review. Clin Oral Implants Res. 2022;33:573–85.CrossRefPubMed
5.
Zurück zum Zitat Schmidt A, Billig JW, Schlenz MA, Wöstmann B. The influence of using different types of scan bodies on the transfer accuracy of Implant position: an in Vitro Study. Int J Prosthodont. 2021;34:254–60.CrossRefPubMed Schmidt A, Billig JW, Schlenz MA, Wöstmann B. The influence of using different types of scan bodies on the transfer accuracy of Implant position: an in Vitro Study. Int J Prosthodont. 2021;34:254–60.CrossRefPubMed
6.
Zurück zum Zitat Schmidt A, Billig JW, Schlenz MA, Rehmann P, Wöstmann B. Influence of the Accuracy of Intraoral Scanbodies on Implant position: differences in Manufacturing Tolerances. Int J Prosthodont. 2019;32:430–2.CrossRefPubMed Schmidt A, Billig JW, Schlenz MA, Rehmann P, Wöstmann B. Influence of the Accuracy of Intraoral Scanbodies on Implant position: differences in Manufacturing Tolerances. Int J Prosthodont. 2019;32:430–2.CrossRefPubMed
7.
Zurück zum Zitat Revilla-León M, Smith Z, Methani MM, Zandinejad A, Özcan M. Influence of scan body design on accuracy of the implant position as transferred to a virtual definitive implant cast. J Prosthet Dent. 2021;125:918–23.CrossRefPubMed Revilla-León M, Smith Z, Methani MM, Zandinejad A, Özcan M. Influence of scan body design on accuracy of the implant position as transferred to a virtual definitive implant cast. J Prosthet Dent. 2021;125:918–23.CrossRefPubMed
8.
Zurück zum Zitat Amin S, Weber HP, Finkelman M, El Rafie K, Kudara Y, Papaspyridakos P. Digital vs. conventional full-arch implant impressions: a comparative study. Clin Oral Implants Res. 2017;28:1360–7.CrossRefPubMed Amin S, Weber HP, Finkelman M, El Rafie K, Kudara Y, Papaspyridakos P. Digital vs. conventional full-arch implant impressions: a comparative study. Clin Oral Implants Res. 2017;28:1360–7.CrossRefPubMed
9.
Zurück zum Zitat Papaspyridakos P, Gallucci GO, Chen CJ, Hanssen S, Naert I, Vandenberghe B. Digital versus conventional implant impressions for edentulous patients: accuracy outcomes. Clin Oral Implants Res. 2016;27:465–72.CrossRefPubMed Papaspyridakos P, Gallucci GO, Chen CJ, Hanssen S, Naert I, Vandenberghe B. Digital versus conventional implant impressions for edentulous patients: accuracy outcomes. Clin Oral Implants Res. 2016;27:465–72.CrossRefPubMed
10.
Zurück zum Zitat Aktas G, Özcan N, Aydin DH, Şahin E, Akça K. Effect of digitizing techniques on the fit of implant-retained crowns with different antirotational abutment features. J Prosthet Dent. 2014;111:367–72.CrossRefPubMed Aktas G, Özcan N, Aydin DH, Şahin E, Akça K. Effect of digitizing techniques on the fit of implant-retained crowns with different antirotational abutment features. J Prosthet Dent. 2014;111:367–72.CrossRefPubMed
11.
Zurück zum Zitat Stimmelmayr M, Güth JF, Erdelt K, Edelhoff D, Beuer F. Digital evaluation of the reproducibility of implant scanbody fit-an in vitro study. Clin Oral Investig. 2012;16:851–6.CrossRefPubMed Stimmelmayr M, Güth JF, Erdelt K, Edelhoff D, Beuer F. Digital evaluation of the reproducibility of implant scanbody fit-an in vitro study. Clin Oral Investig. 2012;16:851–6.CrossRefPubMed
12.
Zurück zum Zitat Lee SJ, Gallucci GO. Digital vs. conventional implant impressions: efficiency outcomes. Clin Oral Implants Res. 2013;24:111–5.CrossRefPubMed Lee SJ, Gallucci GO. Digital vs. conventional implant impressions: efficiency outcomes. Clin Oral Implants Res. 2013;24:111–5.CrossRefPubMed
13.
Zurück zum Zitat Gracis S, Appiani A, Noè G. Digital workflow in implant prosthodontics: the critical aspects for reliable accuracy. J Esthet Restor Dent. 2023;35:250–61.CrossRefPubMed Gracis S, Appiani A, Noè G. Digital workflow in implant prosthodontics: the critical aspects for reliable accuracy. J Esthet Restor Dent. 2023;35:250–61.CrossRefPubMed
14.
Zurück zum Zitat Holst S, Persson A, Wichmann M, Karl M. Digitizing implant position locators on master casts: comparison of a noncontact scanner and a contact-probe scanner. Int J Oral Maxillofac Implants. 2012;27:29–35.PubMed Holst S, Persson A, Wichmann M, Karl M. Digitizing implant position locators on master casts: comparison of a noncontact scanner and a contact-probe scanner. Int J Oral Maxillofac Implants. 2012;27:29–35.PubMed
15.
Zurück zum Zitat Çakmak G, Donmez MB, Atalay S, Yilmaz H, Kökat AM, Yilmaz B. Accuracy of single implant scans with a combined healing abutment-scan body system and different intraoral scanners: an in vitro study. J Dent. 2021;113:103773.CrossRefPubMed Çakmak G, Donmez MB, Atalay S, Yilmaz H, Kökat AM, Yilmaz B. Accuracy of single implant scans with a combined healing abutment-scan body system and different intraoral scanners: an in vitro study. J Dent. 2021;113:103773.CrossRefPubMed
16.
Zurück zum Zitat Yilmaz B, Gouveia D, Marques VR, Diker E, Schimmel M, Abou-Ayash S. The accuracy of single implant scans with a healing abutment-scanpeg system compared with the scans of a scanbody and conventional impressions: an in vitro study. J Dent. 2021;110:103684.CrossRefPubMed Yilmaz B, Gouveia D, Marques VR, Diker E, Schimmel M, Abou-Ayash S. The accuracy of single implant scans with a healing abutment-scanpeg system compared with the scans of a scanbody and conventional impressions: an in vitro study. J Dent. 2021;110:103684.CrossRefPubMed
17.
Zurück zum Zitat Atalay S, Çakmak G, Donmez MB, Yilmaz H, Kökat AM, Yilmaz B. Effect of implant location and operator on the accuracy of implant scans using a combined healing abutment-scan body system. J Dent. 2021;115:103855.CrossRefPubMed Atalay S, Çakmak G, Donmez MB, Yilmaz H, Kökat AM, Yilmaz B. Effect of implant location and operator on the accuracy of implant scans using a combined healing abutment-scan body system. J Dent. 2021;115:103855.CrossRefPubMed
18.
Zurück zum Zitat Donmez MB, Marques VR, Çakmak G, Yilmaz H, Schimmel M, Yilmaz B. Congruence between the meshes of a combined healing abutment-scan body system acquired with four different intraoral scanners and the corresponding library file: an in vitro analysis. J Dent. 2022;118:103938.CrossRefPubMed Donmez MB, Marques VR, Çakmak G, Yilmaz H, Schimmel M, Yilmaz B. Congruence between the meshes of a combined healing abutment-scan body system acquired with four different intraoral scanners and the corresponding library file: an in vitro analysis. J Dent. 2022;118:103938.CrossRefPubMed
19.
Zurück zum Zitat Yilmaz B, Abou-Ayash S. A digital intraoral implant scan technique using a combined healing abutment and scan body system. J Prosthet Dent. 2020;123:206–9.CrossRefPubMed Yilmaz B, Abou-Ayash S. A digital intraoral implant scan technique using a combined healing abutment and scan body system. J Prosthet Dent. 2020;123:206–9.CrossRefPubMed
20.
Zurück zum Zitat Kuhr F, Schmidt A, Rehmann P, Wöstmann B. A new method for assessing the accuracy of full arch impressions in patients. J Dent. 2016;55:68–74.CrossRefPubMed Kuhr F, Schmidt A, Rehmann P, Wöstmann B. A new method for assessing the accuracy of full arch impressions in patients. J Dent. 2016;55:68–74.CrossRefPubMed
21.
Zurück zum Zitat Schmidt A, Rein PE, Wöstmann B, Schlenz MA. A comparative clinical study on the transfer accuracy of conventional and digital implant impressions using a new reference key-based method. Clin Oral Implants Res. 2021;32:460–9.CrossRefPubMed Schmidt A, Rein PE, Wöstmann B, Schlenz MA. A comparative clinical study on the transfer accuracy of conventional and digital implant impressions using a new reference key-based method. Clin Oral Implants Res. 2021;32:460–9.CrossRefPubMed
22.
Zurück zum Zitat International Organization for Standardization. Accuracy (trueness and precision) of measurement methods and results d part 1: General principles and definitions. ISO; 1994. International Organization for Standardization. Accuracy (trueness and precision) of measurement methods and results d part 1: General principles and definitions. ISO; 1994.
23.
Zurück zum Zitat Faul F, Erdfelder E, Lang AG, Buchner A. G*Power 3: a flexible statistical power analysis program for the social, behavioral, and biomedical sciences. Behav Res Methods. 2007;39:175–91.CrossRefPubMed Faul F, Erdfelder E, Lang AG, Buchner A. G*Power 3: a flexible statistical power analysis program for the social, behavioral, and biomedical sciences. Behav Res Methods. 2007;39:175–91.CrossRefPubMed
24.
Zurück zum Zitat Ender A, Mehl A. Accuracy of complete-arch dental impressions: a new method of measuring trueness and precision. J Prosthet Dent. 2013;109:121–8.CrossRefPubMed Ender A, Mehl A. Accuracy of complete-arch dental impressions: a new method of measuring trueness and precision. J Prosthet Dent. 2013;109:121–8.CrossRefPubMed
25.
Zurück zum Zitat Knechtle N, Wiedemeier D, Mehl A, Ender A. Accuracy of digital complete-arch, multi-implant scans made in the edentulous jaw with gingival movement simulation: an in vitro study. J Prosthet Dent. 2022;128:468–78.CrossRefPubMed Knechtle N, Wiedemeier D, Mehl A, Ender A. Accuracy of digital complete-arch, multi-implant scans made in the edentulous jaw with gingival movement simulation: an in vitro study. J Prosthet Dent. 2022;128:468–78.CrossRefPubMed
26.
Zurück zum Zitat Hamm J, Berndt EU, Beuer F, Zachriat C. Evaluation of model materials for CAD/CAM in vitro studies. Int J Comput Dent. 2020;23:49–56.PubMed Hamm J, Berndt EU, Beuer F, Zachriat C. Evaluation of model materials for CAD/CAM in vitro studies. Int J Comput Dent. 2020;23:49–56.PubMed
27.
Zurück zum Zitat Mizumoto RM, Alp G, Özcan M, Yilmaz B. The effect of scanning the palate and scan body position on the accuracy of complete-arch implant scans. Clin Implant Dent Relat Res. 2019;21:987–94.CrossRefPubMed Mizumoto RM, Alp G, Özcan M, Yilmaz B. The effect of scanning the palate and scan body position on the accuracy of complete-arch implant scans. Clin Implant Dent Relat Res. 2019;21:987–94.CrossRefPubMed
28.
Zurück zum Zitat Alhossaini SJ, Neena AF, Issa NO, Abouelkheir HM, Gaweesh YY. Accuracy of markerless registration methods of DICOM and STL files used for computerized surgical guides in mandibles with metal restorations: an in vitro study. J Prosthet Dent 2022:S0022-3913(22)00636-9. Alhossaini SJ, Neena AF, Issa NO, Abouelkheir HM, Gaweesh YY. Accuracy of markerless registration methods of DICOM and STL files used for computerized surgical guides in mandibles with metal restorations: an in vitro study. J Prosthet Dent 2022:S0022-3913(22)00636-9.
29.
Zurück zum Zitat Papaspyridakos P, Benic GI, Hogsett VL, White GS, Lal K, Gallucci GO. Accuracy of implant casts generated with splinted and non-splinted impression techniques for edentulous patients: an optical scanning study. Clin Oral Implants Res. 2012;23:676–81.CrossRefPubMed Papaspyridakos P, Benic GI, Hogsett VL, White GS, Lal K, Gallucci GO. Accuracy of implant casts generated with splinted and non-splinted impression techniques for edentulous patients: an optical scanning study. Clin Oral Implants Res. 2012;23:676–81.CrossRefPubMed
30.
Zurück zum Zitat Çakmak G, Yilmaz H, Treviño A, Kökat AM, Yilmaz B. The effect of scanner type and scan body position on the accuracy of complete-arch digital implant scans. Clin Implant Dent Relat Res. 2020;22:533–41.CrossRefPubMed Çakmak G, Yilmaz H, Treviño A, Kökat AM, Yilmaz B. The effect of scanner type and scan body position on the accuracy of complete-arch digital implant scans. Clin Implant Dent Relat Res. 2020;22:533–41.CrossRefPubMed
31.
Zurück zum Zitat Emir F, Ayyıldız S. Evaluation of the trueness and precision of eight extraoral laboratory scanners with a complete-arch model: a three-dimensional analysis. J Prosthodont Res. 2019;63:434–9.CrossRefPubMed Emir F, Ayyıldız S. Evaluation of the trueness and precision of eight extraoral laboratory scanners with a complete-arch model: a three-dimensional analysis. J Prosthodont Res. 2019;63:434–9.CrossRefPubMed
32.
Zurück zum Zitat Keul C, Güth JF. Accuracy of full-arch digital impressions: an in vitro and in vivo comparison. Clin Oral Investig. 2020;24:735–45.CrossRefPubMed Keul C, Güth JF. Accuracy of full-arch digital impressions: an in vitro and in vivo comparison. Clin Oral Investig. 2020;24:735–45.CrossRefPubMed
33.
Zurück zum Zitat Pacquet W, Tapie L, Mawussi B, Boitelle P. Volumetric and dimensional accuracy assessment of CAD-CAM-manufactured dental prostheses from different materials. J Prosthet Dent. 2023;129:150–9.CrossRefPubMed Pacquet W, Tapie L, Mawussi B, Boitelle P. Volumetric and dimensional accuracy assessment of CAD-CAM-manufactured dental prostheses from different materials. J Prosthet Dent. 2023;129:150–9.CrossRefPubMed
34.
Zurück zum Zitat O’Toole S, Osnes C, Bartlett D, Keeling A. Investigation into the accuracy and measurement methods of sequential 3D dental scan alignment. Dent Mater. 2019;35:495–500.CrossRefPubMed O’Toole S, Osnes C, Bartlett D, Keeling A. Investigation into the accuracy and measurement methods of sequential 3D dental scan alignment. Dent Mater. 2019;35:495–500.CrossRefPubMed
35.
Zurück zum Zitat Lerner H, Nagy K, Luongo F, Luongo G, Admakin O, Mangano FG. Tolerances in the production of six different implant scanbodies: a comparative study. Int J Prosthodont. 2021;34:591–9.CrossRefPubMed Lerner H, Nagy K, Luongo F, Luongo G, Admakin O, Mangano FG. Tolerances in the production of six different implant scanbodies: a comparative study. Int J Prosthodont. 2021;34:591–9.CrossRefPubMed
36.
Zurück zum Zitat Mizumoto RM, Yilmaz B, McGlumphy EA Jr, Seidt J, Johnston WM. Accuracy of different digital scanning techniques and scan bodies for complete-arch implant-supported prostheses. J Prosthet Dent. 2020;123:96–104.CrossRefPubMed Mizumoto RM, Yilmaz B, McGlumphy EA Jr, Seidt J, Johnston WM. Accuracy of different digital scanning techniques and scan bodies for complete-arch implant-supported prostheses. J Prosthet Dent. 2020;123:96–104.CrossRefPubMed
37.
Zurück zum Zitat Yilmaz B, Rizzo Marques V, Guo X, Gouveia D, Abou-Ayash S. The effect of scanned area on the accuracy and time of anterior single implant scans: an in vitro study. J Dent. 2021;109:103620.CrossRefPubMed Yilmaz B, Rizzo Marques V, Guo X, Gouveia D, Abou-Ayash S. The effect of scanned area on the accuracy and time of anterior single implant scans: an in vitro study. J Dent. 2021;109:103620.CrossRefPubMed
Metadaten
Titel
Positional transfer accuracy of titanium base implant abutment provided by two different scan body designs: an invitro study
verfasst von
Rania E. Ramadan
Mahmoud Khamis Abdel Razek
Faten S. Mohamed
Rania A. Fahmy
Mervat E. Abd-Ellah
Publikationsdatum
01.12.2023
Verlag
BioMed Central
Erschienen in
BMC Oral Health / Ausgabe 1/2023
Elektronische ISSN: 1472-6831
DOI
https://doi.org/10.1186/s12903-023-03399-9

Weitere Artikel der Ausgabe 1/2023

BMC Oral Health 1/2023 Zur Ausgabe

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Klinik aktuell Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Update Zahnmedizin

Bestellen Sie unseren kostenlosen Newsletter und bleiben Sie gut informiert – ganz bequem per eMail.