Anatomical and imaging measurements of the angle between the axis of the lumbar pedicle and lateral isthmus margin and its clinical significance

verfasst von: Shuiquan Wang, Dilimulati Aikeremu, Alafate Kahaer, Abulikemu Maimaiti, Yang Xiao, Abudusalamu Tuoheti, Rui Zhang, Xieraili Maimaiti, Hailong Guo, Paerhati Rexiti

Erschienen in: Journal of Orthopaedic Surgery and Research | Ausgabe 1/2023

Abstract

Background

This study aims to explore the measurement of the angle between the axis of the pedicle and the lateral margin of the isthmus on the lumbar spine, and investigate its clinical significance.

Methods

The angle was measured on 120 normal adults’ X-ray and 25 dry anatomical specimens. 60 screws were placed by junior residents on 6 wet specimens through the freehand technique. 30 screws were placed on one side with their original experience. After learning the techniques mentioned in the study, 30 screws were placed on the other side. The specimens were examined by X-ray and CT, and the angles of the screw paths and the integrity of the pedicle were evaluated.

Results

The angles of 120 subjects and 25 anatomical specimens show a gradually increasing trend. The differences among each segment were statistically significant (P < 0.05), but the difference in the same segment between the X-ray and the anatomical specimens was not statistically significant (P > 0.05). Furthermore, the differences in L1, L2, and L3 between the two genders were not statistically significant (P > 0.05). However, the angles were larger in female group than in male group in L4 and L5, and the differences were statistically significant (P < 0.05). The difference in the deviation rate of screw placement before and after the learning was statistically significant only in the L5 segment (P < 0.05). The difference in overall excellence rate was statistically significant (P < 0.05).

Conclusions

The measurement of the angle between the axis of the pedicle and the lateral margin of the isthmus on the lumbar can improve the accuracy of the lumbar sagittal screw angle.
Hinweise
Shuiquan Wang and Dilimulati Aikeremu equally contributed to this work.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
SSA
Sagittal screw angle
PACS
Picture archiving and communication system

Background

The error of the sagittal screw angle (SSA) of the pedicle is one of the risk factors. To reduce complications, repeated fluoroscopy is often used to improve the accuracy. How to reduce intraoperative radiation and improve safety through the inherent anatomic markers of the spine has been the goal of orthopedists [14]. At present, Xiao et al. [5] reported that the SSA could be determined through the included angle between the lateral margin of the lumbar lamina and the axis of pedicle. However, there is no related report on the determination of SSA through the angle between the isthmus and the axis of the pedicle. The author considers it can be used as a reliable reference for spine surgery.

Materials and methods

Acquisition of 3D reconstruction data

Source of the subjects

A total of 362 anterior–posterior and lateral X-rays of normal lumbar vertebrae were collected from the Imaging Center of the First Affiliated Hospital of Xinjiang Medical University. Twenty-five dry specimens of the lumbar vertebrae with the complete structure were selected (Provided by the Anatomy Teaching and Research Office of Xinjiang Medical University). Six wet specimens of the lumbar with a complete structure (3 men and 3 women).

Inclusion and exclusion criteria

Inclusion criteria: (1) The X-ray with good quality and the structures were displayed apparently; (2) Within the age of 20–50 years; (3) Gender equality. Exclusion criteria: (1) The X-rays with inferior quality; (2) Lumbar sacralization or lumbarization. 120 cases (Male: 60, female: 60) were selected according to the inclusion criteria. Each gender group was divided into 5 subgroups, according to five different segments of the lumbar vertebrae.

Experimental instruments and materials

1.
A protractor (accuracy of 1°, estimate of 0.1°).
 
2.
A medical electric drill, and 1-mm, 1.5 mm, and 2-mm Kirschner wires.
 
3.
Titanium alloy pedicle screws with a diameter of 5.5–6.0 mm, with lengths of 40 and 45 mm (produced by Zhengtian Instrument Co., Ltd., Tianjin, China).
 
4.
Imaging instruments: Hitachi 500 mA DR X-ray machine, Japan; Siemens computed tomography (CT) system, Germany; PACS image software system.
 

Imaging data measurement

It is required that the bilateral pedicles of the pedicle and the upper and lower endplates are completely overlapped, without bilateral shadows (Fig. 1a).
1.
Confirmation of the axis of the pedicle: The middle point of the narrowest section of the pedicle (short yellow arrow) and the middle point of the junction between the pedicle and the vertebral body (short red arrow) were first determined along the pedicle. The axis is the connecting line between the two midpoints shown above (the mid-long yellow arrow).
 
2.
Confirmation of the lateral margin of the isthmus: Because only the surface of the isthmus can be seen during the operation, the center axis of the isthmus cannot match the actual demand. The lateral margin of the isthmus was determined on the surface of the lamina (the line between the lower margin of the superior articular process and the upper margin of the inferior articular process on the same segment). To facilitate the determination of the upper margin of the inferior articular process, the upper margin of the superior articular process of the caudal vertebral body was used as a reference.
 
3.
Angle measurement: The angle was measured using the picture archiving and communication system (PACS).
 

Measurement of the anatomical specimen

Pre-experiment

Two 2.0 mm fine Kirschner wires were inserted into the axis of the pedicle and isthmus through the Ebraheim method [6]. Then, the lateral X-ray was performed, and the angle between two Kirschner wires was measured through the PACS.
In addition, the coordinate projection method was used to measure the same angle between the two Kirschner wires in the specimen. The specimen was placed on a plastic foam board with the lateral side perpendicular to the board. Then, 1.0-mm Kirschner wires were vertically inserted into the foam plate at the upper, lower, left, and right sites along the course of the two Kirschner wires in the specimen. Angles formed by connecting the corresponding holes on the board were measured using protractor after removing the specimen..
The angles between the two different methods mentioned above were compared (Fig. 1b–d). It was found that there was no difference between the results. Subsequent specimens were not measured using the Ebraheim method, but the coordinate projection method was used.

Confirmation of the axis of the pedicle

The base of the pedicle is almost vertical to the corresponding vertebral body [7, 8]. The midpoint of the narrowest part of the pedicle was marked. Then, the midpoint of the upper and lower margin of the junction between the pedicle and the posterior wall of the vertebral body was determined, and these two points were connected into a straight line. After the line was drawn, it was confirmed whether the line was vertical to the posterior wall of the vertebral body. The electrocardiogram (ECG) coordinate paper was attached to the foam plastic board, and the specimen was placed on it laterally. The researcher held one Kirschner wire in hand, and placed it over the pedicle and parallel to the surface of the board, instead of inserting these into the bone using an electric drill introduced by Ebraheim [6]. These two marked midpoints of the pedicle and the connecting line were observed, and the Kirschner wires on the hand were finely adjusted to keep parallel to the connecting line and perpendicular to the posterior wall of the vertebral body. The researcher held the Kirschner wire and maintained it still, while the assistant inserted two 1.0-mm fine Kirschner wires along the direction of the Kirschner wire into the plastic foam board covered in coordinate paper and fixed on the same side of the Kirschner that the researcher was holding. With the aid of the small lattices on the coordinate paper, the projection of the axis of the pedicle could be accurately marked on the coordinate paper.Confirmation of the lateral margin of the isthmus was using the same coordinate projection method mentioned above to confirm.

Angle measurement

Four small holes and one big hole stamped by the transverse process were left on the coordinate paper. The big hole brang difficulties in confirming the location of the intersection point. To mark the cross point of these two lines, a piece of transparent plastic film was used to cover the coordinate paper. Then, the small holes left by the Kirschner wires were connected in pairs using a pen to form the angles on the plastic film. The left and right angles of the specimen were recorded respectively, and the mean value was calculated.

Screw placement by residents

The angle of each segment of the six wet specimens was measured by X-ray before placement, which was used to calculate the deviation rate at last. Operators were randomly selected from the junior resident in the department of spine surgery, who subsequently placed screws without the aid of an X-ray. First, the 30 screws were placed on one side based on their original free-hand experience, and then 30 on the other side referring to the author’s new technique (based on the average value on each segment was calculated on the 120 X-rays and 25 dry specimens) (Fig. 2). Nevertheless, the residents did not know exactly the angles used to calculate the deviation rate. The specimens were examined by X-ray and CT.
The angle between the center line of the screw and the anatomic axis of the pedicle was defined as the screw angle (SA) (Fig. 3). These were evaluated according to the screw placement criteria proposed by Tiansi Tang 9:(1) Excellent, the screw is within the pedicle, SA is ≤ 5°;(2) Good, within the pedicle, SA is within 5 and 10°; (3) Acceptable, SA is ≥ 11°; (4) Poor, beyond the pedicle cortex or the vertebral body, and enters into the intervertebral space.
In addition, the ratio of the screw angle to the included angle measured by X-ray was defined as the deviation rate: Screw angle/Included angle × 100%.

Statistical analysis

All data were analyzed using SPSS 18.0. A T test was used to compare the difference between specimen and imaging data. The excellence rate of the screw before and after the learning was compared using the Chi-square test. The inspection level was set at α = 0.05.

Results

Imaging and anatomic specimen measurement results

The results of 120 patients X-ray and 25 dry specimens were demonstrated in Table 1.
Table 1
Measurement of X-ray angles and anatomic specimen angles and comparison in different gender group
Vertebral sequence
X-ray
Anatomic Angle (χ̅ ± s, °)
P1
t
P2
Male
Female
Mean
L1
92.5 ± 1.0
92.7 ± 1.1
92.6 ± 1.1
93.0 ± 1.1
0.523
3.334
0.070
L2
93.7 ± 1.6
93.8 ± 1.6
93.7 ± 1.5
94.1 ± 1.1
0.870
1.439
0.232
L3
97.0 ± 1.8
97.4 ± 1.7
97.2 ± 1.8
97.0 ± 1.8
0.290
0.257
0.613
L4
99.3 ± 1.9
100.2 ± 2.5
99.8 ± 2.2
99.6 ± 1.7
0.017*
0.146
0.703
L5
102.7 ± 0.5
105.0 ± 0.7
103.9 ± 1.3
104.0 ± 1.9
0.000*
0.24
0.625
P1: Statistical comparison of imaging angles between male and female subjects
P2: Statistical comparison of results between imaging and anatomic measurements
*P < 0.05

Results of screw placement by residents

It revealed that 57 screws were placed within the pedicle. One screw was placed in the intervertebral space, one screw was placed out of the pedicle, and one screw perforated the inner wall of the pedicle. Furthermore, all three failed screws were found in the control group (before learning).
According to the evaluation criteria of the SSA proposed by Tang et al. [9], among the 30 screws in the control group, 17 screws were excellent, 9 screws were good, 1 screw was acceptable, and 3 screws were poor. The excellent rate was 57.08%, and the mean deviation rate was 5.17% (Table 2).
Table 2
Comparison of screw placement before and after the learning of resident doctors
Angle of L1 (°)
Control group
Experimental group
Screw angle (°)
Classification
Deviation rate (%)
Screw angle (°)
Classification
Deviation rate (%)
L1
 92
3
Excellent
3.26 (3/92)
2
Excellent
2.17 (2/92)
 92.5
1
Excellent
1.08 (1/92.5)
1
Excellent
1.08 (1/92.5)
 93.5
6
Good
6.42 (6/93.5)
2
Excellent
2.14 (2/93.5)
 93
2
Excellent
2.15 (2/93)
1
Excellent
1.08 (1/93)
 92.3
2
Excellent
2.17 (2/92.3)
0
Excellent
0
 91.8
0
Excellent
0
0
Excellent
0
 
Excellent rate
83.30%
Mean 2.51%
Excellent rate
100%
Mean 1.08%
L2
 93
2
Excellent
2.15 (2/93)
2
Excellent
2.15 (2/93)
 94
6
Good
6.38 (6/94)
1
Excellent
1.06 (1/94)
 93
3
Excellent
3.23 (3/93)
0
Excellent
0
 94.3
4
Excellent
4.24 (4/94.3)
2
Excellent
2.12 (2/94.3)
 93.5
1
Excellent
1.07 (1/93.5)
1
Excellent
1.07 (1/93.5)
 94.2
0
Excellent
0
2
Excellent
2.12 (2/94.2)
 
Excellent rate
67.70%
Mean 2.85%
Excellent rate
100%
Mean 1.42%
L3
 96.5
7
Good
7.25 (7/96.5)
2
Excellent
2.07 (2/96.5)
 97.5
4
Excellent
4.1 (4/97.5)
0
Excellent
0
 97
2
Excellent
2.06 (2/97)
3
Excellent
3.09 (3/97)
 97.2
6
Good
6.17 (6/97.2)
5
Excellent
5.14 (5/97.2)
 96.4
2
Excellent
2.07 (2/96.4)
2
Excellent
2.07 (2/96.4)
 97.9
2
Excellent
2.04 (2/97.9)
2
Excellent
2.04 (2/97.9)
 
Excellent rate
67.70%
Mean 3.95%
Excellent rate
100%
Mean 2.40%
L4
 99
4
Excellent
4.04 (4/99)
6
Good
6.06% (6/99)
 101
7
Good
6.93 (7/101)
3
Excellent
2.97% (3/101)
 100.1
22
Poor
21.98 (22/100.1)
0
Excellent
0
 100.3
7
Good
6.98 (7/100.3)
3
Excellent
2.99 (3/100.3)
 99.5
3
Excellent
3.02 (3/99.5)
2
Excellent
2.01 (2/99.5)
 100
2
Excellent
2 (2/100)
4
Excellent
4 (4/100)
 
Excellent rate
50.00%
Mean 7.50%
Excellent rate
83.30%
Mean 3.01%
L5
 103.5
16
Poor
15.46 (16/103.5)
5
Excellent
4.83 (5/103.5)
 105.5
5
Excellent
4.74 (5/105.5)
4
Excellent
3.79 (4/105.5)
 106
7
Good
6.60 (7/106)
7
Good
6.6 (7/106)
 102.5
11
Acceptable
10.73 (11/102.5)
6
Good
5.85 (6/102.5)
 103
7
Good
6.8 (7/103)
2
Excellent
1.94 (2/103)
 102.8
10
Good
9.73 (10/102.8)
6
Good
5.84 (6/102.8)
 
Excellent rate
16.70%
Mean 9.01%
Excellent rate
50%
Mean 4.81%
Among the 30 screws in the experimental group (after learning), 26 screws were excellent, 4 screws were good, and none of the screws was found to be acceptable or poor. The excellent rate was 86.67%, and the mean deviation rate was 2.68% (Table 2).

Statistical comparison of the results of screw placement by residents

The difference in the deviation rate of screw placement at L1–L4 before and after the learning was not statistically significant, but L5 was significant (t = 5.976, P < 0.05) (Table 3). After learning, the excellent rate of screw placement was significantly higher (χ2 = 15.729, P < 0.05).
Table 3
Comparison of screw placement deviation rates before and after the learning of resident doctors
Vertebral sequence
Control group
Experimental group
T
P
L1
2.51 ± 2.21
1.07 ± 0.96
2.124
0.176
L2
2.85 ± 2.29
1.42 ± 0.87
2.024
0.185
L3
3.95 ± 2.31
2.40 ± 1.68
1.764
0.214
L4
7.5 ± 7.38
3.01 ± 2.02
2.056
0.182
L5
9.01 ± 3.85
4.81 ± 1.71
5.976
0.035*
*P < 0.05

Discussion

Analysis of the results of the X-ray and anatomical specimens

Table 1 demonstrates that in X-ray and anatomical specimens, the angle from the upper to the lower lumbar shows a gradually increasing trend, and the differences among each segment were statistically significant (P < 0.05). This is due to the fact that the closer to the sacrum, the greater the angle of the lateral margin of the isthmus and the axis of the pedicle, allowing more attachment to the strong lower lumbar ligamentous structures and bringing them into line with normal biomechanics. Therefore, the author considers that the lumbar vertebrae can be divided into two categories: L1 and L2 can both belong to the upper lumbar; while L3, L4, and L5 can be classified as the lower lumbar. So in screw placement, the upper region can form a right angle, while in lower at a larger angle which is set at approximately 100°.
The differences between the two genders at L1, L2, and L3 were not statistically significant. However, the angles were larger in women than in men at the L4 and L5 segments, the differences were statistically significant, and the deviation was 2–3°. It may be correlated to the difference in the physiological curvature of the lumbosacral region between genders. This suggests that special attention should be paid to assisted the screw placement in L4 and L5 vertebrae by identifying the above anatomical markers in different gender group.
The differences at each segment between the anatomical and imaging measurements were not statistically significant.

The difference between the lateral margin of isthmus with the lateral margin of lamina and their angles

Figure 4 illustrates the significant difference between them. Because the lateral margin of the isthmus is not affected by the radian of the inferior margin of the lamina and facet joint, the measurement numerical values of its angle between the axis of the pedicle are slightly smaller. If the inferior margin of the lamina or facet joint has severe hyperplasia, the degree of angle between the lateral margin of the lamina and the axis of the pedicle on X-rays would further increase, which is not as constant as the lateral margin of the isthmus mentioned above.
The isthmus and lamina are both parts of the vertebral arch, according to the definition. However, these two are not of the same anatomical structure. The isthmus is most likely to be confused with the lamina [10]. The isthmus refers to the junction of the superior and inferior articular processes, while the lamina is derived from the expansion of the bilateral pedicle of the vertebral arch in the backward and inward direction and convergence in the posterior midline. The isthmus is the inherent anatomical landmark of lumbar vertebrae, it includes two different parts [1113]. Among these, the lateral pars of the isthmus is an osseous pillar that connects the superior and inferior articular processes and is located at the lateral margin of the lamina (Fig. 5a). The true isthmus of the vertebral arch (isthmic pars or pars interarticularis) refers to the location of the isthmus fissure (Fig. 5b), which is the part of the posterior bone structure of the spine located between the pedicle and lamina [1113]. In the present study, the isthmus that forms an included angle with the axis of the pedicle refers to the lateral isthmus, instead of the true isthmus.
The lateral margin of the lamina is the part between both the lower margins of the unilateral superior and inferior articular processes of the same segment [5, 14] (Fig. 5d). But the lateral pars of the isthmus is the part between the lower margin of the superior articular process and the upper margin of the inferior articular process of the same segment (Fig. 5c). These two anatomic definitions have the same starting point but different ending points.
In anatomical morphology, there is a certain degree of radian between the surface of the lateral margin of the isthmus and the lateral margin of the lower lamina 5 (Figs. 4c, d). The surface of the lateral margin of the isthmus is a straight line (Fig. 4b, d), but the lower margin of the lamina is backward and slightly warps, especially in the lower lumbar L4 and L5 segments (Fig. 4a, c). Other scholars choose the starting and ending point of the lateral margin of the lamina to serve as a benchmark line, this should be correspondingly adjusted according to the degree of back warping of the lamina, especially in the lower lumbar. Moreover, this line cannot be observed in the anatomical specimen or operation. This non-present connecting line or plane can only be imagined by the operator, and based on hypothesis, further infers the angle of the screw in the pedicle, causing a certain deviation.
The anatomical adjacent relationship with the screw entrance point: The isthmus is the nearest and fixed anatomical marker for the vertex of the crista lambdoidal (the junction of the lateral margin of the isthmus and accessory process) which is the insertion point for the screw. Along the straight lateral margin of the isthmus, it is easy for the operator looking for it. These also would enable the measurement of its angle conveniently and accurately. The lamina is infero & internal to the isthmus. It is relatively far and a certain radian from the screw entrance point. Therefore, the author did not consider it is an ideal anatomical reference.

Clinical application of the angle

The degeneration of the lumbar facet joint has been found in 89.2% of a 60–69 years old population [1520]. If the connecting line between the lower margins of the superior and inferior articular processes of the same segment was used, the severe degeneration of the articular process would affect the operator in determining the angle.
But the lateral margin of the isthmus will not degenerate, and its angle between the axis of the pedicle in various segments follows a certain rule and does not change with the position of the patient during the operation.
The statistical analysis of the screw placement before and after the junior doctors learned the new method revealed good accuracy (Table 3). The difference in the deviation rate of screw placement before and after the learning was statistically significant only in the L5 segment (P < 0.05). This suggests that it is particularly important to learn this method for the screw placement of the L5 vertebrae. In addition, the difference in the overall excellent rate of screw placement was statistically significant (P < 0.05). In the upper lumbar, the screw was only required to be perpendicular to the surface of the vertebral plate, which was not very difficult. However, it was relatively difficult to place the screw in the lower lumbar due to the presence of the lumbosacral angle. Furthermore, in clinical practice, the number of patients with degenerative diseases in the lower lumbar is relatively more.
The angle should be accurately measured on the X-ray before the operation, so the accuracy of screw placement can be further improved, operation time is shortened, and the dose of X-ray radiation is reduced.

Conclusion

The screw placement that refers to the angle between the axis of the pedicle and the lateral margin of the isthmus would not be affected by the degeneration of the facet joint, position, and lumbosacral angle, and can be controlled more accurately.

Acknowledgements

Shuiquan Wang and Dilimulati Aikeremu contributed equally in this study.

Declarations

The study design was approved by the Ethics Committee of Xinjiang Medical University before data collection and analysis. Informed consent was obtained for experimentation with human subjects.

Competing interests

The authors declare that the research was conducted in the absence of any commercial or financial relationships that could be construed as a potential conflict of interest.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Wu C, Tan L. Study and application of the relationship between lateral margin and upper margin and pedicle of vertebral arch in the upper thoracic and lumbar spine of X-ray. Orthop J China. 2007;15:1559–61. Wu C, Tan L. Study and application of the relationship between lateral margin and upper margin and pedicle of vertebral arch in the upper thoracic and lumbar spine of X-ray. Orthop J China. 2007;15:1559–61.
2.
Zurück zum Zitat Yu X, Pang QJ, Yu GR. Measurement of the angle between the axis of lumbar vertebral arch and the outer edge of vertebral lamina and its clinical significance. Chin J Clin Anat. 2010;28:478–82. Yu X, Pang QJ, Yu GR. Measurement of the angle between the axis of lumbar vertebral arch and the outer edge of vertebral lamina and its clinical significance. Chin J Clin Anat. 2010;28:478–82.
3.
Zurück zum Zitat Ebraheim NA, Hassan A, Lee M, Xu RM. Functional anatomy of the lumbar spine. Seminars Pain Med. 2004;2:131–7.CrossRef Ebraheim NA, Hassan A, Lee M, Xu RM. Functional anatomy of the lumbar spine. Seminars Pain Med. 2004;2:131–7.CrossRef
4.
Zurück zum Zitat Guo SF. Clinical Anatomy of Orthopaedics. 1st ed. Jinan: Shandong Science and Technology Press; 2002. p. 151. Guo SF. Clinical Anatomy of Orthopaedics. 1st ed. Jinan: Shandong Science and Technology Press; 2002. p. 151.
5.
Zurück zum Zitat Karapinar L, Erel N, Ozturk H, et al. Pedicle screw placement with a free hand technique in thoracolumbar spine: is it safe? J Spinal Disord Tech. 2008;21:63–7.CrossRefPubMed Karapinar L, Erel N, Ozturk H, et al. Pedicle screw placement with a free hand technique in thoracolumbar spine: is it safe? J Spinal Disord Tech. 2008;21:63–7.CrossRefPubMed
6.
Zurück zum Zitat Chan CY, Kwan MK, Saw LB. Safety of thoracic pedicle screw application using the funnel technique in Asians: a cadaveric evaluation. Eur Spine J. 2010;19:78–84.CrossRefPubMed Chan CY, Kwan MK, Saw LB. Safety of thoracic pedicle screw application using the funnel technique in Asians: a cadaveric evaluation. Eur Spine J. 2010;19:78–84.CrossRefPubMed
7.
Zurück zum Zitat Zou HB. Essentials of spinal microsurgery. 1st ed. Xi’an: World Publishing Corporation; 2014. p. 251–2. Zou HB. Essentials of spinal microsurgery. 1st ed. Xi’an: World Publishing Corporation; 2014. p. 251–2.
8.
Zurück zum Zitat Bogduk N. Clinical Anatomy of the lumbar spine and sacrum. 4th ed. Edinburgh: Elsevier Churchill Livingstone; 2005. Bogduk N. Clinical Anatomy of the lumbar spine and sacrum. 4th ed. Edinburgh: Elsevier Churchill Livingstone; 2005.
9.
Zurück zum Zitat Tang TS, Yang HL, Ni CF, Liu D. Analysis of the position of pedicle screws in 112 cases. Chin J Surg. 1993;31:411–3.PubMed Tang TS, Yang HL, Ni CF, Liu D. Analysis of the position of pedicle screws in 112 cases. Chin J Surg. 1993;31:411–3.PubMed
10.
Zurück zum Zitat More KL, Dalley AF, Agur AMR. Clinically oriented anatomy. 6th ed. Philadelphia: Lippincott Williams&Wilkins; 2010. p. 478–9. More KL, Dalley AF, Agur AMR. Clinically oriented anatomy. 6th ed. Philadelphia: Lippincott Williams&Wilkins; 2010. p. 478–9.
11.
Zurück zum Zitat Torun F, Tuna H, Buyukmumcu M, et al. The lumbar roots and pedicles: a morphometric analysis and anatomical features. J Clin Neurosci. 2008;15:895–9.CrossRefPubMed Torun F, Tuna H, Buyukmumcu M, et al. The lumbar roots and pedicles: a morphometric analysis and anatomical features. J Clin Neurosci. 2008;15:895–9.CrossRefPubMed
12.
Zurück zum Zitat Mogora A, Schwartz A. Relation between the low back pain syndrome and x-ray findings. I. Degenerative osteoarthritis. Scand J Rehab Med. 1976;8:115–25. Mogora A, Schwartz A. Relation between the low back pain syndrome and x-ray findings. I. Degenerative osteoarthritis. Scand J Rehab Med. 1976;8:115–25.
13.
Zurück zum Zitat Kirkaldy-Willis WH, Farfan HF. Instability of the lumbar spine. Clin Orthop Relat Res. 1982;165:110–23.CrossRef Kirkaldy-Willis WH, Farfan HF. Instability of the lumbar spine. Clin Orthop Relat Res. 1982;165:110–23.CrossRef
14.
Zurück zum Zitat Yong-Hing K, Kirkaldy-Willis WH. The pathophysiology of degenerative disease of the lumbar spine. Orthop Clin N Am. 1983;14:491–504.CrossRef Yong-Hing K, Kirkaldy-Willis WH. The pathophysiology of degenerative disease of the lumbar spine. Orthop Clin N Am. 1983;14:491–504.CrossRef
15.
Zurück zum Zitat Jaumard NV, Welch WC, Winkelstein BA. Spinal facet joint biomechanics and mechanotransduction in normal, injury and degenerative conditions. J Biomech Eng. 2011;133:071010.CrossRefPubMed Jaumard NV, Welch WC, Winkelstein BA. Spinal facet joint biomechanics and mechanotransduction in normal, injury and degenerative conditions. J Biomech Eng. 2011;133:071010.CrossRefPubMed
16.
Zurück zum Zitat Varlotta GP, Lefkowitz TR, Schweitzer M, et al. The lumbar facet joint :a review of current Knowledge: part1: anatomy, biomechanics and grading. Skeletal Radiol. 2011;40:13–23.CrossRefPubMed Varlotta GP, Lefkowitz TR, Schweitzer M, et al. The lumbar facet joint :a review of current Knowledge: part1: anatomy, biomechanics and grading. Skeletal Radiol. 2011;40:13–23.CrossRefPubMed
17.
Zurück zum Zitat Kalichman L, Hunter DJ. Lumbar facet joint osteoarthritis: a review. Semin Arthritis Rheum. 2007;37:69–80.CrossRefPubMed Kalichman L, Hunter DJ. Lumbar facet joint osteoarthritis: a review. Semin Arthritis Rheum. 2007;37:69–80.CrossRefPubMed
18.
Zurück zum Zitat Gellhorn AC, Katz JN, Suri P. Osteoarthritis of the spine: the facet joints. Nat Rev Rheumatol. 2013;9:216–24.CrossRefPubMed Gellhorn AC, Katz JN, Suri P. Osteoarthritis of the spine: the facet joints. Nat Rev Rheumatol. 2013;9:216–24.CrossRefPubMed
19.
Zurück zum Zitat Izzo R, Guarnieeri G, Guglielmi G, et al. Biomechanics of the spine. Part I: spinal stability. Eur J Radiol. 2013;82:118–26.CrossRefPubMed Izzo R, Guarnieeri G, Guglielmi G, et al. Biomechanics of the spine. Part I: spinal stability. Eur J Radiol. 2013;82:118–26.CrossRefPubMed
20.
Metadaten
Titel
Anatomical and imaging measurements of the angle between the axis of the lumbar pedicle and lateral isthmus margin and its clinical significance
verfasst von
Shuiquan Wang
Dilimulati Aikeremu
Alafate Kahaer
Abulikemu Maimaiti
Yang Xiao
Abudusalamu Tuoheti
Rui Zhang
Xieraili Maimaiti
Hailong Guo
Paerhati Rexiti
Publikationsdatum
01.12.2023
Verlag
BioMed Central
Erschienen in
Journal of Orthopaedic Surgery and Research / Ausgabe 1/2023
Elektronische ISSN: 1749-799X
DOI
https://doi.org/10.1186/s13018-023-03983-3

Weitere Artikel der Ausgabe 1/2023

Journal of Orthopaedic Surgery and Research 1/2023 Zur Ausgabe

Arthropedia

Grundlagenwissen der Arthroskopie und Gelenkchirurgie. Erweitert durch Fallbeispiele, Videos und Abbildungen. 
» Jetzt entdecken

Knie-TEP: Kein Vorteil durch antibiotikahaltigen Knochenzement

29.05.2024 Periprothetische Infektionen Nachrichten

Zur Zementierung einer Knie-TEP wird in Deutschland zu über 98% Knochenzement verwendet, der mit einem Antibiotikum beladen ist. Ob er wirklich besser ist als Zement ohne Antibiotikum, kann laut Registerdaten bezweifelt werden.

Häusliche Gewalt in der orthopädischen Notaufnahme oft nicht erkannt

28.05.2024 Häusliche Gewalt Nachrichten

In der Notaufnahme wird die Chance, Opfer von häuslicher Gewalt zu identifizieren, von Orthopäden und Orthopädinnen offenbar zu wenig genutzt. Darauf deuten die Ergebnisse einer Fragebogenstudie an der Sahlgrenska-Universität in Schweden hin.

Fehlerkultur in der Medizin – Offenheit zählt!

28.05.2024 Fehlerkultur Podcast

Darüber reden und aus Fehlern lernen, sollte das Motto in der Medizin lauten. Und zwar nicht nur im Sinne der Patientensicherheit. Eine negative Fehlerkultur kann auch die Behandelnden ernsthaft krank machen, warnt Prof. Dr. Reinhard Strametz. Ein Plädoyer und ein Leitfaden für den offenen Umgang mit kritischen Ereignissen in Medizin und Pflege.

Mehr Frauen im OP – weniger postoperative Komplikationen

21.05.2024 Allgemeine Chirurgie Nachrichten

Ein Frauenanteil von mindestens einem Drittel im ärztlichen Op.-Team war in einer großen retrospektiven Studie aus Kanada mit einer signifikanten Reduktion der postoperativen Morbidität assoziiert.

Update Orthopädie und Unfallchirurgie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.