Evaluation of left ventricular and left atrial volumetric function from native MR multislice 4D flow magnitude data

verfasst von: Clemens Reiter, Gert Reiter, Corina Kräuter, Daniel Scherr, Albrecht Schmidt, Michael Fuchsjäger, Ursula Reiter

Erschienen in: European Radiology | Ausgabe 2/2024

Abstract

Objectives

To assess the feasibility, precision, and accuracy of left ventricular (LV) and left atrial (LA) volumetric function evaluation from native magnetic resonance (MR) multislice 4D flow magnitude images.

Materials & Methods

In this prospective study, 60 subjects without signs or symptoms of heart failure underwent 3T native cardiac MR multislice 4D flow and bSSFP-cine realtime imaging. LV and LA volumetric function parameters were evaluated from 4D flow magnitude (4D flow-cine) and bSSFP-cine data using standard software to obtain end-diastolic volume (EDV), end-systolic volume (ESV), ejection-fraction (EF), stroke-volume (SV), LV muscle mass (LVM), LA maximum volume, LA minimum volume, and LA total ejection fraction (LATEF). Stroke volumes derived from both imaging methods were further compared to 4D pulmonary artery flow-derived net forward volumes (NFV). Methods were compared by correlation and Bland-Altman analysis.

Results

Volumetric function parameters from 4D flow-cine and bSSFP-cine showed high to very high correlations (r = 0.83-0.98). SV, LA volumes and LATEF did not differ between methods. LV end-diastolic and end-systolic volumes were slightly underestimated (EDV: –2.9 ± 5.8 mL; ESV: -2.3 ± 3.8 mL), EF was slightly overestimated (EF: 0.9 ± 2.6%), and LV mass was considerably overestimated (LVM: 39.0 ± 11.4 g) by 4D flow-cine imaging. SVs from both methods correlated very highly with NFV (r = 0.91 in both cases) and did not differ from NFV.

Conclusion

Native multislice 4D flow magnitude data allows precise evaluation of LV and LA volumetric parameters; however, apart from SV, LV volumetric parameters demonstrate bias and need to be referred to their respective normal values.

Clinical relevance statement

Volumetric function assessment from native multislice 4D flow magnitude images can be performed with routinely used clinical software, facilitating the application of 4D flow as a one-stop-shop functional cardiac MR exam, providing consistent, simultaneously acquired, volume and flow data.

Key points

Native multislice 4D flow imaging allows evaluation of volumetric left ventricular and atrial function parameters.
Left ventricular and left atrial function parameters derived from native multislice 4D flow data correlate highly with corresponding standard cine-derived parameters.
Multislice 4D flow-derived volumetric stroke volume and net forward volume do not differ.
Hinweise

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
2D
Two-dimensional
3D
Three-dimensional
4D flow
Time-resolved, three-dimensional, three-directional phase contrast imaging
BSA
Body surface area
bSSFP
Balanced steady-state free precession
ECG
Electrographically
ED
End-diastole
EDV
End-diastolic volume
EF
Ejection fraction
ES
End-systole
ESV
End-systolic volume
FLASH
Fast low angle shot
ICC
Intra-class correlation coefficient
LA
Left atrium
LATEF
Total left atrial ejection fraction
LAVmax
Maximal left atrial volume
LAVmin
Minimal left atrial volume
LV
Left ventricle
LVM
Left ventricular myocardial mass
NFV
Net forward volume
r
Pearson correlation coefficient
SD
Standard deviation
SV
Stroke volume

Introduction

Cardiac magnetic resonance (MR) two-dimensional (2D) balanced steady-state free precession (bSSFP) cine imaging has been established as the standard reference technique for the assessment of ventricular systolic function and myocardial mass [13]. Time-resolved, three-dimensional (3D), three-directional MR phase contrast (4D flow) imaging provides – in addition to 3D velocity fields – three-dimensional anatomical (magnitude) cine images. While numerous novel velocity-related cardiac functional 4D flow parameters have been explored [46], the volumetric assessment of cardiac function using 4D flow magnitude images has rarely been investigated [711]. However, acquisition of both volumetric function and flow-based hemodynamic parameters from one measurement would not only allow to simplify and shorten MR imaging protocols but would also enable the comparison of parameters without physiological cycle-to-cycle variations. The latter aspect could significantly improve cross-check evaluation between volumetric and flow parameters, e.g. when applying the conservation-of-mass principle to evaluate mitral valve regurgitation volumes using LV volumetric stroke volume and aortic net flow volume, or when controlling segmentations quality [6, 12, 13].
Compared to bSSFP-cine sequences, 4D flow magnitude images are based on fast low-angle shot (FLASH) readout, which intrinsically has lower blood-to-myocardium contrast [14, 15]. Furthermore, 4D flow techniques are typically implemented as 3D sequences, which suffer from reduced in-flow blood enhancement and consequently lower blood-to-myocardium contrast than 2D-based measurements [6, 14, 16]. To address the latter problem in 3D acquisitions, data for volumetric evaluation from 4D flow measurements have been acquired after application of gadobenate dimeglumine, gadopentetate dimeglumine, or gadoterate meglumine [911], or following administration of ferumoxytol (used off-label) as a contrast agent [7, 8]. This has allowed the extraction of left and right ventricular volumetric function parameters comparable to standard 2D bSSFP-cine-derived measurements [711]. However, because of the controversy regarding the safety of gadolinium and ferumoxytol as contrast agents, their use, when not indicated by the referral diagnosis, is difficult to justify [17, 18].
Using a multislice time-resolved 2D phase contrast sequence with three-directional velocity encoding to acquire multislice 4D flow data, the corresponding native (non-contrast enhanced) magnitude series could potentially be applied for volumetric function evaluation. Therefore, the aim of the present study was to investigate the applicability of native multislice 4D flow magnitude images for the evaluation of left ventricular (LV) and left atrial (LA) volumetric function parameters, and to validate the results by comparison with bSSFP-cine imaging-derived LV and LA volumetric function parameters as well as 4D flow-derived net forward volume.

Materials and methods

Study population

This prospective study was approved by the local ethical review board and complied with the Declaration of Helsinki. All participants provided written informed consent. Sixty-one subjects without signs or symptoms of heart failure were recruited between October 2016 and March 2017 for native cardiac MR imaging. Exclusion criteria were abnormal heart rhythm, known cardiac shunts, and contraindications to MR. One subject did not undergo cardiac MR imaging because of claustrophobia. Therefore, 60 subjects were included in the data analysis. Figure 1 presents the subject flowchart for the study.

Cardiac magnetic resonance imaging

All subjects underwent comprehensive electrographically (ECG)-gated 3T cardiac MR imaging (Magnetom Skyra, Siemens Healthineers) in the supine position using a phased-array 18-channel body matrix and a spine matrix coil. The protocol included free breathing bSSFP-cine realtime and 4D flow imaging.
bSSFP-cine realtime series were acquired in LV 2-chamber, 4-chamber, and 3-chamber views, as well as in a stack of contiguous slices in short-axis orientation covering the LV cavity. To cover all phases of the cardiac cycle, series were acquired for approximately 1.5 heartbeats. Typical protocol parameters were as follows: spatial resolution, 2.3 × 3.9 × 7.0 mm3 for long-axis and 2.5 × 4.2 × 8.0 mm3 for short-axis images; echo/repetition time, 1.1/2.5 ms; flip angle, 40°; parallel acquisition factor, 3; temporal resolution, 36 ms. Cardiac shimming and transmission frequency optimization were employed to minimize bSSFP-related dark band artifacts.
Multislice 4D flow data were acquired in 3-chamber orientation using a retrospectively ECG-gated, FLASH-based 2D phase-contrast sequence with simple three-directional velocity encoding [19, 20] covering the heart. Typical protocol parameters were as follows: spatial resolution, 1.8 x 2.5 x 4 mm3; echo/repetition time, 3.1/5.2 ms; parallel acquisition factor, 2; temporal resolution = 41.8 ms interpolated to 30 cardiac phases; number of averages, 2; 21-39 gapless slices; 36-62 heartbeats per slice; typical scan time, 22 minutes. Velocity encoding between 100 and 140 cm/s was chosen to prevent aliasing in the cardiac chambers. Arrythmia rejection was used to suppress artefacts from ECG miss-triggering or irregular heartbeats.

Image analysis

Preprocessing of 4D flow magnitude data and image quality scoring

4D flow magnitude datasets were multiplanar reconstructed in LV 2-chamber and 4-chamber view cut planes, as well as a stack of 11-14 contiguous short-axis slices (slice thickness, 8 mm) covering the entire LV cavity using standard software (cvi42, Circle Cardiovascular imaging) to obtain 4D flow-cine series for further evaluation (Fig. 2a).
Image quality of end-systolic and end-diastolic 4D flow- and bSSFP-cine images in the stack of cine short-axis as well as in cine long-axis images were evaluated by an experienced reader (C.R., 7 years of experience) using the following 5-point Likert scale based on the visibility of endocardial borders [21, 22]: 5 = excellent (all borders can clearly be delineated), 4 = good (mild artefacts, borders have to be interpolated minimally), 3 = adequate (moderate artefacts, borders have to interpolated over short distances), 2 = fair (significant artefacts, borders have to be interpolated for larger parts of contours), 1 = inadequate (borders cannot be reliably identified).

Volumetric evaluation

LV and LA volumetric function parameters were evaluated from 4D flow-cine and bSSFP-cine series using standard software (cvi42, Circle Cardiovascular imaging). Heart rate was determined as an average throughout the measurements. Whereas on 4D flow-cine images contours were drawn manually, for bSSFP-cine images the segmentations of LV and LA automatically suggested by the software were inspected and corrected where necessary.
LV end-diastole (ED) and end-systole (ES) were each defined visually in a midventricular short-axis slice as the phases with the largest and smallest LV cross-sectional areas, respectively [23, 24]. LV end-diastolic volume (EDV), end-systolic volume (ESV), stroke volume (SV), ejection fraction (EF) and LV myocardial mass (LVM) were evaluated by segmentation of the ED and ES endo- and epicardial short-axis contours. Window/level was adjusted to optimize the blood-to-myocardial contrast. The most basal LV short-axis slice was selected as the one with at least 50% of the LV cavity surrounded by myocardium [3]. Papillary muscles and trabeculations were excluded from the LV blood pool, while the LV outflow tract was included in the LV cavity (Fig. 2b). A discrepancy > 5% between end-diastolic and end-systolic LVM was used to identify incorrect segmentation, which was then addressed by correction of the endo- and/or epicardial borders as needed. The reported LVM is the average of the systolic and diastolic results.
LA minimal volume (LAVmin), LA maximal volume (LAVmax), and the total LA ejection fraction (LATEF) were evaluated using the bi-planar area-length method [25]. Volumes were derived from manual segmentation of the LA cavity in the cardiac phases with the largest and smallest LA cross-sectional areas with a closed mitral valve as visualized in the LV 2- and 4-chamber view series. The mitral valve plane was approximated by a straight line. The LA appendage was included, and pulmonary veins were excluded from the LA cavity (Fig. 2c) [26].
To investigate the intra- and inter-observer variability of volumetric function parameters from 4D flow-cine and bSSFP-cine measurements, the first 10 consecutive female and 10 consecutive male subjects were evaluated twice by one observer and once by a second observer (C.R. and U.R., 7 and 20 years of experience, respectively). Data were analyzed blinded to prior evaluations.

4D flow phase contrast evaluation

4D flow velocity fields were analyzed employing prototype software (4Dflow, Siemens Healthineers). LV and LA velocity vector fields were visually inspected to detect unknown cardiac shunts or significant valve regurgitations. Pulmonary artery net forward volume (NFV) was assessed from the phase-offset-corrected 4D flow velocity field by multiplanar reconstruction of an evaluation plane through the main pulmonary artery above the pulmonary valve. Pulmonary artery cross-sectional area was automatically segmented and manually corrected if necessary (Fig. 2d).

Statistical analysis

Statistical analysis was performed using SPSS® (SPSS Software v28). Distributions of parameters are specified as means and standard deviations (SD) as well as modus in case of image scores; 95% confidence intervals are given in brackets.
Normality of distributions was tested with the Shapiro–Wilk test. As appropriate, differences of non-paired samples were compared by t-test or Mann–Whitney U test, differences of paired samples by paired t-test or Wilcoxon signed-rank test. Pearson correlation (r) and Bland-Altman analysis were employed to study the relationship between continuous parameters from 4D flow and bSSFP-cine measurements. Two-way mixed effect model, single measure, absolute agreement intra-class correlation coefficients (ICC) were used to quantify inter- and intra-observer variability. Correlations were classified according to the correlation coefficients as low (0.3–0.5), moderate (0.5–0.7), high (0.7–0.9), or very high (0.9–1.0) [27].
A value < 0.05 was regarded as statistically significant. For ICCs, non-overlapping confidence intervals were regarded as indicating a significant difference.

Results

The baseline characteristics of the study subjects are summarized in Table 1. None of the subjects analyzed demonstrated a cardiac shunt or significant valve regurgitation in the 4D flow velocity field.
Table 1
Baseline characteristics of the study population. BSA, body surface area; sBP, systolic blood pressure; dBP, diastolic blood pressure
parameter
total
female
male
p
demographic data
subjects (number)
60
35
25
 
age (years)
61 ± 9
60 ± 8
62 ± 10
0.371
size (cm)
171 ± 9
165 ± 7
178 ± 7
< 0.001
weight (kg)
75 ± 14
67 ± 10
85 ± 12
< 0.001
BSA (m2)
1.87 ± 0.21
1.75 ± 0.15
2.05 ± 0.17
< 0.001
sBP (mmHg)
135 ± 17
130 ± 17
143 ± 15
0.003
dBP (mmHg)
76 ± 9
73 ± 10
80 ± 6
0.001
medical history
no prior medical history
23 (38%)
9 (26%)
14 (56%)
 
hypertension
16 (27%)
9 (26%)
7 (28%)
 
hypothyroidism
19 (32%)
17 (49%)
2 (8%)
 
hyperlipidemia
5 (8%)
4 (11%)
1 (4%)
 
osteoporosis
6 (10%)
6 (17%)
  
gout
3 (5%)
1 (3%)
2 (8%)
 
rheumatoid arthritis
2 (3%)
2 (6%)
  

Image quality

All evaluated images received a quality score ≥ 3. Examples with overall excellent image quality scores are shown in Fig. 3a and b. Average image quality scores for 4D flow-cine and bSSFP-cine images are given in Table 2 and demonstrate differences in systolic short-axis as well as systolic 4-chamber view images. Representative examples of systolic 4D flow-cine short-axis and systolic 4-chamber bSSFP-cine images with lower image quality scores are shown in Figures 3c and 3d, respectively.
Table 2
Image quality scores for MSL-4D flow-cine and bSSFP-cine images. Data are reported as mean and standard deviations (SD) together with the modal value given in parentheses. The p value corresponds to the Wilcoxon rank-sum test between 4D flow-cine and bSSFP-cine quality scores. p, significance level; 4ch, 4-chamber view; 2ch, left ventricular 2-chamber view; SA, short-axis view
parameter
4D flow-cine
bSSFP-cine
p
4ch diastolic
4.98 ± 0.13 (5.0)
4.92 ± 0.28 (5.0)
0.104
4ch systole
4.98 ± 0.13 (5.0)
4.92 ± 0.28 (5.0)
0.046
2ch diastolic
4.98 ± 0.13 (5.0)
5.00 ± 0.00 (5.0)
0.325
2ch systolic
4.97 ± 0.18 (5.0)
4.98 ± 0.13 (5.0)
0.325
SA diastolic
4.95 ± 0.29 (5.0)
5.00 ± 0.00 (5.0)
0.161
SA systolic
3.88 ± 0.52 (4.0)
5.00 ± 0.00 (5.0)
< 0.001

Volumetric function parameters

The average heart rate during 4D flow (68 ± 10 min-1) and bSSFP-cine imaging (67 ± 10 min-1) did not differ (p = 0.357). There were high to very high correlations between the volumetric parameters evaluated from 4D flow-cine and bSSFP-cine series. Results for the parameters are given in Table 3 with corresponding scatter plots in Fig. 4 and Bland-Altman plots in Fig. 5. Significant biases were present for all LV volumetric function parameters except the stroke volume; all biases were small (< 5% of the parameter’s mean value) except the ones for LVM. LVM differences between 4D flow-cine and bSSFP-cine evaluation correlated moderately (r = 0.61) with LVM average values, and only LVM (46.2 ± 11.5 g vs. 33.9 ± 8.3 g) showed a significant difference in bias between males and females (p < 0.001).
Table 3
Relationship of left ventricular and atrial volumetric parameters derived from 4D flow-cine and bSSFP-cine images. r is the Pearson-correlation coefficient between 4D flow-cine and bSSFP-cine parameters. In case of all patients p refers to the significance of the bias of a parameter measured with both methods, in case of the gender comparison p refers to the significance of bias differences between female (f) and male (m). EDV, end-diastolic volume; ESV, end-systolic volume; SV, stroke volume; LVM, left ventricular mass; EF; ejection fraction; LAVmax, maximal left atrial volume; LAVmin, minimal left atrial volume; LATEF, total left atrial ejection fraction
parameter
4D flow-cine
bSSFP-cine
r
bias
p
All
EDV (mL)
137.9 ± 28.6
140.8 ± 28.8
0.98
−2.9 ± 5.8
< 0.001
ESV (mL)
46.2 ± 12.7
48.6 ± 13.2
0.96
−2.3 ± 3.8
< 0.001
SV (mL)
91.7 ± 18.4
92.3 ± 18.1
0.95
−0.6 ± 5.8
0.433
LVM (g)
157.8 ± 37.6
118.8 ± 30.6
0.96
39.0 ± 11.4
< 0.001
EF (%)
66.7 ± 4.6
65.8 ± 4.4
0.84
0.9 ± 2.6
0.005
LAVmax (mL)
78.6 ± 17.5
77.8 ± 17.9
0.96
0.8 ± 4.4
0.152
LAVmin (mL)
36.3 ± 11.1
35.6 ± 12.0
0.97
0.7 ± 3.4
0.143
LATEF (%)
54.2 ± 7.1
54.7 ± 8.1
0.83
−0.5 ± 4.6
0.380
Gender specific
EDV (mL)
f
121.4 ± 19.1
125.1 ± 21.0
0.95
−3.7 ± 6.4
0.514
m
161.0 ± 23.1
162.9 ± 23.2
0.98
−1.9 ± 4.8
ESV (mL)
f
39.5 ± 8.9
41.9 ± 9.6
0.93
−2.5 ± 3.6
0.737
m
55.7 ± 11
57.8 ± 12.1
0.94
−2.1 ± 4.0
SV (mL)
f
82.0 ± 13.6
83.2 ± 14.4
0.93
−1.2 ± 5.4
0.334
m
105.3 ± 15.5
105.0 ± 15.0
0.91
0.3 ± 6.4
LVM (g)
f
132.9 ± 20.6
99.0 ± 18.2
0.92
33.9 ± 8.3
0.000
m
192.6 ± 26.6
146.5 ± 21.5
0.91
46.1 ± 11.5
EF (%)
f
67.6 ± 4.9
66.6 ± 4.6
0.85
1.0 ± 2.6
0.653
m
65.5 ± 4.0
64.6 ± 3.9
0.77
0.9 ± 2.6
LAVmax (mL)
f
72.9 ± 16.8
72.2 ± 17.6
0.96
0.7 ± 4.7
0.777
m
86.6 ± 15.4
85.6 ± 15.4
0.96
1.0 ± 4.1
LAVmin (mL)
f
31.6 ± 8.7
30.8 ± 9.5
0.93
0.8 ± 3.5
0.648
mm
42.8 ± 10.9
42.4 ± 12.0
0.96
0.4 ± 3.3
LATEF (%)
f
56.7 ± 5.2
57.7 ± 5.5
0.66
−0.9 ± 4.4
0.406
m
50.7 ± 7.9
50.6 ± 9.5
0.86
0.1 ± 4.8

Observer variability

Inter- and intraobserver agreement of LV volumetric function parameters were excellent and did not differ between 4D flow-cine and bSSFP-cine images (Table 4).
Table 4
Intraclass correlation coefficients (ICCs) and their 95% confidence intervals in brackets for LV and LA volumetric function parameters evaluated from 4D flow-cine and bSSFP-cine series. EDV, end-diastolic volume; ESV, end-systolic volume; SV, stroke volume; LVM, left ventricular mass; EF; ejection fraction; LAVmax, maximal left atrial volume; LAVmin, minimal left atrial volume; LATEF, total left atrial ejection fraction
parameter
Inter-observer variability (n = 20)
Intra-observer variability (n = 20)
4D flow-cine
bSSFP-cine
4D flow-cine
bSSFP-cine
EDV (mL)
0.99
1.00
0.97
0.99
[0.966–0.994]
[0.988–0.998]
[0.930–0.988]
[0.978–0.998]
ESV (mL)
0.98
0.98
0.97
0.99
[0.950–0.992]
[0.957–0.993]
[0.926–0.990]
[0.985–0.998]
SV (mL)
0.96
0.99
0.96
0.98
[0.913–0.986]
[0.969–0.995]
[0.900–0.984]
[0.946–0.994]
LVM (g)
0.99
1.00
1.00
0.99
[0.984–0.997]
[0.989–0.998]
[0.989–0.998]
[0.977–0.996]
EF (%)
0.87
0.87
0.91
0.89
[0.708–0.948]
[0.704–0.946]
[0.792–0.963]
[0.738–0.955]
LAVmax (mL)
1.00
0.98
1.00
1.00
[0.985–0.998]
[0.959–0.993]
[0.990–0.999]
[0.989–0.998]
LAVmin (mL)
1.00
0.99
1.00
1.00
[0.988–0.998]
[0.980–0.997]
[0.992–0.999]
[0.988–0.999]
LAVTEF (%)
0.97
0.90
0.97
0.95
[0.936–0.989]
[0.763–0.959]
[0.936–0.989]
[0.882–0.981]

Validation of stroke volumes

Net forward volumes derived from 4D flow imaging correlated very highly with volumetric stroke volumes derived from 4D flow-cine and bSSFP-cine imaging (r = 0.91 in both cases). There was no bias between net forward volume and volumetric stroke volumes (p = 0.218 in case of 4D flow-cine imaging, p = 0.058 in case of bSSFP-cine imaging). Corresponding Bland Altman plots are shown in Fig. 6.

Discussion

The main findings of the present study are as follows: 1) LV and LA volumetric function parameters can be evaluated from multiplanar reformatted native multislice 4D flow magnitude images using standard software for cardiac evaluation; 2) all LV and LA volumetric function parameters derived from 4D flow-cine imaging showed high to very high correlations with corresponding bSSFP-cine-derived metrics; 3) LV SV, LA volumes as well as LATEF did not differ between 4D flow-cine imaging and bSSFP-cine imaging; and 4) SVs derived from both of the latter techniques did not differ from NFV derived from 4D flow.
While the assessment of ventricular volumetric function and myocardial mass from contrast-enhanced 4D flow magnitude data has been reported, the assessment of LA volumes and function from native multislice 4D flow magnitude imaging has not yet been analyzed. Image quality of 4D flow-cine images was sufficient to evaluate LV and LA volumetric parameters with excellent inter- and intra-observer variability, comparable to bSSFP-cine images.
Comparison of LV volumetric function parameters derived from multislice 4D flow magnitude and bSSFP-cine imaging showed strong correlations, similar to those found in studies evaluating LV volumetric function parameters from contrast-enhanced 4D flow magnitude data [711]. In accordance with those studies, LV SV did not differ between techniques; however, unlike studies employing contrast-enhanced 4D flow, the current study indicates that when using native multislice 4D flow, LV volumes were slightly underestimated and LV EF was slightly overestimated, while LVM was considerably overestimated. It can be speculated that the observed biases are most likely due to FLASH- based 4D flow magnitude data. Overestimation of LVM by up to 19% based on native 2D cine FLASH as compared to bSSFP-cine imaging was previously observed and was attributed to the lower contrast in FLASH images, which was thought to impair differentiation of endocardial voxels containing blood or epicardial voxels containing fat from myocardium [15, 28, 29]. The larger bias in our study might have been caused by the in-plane resolution of the reformatted magnitude cine images being lower than that of the 2D cine FLASH images used in the above-mentioned studies. This aspect might be less relevant for contrast-enhanced data.
In the present study LA volumes as well as LATEF were highly correlated between the two techniques and did not differ significantly. Unlike the LV myocardial wall, the LA wall is smooth and allows clear and reproducible delineation on reformatted 4D flow series at least comparable to that possible on bSSFP-cine series. Biplanar area-length assessment of LA volumetric function especially benefits from retrospective reconstruction, enabling optimized LV 2-chamber and 4-chamber view cut planes [30]. In the present study, angulation of bSSFP-cine long-axis images was thoroughly inspected after acquisition and optimized if necessary—a step that might be overlooked in routine clinical cardiac MR investigations.
As our study cohort was free from cardiac shunts or significant valve regurgitations, mass conservation could be employed for validation of stroke volumes [6, 31]. Both volumetric stroke volumes demonstrated very high correlation to 4D flow-derived NFV with no bias, which can be interpreted as a reciprocal validation between free-breathing bSSFP-cine volumetry and 4D flow velocity data as well as an additional validation of the 4D flow-cine volumetry. Interestingly, there was no difference in the correlation between bSSFP-cine and 4D flow SV with the 4D flow-derived NFV, although the 4D flow results were derived from the same measurement and a poorer correlation between bSSFP-cine SV and 4D flow NFV could be expected due to their sequential measurement. The fact that there was no difference in heart rate between measurements (and also no change in respiratory state) could be interpreted as indication that there was no substantial difference in the physiological state as commonly observed between different sequences due to patient adaption to the scanner environment, discomfort, nervousness or the application of contrast agent between measurements [13].
Some limitations of the present study must be addressed. The investigated population consisted of individuals without symptoms of cardiovascular disease, and image quality and/or applicability of mass conservation could differ for individuals with arrythmia. Similarly, patients without arrhythmia might demonstrate different results, although comparisons between cines with bSSFP- and FLASH readout suggest that the current volumetric results can be transferred to these situations [32]. The velocity encoding of the 4D flow measurement was optimized for intracardiac blood flow, such that the aorta was not free of aliasing in most cases and the pulmonary artery NFV had to be used for mass conservation analysis. However, this has the particular advantage of including coronary blood flow [33] and should therefore yield measurements even better comparable to volumetric stroke volumes. Notability the choice of velocity encoding does not have a direct effect on the 4D flow magnitude images [20] such that the observed volumetric results should not change when choosing a higher velocity encoding. Respiratory gating was not available for the employed multislice 4D flow sequence. Therefore, two-fold averaging was used to compensate for the breathing motion, which might have affected image quality. Image quality was, however, adequate, and moreover, 4D flow without respiratory gating provides higher signal-to-noise ratio on magnitude images and has been shown to be accurate [6]. The free-breathing bSSFP-cine realtime imaging that was used as the standard of reference typically has lower spatial resolution than the standard 2D-segmented bSSFP-cine technique. However, breath-holding might have impacted volumetric function parameters [34], guidelines recommend cine realtime imaging for LV functional assessment in patients who cannot hold their breath [3, 35], and realtime functional parameters have been validated against the 2D segmented approach [36].
In conclusion, native multislice 4D flow magnitude data allows precise evaluation of LV and LA volumetric parameters; however, apart from SV, LV volumetric parameters demonstrate bias and need to be referred to their respective normal values. The evaluation can be performed with standard software and may allow native multislice 4D flow to be applied as a one-stop-shop method of functional cardiac MR imaging, providing consistent, simultaneously acquired volume and flow data.

Acknowledgements

The authors thank Ada Muellner, MS, for editing the manuscript.

Declarations

Guarantor

The scientific guarantor of this publication is Ursula Reiter.

Conflict of interest

The authors of this manuscript declare relationships with the following companies:
1. Gert Reiter is an employee of Siemens Healthcare Diagnostics GmbH, Austria. The study was performed under a Master Research Agreement between the Medical University of Graz, Graz University of Technology, and Siemens Healthcare Diagnostics GmbH.
2. Ursula Reiter is a member of the editorial board of European Radiology.
The other authors of this manuscript declare no relationships with any companies whose products or services may be related to the subject matter of the article.

Statistics and biometry

One of the authors (Gert Reiter) has significant statistical expertise. No complex statistical methods were necessary for this paper.
Written informed consent was obtained from all patients in this study.

Ethical approval

Institutional Review Board approval was obtained.

Ethical committee

Medical University of Graz, Austria.
ClinicalTrials.gov: NCT01728597

Study subjects or cohorts overlap

4D flow data of 10 subjects of the study population were included in the definition of an algorithm for automated mitral valve vortex ring extraction from 4D-flow MRI (Kräuter C et al., MRM 84 (2020). https://​doi.​org/​10.​1002/​mrm.​28361). The impact of breath-holding on cine realtime volumetric function parameters was analyzed previously from 56 subjects of this study population (Reiter C et al., Eur J Rad 141 (2021). https://​doi.​org/​10.​1016/​j.​ejrad.​2021.​109756).

Methodology

• prospective study
• performed at one institution
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Jetzt e.Med zum Sonderpreis bestellen!

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Jetzt bestellen und 100 € sparen!

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Literatur
9.
Zurück zum Zitat Vial J, Bouzerar R, Pichois R et al (2020) MRI Assessment of Right Ventricular Volumes and Function in Patients With Repaired Tetralogy of Fallot Using kat-ARC Accelerated Sequences. AJR Am J Roentgenol 215:807–817. https://doi.org/10.2214/AJR.19.22726 Vial J, Bouzerar R, Pichois R et al (2020) MRI Assessment of Right Ventricular Volumes and Function in Patients With Repaired Tetralogy of Fallot Using kat-ARC Accelerated Sequences. AJR Am J Roentgenol 215:807–817. https://​doi.​org/​10.​2214/​AJR.​19.​22726
10.
Zurück zum Zitat Hsiao A, Lustig M, Alley MT et al (2012) Rapid Pediatric Cardiac Assessment of Flow and Ventricular Volume With Compressed Sensing Parallel Imaging Volumetric Cine Phase-Contrast MRI. AJR Am J Roentgenol 198:W250–W259. https://doi.org/10.2214/AJR.11.6969 Hsiao A, Lustig M, Alley MT et al (2012) Rapid Pediatric Cardiac Assessment of Flow and Ventricular Volume With Compressed Sensing Parallel Imaging Volumetric Cine Phase-Contrast MRI. AJR Am J Roentgenol 198:W250–W259. https://​doi.​org/​10.​2214/​AJR.​11.​6969
27.
Zurück zum Zitat Mukaka MM (2012) Statistics corner: A guide to appropriate use of correlation coefficient in medical research. Malawi Med J 24:69–71 Mukaka MM (2012) Statistics corner: A guide to appropriate use of correlation coefficient in medical research. Malawi Med J 24:69–71
30.
Metadaten
Titel
Evaluation of left ventricular and left atrial volumetric function from native MR multislice 4D flow magnitude data
verfasst von
Clemens Reiter
Gert Reiter
Corina Kräuter
Daniel Scherr
Albrecht Schmidt
Michael Fuchsjäger
Ursula Reiter
Publikationsdatum
15.08.2023
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 2/2024
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-023-10017-3

Weitere Artikel der Ausgabe 2/2024

European Radiology 2/2024 Zur Ausgabe

Mammakarzinom: Brustdichte beeinflusst rezidivfreies Überleben

26.05.2024 Mammakarzinom Nachrichten

Frauen, die zum Zeitpunkt der Brustkrebsdiagnose eine hohe mammografische Brustdichte aufweisen, haben ein erhöhtes Risiko für ein baldiges Rezidiv, legen neue Daten nahe.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.