Breast elastography—ready for prime time?

verfasst von: André Pfob, Michael Golatta

Erschienen in: European Radiology | Ausgabe 2/2024

Hinweise
This comment refers to the article available at https://​doi.​org/​10.​1007/​s00330-023-10057-9.

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
The imaging community has evaluated the use of breast elastography for over 20 years. Elastography is an imaging technique that assesses the stiffness of a lesion and builds on the observation that, generally, carcinomas are stiffer than benign tissue [1]. However, breast elastography has not yet been widely accepted by the breast imaging community although the potential has been shown in several studies.
The non-adoption of breast elastography is largely associated with differences in methodology between vendors, variations in technique, and unclear clinical implications. Two general techniques in breast elastography can be distinguished: shear-wave elastography, a quantitative method to measure the lesion stiffness via the propagation of shear-waves, and strain elastography, a relative method that compares the lesion stiffness with the surrounding tissue [2]. Both shear-wave and strain elastography can be interpreted in different ways. For strain elastography, the elastography to B-mode length ratio (E/B ratio) has been established to be the most accurate interpretation method [3]. For shear-wave elastography, however, the appropriate interpretation technique with regard to the optimal cutoff is yet to be defined.
We observed a similar process with the introduction of 2D shear-wave elastography for liver stiffness evaluation in chronic liver diseases, where difficulties were finally overcome by a combination of standardizations between vendors and large prospective trials to drive clinical implications.
In this issue of European Radiology, Xu et al report the results of a large prospective, multicenter trial that evaluates the diagnostic performance of shear-wave elastography in combination with standard breast ultrasound for breast cancer diagnosis [4]. A total of 897 patients with BI-RADS 3 to 5 breast masses underwent both B-mode breast ultrasound and shear-wave elastography; histopathologic evaluation was conducted in those with BI-RADS 4 or 5 masses and a 2-year follow-up for those with BI-RADS 3 masses (upon standard B-mode breast ultrasound). The additional use of shear-wave elastography was evaluated by hypothetically reclassifying participants into the respective BI-RADS categories. The results confirm that a large proportion of patients, 46%, receives an (ultimately) unnecessary biopsy based on B-mode ultrasound as histopathology turns out to be benign. (Hypothetically) adding 2D + 3D shear-wave elastography reduces the proportion of benign biopsies compared to standard breast ultrasound by 54%. Specifically, BI-RADS 4a masses were downgraded to BI-RADS 3 with a cutoff of 90 kPa (= 5.5 m/s or less) and BI-RADS 3 masses were upgraded to BI-RADS 4a with a cutoff of 120 kPa (= 6.3 m/s or more); positive-predicative values (PPV) were 53.9% for B-mode breast ultrasound and 71.4% for the combination with shear-wave elastography. The proportion of missed malignancies was kept below the 2% threshold of the BI-RADS 3 category.
These results are in line with two other prospective multicenter trials. The multicenter BE1 study by Berg et al suggested cutoff values for downgrading BI-RADS 4a lesions at 80 kPa (= 5.2 m/s or less) and for upgrading BI-RADS 3 at 160 kPa (= 7.3 m/s or more); an improvement in specificity with the same sensitivity was observed (PPV 52.6% vs. 65.7%) [5]. A later prospective, multicenter, international study by Golatta et al could not confirm the exploratory cutoff values of previous studies, since the rate of false-positive findings was reduced, but at the expense of an increased rate of missed cancers. Secondary analyses suggest to downgrade BI-RADS 4a lesions at a cutoff of 2.55 m/s or less (= 19.5 kPa or less) could result in a 24% reduction of false positives (and therefore unnecessary biopsies) while maintaining cancer detection rates consistent with current guidelines [6]. In contrast to these two previous studies, the study by Xu et al is the first prospective multicenter study to use 3D shear-wave elastography. While the reported diagnostic performance of 3D shear-wave elastography is descriptively higher compared to 2D shear-wave elastography, measurement reliability is (currently) slightly lower. Although the widespread adoption of 3D elastography may improve its reliability, it remains unclear whether to recommend 2D or 3D systems. Nonetheless, the study by Xu et al adds to the growing evidence that breast elastography improves conventional breast ultrasound. However, two issues remain:
(1)
To date, no trial could prospectively confirm the cutoffs to up- or downgrade suspicious breast masses which are solely based on exploratory analyses conducted in retrospect (and which are remarkably different among the three large trials mentioned above). Change of practice is difficult to be based on hypothetical re-classifications.
 
(2)
Variations in vendors and interpretation techniques continue to represent a challenge for users and scientific societies. Further standardization is needed in order to recommend breast elastography on a large scale.
 
Novel research in the field of breast elastography is ongoing: Recently, an improved software has been evaluated that showed a drastic increase in diagnostic performance in a prospective single-center study [7]. Other studies by Pfob et al are evaluating the use of artificial intelligence in breast elastography with promising results [8]. Although breast elastography may not (yet) be ready for prime time, we need high-quality studies in line with previously published ones [46] to drive innovation and research within the imaging community and to ultimately improve clinical care.

Declarations

Guarantor

The scientific guarantor of this publication is André Pfob.

Conflict of interest

The authors of this manuscript declare to have received institutional funding by Siemens Medical Solutions USA, Inc. The funding body played no role in the writing of the commentary.

Statistics and biometry

No complex statistical methods were necessary for this paper.
Written informed consent was not required.

Ethical approval

Institutional Review Board approval was not required.

Study subjects or cohorts overlap

Not applicable.

Methodology

• commentary
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​.

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Jetzt e.Med zum Sonderpreis bestellen!

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Jetzt bestellen und 100 € sparen!

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Literatur
Metadaten
Titel
Breast elastography—ready for prime time?
verfasst von
André Pfob
Michael Golatta
Publikationsdatum
26.10.2023
Verlag
Springer Berlin Heidelberg
Erschienen in
European Radiology / Ausgabe 2/2024
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-023-10329-4

Weitere Artikel der Ausgabe 2/2024

European Radiology 2/2024 Zur Ausgabe

Mammakarzinom: Brustdichte beeinflusst rezidivfreies Überleben

26.05.2024 Mammakarzinom Nachrichten

Frauen, die zum Zeitpunkt der Brustkrebsdiagnose eine hohe mammografische Brustdichte aufweisen, haben ein erhöhtes Risiko für ein baldiges Rezidiv, legen neue Daten nahe.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.