First clinical experience with a novel, mobile cone-beam CT system for treatment quality assurance in brachytherapy

verfasst von: Andre Karius, Vratislav Strnad, Michael Lotter, Stephan Kreppner, Christoph Bert

Erschienen in: Strahlentherapie und Onkologie | Ausgabe 6/2022

Abstract

Background and purpose

On-site cone-beam computed tomography (CBCT) has gained in importance in adaptive brachytherapy during recent years. Besides treatment planning, there is increased need particularly for image-guidance during interventional procedures and for image-guided treatment quality assurance (QA). For this purpose, an innovative CBCT device was rolled out at our hospital as the first site worldwide. We present the first clinical images and experiences.

Materials and methods

The novel CBCT system is constructed of a 121 cm diameter ring gantry, and features a 43.2 × 43.2 cm2 flat-panel detector, wireless remote-control via tablet-PC, and battery-powered maneuverability. Within the first months of clinical operation, we performed CBCT-based treatment QA for a total of 26 patients (8 with breast, 16 with cervix, and 2 with vaginal cancer). CBCT scans were analyzed regarding potential movements of implanted applicators in-situ during the brachytherapy course.

Results

With the presented device, treatment QA was feasible for the majority of patients. The CBCT scans of breast patients showed sufficient contrast between implanted catheters and tissue. For gynecologic patients, a distinct visualization of applicators was achieved in general. However, reasonable differentiations of organic soft tissues were not feasible.

Conclusion

The CBCT system allowed basic treatment QA measures for breast and gynecologic patients. For image-guidance during interventional brachytherapy procedures, the current image quality is not adequate. Substantial performance enhancements are required for intraoperative image-guidance.
Hinweise
The data that support the findings of this work are available from the corresponding author upon reasonable request.

Introduction

Computed tomography (CT) is a well-established imaging modality and integral part of the imaging workflow in several medical specialties [19]. In recent years, particularly cone-beam CT (CBCT) has found increasing applications for interventional [1] and intraoperative [2, 3] purposes. One factor for this is that corresponding mobile devices as C‑arms [14] or O‑arms [5] allow working flexible at the patient, even in imaging position.
Potential for CBCT in brachytherapy is undoubted. Due to the modality’s good high-contrast capability, resolution, and geometric accuracy, CBCT allows in general distinct visualizations of inserted applicators [6]. As examples, Peters et al. [7] performed CBCT-based adaptive planning of seed implantations into the prostate. Al-Halabi et al. [8] reported respective brachytherapy planning for cervical cancer. Besides treatment planning, there is increased need also for improvements in treatment quality assurance (QA), both during the surgical procedure and throughout the therapy course. This is crucial for therapy success, since interfractional applicator position changes might have significant effects on the dose delivered to target volumes and organs at risk (OARs) [1014]. Image-guided real-time checks of applicator positions appear essential to account for patient-specific clinical requirements and to ensure accurate applicator arrangements. CBCT represents a suitable modality for the aforementioned imaging tasks.
CBCT scanners for interventional environments usually feature mobile designs [1518]. Concurrently, they show significant image quality limitations compared to conventional CT, especially regarding image artifacts [1518]. The clinical use of respective devices may require more effort than three-dimensional (3D) C‑arms [19], due to larger space requirements. In this respect, a compact device design is considered important for the smooth integration in interventional workflows. Furthermore, to enable patient examinations in the lithotomy position required for intraoperative brachytherapy, a large gantry clearance is crucial. One way to combine these requirements is the usage of vast flat-panel detectors (FPDs). These enable examinations of large anatomical regions of interests (ROIs) with one single gantry rotation and non-moving patient table. In addition to CBCT, radiography and fluoroscopy can be performed with FPDs in general. The combination of these imaging modalities particularly provides the possibility of 2D–3D image registration. This has shown to enable rapid assessments of surgical outcomes such as implant stability [20, 21].
On-site CBCT enables the cross-sectional visualization of the patient anatomy with applicators in situ directly at the brachytherapy ward. This allows both treatment planning as well as intraoperative and interfractional treatment QA measures. For this purpose, we rolled out the innovative CBCT system ImagingRing m (medPhoton, Salzburg, Austria) as the first site worldwide. The clinical introduction started in February 2021 and focused on patients with breast cancer and gynecologic malignancies. We present the first clinical images and experiences.

Materials and methods

ImagingRing m

The ImagingRing m (IRm) features CBCT, radiography, and fluoroscopy as imaging modalities. It is constructed of a 121 cm ring gantry, on which X‑ray source and 43.2 × 43.2 cm2 flat-panel detector (FPD) are mounted (Fig. 1). Source and detector can rotate independently along the gantry, thus, allowing for non-isocentric imaging. The FPD is by default operated with a frame rate of 12 Hz and 300 µm pixel size. For detailed technical specifications of the system, please refer to its physical characterization published previously [22].
Within a so-called volume–definition workflow (VDW), four independently, dynamically collimating jaws allow non-isocentric imaging and user-controlled beam-shaping. In this VDW, anterior–posterior and lateral topograms are taken first. Based on these images, the 3D field of view (FOV) of a CBCT scan can then be specified prior its acquisition. Fully opened, the jaws provide for each projection a maximum isocentric planar FOV of 25.4 × 25.4 cm2. For beam quality adjustments, six different prefilters (air; 1.5 mm Al; 0.2 mm, 0.5 mm, and 1.5 mm Cu; 0.3 mm Cu BowTie) can be inserted into the beam path. Four built-in cameras enable patient monitoring during examinations.
The IRm is operated via WLAN remote-control by means of a portable so-called Human Machine Interface (HMI, Fig. 1) that is established on a Windows®(Microsoft, Redmond, Washington, USA)-based tablet-PC. Motorized wheels at the bottom of each leg (Fig. 1) as well as gearboxes enable lateral and longitudinal movements, free rotations, and ±30° gantry tilt. All movements can be performed battery-powered for up to 30 min. This facilitates changing the premises in which the system is used.
The IRm distinguishes itself from previous CBCT systems [15] particularly by the battery-powered mobility, tablet-PC operation, dynamic collimation, non-isocentric imaging, as well as large gantry and FPD. Prior to the device’s clinical roll-out, we conducted a profound technical characterization of its imaging performance using standard QA phantoms. We have reported our results recently [22] (please refer to this publication for detailed information). Due to the good high-contrast visualization achieved in these studies, the IRm was considered suitable for the detection of brachytherapy applicators.

Workflow for breast/gynecologic brachytherapy

At our institution, interstitial brachytherapy as sole therapy modality for breast cancer is performed as high-dose-rate (HDR) Accelerated Partial Breast Irradiation (APBI) [23, 24]. Cervix and vaginal cancers are treated with intracavitary and interstitial techniques. Treatment is performed either in the HDR or pulsed-dose-rate (PDR) regime, depending on specific entity and case.
For gynecologic tumors, brachytherapy starts with the intrauterine, intravaginal, and/or interstitial insertion of appropriate applicators, which are chosen adapted to the prevailing tumor characteristic. This is performed under general anesthesia and guided by transrectal ultrasound. For breast brachytherapy, the implantation of flexible plastic catheters into the breast tissue is performed under radiography control. Following these steps, the irradiation is planned based on CT scans with applicators in situ. The full treatment usually lasts one week for PDR gynecologic and HDR breast patients, and a few days to weeks for HDR gynecologic patients (depending on the case).
About halfway through the treatment period, we acquire a control-CBCT with the IRm (see scheme in Table 1). Patients at high risk of applicator displacement due to repositioning are excluded. For breast patients, we perform this CBCT check to detect potential catheter arrangement variations in situ and to assess the necessity for treatment adaptions. These proceedings follow the methodologies described in detail by Kallis et al. [25].
Table 1
Time scheme for performing control-CBCT acquisitions
Tumor entity
Treatment modality
Slot for control-CBCT
Breast cancer
HDR, 9 × 3.8 Gy
After 4th fraction
Cervix cancer
HDR, 2 × 7 Gy
Prior to 2nd fraction
PDR, 70 × 0.5 Gy
After 35th fraction ± 4 h
Vaginal cancer
HDR, 6 × 5 Gy
After 3rd fraction
PDR, 70 × 0.5 Gy
After 35th fraction ± 4 h
CBCT cone-beam computed tomography, HDR high-dose-rate, PDR pulsed-dose-rate
For the QA of gynecologic brachytherapy, we analyze the CBCT scans immediately after acquisition in search for potential applicator position changes. Senior physician and medical physicist assess the relative distances between needles and applicators in all three main planes as well as multiplanar reconstructions. Both the distances between the individual applicator and needle tips as well as the in-plane distances between applicators and needles at the symphysis level are measured. The measured distances are then compared to the intended distances, which are documented on the (day 0) planning-CT. Moreover, we visually inspect the exact positions of needle tips and the centered location of the probe within the uterus. For patients who may be subject to swaps in afterloader transfer tubes, X‑ray markers are inserted into the applicators to verify the proper applicator–transfer tube assignment. All these factors ensure that the intended dose distribution can be delivered technically correct to the tissue. If discrepancies are found in any of the above, treatment adaption by re-planning or applicator adjustment is foreseen.

Patient and imaging parameters

Within the first 3 months of the IRm’s clinical operation, we treated 8 breast, 16 cervix, and 2 vaginal cancer patients with the described workflow. All gynecologic patients were treated with PDR brachytherapy. Patient characteristics are listed in Table 2.
Table 2
Characteristics of the examined patients. Furthermore, the contrast–noise ratio (CNR) measured for each patient between catheter and surrounding tissue (for breast patients) and exemplary between uterus and bladder (for gynecological patients) are provided
Breast patients
Gynecologic patients
ID
Age
BMI
No. of catheters
CNR
ID
Age
BMI
Cancer
Applicators
CNR
PB1
48
24
19
7.7
PG1
52
29
Cervix
Fletcher
2.4
PB2
70
29
18
5.9
PG2
36
23
Cervix
Fletcher
0.2
PB3
63
41
20
a
PG3
44
24
Cervix
Fletcher + 6 needles
4.8
PB4
83
24
22
9.2
PG4
40
21
Cervix
Fletcher + 3 needles
2.5
PB5
55
34
19
b
PG5
62
31
Cervix
Fletcher
0.2
PB6
61
31
13
8.4
PG6
78
20
Cervix
Fletcher + 3 needles
2.1
PB7
64
27
18
6.1
PG7
62
23
Vagina
Papillon template + 4 needles
3.2
PB8
63
25
20
7.1
PG8
86
20
Vagina
Prostate template + 7 needles
0.5
PG9
82
23
Cervix
Syed template + 6 needles
0.7
PG10
80
23
Cervix
9 Heymann applicators
2.3
PG11
41
22
Cervix
Fletcher
3.9
PG12
84
36
Cervix
Fletcher + 6 needles
0.5
PG13
55
23
Cervix
Fletcher
1.4
PG14
25
24
Cervix
Fletcher + 2 needles
1.2
PG15
48
23
Cervix
Syed template + 4 needles
0.9
PG16
39
23
Cervix
Fletcher + 2 needles
3.2
PG17
61
20
Cervix
Syed template + 10 needles
3.5
PG18
64
35
Cervix
Fletcher
1.6
Median
63
28
7.7
58
23
2.1
Range
48–83
24–41
5.9–9.2
25–86
20–36
0.2–4.8
BMI body mass index in kg/m2, ID identification
aNo scan, due to technical problems
bNo compliance with breathing commands, blurred catheter paths
The IRm did not feature scan protocols for breast and pelvis imaging sufficient for our clinical requirements nor an automatic exposure control. Therefore, the in-house development of protocols and the manual adaption of protocol parameters (prefiltering, current–time product, etc.) to varying patient characteristics was necessary. These proceedings were based on our experience and particularly on the results of the extensive physical characterization of the device [22].
All examinations were performed using the VDW. Patients were always lying on a carbon fiber table (Fig. 1). Breast patients were scanned head-first supine in breath-hold with an acquisition time of 16–18 seconds. Gynecologic patients were examined feet-first supine with flat-lying/non-bent knees (Fig. 1) and filled bladder. The median acquisition time was 40 seconds (range 19–70 seconds). Starting with the eleventh gynecologic patient (named PG11, Table 2), an abdominal belt was tightly applied and butylscopolaminium bromide was administered 15 minutes before each scan. This aimed to reduce image quality deteriorations which may result from respiration or bowel peristalsis.

Assessment of structure differentiability

To provide a quantitative measure for the delimitability of individual tissues as target volumes and OARs, we calculated the contrast–noise ratio (CNR) for each scan. This is a decisive metric for the distinct differentiation of structures:
$$\mathrm{CNR}_{1,2}=\frac{| \mathrm{ROI}_{1}-\mathrm{ROI}_{2}| }{\sqrt{\frac{1}{2}\cdot \left({\sigma _{1}}^{2}+{\sigma _{2}}^{2}\right)}}$$
(1)
ROI1  and ROI2  denote the mean CT numbers, σ1 and σ2  the CT-number standard deviations of the considered ROIs, respectively. For breast patients, CNR was measured between catheters (ROI1 ) and surrounding tissue (ROI2 ) (Fig. 2). For gynecologic patients, CNR was exemplarily assessed between uterus (ROI1 ) and bladder (ROI2 ). These regions were selected, since they usually show a good CT contrast. The ROIs were in each case placed in the immediate vicinity of bladder catheter and applicator tip, respectively (Fig. 2).

Results

For six breast patients, a control-CBCT to planning-CT registration and reconstruction of catheter tracks was performed according to Kallis et al. [25]. In none of these cases, treatment adaption became necessary. All patients underwent tumor resection prior to brachytherapy and therefore had surgical clips marking the tumor bed. These were clearly identifiable on the CBCT scans (Fig. 3). A reasonable median CNR of 7.7 (range 5.9–9.2; Table 2) between catheters and surrounding tissue was obtained. Hence, the image quality was sufficient for treatment QA in each case. This was achieved at reasonable dose levels, with a median X‑ray exposure of 5.1 mGy (weighted cone-beam dose index CBDIw, range 2.7–8.3 mGy). One breast patient did not comply with the given breathing commands, which led to severe motion artifacts. Another patient could not be examined due to technical problems.
The CBCT scans of gynecologic patients allowed in general distinct visualizations of implanted applicators, due to their high contrast to the tissue (Fig. 4). The courses of vaginal and intrauterine applicators were well identifiable for the majority of patients. In each of these cases, applicator position control could be performed. In none of these cases, relevant positional changes, shifts, or slippage were identified. The deviations of the applicator and needle distances measured on the control-CBCT to the ones documented on the planning-CT are provided in Table 3 for each patient. The applicator–transfer tube assignment was feasible by following the courses of the inserted X‑ray markers (Fig. 4, PG3). However, the exact position verification of implanted needles was not always possible. This was specifically the case when needles and particularly needle tips were in close proximity to sigma or rectum filled with intestinal gas. Here, streaking and shadow artifacts originating from these air–tissue transitions substantially impaired the image quality (Fig. 4). Under these circumstances, the CBCT scans were unsuitable for a complete treatment QA. This was the case for 5 of the 18 examined patients (Table 3).
Table 3
Mean and maximum distance deviations between applicator/needle positions, measured between planning-CT and control-CBCT for gynecologic patients. The distances between the individual applicator/needle tips and the in-plane distances at the symphysis level were considered. We also indicated if the treatment quality assurance (QA) was significantly affected by image artifacts
ID
Mean deviation at symphysis level [mm]
Maximum deviation at symphysis level [mm]
Mean deviation between tips [mm]
Maximum deviation between tips [mm]
Needles/applicators significantly affected
PG1
0.7
1.2
0.9
1.9
PG2
0.9
1.6
2.0
2.7
PG3
0.5
1.5
0.5
0.8
PG4
0.4
0.8
1.2
2.3
2
PG5
0.2
0.6
2.0
2.9
PG6
0.4
0.7
1.3
2.8
1
PG7
0.4
0.7
0.3
0.6
2
PG8
0.7
1.5
1.1
2.9
PG9
0.7
1.7
0.8
1.7
5
PG10
Not measured, Heymann-Packing
Not measured, Heymann-Packing
1.4
2.4
PG11
0.3
0.7
1.5
2.5
PG12
0.9
2.0
2.0
2.7
PG13
0.6
1.1
1.2
2.4
PG14
0.5
0.8
0.7
1.3
1
PG15
0.7
1.2
0.9
1.9
PG16
0.5
1.0
0.8
2.8
PG17
0.6
1.5
1.4
2.9
PG18
1.1
1.3
1.5
2.3
ID identification, PGx Gynecologic patient number x
Although the IRm was principally suitable for validating applicator positions, the device showed distinct weaknesses in the imaging of gynecologic patients. As mentioned, severe artifacts occurred at air–tissue transitions. In addition, directional noise originating from pelvic bones and hardening artifacts in the applicators’ vicinity further impaired image quality (Fig. 4). A clear delineation of organic structures (especially of cervix, uterus, and intestine) as well as the identification of vessels and ureters was not possible in general, due to insufficient contrast differences (Fig. 4). This assessment was quantitatively confirmed by the low CNR (median 2.1, range 0.2–4.8; Table 2) measured exemplarily between uterus and bladder. Thus, despite a high median CBDIw of the examinations of 26 mGy (range 9–50 mGy), individual soft tissues could not be differentiated sufficiently. Note, that the examined patients had a median BMI of 23 (range 20–36; Table 2) and thus represented in general normal weighted patients [26].
Regarding the workflow, the gantry’s large diameter allowed working free and flexible at the patient, e.g., for connecting afterloader transfer tubes. The vast FPD combined with the VDW facilitated the acquisition of FOVs sufficiently dimensioned for applicator arrangement control, as well as exact FOV adjustments even for non-isocentric patient positions. Due to its motorized mobility, the system can be maneuvered time-efficiently into imaging position regardless of the exact patient table position within the room. The total time from patient positioning (including repositioning of immobile patients from bedside) to the examination end was <5 minutes for all patients. In each case, the CBCT scans were available fully reconstructed in <50 seconds after acquisition completion. Hence, the fast assessment of all images could always be performed directly on the HMI.

Discussion

By means of the IRm, we created a time-efficient workflow for the QA of breast and gynecologic brachytherapy. The versatile battery-powered mobility allowed fast maneuvers of the device. The reconstructed images were quickly available, which enabled direct and fast assessments regarding decisions on potential therapy adaptions. In this respect, a mobile CBCT system on-ward strongly facilitates the brachytherapy workflow, since neither patients nor staff have to be sent to more distant CT scanners. This leads particularly to the technical feasibility of performing a treatment QA also for gynecologic patients, which was previously not practicable due to the high effort involved (e.g. transport to distant CT scanners). Such a high effort in particular also precludes the acquisition of magnetic resonance imaging (MRI) scans at off-site locations. On-site MRI for interventional purposes, which would be most desirable due to the superior soft tissue delineability [2729], is logistically still not implementable at many brachytherapy sites at all.
For none of the examined patients, the necessity for therapy adaption was diagnosed. Despite this positive experience, performing a treatment QA is considered important for therapy success. For breast patients, Kallis et al. [25] showed that treatment adaptions are indicated in 4% of all cases. For gynecologic brachytherapy, other authors already found substantial variations of the dose delivered to target volumes and OARs throughout the treatment course [1014]. These were related to both inter- and intrafractional anatomical variations (e.g., organ and patient motion), which may lead to altered applicator positions with significant dosimetric impact [1014]. It is one very important precondition of adaptive brachytherapy to detect such applicator arrangement changes and to adapt the treatment instantaneously to the altered conditions. In this respect, the implementation of automated procedures for treatment plan verification and immediate re-planning (in case of plan invalidity) is most desirable. Hence, this is the subject of ongoing research [10, 30, 31]. In summary, it is important for the precise therapy of patients to perform a treatment QA. CBCT represents a functional and smooth modality for this purpose.
For breast patients, the IRm’s image quality fulfilled clinical requirements at reasonable dose levels. The obtained CNR between catheters and tissue of at median 7.7 was approximately in the range of planning-CTs (median CNR of about 9, measured as described in the “Assessment of structure differentiability” section). Thus, the CBCT scans enabled distinct catheter reconstructions. Treatment QA for breast patients was therefore feasible by using the IRm scans without major drawbacks compared to conventional CT images. However, the current scan times of the IRm of 16–18 seconds are a disadvantage compared to conventional CT (acquisition time in our clinic: <6 seconds). This was also evident from the fact that one examined patient was not able to hold her breath. In particular, based on our experience, not all patients are able to follow correspondingly long breathing commands immediately after surgery. However, this would be required for the acquisition of planning-CTs. To avoid potential motion or breathing artifacts, it is therefore desired to reduce the IRm’s acquisition time. Currently, this associated with enhanced artifacts and decreased circularity of catheters due to partial undersampling. The scan time reduction with undiminished image quality is therefore subject of ongoing work.
The CBCT scans of gynecologic patients were acquired with a median CBDIw of 26 mGy. This exceeds the dose of comparable devices (CBDIw about 20 mGy) [32] and also conventional CT systems (CTDIvol ≤15 mGy) [33, 34]. Despite this increased radiation exposure, distinct differentiations of pelvic organs and tissues were not achievable with the IRm. Target structures and OARs could only be inadequately separated. This was quantitatively confirmed by the CNRs calculated exemplarily between uterus and bladder. With a median CNR of only 2.1, the scans did generally not fulfill Rose’s criterion (Rose’s criterion states that a structure must show a CNR of at least 3 to 5 to be detectable) [35]. This was clear evidence for insufficient low-contrast differences. Please note again that the examined patients represented with a median BMI of 23 in general normal weighted patients, as reported in the “Results” section [26]. The inadequate CNR is therefore not directly attributable to extreme patient characteristics (such as it might be the case for very obese patients) accompanied by strong object scatter. Rather, it indicates a lack of CT-number contrast in the device’s imaging performance. For comparison, planning-CTs yield a respective median CNR of about 6 (again measured as described above). Furthermore, as mentioned, the IRm showed severe artifacts in the imaging of gynecologic patients.
Based on these observations, we conclude that the routine use of the IRm for the purpose of differentiating pelvic soft tissues is currently not justifiable. In particular, the IRm can currently also not support the intraoperative, image-guided insertion of applicators. This is because (due to the lack of organ differentiability) no reliable prediction can be made in which direction and how far applicators and needles can be inserted without risking injuries of corresponding OARs. In this respect, the imaging performance of the IRm is inferior to conventional CT. However, as reported, the device can be used for the verification of the implant stability in-situ and showed to be in principle applicable for the treatment QA of gynecological patients. Although for some patients (Table 3) image artifacts substantially impaired the precise detection of individual needle tips, the IRm enables imaging for treatment verification directly on the ward. This is a procedure that was previously very impractical in our hospital due to the high effort involved, as described above. We think that the advantage of such an on-site treatment verification outweighs the disadvantage of a suboptimal image quality and can have a large benefit for individual patients.
From our perspective, intraoperative CBCT-guided applicator insertion might represent an important step for adaptive gynecologic brachytherapy at many brachytherapy wards. However, to support this by the IRm, significant improvements of its imaging performance are imperative. This relates in particular to improvements in image contrast and soft tissue delineation. Nonetheless, with CBCT it appears feasible to detect the position of applicators precisely, especially around the fundus uteri or deep in the pelvis (in regions for which transrectal sonography is not or only limited applicable). For this, CBCT imaging is appropriate and beneficial. The principle of mobile, non-isocentric imaging also enables patient examinations directly in the surgical theater. This offers a possibility to improve the accuracy of implantations and consequently of treatment quality. A large gantry, as provided by the IRm, further enables imaging in the lithotomy position. This procedure was previously only possible to a limited extent with other gantry-based CBCT-CT/systems. Hence, we think that mobile CBCT imaging with the IRm has in principle high potential for image-guided adaptive brachytherapy of gynecologic malignancies. However, as mentioned, improvements of the device’s imaging performance are imperative.

Conclusion

By means of a mobile device for on-site CBCT imaging, we implemented a time-efficient workflow for the QA of breast and gynecologic brachytherapy. However, significant performance enhancements of the IRm are strictly required for treatment planning imaging and, most desirable, intraoperative image-guidance.

Acknowledgements

The present work was performed by Andre Karius in (partial) fulfillment of the requirements for obtaining the degree “Dr. rer. biol. hum.” at the Friedrich-Alexander Universität Erlangen-Nürnberg.

Conflict of interest

A. Karius, V. Strnad, M. Lotter, S. Kreppner and C. Bert declare that they have no competing interests.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Jetzt e.Med zum Sonderpreis bestellen!

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Jetzt bestellen und 100 € sparen!

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Strahlentherapie und Onkologie

Print-Titel

•Übersichten, Originalien, Kasuistiken

•Kommentierte Literatur aus der Radioonkologie, Strahlenbiologie und -physik

Literatur
1.
Zurück zum Zitat Orth RC, Wallace MJ, Kuo MD (2008) C‑arm cone-beam CT: general principles and technical considerations for use in interventional radiology. J Vasc Interv Radiol 19(6):814–820CrossRef Orth RC, Wallace MJ, Kuo MD (2008) C‑arm cone-beam CT: general principles and technical considerations for use in interventional radiology. J Vasc Interv Radiol 19(6):814–820CrossRef
6.
Zurück zum Zitat Song WY, Tanderup K, Pieters BR (eds) (2017) Emerging technologies in Brachytherapy, 1st edn. CRC press, , pp 233–242CrossRef Song WY, Tanderup K, Pieters BR (eds) (2017) Emerging technologies in Brachytherapy, 1st edn. CRC press, , pp 233–242CrossRef
23.
24.
Zurück zum Zitat Strnad V, Ott OJ, Hildebrandt G et al (2016) 5‑year results of accelerated partial breast irradiation using sole interstitial multicatheter brachytherapy versus whole-breast irradiation with boost after breast-conserving surgery for low-risk invasive and in-situ carcinoma of the female breast: a randomised, phase 3, non-inferiority trial. Lancet 387(10015):223–238. https://doi.org/10.1016/S0140-6736(15)00471-7CrossRef Strnad V, Ott OJ, Hildebrandt G et al (2016) 5‑year results of accelerated partial breast irradiation using sole interstitial multicatheter brachytherapy versus whole-breast irradiation with boost after breast-conserving surgery for low-risk invasive and in-situ carcinoma of the female breast: a randomised, phase 3, non-inferiority trial. Lancet 387(10015):223–238. https://​doi.​org/​10.​1016/​S0140-6736(15)00471-7CrossRef
26.
Zurück zum Zitat WHO (2000) Obesity: preventing and managing the global epidemic. WHO technical report series 894. WHO, Genf WHO (2000) Obesity: preventing and managing the global epidemic. WHO technical report series 894. WHO, Genf
34.
Zurück zum Zitat Herts BR, Baker ME, Obuchowski N et al (2013) Dose reduction for abdominal and pelvic MDCT after change to graduated weight-based protocol for selecting quality reference tube current, peak Kilovoltage, and slice collimation. Med Phys Inform 200(6):1298–1303. https://doi.org/10.2214/AJR.12.9387CrossRef Herts BR, Baker ME, Obuchowski N et al (2013) Dose reduction for abdominal and pelvic MDCT after change to graduated weight-based protocol for selecting quality reference tube current, peak Kilovoltage, and slice collimation. Med Phys Inform 200(6):1298–1303. https://​doi.​org/​10.​2214/​AJR.​12.​9387CrossRef
35.
Zurück zum Zitat Cherry SR, Sorenson JA, Phelps ME (2021) Physics in nuclear medicine. Elsevier, p 244 Cherry SR, Sorenson JA, Phelps ME (2021) Physics in nuclear medicine. Elsevier, p 244
Metadaten
Titel
First clinical experience with a novel, mobile cone-beam CT system for treatment quality assurance in brachytherapy
verfasst von
Andre Karius
Vratislav Strnad
Michael Lotter
Stephan Kreppner
Christoph Bert
Publikationsdatum
12.03.2022
Verlag
Springer Berlin Heidelberg
Erschienen in
Strahlentherapie und Onkologie / Ausgabe 6/2022
Print ISSN: 0179-7158
Elektronische ISSN: 1439-099X
DOI
https://doi.org/10.1007/s00066-022-01912-7

Weitere Artikel der Ausgabe 6/2022

Strahlentherapie und Onkologie 6/2022 Zur Ausgabe

Erhebliches Risiko für Kehlkopfkrebs bei mäßiger Dysplasie

29.05.2024 Larynxkarzinom Nachrichten

Fast ein Viertel der Personen mit mäßig dysplastischen Stimmlippenläsionen entwickelt einen Kehlkopftumor. Solche Personen benötigen daher eine besonders enge ärztliche Überwachung.

15% bedauern gewählte Blasenkrebs-Therapie

29.05.2024 Urothelkarzinom Nachrichten

Ob Patienten und Patientinnen mit neu diagnostiziertem Blasenkrebs ein Jahr später Bedauern über die Therapieentscheidung empfinden, wird einer Studie aus England zufolge von der Radikalität und dem Erfolg des Eingriffs beeinflusst.

Erhöhtes Risiko fürs Herz unter Checkpointhemmer-Therapie

28.05.2024 Nebenwirkungen der Krebstherapie Nachrichten

Kardiotoxische Nebenwirkungen einer Therapie mit Immuncheckpointhemmern mögen selten sein – wenn sie aber auftreten, wird es für Patienten oft lebensgefährlich. Voruntersuchung und Monitoring sind daher obligat.

Costims – das nächste heiße Ding in der Krebstherapie?

28.05.2024 Onkologische Immuntherapie Nachrichten

„Kalte“ Tumoren werden heiß – CD28-kostimulatorische Antikörper sollen dies ermöglichen. Am besten könnten diese in Kombination mit BiTEs und Checkpointhemmern wirken. Erste klinische Studien laufen bereits.

Update Onkologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.