Skip to main content
Erschienen in: BMC Oral Health 1/2022

Open Access 01.12.2022 | Research

Validation of an implant stability measurement device using the percussion response: a clinical research study

verfasst von: Yurie Okuhama, Koudai Nagata, Hyunjin Kim, Hayato Tsuruoka, Mihoko Atsumi, Hiromasa Kawana

Erschienen in: BMC Oral Health | Ausgabe 1/2022

Abstract

Background

Several devices have been developed to measure implant-bone stability as an indicator of successful implant treatment; these include Osstell®, which measures the implant stability quotient (ISQ), and the more recent AnyCheck®, which relies on percussion for the implant stability test (IST). These devices make it possible to measure implant stability. However, no studies have compared the performance of AnyCheck® and Osstell® (i.e., IST and ISQ values) in clinical practice. Therefore, this study aimed to determine the correlation between primary and secondary implant stability using the Osstell® and AnyCheck® devices.

Methods

Ten patients (7 women; age [mean ± standard deviation]: 49.1 ± 13.3 years) with partially edentulous jaws who received a total of 15 implants were included. IST (AnyCheck®) and ISQ (Osstell®) values were measured immediately after implantation and at 1, 2, 3, 4, and 6 weeks post-implantation. Each measurement was performed three times, and the average value was used as the result. The correlation between measurements obtained using the two devices was determined using Spearman's rank correlation coefficient.

Results

The IST values ranged from 79.1 ± 2.87 to 82.4 ± 2.65. The ISQ values ranged from 76.0 ± 2.8 to 80.2 ± 2.35. Spearman's rank correlation coefficient was r = 0.64 immediately after implantation, r = 0.29 at 1 week, r = 0.68 at 2 weeks, r = 0.53 at 3 weeks, r = 0.68 at 4 weeks, and r = 0.56 at 6 weeks. A positive correlation was found in all cases, except at week 1 when the correlation was weak; the IST and ISQ values decreased the most during the first postoperative week and increased during the second week. The IST values were also slightly higher at all measurement points.

Conclusion

The ability to assess implant stability without removing the abutment during healing is essential for determining the timing of loading without the risk of bone resorption. The results of this study suggest that AnyCheck® is useful for determining primary and secondary implant stability.
Hinweise

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
ISQ
Implant stability quotient
RFA
Resonance frequency analysis
IT
Insertion torque
IST
Implant stability test

Background

In recent years, the use of dental implants has become widespread in the field of dentistry, and various technological advancements have been proposed to improve treatment outcomes [13]. For instance, several devices have been developed to measure implant stability as an indicator of the success of implant treatment. The Osstell® device [4] allows the measurement of the implant stability quotient (ISQ) using the resonance frequency analysis (RFA) method, whereas the Periotest® device [5] uses the percussion method. More recently, the AnyCheck® device, which also relies on the percussion method, has been developed [6]. Importantly, the insertion torque (IT) of the implant into the bone influences the success of implant treatment; therefore, the ability of these devices to quantify and evaluate implant stability has contributed greatly to the success of implant treatments [7, 8], benefitting both dentist and patients. There are two types of implant surgery: those that allow submerged implant healing and those with non-submerged implant healing. Submerged implant healing is often considered when the primary stability is poor or when bone grafting has been performed [9]. In non-submerged implant healing, removal of healing abutments prior to superstructure placement has been reported to be a cause of accelerated bone resorption [10]. Therefore, the concept of “one abutment–one time,” in which the abutment is placed immediately after implantation to control bone resorption, is popular [11]. Despite its long history of use, the Osstell® device requires removal of the healing abutment and the attachment of smart pegs. Of note, AnyCheck® does not require the healing abutment to be attached or removed; therefore, it can measure implant stability without promoting bone resorption. Although there have been various reports on implant stability, thus far, no study has compared the ISQ and implant stability test (IST) values in clinical practice [12]. To address this gap in knowledge, the present study aimed to investigate the correlation between implant stability for the Osstell® and AnyCheck® devices.

Materials and methods

Patients

Ten patients (7 women, 3 men) with partially edentulous jaws who underwent implant treatment at our university hospital (n = 15 implants) were included in this study. The mean age (± standard deviation) was 49.1 ± 13.3 years. Patients were selected based on absence of systemic diseases, smoking status (non-smokers), and non-requirement of bone grafting. The IT was set at 35 Ncm using micromotor and torque wrench for all patients. Healing abutments of the following diameters were attached to the implants: 2 mm in one implant, 4 mm in nine, and 6 mm in five implants. This study was approved by the institutional ethics committee of our hospital (approval #739), and written informed consent was obtained from all patients.

Surgical procedure

All patients were instructed to take an oral dose (1 g) of amoxicillin hydrate (Sawacillin Capsules®; LTL Pharma, Tokyo, Japan) 1 h before surgery. After administration of the anesthetic (Lidocaine/Adrenaline bitartrate®; Showa Yakuhin Kako Co., Ltd., Tokyo, Japan), the alveolar mucosa, including the periosteum, was incised at the top of the ridge and separated. After drilling, implants were placed according to the implant system protocol; the torque and depth of placement were adjusted with a torque ratchet. All implant placements were performed via freehand insertion; additionally, all surgeries were performed in a non-submerged fashion. The implant system used was Straumann® SLActive φ 4.1 × 10 (bone level tapered implant; Basel, Switzerland). All surgeries were performed by the same doctor, a teaching Associate in the Department of Implantology at our university hospital.

Measurement of the IST and ISQ values

The IST values were measured using the AnyCheck® device (Neobiotech Co., Ltd., Seoul, South Korea) (Fig. 1). The bone-to-implant stability index was set based on the ISQ values (0–59, not recommended for loading; 60–99, good stability, recommended for loading); the IST and ISQ have similar reference values. Osstell® was used instead of Periotest® in this study. Briefly, to determine the IST value, the healing abutment was struck six times over 3 s, and the contact time with the healing abutment was measured to calculate the stability. Notably, in accordance with the manufacturer’s recommendations, the patient was placed in an upright position during measurement, and the contact angle was set at 0°–30°. Since AnyCheck® uses a standard healing abutment height of 4 mm, values for healing abutment heights other than 4 mm were corrected as recommended by the manufacturer (Table 1).
Table 1
Corrected IST values, measured using the AnyCheck® device, based on the healing abutment height
Healing abutment height (mm)
IST value
7
+ 6
6
+ 4
5
+ 2
4
± 0
3
− 2
2
− 4
1
− 6
IST, implant stability test
The ISQ values were determined using the Osstell® ISQ device (Integration Diagnostics Ltd., Goteborgsvagen, Sweden) (Fig. 2). In principle, magnetic pulses based on the RFA method stimulate and resonate the smart peg (Integration Diagnostics Ltd.) attached to the implant body in the patient’s mouth, making it possible to quantify stability. At the time of measurement, the intraoral healing abutment was removed, and the smart peg was attached to the implant body via hand tightening.
Both the IST and ISQ values were measured immediately after implantation and at 1-, 2-, 3-, 4-, and 6-weeks post-implantation. Each measurement was taken three times, and the mean was used as the definitive result. The ISQ was measured following assessment of the IST. For all implants, impressions were obtained at 4 weeks after placement, and provisional restorations were placed at 6 weeks. All measurements were taken by the same dental surgeon.

Statistical analyses

Correlations between the IST and ISQ values were assessed using BellCurve for Excel (Social Survey Research Information, Inc., Tokyo). Spearman's rank correlation coefficients were used to determine correlations.
Sample size was calculated by one-way analysis of variance using G-Power (version 3.1.9.2). The sample size required to obtain 80% of the effect size of 0.4 at α = 0.05 was calculated.

Results

The IST values immediately, 1 week, 2 weeks, 3 weeks, 4 weeks, and 6 weeks after implantation were 81.0 ± 2.82, 79.1 ± 2.87, 79.7 ± 2.83, 80.5 ± 2.71, 80.9 ± 4.0, and 82.4 ± 2.65, respectively. The ISQ values immediately, 1 week, 2 weeks, 3 weeks, 4 weeks, and 6 weeks after implantation were 79.8 ± 2.89, 76.0 ± 2.8, 77.8 ± 2.63, 79.2 ± 2.44, 79.7 ± 2.77, and 80.2 ± 2.35, respectively (Fig. 3). Of note, both the IST and ISQ values decreased the most in the first week after surgery and increased in the second week; additionally, the IST value was slightly higher at all measurement points. The Spearman’s rank correlation coefficients for each measurement period were as follows: r = 0.64 immediately after implantation; r = 0.29 at 1 week; r = 0.68 at 2 weeks; r = 0.53 at 3 weeks, r = 0.68 at 4 weeks, and r = 0.56 at 6 weeks. A positive correlation was found in all cases, except at 1 week when the correlation was weak (Fig. 4).

Discussion

This study compared the changes in implant stability using the Osstell® and AnyCheck® devices. Our analysis indicated that the measurements exhibited a positive correlation of > 0.5, except after 2 weeks. This suggested that AnyCheck® had the same performance as Osstell®.
When the IT is high, bone resorption is promoted. Optimization of the IT is considered the key to successful implant treatment [1315]. In this study, all the implants had an IT of 35 Ncm. However, even in cases of low IT, the use of AnyCheck® allows safe assessment of implant stability. The IST and ISQ values in this study were high. Zwaan et al. [16] placed 163 implants in the maxilla and compared the IT at 50 Ncm, 40–45 Ncm, 30–35 Ncm, and ≤ 30 Ncm and found that the ISQ values were 76.2 ± 5.3, 72.3 ± 5.3, 70.0 ± 6.7, and 68.1 ± 6.2, respectively. The ISQ values were also reported to be higher for tapered implants than for straight implants. Van Eekeren et al. [17] compared bone-level with tissue-level implants and revealed that the ISQ values (at the time of placement and 2, 3, and 12 weeks postoperatively) were 77.8, 75.6, 76.3, and 79.1, and 74.0, 71.8, 72.6, and 76.8, respectively. Importantly, the above results suggest that ISQ values tend to vary according to bone quality, implantation site, and implant shape, in line with the findings reported elsewhere [18]. As reported above, the authors of this study think that the high value was due to the use of bone-level and tapered implants. Oates et al. [19] reported that the stability of SLActive® implants changed from a decrease to an increase at 2 weeks after placement, in line with our results. In the present study, the weakest correlation was observed after 2 weeks. This may be explained by individual differences in the decline of primary stability, resulting in large differences in IST and ISQ.
Park et al. [6] placed an implant into an artificial bone block to verify the accuracy of AnyCheck®; interestingly, the stability decreased as the height of the healing abutment increased and as the contact angle decreased from 30° to 0° (perpendicular to the long axis of the implant and parallel to the ground). Subsequently, Lee et al. [20] placed implants at 10 N, 15 N, and 35 N into artificial bone blocks together with five different diameters of healing abutments of the same height, measured the IST values using AnyCheck®, and compared them with the ISQ values determined using Osstell®. Importantly, they reported that the diameter of the healing abutment did not affect the ISQ and IST values, which exhibited a strong correlation. Consistent with these results, Lee et al. [21] also found that the results for the AnyCheck® and Osstell® devices were correlated in the context of both internal-connection and external-connection implants (within pig bone). Of note, they also reported that the IST values were higher for both implants and that there was no significant difference between the IST and ISQ values. However, neither the IST nor the ISQ values are known to be accurate; they should only be considered as one among several indicators.
In clinical practice, Al-Jamal et al. [22] demonstrated that there was a significant correlation between primary stability and IT using the AnyCheck® device in the context of 40 implants. However, they did not compare their findings with measurements obtained using the Osstell® device. The present study is the first in which the IST and ISQ values were measured and compared weekly in clinical practice, from immediately after implantation to 4 weeks later. While the Osstell® is a device with a long history of use and has been explored in many studies to date, its use requires removal of the healing abutment and attachment of the smart peg. The recently released Osstell Beacon® is cordless. However, as before, it still requires a smart peg, and the healing abutment must be attached and removed. Esposito et al. [23] reported that the removal of the healing abutment (three times after implantation until the time of superstructure attachment) led to 0.16 mm of bone resorption per year (versus non-removal of the healing abutment). Similar results were obtained by Bressan et al. [24]—0.43 mm of bone resorption over 3 years in healing abutment removal versus non-removal contexts—as well as by Koutouzis et al. [25]—0.13 mm versus 0.28 mm bone resorption in 6 months after implantation in the without versus with healing abutment removal context). Importantly, AnyCheck®, which allows the measurement of stability without the need to attach or to detach the healing abutment, reduces bone resorption and can be applied to low-torque cases. In the present study, a positive correlation of > 0.5 was observed at all measurement points, except after 2 weeks. Considering the risk of bone resorption and other factors, the AnyCheck® is expected to perform as well or better than the Osstell®. Since there are no reports comparing the two devices in clinical practice, further validation of this matter is necessary. Furthermore, this study has some limitations. The sample size for this study was small. This was due to the limited number of patients in whom implants of the same system, diameter, and length were placed. In addition, in vitro studies cannot assess changes in implant stability over time. Therefore, studies using models could not be conducted previously. In the future, it is necessary to distinguish between bone quality and implant diameter to obtain more detailed data.

Conclusion

The ability to assess implant stability without removing the abutment during healing is essential for determining the time at which load can be applied without the risk of bone resorption. Altogether, our results suggest the similar performance of Osstell® and AnyCheck®, and, consequently, the usefulness of the latter for the determination of implant stability.

Acknowledgements

Not applicable.

Declarations

This study was approved by the Kanagawa Dental University Ethics Committee (approval # 739), and written informed consent was obtained from all patients. All methods were performed in accordance with the 1964 Helsinki Declaration.
Not applicable.

Competing interests

The authors declare that they have no competing interests.
Open AccessThis article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://​creativecommons.​org/​licenses/​by/​4.​0/​. The Creative Commons Public Domain Dedication waiver (http://​creativecommons.​org/​publicdomain/​zero/​1.​0/​) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
2.
Zurück zum Zitat Makowiecki A, Hadzik J, Błaszczyszyn A, Gedrange T, Dominiak M. An evaluation of superhydrophilic surfaces of dental implants - a systematic review and meta-analysis. BMC Oral Health. 2019;19:79.CrossRef Makowiecki A, Hadzik J, Błaszczyszyn A, Gedrange T, Dominiak M. An evaluation of superhydrophilic surfaces of dental implants - a systematic review and meta-analysis. BMC Oral Health. 2019;19:79.CrossRef
3.
Zurück zum Zitat Papaspyridakos P, De Souza A, Vazouras K, Gholami H, Pagni S, Weber HP. Survival rates of short dental implants (≤6 mm) compared with implants longer than 6 mm in posterior jaw areas: a meta-analysis. Clin Oral Implants Res. 2018;29(Suppl 16):8–20.CrossRef Papaspyridakos P, De Souza A, Vazouras K, Gholami H, Pagni S, Weber HP. Survival rates of short dental implants (≤6 mm) compared with implants longer than 6 mm in posterior jaw areas: a meta-analysis. Clin Oral Implants Res. 2018;29(Suppl 16):8–20.CrossRef
4.
Zurück zum Zitat Staedt H, Kämmerer PW, Goetze E, Thiem DGE, Al-Nawas B, Heimes D. Implant primary stability depending on protocol and insertion mode - an ex vivo study. Int J Implant Dent. 2020;6:1–9.CrossRef Staedt H, Kämmerer PW, Goetze E, Thiem DGE, Al-Nawas B, Heimes D. Implant primary stability depending on protocol and insertion mode - an ex vivo study. Int J Implant Dent. 2020;6:1–9.CrossRef
5.
Zurück zum Zitat Olivé J, Aparicio C. Periotest method as a measure of osseointegrated oral implant stability. Int J Oral Maxillofac Implants. 1990;5:390–400.PubMed Olivé J, Aparicio C. Periotest method as a measure of osseointegrated oral implant stability. Int J Oral Maxillofac Implants. 1990;5:390–400.PubMed
6.
Zurück zum Zitat Park YH, Leesungbok R, Lee SW, Paek J, Lee JY. Differences in percussion-type measurements of implant stability according to height of healing abutments and measurement angle. J Korean Acad Prosthodont. 2018;56:278–86.CrossRef Park YH, Leesungbok R, Lee SW, Paek J, Lee JY. Differences in percussion-type measurements of implant stability according to height of healing abutments and measurement angle. J Korean Acad Prosthodont. 2018;56:278–86.CrossRef
7.
Zurück zum Zitat Atsumi M, Park SH, Wang HL. Methods used to assess implant stability: current status. Int J Oral Maxillofac Implants. 2007;22:743–54.PubMed Atsumi M, Park SH, Wang HL. Methods used to assess implant stability: current status. Int J Oral Maxillofac Implants. 2007;22:743–54.PubMed
8.
Zurück zum Zitat Lages FS, Douglas-de Oliveira DW, Costa FO. Relationship between implant stability measurements obtained by insertion torque and resonance frequency analysis: a systematic review. Clin Implant Dent Relat Res. 2018;20:26–33.CrossRef Lages FS, Douglas-de Oliveira DW, Costa FO. Relationship between implant stability measurements obtained by insertion torque and resonance frequency analysis: a systematic review. Clin Implant Dent Relat Res. 2018;20:26–33.CrossRef
9.
Zurück zum Zitat Troiano G, Lo Russo L, Canullo L, Ciavarella D, Lo Muzio L, Laino L. Early and late implant failure of submerged versus non-submerged implant healing: a systematic review, meta-analysis and trial sequential analysis. J Clin Periodontol. 2018;45:613–23.CrossRef Troiano G, Lo Russo L, Canullo L, Ciavarella D, Lo Muzio L, Laino L. Early and late implant failure of submerged versus non-submerged implant healing: a systematic review, meta-analysis and trial sequential analysis. J Clin Periodontol. 2018;45:613–23.CrossRef
10.
Zurück zum Zitat Degidi M, Nardi D, Piattelli A. One abutment at one time: non-removal of an immediate abutment and its effect on bone healing around subcrestal tapered implants. Clin Oral Implants Res. 2011;22:1303–7.CrossRef Degidi M, Nardi D, Piattelli A. One abutment at one time: non-removal of an immediate abutment and its effect on bone healing around subcrestal tapered implants. Clin Oral Implants Res. 2011;22:1303–7.CrossRef
11.
Zurück zum Zitat Molina A, Sanz-Sánchez I, Martín C, Blanco J, Sanz M. The effect of one-time abutment placement on interproximal bone levels and peri-implant soft tissues: a prospective randomized clinical trial. Clin Oral Implants Res. 2017;28:443–52.CrossRef Molina A, Sanz-Sánchez I, Martín C, Blanco J, Sanz M. The effect of one-time abutment placement on interproximal bone levels and peri-implant soft tissues: a prospective randomized clinical trial. Clin Oral Implants Res. 2017;28:443–52.CrossRef
12.
Zurück zum Zitat Hemanth H, Mallajosyula SS. Polarization influences the evolution of nucleobase-graphene interactions. Nanoscale. 2021;13:4060–72.CrossRef Hemanth H, Mallajosyula SS. Polarization influences the evolution of nucleobase-graphene interactions. Nanoscale. 2021;13:4060–72.CrossRef
13.
Zurück zum Zitat Cha JY, Pereira MD, Smith AA, Houschyar KS, Yin X, Mouraret S, et al. Multiscale analyses of the bone-implant interface. J Dent Res. 2015;94:482–90.CrossRef Cha JY, Pereira MD, Smith AA, Houschyar KS, Yin X, Mouraret S, et al. Multiscale analyses of the bone-implant interface. J Dent Res. 2015;94:482–90.CrossRef
14.
Zurück zum Zitat Kotsu M, Urbizo Velez J, Bengazi F, Tumedei M, Fujiwara S, Kato S, et al. Healing at implants installed from ~ 70- to < 10-Ncm insertion torques: an experimental study in dogs. Oral Maxillofac Surg. 2021;25:55–64.CrossRef Kotsu M, Urbizo Velez J, Bengazi F, Tumedei M, Fujiwara S, Kato S, et al. Healing at implants installed from ~ 70- to < 10-Ncm insertion torques: an experimental study in dogs. Oral Maxillofac Surg. 2021;25:55–64.CrossRef
15.
Zurück zum Zitat Amari Y, Piattelli A, Apaza Alccayhuaman KA, Mesa NF, Ferri M, Iezzi G, et al. Bone healing at non-submerged implants installed with different insertion torques: a split-mouth histomorphometric randomized controlled trial. Int J Implant Dent. 2019;5:39.CrossRef Amari Y, Piattelli A, Apaza Alccayhuaman KA, Mesa NF, Ferri M, Iezzi G, et al. Bone healing at non-submerged implants installed with different insertion torques: a split-mouth histomorphometric randomized controlled trial. Int J Implant Dent. 2019;5:39.CrossRef
16.
Zurück zum Zitat Zwaan J, Vanden Bogaerde L, Sahlin H, Sennerby L. A one-year follow-up study of a tapered hydrophilic implant design using various placement protocols in the maxilla. Open Dent J. 2016;10:680–91.CrossRef Zwaan J, Vanden Bogaerde L, Sahlin H, Sennerby L. A one-year follow-up study of a tapered hydrophilic implant design using various placement protocols in the maxilla. Open Dent J. 2016;10:680–91.CrossRef
17.
Zurück zum Zitat van Eekeren P, Said C, Tahmaseb A, Wismeijer D. Resonance frequency analysis of thermal acid-etched, hydrophilic implants during first 3 months of healing and osseointegration in an early-loading protocol. Int J Oral Maxillofac Implants. 2015;30:843–50.CrossRef van Eekeren P, Said C, Tahmaseb A, Wismeijer D. Resonance frequency analysis of thermal acid-etched, hydrophilic implants during first 3 months of healing and osseointegration in an early-loading protocol. Int J Oral Maxillofac Implants. 2015;30:843–50.CrossRef
18.
Zurück zum Zitat Huang H, Wu G, Hunziker E. The clinical significance of implant stability quotient (ISQ) measurements: a literature review. J Oral Biol Craniofac Res. 2020;10:629–38.CrossRef Huang H, Wu G, Hunziker E. The clinical significance of implant stability quotient (ISQ) measurements: a literature review. J Oral Biol Craniofac Res. 2020;10:629–38.CrossRef
19.
Zurück zum Zitat Oates TW, Valderrama P, Bischof M, Nedir R, Jones A, Simpson J, et al. Enhanced implant stability with a chemically modified SLA surface: a randomized pilot study. Int J Oral Maxillofac Implants. 2007;22:755–60.PubMed Oates TW, Valderrama P, Bischof M, Nedir R, Jones A, Simpson J, et al. Enhanced implant stability with a chemically modified SLA surface: a randomized pilot study. Int J Oral Maxillofac Implants. 2007;22:755–60.PubMed
20.
Zurück zum Zitat Lee DH, Shin YH, Park JH, Shim JS, Shin SW, Lee JY. The reliability of Anycheck device related to healing abutment diameter. J Adv Prosthodont. 2020;12:83–8.CrossRef Lee DH, Shin YH, Park JH, Shim JS, Shin SW, Lee JY. The reliability of Anycheck device related to healing abutment diameter. J Adv Prosthodont. 2020;12:83–8.CrossRef
21.
Zurück zum Zitat Lee J, Pyo SW, Cho HJ, An JS, Lee JH, Koo KT, et al. Comparison of implant stability measurements between a resonance frequency analysis device and a modified damping capacity analysis device: an in vitro study. J Periodontal Implant Sci. 2020;50:56–66.CrossRef Lee J, Pyo SW, Cho HJ, An JS, Lee JH, Koo KT, et al. Comparison of implant stability measurements between a resonance frequency analysis device and a modified damping capacity analysis device: an in vitro study. J Periodontal Implant Sci. 2020;50:56–66.CrossRef
22.
Zurück zum Zitat Al-Jamal MFJ, Al-Jumaily HA. Can the bone density estimated by CBCT predict the primary stability of dental implants? A new measurement protocol. J Craniofac Surg. 2021;32:e171–4.CrossRef Al-Jamal MFJ, Al-Jumaily HA. Can the bone density estimated by CBCT predict the primary stability of dental implants? A new measurement protocol. J Craniofac Surg. 2021;32:e171–4.CrossRef
23.
Zurück zum Zitat Esposito M, Bressan E, Grusovin MG, D’Avenia F, Neumann K, Sbricoli L, et al. Do repeated changes of abutments have any influence on the stability of peri-implant tissues? One-year post-loading results from a multicentre randomised controlled trial. Eur J Oral Implantol. 2017;10:57–72.PubMed Esposito M, Bressan E, Grusovin MG, D’Avenia F, Neumann K, Sbricoli L, et al. Do repeated changes of abutments have any influence on the stability of peri-implant tissues? One-year post-loading results from a multicentre randomised controlled trial. Eur J Oral Implantol. 2017;10:57–72.PubMed
24.
Zurück zum Zitat Bressan E, Grusovin MG, D’Avenia F, Neumann K, Sbricoli L, Luongo G, et al. The influence of repeated abutment changes on peri-implant tissue stability: 3-year post-loading results from a multicentre randomised controlled trial. Eur J Oral Implantol. 2017;10:373–90.PubMed Bressan E, Grusovin MG, D’Avenia F, Neumann K, Sbricoli L, Luongo G, et al. The influence of repeated abutment changes on peri-implant tissue stability: 3-year post-loading results from a multicentre randomised controlled trial. Eur J Oral Implantol. 2017;10:373–90.PubMed
25.
Zurück zum Zitat Koutouzis T, Koutouzis G, Gadalla H, Neiva R. The effect of healing abutment reconnection and disconnection on soft and hard peri-implant tissues: a short-term randomized controlled clinical trial. Int J Oral Maxillofac Implants. 2013;28:807–14.CrossRef Koutouzis T, Koutouzis G, Gadalla H, Neiva R. The effect of healing abutment reconnection and disconnection on soft and hard peri-implant tissues: a short-term randomized controlled clinical trial. Int J Oral Maxillofac Implants. 2013;28:807–14.CrossRef
Metadaten
Titel
Validation of an implant stability measurement device using the percussion response: a clinical research study
verfasst von
Yurie Okuhama
Koudai Nagata
Hyunjin Kim
Hayato Tsuruoka
Mihoko Atsumi
Hiromasa Kawana
Publikationsdatum
01.12.2022
Verlag
BioMed Central
Erschienen in
BMC Oral Health / Ausgabe 1/2022
Elektronische ISSN: 1472-6831
DOI
https://doi.org/10.1186/s12903-022-02320-0

Weitere Artikel der Ausgabe 1/2022

BMC Oral Health 1/2022 Zur Ausgabe

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Klinik aktuell Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Update Zahnmedizin

Bestellen Sie unseren kostenlosen Newsletter und bleiben Sie gut informiert – ganz bequem per eMail.