Technique, protocols and adverse reactions for contrast-enhanced spectral mammography (CESM): a systematic review

verfasst von: Moreno Zanardo, Andrea Cozzi, Rubina Manuela Trimboli, Olgerta Labaj, Caterina Beatrice Monti, Simone Schiaffino, Luca Alessandro Carbonaro, Francesco Sardanelli

Erschienen in: Insights into Imaging | Ausgabe 1/2019

Abstract

We reviewed technical parameters, acquisition protocols and adverse reactions (ARs) for contrast-enhanced spectral mammography (CESM). A systematic search in databases, including MEDLINE/EMBASE, was performed to extract publication year, country of origin, study design; patients; mammography unit/vendor, radiation dose, low-/high-energy tube voltage; contrast molecule, concentration and dose; injection modality, ARs and acquisition delay; order of views; examination time. Of 120 retrieved articles, 84 were included from 22 countries (September 2003–January 2019), totalling 14012 patients. Design was prospective in 44/84 studies (52%); in 70/84 articles (83%), a General Electric unit with factory-set kVp was used. Per-view average glandular dose, reported in 12/84 studies (14%), ranged 0.43–2.65 mGy. Contrast type/concentration was reported in 79/84 studies (94%), with Iohexol 350 mgI/mL mostly used (25/79, 32%), dose and flow rate in 72/84 (86%), with 1.5 mL/kg dose at 3 mL/s in 62/72 studies (86%). Injection was described in 69/84 articles (82%), automated in 59/69 (85%), manual in 10/69 (15%) and flush in 35/84 (42%), with 10–30 mL dose in 19/35 (54%). An examination time < 10 min was reported in 65/84 studies (77%), 120 s acquisition delay in 65/84 (77%) and order of views in 42/84 (50%) studies, beginning with the craniocaudal view of the non-suspected breast in 7/42 (17%). Thirty ARs were reported by 14/84 (17%) studies (26 mild, 3 moderate, 1 severe non-fatal) with a pooled rate of 0.82% (fixed-effect model). Only half of CESM studies were prospective; factory-set kVp, contrast 1.5 mL/kg at 3 mL/s and 120 s acquisition delay were mostly used; only 1 severe AR was reported. CESM protocol standardisation is advisable.
Hinweise

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Abkürzungen
AGD
Average glandular dose
CC
Craniocaudal
CESM
Contrast-enhanced spectral mammography
CI
Confidence interval
CT
Computed tomography
DM
Digital mammography
ICA
Iodinated contrast agent
kVp
Peak kilovoltage
MLO
Mediolateral oblique
MRI
Magnetic resonance imaging
PRISMA
Preferred Reporting Items for Systematic Reviews and Meta-Analyses

Key points

  • Eighty-four articles on CESM totalling 14012 patients were reviewed
  • A 1.5 mL/kg contrast dose automatically injected at 3 mL/s was generally adopted
  • Per-view average glandular dose ranged from 0.43 to 2.65 mGy
  • Studies for contrast agent dose-finding and view acquisition ordering are lacking
  • Adverse reaction rate (only one severe) was similar to that reported for CT

Background

During the 1960s and 1970s, randomised controlled trials proved that screen-film mammography for breast cancer screening yields a reduction in breast cancer mortality [1]. Since the early 2000s, screen-film mammography was progressively replaced by digital mammography (DM), which improved performance especially in women under 50 years of age and in case of dense breasts, even though providing an intrinsically inferior spatial resolution [2]. In the last two decades, digital breast tomosynthesis brought substantial further improvements [3, 4], increasing cancer detection rate and reducing the recall rate [5].
Contrast-enhanced mammography is the combination of X-ray mammography with intravenous administration of iodinated contrast agent (ICA) [6]. It was first attempted using a digital subtraction technique [79], but this approach was soon abandoned due to difficulties in co-registration of unenhanced and contrast-enhanced images [10, 11]. In the last two decades, contrast-enhanced spectral mammography (CESM) has been introduced, based on dual-energy breast exposure (about 26–33 kVp and 44–50 kVp) after contrast administration, so that the pre-contrast exposure was no longer needed [10, 12]. CESM allows for the visualisation of enhancing findings over the normal unenhancing breast tissue, exploiting the increased contrast uptake of malignancies [6, 10, 13].
Original studies have investigated the use of CESM in a number of settings, such as evaluation of symptomatic women [1417], screening recalls [1822], local staging [2332], pre- and post-operative evaluations [23, 24, 3336] and neoadjuvant chemotherapy response monitoring [3740]. In 2016, a first meta-analysis on CESM described a high pooled sensitivity (98%) albeit with a relatively low specificity (58%) [41], the latter partly caused by inexperience. A more recent meta-analysis [42] reported globally satisfying data for CESM-pooled sensitivity (89%) and specificity (84%), proposing it as an alternative to contrast-enhanced magnetic resonance imaging (MRI) and even suggesting CESM as a “useful triage test for initial breast lesions assessment” [41].
A time delay between the first appearance of new imaging techniques and their implementation in diagnostic routine is expected for many reasons, including not only the definition of indications but also the reproducibility of results. The latter is strongly influenced by technique details, such as contrast agent concentration, dose and injection rate, breast compression and positioning, exposure parameters and acquisition protocol. Indeed, the fact that CESM is variably performed across different centres, without an agreed and standardised technique, does not come as a surprise: this circumstance echoes the one observed for contrast-enhanced breast MRI in the 1990s, now settled by the publication of detailed international guidelines [4346].
Therefore, the aim of this work was to review CESM studies focusing on adopted technique, contrast agent issues and acquisition workflow. This effort is crucial for future CESM investigations to be reproducible and comparable.

Methods

Study protocol

No ethics committee approval was needed for this systematic review. The study protocol was registered on PROSPERO (protocol CRD42018118554), the international prospective register of systematic reviews [47]. This systematic review was reported according to the Preferred Reporting Items for Systematic Reviews and Meta-Analyses (PRISMA) statement [48].

Search strategy and eligibility criteria

In February 2019, a systematic search was performed on MEDLINE (PubMed, https://​www.​ncbi.​nlm.​nih.​gov/​pubmed/​), EMBASE (Elsevier), the Cochrane Library (Cochrane Database of Systematic Reviews) and the Cochrane Central Register of Controlled Trials for articles that reported or may have reported CESM technique. A controlled vocabulary (medical subject headings in PubMed and EMBASE thesaurus keywords in EMBASE) was used. The search string was (cesm OR ‘contrast enhanced spectral mammography’/exp. OR ‘dual energy mammography’ OR ‘contrast enhanced digital mammography’/exp. OR ‘contrast-enhanced mammography’ OR ‘dual-energy subtraction mammography’ OR cedm OR cedsm OR ‘contrast enhanced spectral imaging’ OR ‘high energy and low energy digital mammography’) AND (‘procedures’/exp. OR ‘method’ OR ‘methods’ OR ‘procedure’ OR ‘procedures’ OR ‘technique’ OR ‘acquisition’/exp. OR ‘contrast medium’/exp. OR ‘contrast agent’ OR ‘contrast dye’ OR ‘contrast material’ OR ‘contrast media’ OR ‘contrast medium’ OR ‘radiocontrast medium’ OR ‘radiography contrast medium’ OR ‘roentgen contrast medium’ OR ‘image processing’/exp. OR ‘image processing’ OR ‘image processing, computer-assisted’ OR ‘processing, image’).
The search was limited to original studies on humans published in English, French and Spanish on peer-reviewed journals, with an available abstract. No publication date limits were applied. First article screening was performed by two independent readers (A.C. and M.Z., with 1- and 3-year experience in breast imaging, respectively) considering only title and abstract. Eligible articles were those that reported in the title or in the abstract the use of CESM technique or that could have contained these data in the manuscript. After downloading eligible articles, the full text was read for a complete assessment. Finally, references of included articles were hand-searched to check for further eligible studies.

Data extraction

Data extraction was performed independently by the same two readers who performed the literature search. Disagreements were settled by consensus. For each analysed article, year of publication, institution (such as hospitals, imaging facilities, breast units including radiology sections or any other type of centre in which CESM is performed) and country origin as well as research groups, design, number of patients and demographics were retrieved. Mammography unit, vendor, radiation dose and technical features such as low- and high-energy peak kilovoltage (kVp), anode/filter combinations and exposure parameters were also extracted. Moreover, contrast agent type, dose and concentration were retrieved, as well as injection modality, if manual or automated, flow rate and additional post-contrast saline flush or “bolus chaser” if present. Furthermore, mild, moderate or severe adverse reactions to ICAs were extracted alongside strategies for their prevention. Regarding the acquisition protocol, time between contrast injection and first image acquisition and maximum examination duration were extracted. Regarding the order of views, we reported the acquisition sequence of the standard mammographic projections considering the craniocaudal (CC) and the mediolateral oblique (MLO) views, including the first side acquired. Missing data were requested to authors.

Evidence synthesis

To avoid risk of data duplication bias, in case of articles published by the same research group, we considered the possibility of performing subgroup analysis: therefore, before delving into further analysis of protocol description, we chose to change our viewpoint from the number of articles reporting a specific protocol to the minimum number of times a protocol was reported by a single research group.
Regarding the pooled rate of adverse reactions related to ICA administration across studies, statistical analysis was performed using Comprehensive Meta-Analysis v2.2.057 (Biostat, Englewood, NJ, USA) using the meta-analysis model “Number of events and study population”. I2 statistics was first calculated to assess heterogeneity and the fixed-effect model was used to provide the rate of adverse reactions and 95% of confidence intervals (CI). The risk of publication bias was assessed by visually inspecting funnel plot and performing the Egger test [49].

Results

Studies

A flowchart of study selection is shown in Fig. 1. Of 120 retrieved articles, 84 (70%), published between September 2003 and January 2019, were analysed [710, 1340, 50101]; 40/84 (48%) being retrospective and 44/84 (52%) prospective (43/44 monocentric (98%) and 1/44 multicentric (2%); 54/84 (64%) articles investigated CESM diagnostic performance, whereas 30/84 (36%) focused on technical features. The geographic distribution of research groups is depicted in Fig. 2.

Populations and settings

Data synthesis is reported in Table 1. The number of patients ranged from 5 [63] to 2303 [13], for a total of 14,012 patients, with mean or median age ranging from 45 years [40] to 66 years [23]. In 29/84 studies (35%), CESM was performed on patients from comprehensive databases of heterogeneous settings, such as pre- or post-operative evaluation, adjuvant or neoadjuvant chemotherapy response monitoring and equivocal findings at conventional imaging. The remaining 55 studies (65%) were individually centred on a unique setting. Twenty-seven studies (32%) performed CESM on suspicious cases from conventional imaging and screening recalls, 11 studies (13%) in a first-line screening setting, 7 (8%) performed CESM exclusively for known cancer staging, 4 (5%) in a pre-operative setting, 4 (5%) to assess and monitor the response to adjuvant chemotherapy and 2 (2%) in a post-operative setting.
Table 1
Main characteristics of the 84 analysed studies
Author/year
Ref.
Study design
Country of research group
Number of patients
Mean or median age (years)
Contrast agent type
Concentration (mgI/mL)
Dose (mL/kg)
Flow rate (mL/s)
Delay after injection (s)
Total exam time
Houben 2019
[22]
R
The Netherlands
147
61
Iopromide
300
1.5
3
120
 
Barra 2018
[40]
P mono
Brazil
33
45
Iohexol
300
1.5
3
120
B
Bicchierai 2018
[93]
R
Italy
40
50
Iopromide
370
1.5
3
120
B
Danala 2018
[69]
R
USA
111
 
Iohexol
350
1.5
3
120
B
Deng 2018
[78]
R
Taiwan
141
48
Iohexol
350
1.5
3
120
B
Helal 2018
[25]
P mono
Egypt
300
54
Iohexol
300
1.5
3
120
B
Kim 2018
[87]
P mono
South Korea
84
51
Iohexol
350
1.5
2
120
B
Klang 2018
[88]
R
Israel
953
51
Iopamidol
370
1.5
3
120
B
Łuczyńska 2018
[36]
R
Poland
82
57
Iopromide
370
1.5
3
120
B
Moustafa 2018
[17]
P mono
Egypt
160
 
Iohexol
300
1.5
3
120
B
Navarro 2018
[90]
P mono
Chile
465
53
Ioversol
320
1.5
  
B
Patel 2018 (01)
[38]
P mono
USA
65
53
Iohexol
350
1.5
3
120
A
Patel 2018 (02)
[34]
R
USA
50
57
Iohexol
350
1.5
3
120
B
Patel 2018 (03)
[23]
R
USA
30
66
Iohexol
350
1.5
3
120
B
Phillips 2018
[82]
R
USA
45
53
Iohexol
350
1.5
3
120
 
Sorin 2018
[92]
R
Israel
611
54
Iopamidol
370
1.5
3
120
B
Tohamey 2018
[51]
P mono
Egypt
178
46
Iohexol
300
1.5
3
120
B
Travieso-Aja 2018
[24]
R
Spain
158
51
  
1.5
3
120
B
Xing 2018
[84]
P mono
China
235
51
Iohexol
350
1.5
3
120
B
Barra 2017
[39]
R
Brazil
11
46
Iohexol
300
1–2
3
120
B
Bhimani 2017
[13]
R
USA
2303
 
Iopamidol
370
1.5
2
120
B
Fallenberg 2017
[76]
P multi
Germany
155
53
Iobitridol
300
1.5
3
120
A
Gluskin 2017
[63]
R
USA
5
59
Iohexol
350
1.5
3
150–180
A
Helal 2017 (01)
[28]
P mono
Egypt
98
50
Iohexol
300
1.5
3
120
B
Helal 2017 (02)
[99]
P mono
Egypt
30
47
Iohexol
300
1.5
 
120
 
Houben 2017
[58]
R
The Netherlands
839
60
Iopromide
300
1.5
3
120
 
Iotti 2017
[37]
P mono
Italy
54
54
Ioversol
350
1.5
 
120
 
James 2017
[56]
R
USA
173
 
Iohexol
350
1.5
3
120
A
Jochelson 2017
[54]
P mono
USA
309
51
Iohexol
350
1.5
3
150–180
B
Knogler 2017
[94]
P mono
Austria
11
58
Iomeprol
400
2
3.5
90
 
Lee-Felker 2017
[26]
R
USA
52
50
Iohexol
350
 
3
120
B
Lewis 2017
[16]
R
USA
208
 
Iohexol
350
1.5
3
120
B
Li 2017
[100]
R
USA
48
56
Iopamidol
370
1.5
1.5–2
 
B
Mori 2017
[74]
P mono
Japan
72
48
Iohexol
300
1.5
3
120
 
Patel 2017 (01)
[27]
R
USA
88
62
Iohexol
350
1.5
3
120
B
Patel 2017 (02)
[65]
R
USA
410
 
Iohexol
350
1.5
3
120
B
Phillips 2017
[70]
P mono
USA
38
53
Iohexol
350
1.5
3
120
B
Richter 2017
[62]
R
Germany
118
58
Iopromide
300
1.5
2–3
120
 
Saraya 2017
[18]
P mono
Egypt
34
54
Iohexol
300
1.5
4
 
C
Savaridas 2017
[75]
P mono
Australia
66
54
  
1.5
3
120
B
Sogani 2017
[80]
R
USA
278
51
Iohexol
350
1.5
3
150
A
Ali-Mucheru 2016
[33]
R
USA
351
62
Iohexol
350
1.5
3
120
B
Ambicka 2016
[29]
R
Poland
82
57
Iopromide
370
1.5
3
120
B
Brandan 2016
[77]
P mono
Mexico
18
51
Ioversol
300
 
4
60
B
Cheung 2016 (01)
[72]
R
Taiwan
256
48
Iohexol
350
1.5
3
120
A
Cheung 2016 (02)
[98]
R
Taiwan
87
54
Iohexol
350
1.5
3
120
B
Kamal 2016
[95]
R
Egypt
239
48
Iohexol
300
1.5
3
120
B
Kariyappa 2016
[68]
P mono
India
44
 
Iomeprol
350
1.5
3
120
B
Knogler 2016
[83]
P mono
Austria
15
58
Iomeprol
400
2
3.5
60–90
 
Lalji 2016
[21]
R
The Netherlands
199
58
Iopromide
300
1.5
3
120
 
Łuczyńska 2016 (01)
[50]
P mono
Poland
116
55
Iopromide
370
1.5
3
120
B
Łuczyńska 2016 (02)
[67]
P mono
Poland
193
55
Iopromide
370
1.5
3
120
B
Tardivel 2016
[19]
R
France
195
56
Iobitridol
300
1.5
3
120
B
Tennant 2016
[15]
R
UK
99
49
      
Tsigginou 2016
[89]
P mono
Greece
216
55
Iopromide
300
1.5
2–3
120
B
Wang 2016
[97]
P mono
China
68
53
Iohexol
350
1.5
3
120
A
Yagil 2016
[71]
R
Israel
200
51
Iopamidol
370
1.5
3
120
B
Chou 2015
[14]
P mono
Taiwan
185
51
Iohexol
300
1.5
2
120
B
Elsaid 2015
[73]
P mono
Egypt
34
55
Iohexol
300
1.5
3
 
B
Hobbs 2015
[81]
P mono
Australia
49
55
Iohexol
350
1.5
3
120
B
Kamal 2015
[79]
R
Egypt
168
 
Iohexol
300
1.5
3
120
B
Lobbes 2015
[30]
R
The Netherlands
87
62
Iopromide
300
1.5
3
120
 
Łuczyńska 2015 (01)
[91]
P mono
Poland
174
56
Iopromide
370
1.5
3
120
B
Łuczyńska 2015 (02)
[53]
P mono
Poland
102
 
Iopromide
370
1.5
3
120
 
Badr 2014
[101]
P mono
France
75
54
Iohexol
300
1.5
 
120
B
Blum 2014
[31]
P mono
Germany
20
57
Iopamidol
300
1.5
3
120
 
Cheung 2014
[86]
R
Taiwan
89
48
Iohexol
350
1.5
3
120–180
B
Fallenberg 2014 (01)
[85]
P mono
Germany
118
53
Iobitridol
300
1.5
3
120
B
Fallenberg 2014 (02)
[32]
P mono
Germany
80
54
Iobitridol
300
1.5
3
120
B
Francescone 2014
[66]
R
USA
88
50
      
Jeukens 2014
[60]
R
The Netherlands
47
58
Iopromide
300
1.5
3
120
 
Lobbes 2014
[20]
R
The Netherlands
113
57
Iopromide
300
1.5
3
120
 
Łuczyńska 2014
[35]
P mono
Poland
152
56
Iopromide
370
1.5
3
120
B
Mokhtar 2014
[57]
P mono
Egypt
60
 
Iohexol
300
1.5
 
120
A
Travieso-Aja 2014
[64]
R
Spain
136
49
  
1.5
3
120
B
Hill 2013
[10]
R
Canada
98
57
Iobitridol
300
1.5
3
120
B
Jochelson 2013
[55]
P mono
USA
82
50
Iohexol
350
1.5
3
150–300
B
Dromain 2012
[52]
P mono
France
110
57
Iobitridol
300
1.5
3
120
A
Diekmann 2011
[61]
P mono
Germany
70
55
Iopromide
370
1
4
60/120/180
A
Dromain 2011
[59]
P mono
France
120
56
Iobitridol
300
1.5
3
120
A
Dromain 2006
[9]
P mono
France
20
63
Iohexol
300
 
3
30
B
Diekmann 2005
[8]
P mono
Germany
21
 
Iopromide
370
1
4
60/120/180
A
Jong 2003
[7]
P mono
Canada
22
 
Iohexol
300
  
60
B
Lewin 2003
[96]
P mono
USA
26
51
Iohexol
350
 
4–5
150
 
R retrospective, P mono prospective monocentric, P multi prospective multicentric, A = total exam time < 5 min, B = total exam time between 5 and 10 min, C = total exam time > 10 min
Timing of CESM examination with menstrual cycle was reported only in 18/84 studies (21%). In 10/18 (56%) articles, it was mentioned but not applied; in 6/18 (33%), it was applied with a feasibility window between the 5th and 14th day of menstrual cycle; in 2/18 (11%), CESM was synchronously performed with MRI in different phases of menstrual cycle to evaluate and compare background parenchymal enhancement.

Technical features and parameters

In 70 out of 84 studies (83%), different systems from General Electric Healthcare (Chicago, IL, USA) were used, all with a prototype or a commercial release of the SenoBright upgrade which is required to perform dual-energy contrast-enhanced imaging. Twelve out of 84 articles (14%) reported the adoption of Selenia Dimensions Mammography Unit (Hologic Inc., Marlborough, MA, USA), while the remaining 2/84 (3%) studies were conducted with a Siemens Healthineers (Erlangen, Germany) Mammography System (Mammomat or Mammomat Inspiration).
The type of ICA used was not reported in five articles [15, 24, 64, 66, 75], while in the remaining 79 studies (94%), for a total of 13465 patients (96%), six different molecules were used: Iohexol was the most frequently employed, being used in 42/79 studies (53%) for a total of 5049/13465 patients (37%), followed by Iopromide (18/79 studies, 23%; 2798/13465 patients, 21%), while Iobitridol, Iomeprol, Iopamidol and Ioversol were administered in the remaining studies (19/79 studies, 24%; 5618/13465 patients, 42%). Iohexol was utilised at a concentration of 350 mg iodine/mL (25/42 studies, 60%; 3330/5049 patients, 66%) or 300 mg iodine/mL (17/42 studies, 40%; 1719/5049 patients, 34%). Iopromide was also administered at two different concentrations: 370 mg iodine/mL (10/18 studies, 56%; 1032/2798 patients, 37%) and 300 mg iodine/mL (8/18 studies, 44%; 1766/2798 patients, 63%).
Of the 69 studies including a specification of the contrast injection modality, 59 (85%) utilised an automated power injector (10584/11725 patients, 90%) while manual contrast injection was carried out in the remaining 10 (15%) [7, 9, 17, 25, 28, 51, 57, 73, 95, 99] for a total of 1141/11725 patients (10%).
Contrast agent dose, detailed in 77 studies, was fixed at 1.5 mL/kg in 72 (93%) of them for a total of 13559/13687 (99%) patients. Contrast agent flow rate, reported in 76/84 studies (90%), was most frequently fixed at 3 mL/s (65/76 studies, 86%); the 11 remaining articles detailed a flow rate ranging from 2 to 5 mL/s. Thirty-five out of 84 (42%) articles for a total 8734/14012 patients (62%) also mentioned the use of additional post-contrast saline flush or “bolus chaser,” 19 of them (54%, for a total 4477/8734 patients, 51%) likewise detailing a saline amount ranging from 10 to 30 mL.
Of 69 studies detailing the tube voltage of both low- and high-energy acquisitions, all but one (99%) acquired low-energy images between 26 and 33.2 kVp, which is the peak kilovoltage threshold of iodine, while all 69 acquired high-energy images well above this threshold, i.e. between 44 and 50 kVp. The anode/filter combination was reported by 42/84 studies. Exposure parameters were unambiguously reported only in one study [10], whereas in 5 early studies [7, 8, 32, 59, 85], they were manually adjusted according to breast thickness and density; thirty-five other studies declared an automatic regulation of these parameters performed by the mammography unit.
Regarding radiation dose, data were scarcer: even though 45/84 articles (54%) mentioned this aspect, 17/45 (31%) did it without exhibiting original information but reporting observations from previous studies, therefore restricting the number of studies with new data to 28/84 (33%). Of these 28 studies, 19 (68%) provided an average glandular dose (AGD), 3 (16%) of them calculating it per-patient and ranging 1.5–6.9 mGy [8, 9, 58], 5/19 (26%) calculating it per-breast ranging 2.19–7.15 mGy and the remaining 11 (58%) reporting a per-view AGD ranging from 0.43 [61] to 2.65 mGy [101]. A comparison with DM was mentioned in 17 studies: only 1 (6%) documented a dose reduction (− 2%) for CESM compared to DM [32], while other 16 (94%) reported an increase in AGD ranging between 6.2% [85] and 100% [77]. However, it is worth to notice that 3 studies specifically contrived to assess CESM radiation doses reported an AGD increase of 42% [56], 78% [82] and 80% [60].

Acquisition protocols

Studies reporting the time interval between contrast injection and the first image acquisition were 78 out of 84 (93%), for a total 13244/14012 patients (95%) and 65 (83%) of them (12278/13244 patients, 93%) had it fixed at 120 s.
Sixty-six out of 84 articles (79%, 11900/14012 patients, 85%) gave an indication of the acquisition time after contrast injection: in 12/66 (18%, 1381/11900 patients, 11.6%), the exam was completed in less than 5 min; in 52/66 (80%, for total of 10485/11900 patients, 88.1%) between 5 and 10 min, while in 1/66 (2%, 34/11900 patients, 0.3%) the duration exceeded 10 min.
The outline of the image acquisition sequence remains more variable. Ten out of 84 studies (12%), accounting for 2734 patients (19%) did not clearly describe it and did not provide a reference to other protocols, while 3/84 (4%, 103/14012 patients, 1%) employed a curtailed and side-insensitive acquisition sequence. Adherence to standard but unspecified digital mammography protocols was declared by 29/84 (34%) studies, for total 3741/14012 patients (27%). The other half of the articles analysed (42/84, accounting for 7434/14012 patients, 53%) unequivocally detailed an acquisition sequence. Of these 42 studies, 14 (34%, 2048/7434 patients, 28%) adopted a projection order that was conventionally agreed upon, while the other 28 (66%, accounting for 5386/7434 patients, 72%) based their acquisition sequence on the presence of previous suspect or clearly pathologic findings.
Eighty-four articles came from 38 different research groups. Subgroup analysis according to research groups showed that 17 acquisition sequences based on a conventionally agreed projection order were executed in 15 research groups. As described in Fig. 3, the most common sequence description, reported by 6/17 (35%) institutions, was MLO - MLO - CC - CC (in order of acquisition), without any further indication about the first side to be examined (right or left or side with/without suspicious lesion or already diagnosed cancer). The second most common sequence (4/17, 24%) was CC - CC - MLO - MLO with the first projection standardised on the right side (independently of pathology or with suspected pathology).
Among the 22 acquisition sequences (coming from 20 institutions) centred on the presence of previous suspect or clearly pathologic findings, we found substantial variability between different orders of acquisition, as shown in Fig. 4. However, the most common sequence, adopted by 4/22 (19%) research groups, was 1) CC, suspected side; 2) CC, non-suspected side; 3) MLO, suspected side; and 4) MLO, non-suspected side.

Contrast agent adverse reaction rate meta-analysis

Regarding side effects from ICA administration, 48/84 studies (57%) declared a preventive anamnestic screening for previous adverse reactions or general contraindications to ICA administration. Pre-examination tests of renal function was mentioned in 39/84 studies (46%). Of note, 14/84 studies (29%) reported 30 adverse reactions out of 14012 patients, of which 26/30 (87%) were mild reactions limited to pruritus, hives, “scratchy throat” or other minor skin flushing that resolved promptly even when antihistamines or corticosteroids were not administered. In 3/30 (10%) cases [54, 58, 87], side effects were of moderate importance with nausea and vomiting, widespread urticaria resolved only after antihistamines and corticosteroids per os, and dyspnea that equally responded to oral antihistamine administration. Only 1/30 (3%) severe adverse reaction, requiring “intensive care” but resolved after short time, occurred in 14012 patients (0.007%) [61].
Therefore, the number of adverse reactions related to ICA administration ranged from 0, reported by 70 (88%) studies, to a maximum of 6 adverse reactions [14] with a total of 30 adverse reactions, showing no heterogeneity (Q = 64, degree of freedom 83, τ = 2.0972, I2 = 0%, p = 0.931). As shown in the forest plot of Fig. 5, using fixed-effect model, the pooled rate of adverse reactions across studies was 0.82%, with 0.64% and 1.05% as 95% CI.
Visually inspecting the funnel plot in Fig. 6, risk of publication bias was found, as confirmed by the Egger test (p = 0.00028).

Discussion

Our systematic review included 84 articles, accounting for 14012 patients, reporting the use of CESM in various settings. The sheer number of studies and, as depicted in Fig. 7, their increase in the last 3 years (27 studies between 2003 and December 2015, 57 from January 2016 to January 2019) points out a considerable interest in this emerging breast imaging modality.
A number of narrative reviews [6, 42, 102106] favourably outlined CESM future perspectives in several clinical settings (e.g. recall work-up, pre-operative staging, and monitoring the effect of neoadjuvant therapy) as a potential alternative to MRI.
In the first phase of CESM development, some non-fixed parameters regarding contrast agent administration (i.e. contrast agent molecule, concentration, dose, flow rate, and injection modality) and some acquisition features (i.e. time between contrast injection and first acquisition, kVp ranges for low- and high-energy acquisitions) gained an international agreement. However, in the framework of comprehensive optimisation and standardisation of CESM, large-scale studies are undoubtedly needed to address the knowledge gap concerning the choice of technical parameters.
Our data show a consensus among studies (93%) on the choice of 1.5 mL/kg contrast dose administered with a 3 mL/s flow rate (74%) and a less extensive agreement on the use of Iohexol (53% of all studies) at a concentration of 350 mg iodine/mL (30% of all studies). However, these parameters have probably been empirically adopted from CT protocols, as the first investigators plainly stated [7], without any other particular explication or justification. No dose-finding studies have been published yet.
Similarly, the common use of a power contrast injector (87% of all studies, with the remaining 13% coming from a single research group) is assumed from CT and MRI protocols in which it has been demonstrated to be effective in obtaining a stable contrast inflow and bolus shape [107109]. Moreover, the use of a power injector allows for the administration of a bolus chaser, reported only in 42% of all articles, a technical refinement that has shown good results in CT [110, 111].
Two other points need to be mentioned. The first one is the correlation between menstrual cycle phase and background parenchymal enhancement, explored in a few studies [10, 75, 80] and/or fluctuations of lesion contrast uptake. Secondly, since CESM is based on a dual X-ray exposure, of which the low-energy one has been demonstrated to be equal to standard DM [66], an increase in radiation dose is expected. However, while preliminary studies estimated a negligible [7] or curtailed AGD increase, studies specifically devised to ascertain CESM effective AGD found a substantial AGD increment ranging 42–80% [56, 60, 82]. While CESM AGDs remain under the threshold stated by European guidelines for screening mammography [112], further studies are needed to investigate CESM AGD [56, 82].
Furthermore, we remark the absence of standardised protocols. This methodological void, especially regarding the acquisition workflow, represents a threat to reproducibility and comparison of imaging results. While 98% of all studies reporting the total examination time completed the examination before 10 min from contrast administration, and while some studies presented evidence on the irrelevance of the acquisition order [55, 64], there are no studies comparing different approaches.
The pooled rate of adverse reactions to ICA administration was 0.82% (0.64–1.05% 95% CI) with a total of 30 adverse reactions in 14012 patients, a rate similar to that reported for CT 0.6% [113] in 84928 adult patients or 0.7% [114] in 29508 patients (given Iopromide, which is also used for CESM). Particularly, considering only severe adverse reactions in CT, Wang et al. [113] reported 11/84928 (0.0129%) reactions, as well as Mortelé et al. [114] 4/29508 (0.0135%). These rates seem to be higher than that found in our meta-analysis 1/14012 (0.007%), a comparison to consider with caution due to the nature of rare events such as severe reactions to ICA. One aspect to consider is the different profile of patients undergoing CESM compared to those requiring contrast-enhanced CT, the former being that of basically “healthy” subjects, the latter implying the possibility of relevant disease, including also serious emergency conditions.
This review has limitations. Patient data are probably shared and duplicate among some studies from the same research group. This has been shown to negatively impact on review quality [115, 116] and could only be prevented via individual patient data sharing [117]. However, for technical aspects of this systematic review, our choice to evaluate study groups rather than single articles should have mitigated this bias. Conversely, our pooled rate of adverse reactions could be underestimated.
In conclusion, our review shows that CESM is unevenly performed across different centres, in terms of contrast agent type and concentration and order of view acquisition. However, most research groups performed CESM using a contrast dose of 1.5 mL/kg, factory-set kVp ranges for low- and high-energy acquisitions, beginning image acquisition after 120 s from contrast agent injection and completing the examination within 10 min. Further studies are needed to investigate the role of background parenchymal enhancement and to harvest data that can firmly back up subsequent technical guidelines and consensus statements for standardised CESM protocols.
Not applicable
Not applicable

Competing interests

FS declares to have received grants from or to be member of speakers’ bureau/advisory board for Bayer, Bracco, and General Electric. All other authors declare that they have no competing interests.
Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://​creativecommons.​org/​licenses/​by/​4.​0/​), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Literatur
1.
Zurück zum Zitat Frigerio A, Sardanelli F, Podo F (2017) Radiological screening of breast cancer: evolution. In: Veronesi U, Goldhirsch A, Veronesi P, Gentilini OD, Leonardi MC (eds) Breast Cancer. Springer International Publishing, Cham, pp 171–203CrossRef Frigerio A, Sardanelli F, Podo F (2017) Radiological screening of breast cancer: evolution. In: Veronesi U, Goldhirsch A, Veronesi P, Gentilini OD, Leonardi MC (eds) Breast Cancer. Springer International Publishing, Cham, pp 171–203CrossRef
2.
Zurück zum Zitat Pisano ED, Gatsonis C, Hendrick E et al (2005) Diagnostic performance of digital versus film mammography for breast-cancer screening. N Engl J Med 353:1773–1783PubMedCrossRef Pisano ED, Gatsonis C, Hendrick E et al (2005) Diagnostic performance of digital versus film mammography for breast-cancer screening. N Engl J Med 353:1773–1783PubMedCrossRef
3.
Zurück zum Zitat Gilbert FJ, Tucker L, Young KC (2016) Digital breast tomosynthesis (DBT): a review of the evidence for use as a screening tool. Clin Radiol 71:141–150PubMedCrossRef Gilbert FJ, Tucker L, Young KC (2016) Digital breast tomosynthesis (DBT): a review of the evidence for use as a screening tool. Clin Radiol 71:141–150PubMedCrossRef
4.
Zurück zum Zitat Pattacini P, Nitrosi A, Giorgi Rossi P et al (2018) Digital mammography versus digital mammography plus tomosynthesis for breast cancer screening: the Reggio Emilia Tomosynthesis randomized trial. Radiology 288:375–385PubMedCrossRef Pattacini P, Nitrosi A, Giorgi Rossi P et al (2018) Digital mammography versus digital mammography plus tomosynthesis for breast cancer screening: the Reggio Emilia Tomosynthesis randomized trial. Radiology 288:375–385PubMedCrossRef
5.
Zurück zum Zitat Marinovich ML, Hunter KE, Macaskill P, Houssami N (2018) Breast cancer screening using tomosynthesis or mammography: a meta-analysis of cancer detection and recall. J Natl Cancer Inst 110:942–949PubMedCrossRef Marinovich ML, Hunter KE, Macaskill P, Houssami N (2018) Breast cancer screening using tomosynthesis or mammography: a meta-analysis of cancer detection and recall. J Natl Cancer Inst 110:942–949PubMedCrossRef
6.
Zurück zum Zitat Patel BK, Lobbes MBI, Lewin J (2018) Contrast enhanced spectral mammography: a review. Semin Ultrasound CT MRI 39:70–79PubMedCrossRef Patel BK, Lobbes MBI, Lewin J (2018) Contrast enhanced spectral mammography: a review. Semin Ultrasound CT MRI 39:70–79PubMedCrossRef
7.
Zurück zum Zitat Jong RA, Yaffe MJ, Skarpathiotakis M et al (2003) Contrast-enhanced digital mammography: initial clinical experience. Radiology 228:842–850PubMedCrossRef Jong RA, Yaffe MJ, Skarpathiotakis M et al (2003) Contrast-enhanced digital mammography: initial clinical experience. Radiology 228:842–850PubMedCrossRef
8.
Zurück zum Zitat Diekmann F, Diekmann S, Jeunehomme F, Muller S, Hamm B, Bick U (2005) Digital mammography using iodine-based contrast media. Invest Radiol 40:397–404PubMedCrossRef Diekmann F, Diekmann S, Jeunehomme F, Muller S, Hamm B, Bick U (2005) Digital mammography using iodine-based contrast media. Invest Radiol 40:397–404PubMedCrossRef
9.
Zurück zum Zitat Dromain C, Balleyguier C, Muller S et al (2006) Evaluation of tumor angiogenesis of breast carcinoma using contrast-enhanced digital mammography. AJR Am J Roentgenol 187:W528–W537PubMedCrossRef Dromain C, Balleyguier C, Muller S et al (2006) Evaluation of tumor angiogenesis of breast carcinoma using contrast-enhanced digital mammography. AJR Am J Roentgenol 187:W528–W537PubMedCrossRef
10.
Zurück zum Zitat Hill ML, Mainprize JG, Carton A-K et al (2013) Anatomical noise in contrast-enhanced digital mammography. Part II. Dual-energy imaging. Med Phys 40:081907PubMedCrossRef Hill ML, Mainprize JG, Carton A-K et al (2013) Anatomical noise in contrast-enhanced digital mammography. Part II. Dual-energy imaging. Med Phys 40:081907PubMedCrossRef
11.
Zurück zum Zitat Dromain C, Balleyguier C, Adler G, Garbay JR, Delaloge S (2009) Contrast-enhanced digital mammography. Eur J Radiol 69:34–42PubMedCrossRef Dromain C, Balleyguier C, Adler G, Garbay JR, Delaloge S (2009) Contrast-enhanced digital mammography. Eur J Radiol 69:34–42PubMedCrossRef
12.
Zurück zum Zitat Skarpathiotakis M, Yaffe MJ, Bloomquist AK et al (2002) Development of contrast digital mammography. Med Phys 29:2419–2426PubMedCrossRef Skarpathiotakis M, Yaffe MJ, Bloomquist AK et al (2002) Development of contrast digital mammography. Med Phys 29:2419–2426PubMedCrossRef
13.
Zurück zum Zitat Bhimani C, Matta D, Roth RG et al (2017) Contrast-enhanced spectral mammography. Acad Radiol 24:84–88PubMedCrossRef Bhimani C, Matta D, Roth RG et al (2017) Contrast-enhanced spectral mammography. Acad Radiol 24:84–88PubMedCrossRef
14.
Zurück zum Zitat Chou C-P, Lewin JM, Chiang C-L et al (2015) Clinical evaluation of contrast-enhanced digital mammography and contrast enhanced tomosynthesis—comparison to contrast-enhanced breast MRI. Eur J Radiol 84:2501–2508PubMedCrossRef Chou C-P, Lewin JM, Chiang C-L et al (2015) Clinical evaluation of contrast-enhanced digital mammography and contrast enhanced tomosynthesis—comparison to contrast-enhanced breast MRI. Eur J Radiol 84:2501–2508PubMedCrossRef
15.
Zurück zum Zitat Tennant SL, James JJ, Cornford EJ et al (2016) Contrast-enhanced spectral mammography improves diagnostic accuracy in the symptomatic setting. Clin Radiol 71:1148–1155PubMedCrossRef Tennant SL, James JJ, Cornford EJ et al (2016) Contrast-enhanced spectral mammography improves diagnostic accuracy in the symptomatic setting. Clin Radiol 71:1148–1155PubMedCrossRef
16.
Zurück zum Zitat Lewis TC, Pizzitola VJ, Giurescu ME et al (2017) Contrast-enhanced digital mammography: a single-institution experience of the first 208 cases. Breast J 23:67–76PubMedCrossRef Lewis TC, Pizzitola VJ, Giurescu ME et al (2017) Contrast-enhanced digital mammography: a single-institution experience of the first 208 cases. Breast J 23:67–76PubMedCrossRef
17.
Zurück zum Zitat Moustafa AFI, Kamal EF, Hassan MM, Sakr M, Gomaa MMM (2018) The added value of contrast enhanced spectral mammography in identification of multiplicity of suspicious lesions in dense breast. Egypt J Radiol Nucl Med 49:259–264CrossRef Moustafa AFI, Kamal EF, Hassan MM, Sakr M, Gomaa MMM (2018) The added value of contrast enhanced spectral mammography in identification of multiplicity of suspicious lesions in dense breast. Egypt J Radiol Nucl Med 49:259–264CrossRef
18.
Zurück zum Zitat Saraya S, Adel L, Mahmoud A (2017) Indeterminate breast lesions: can contrast enhanced digital mammography change our decisions? Egypt J Radiol Nucl Med 48:547–552CrossRef Saraya S, Adel L, Mahmoud A (2017) Indeterminate breast lesions: can contrast enhanced digital mammography change our decisions? Egypt J Radiol Nucl Med 48:547–552CrossRef
19.
Zurück zum Zitat Tardivel A-M, Balleyguier C, Dunant A et al (2016) Added value of contrast-enhanced spectral mammography in postscreening assessment. Breast J 22:520–528PubMedCrossRef Tardivel A-M, Balleyguier C, Dunant A et al (2016) Added value of contrast-enhanced spectral mammography in postscreening assessment. Breast J 22:520–528PubMedCrossRef
20.
Zurück zum Zitat Lobbes MB, Lalji U, Houwers J et al (2014) Contrast-enhanced spectral mammography in patients referred from the breast cancer screening programme. Eur Radiol 24:1668–1676 Lobbes MB, Lalji U, Houwers J et al (2014) Contrast-enhanced spectral mammography in patients referred from the breast cancer screening programme. Eur Radiol 24:1668–1676
21.
Zurück zum Zitat Lalji UC, Houben IP, Prevos R et al (2016) Contrast-enhanced spectral mammography in recalls from the Dutch breast cancer screening program: validation of results in a large multireader, multicase study. Eur Radiol 26:4371–4379PubMedPubMedCentralCrossRef Lalji UC, Houben IP, Prevos R et al (2016) Contrast-enhanced spectral mammography in recalls from the Dutch breast cancer screening program: validation of results in a large multireader, multicase study. Eur Radiol 26:4371–4379PubMedPubMedCentralCrossRef
22.
Zurück zum Zitat Houben IP, Vanwetswinkel S, Kalia V et al (2019) Contrast-enhanced spectral mammography in the evaluation of breast suspicious calcifications: diagnostic accuracy and impact on surgical management. Acta Radiol. [Epub ahead of print] Houben IP, Vanwetswinkel S, Kalia V et al (2019) Contrast-enhanced spectral mammography in the evaluation of breast suspicious calcifications: diagnostic accuracy and impact on surgical management. Acta Radiol. [Epub ahead of print]
23.
Zurück zum Zitat Patel BK, Davis J, Ferraro C et al (2018) Value added of preoperative contrast-enhanced digital mammography in patients with invasive lobular carcinoma of the breast. Clin Breast Cancer 18:e1339–e1345PubMedCrossRef Patel BK, Davis J, Ferraro C et al (2018) Value added of preoperative contrast-enhanced digital mammography in patients with invasive lobular carcinoma of the breast. Clin Breast Cancer 18:e1339–e1345PubMedCrossRef
24.
Zurück zum Zitat Travieso-Aja MDM, Naranjo-Santana P, Fernández-Ruiz C et al (2018) Factors affecting the precision of lesion sizing with contrast-enhanced spectral mammography. Clin Radiol 73:296–303PubMedCrossRef Travieso-Aja MDM, Naranjo-Santana P, Fernández-Ruiz C et al (2018) Factors affecting the precision of lesion sizing with contrast-enhanced spectral mammography. Clin Radiol 73:296–303PubMedCrossRef
25.
Zurück zum Zitat Helal MH, Mansour SM, Salaleldin LA, Alkalaawy BM, Salem DS, Mokhtar NM (2018) The impact of contrast-enhanced spectral mammogram (CESM) and three-dimensional breast ultrasound (3DUS) on the characterization of the disease extend in cancer patients. Br J Radiol 91:20170977 Helal MH, Mansour SM, Salaleldin LA, Alkalaawy BM, Salem DS, Mokhtar NM (2018) The impact of contrast-enhanced spectral mammogram (CESM) and three-dimensional breast ultrasound (3DUS) on the characterization of the disease extend in cancer patients. Br J Radiol 91:20170977
26.
Zurück zum Zitat Lee-Felker SA, Tekchandani L, Thomas M et al (2017) Newly diagnosed breast cancer: comparison of contrast-enhanced spectral mammography and breast MR imaging in the evaluation of extent of disease. Radiology 285:389–400PubMedCrossRef Lee-Felker SA, Tekchandani L, Thomas M et al (2017) Newly diagnosed breast cancer: comparison of contrast-enhanced spectral mammography and breast MR imaging in the evaluation of extent of disease. Radiology 285:389–400PubMedCrossRef
27.
Zurück zum Zitat Patel BK, Garza SA, Eversman S, Lopez-Alvarez Y, Kosiorek H, Pockaj BA (2017) Assessing tumor extent on contrast-enhanced spectral mammography versus full-field digital mammography and ultrasound. Clin Imaging 46:78–84PubMedCrossRef Patel BK, Garza SA, Eversman S, Lopez-Alvarez Y, Kosiorek H, Pockaj BA (2017) Assessing tumor extent on contrast-enhanced spectral mammography versus full-field digital mammography and ultrasound. Clin Imaging 46:78–84PubMedCrossRef
28.
Zurück zum Zitat Helal MH, Mansour SM, Zaglol M, Salaleldin LA, Nada OM, Haggag MA (2017) Staging of breast cancer and the advanced applications of digital mammogram: what the physician needs to know? Br J Radiol 90:20160717PubMedPubMedCentralCrossRef Helal MH, Mansour SM, Zaglol M, Salaleldin LA, Nada OM, Haggag MA (2017) Staging of breast cancer and the advanced applications of digital mammogram: what the physician needs to know? Br J Radiol 90:20160717PubMedPubMedCentralCrossRef
29.
Zurück zum Zitat Ambicka A, Luczynska E, Adamczyk A, Harazin-Lechowska A, Sas-Korczynska B, Niemiec J (2016) The tumour border on contrast-enhanced spectral mammography and its relation to histological characteristics of invasive breast cancer. Pol J Pathol 3:295–299CrossRef Ambicka A, Luczynska E, Adamczyk A, Harazin-Lechowska A, Sas-Korczynska B, Niemiec J (2016) The tumour border on contrast-enhanced spectral mammography and its relation to histological characteristics of invasive breast cancer. Pol J Pathol 3:295–299CrossRef
30.
Zurück zum Zitat Lobbes MB, Lalji UC, Nelemans PJ et al (2015) The quality of tumor size assessment by contrast-enhanced spectral mammography and the benefit of additional breast MRI. J Cancer 6:144–150PubMedPubMedCentralCrossRef Lobbes MB, Lalji UC, Nelemans PJ et al (2015) The quality of tumor size assessment by contrast-enhanced spectral mammography and the benefit of additional breast MRI. J Cancer 6:144–150PubMedPubMedCentralCrossRef
31.
Zurück zum Zitat Blum KS, Rubbert C, Mathys B, Antoch G, Mohrmann S, Obenauer S (2014) Use of contrast-enhanced spectral mammography for intramammary cancer staging. Acad Radiol 21:1363–1369PubMedCrossRef Blum KS, Rubbert C, Mathys B, Antoch G, Mohrmann S, Obenauer S (2014) Use of contrast-enhanced spectral mammography for intramammary cancer staging. Acad Radiol 21:1363–1369PubMedCrossRef
32.
Zurück zum Zitat Fallenberg EM, Dromain C, Diekmann F et al (2014) Contrast-enhanced spectral mammography versus MRI: initial results in the detection of breast cancer and assessment of tumour size. Eur Radiol 24:256–264PubMedCrossRef Fallenberg EM, Dromain C, Diekmann F et al (2014) Contrast-enhanced spectral mammography versus MRI: initial results in the detection of breast cancer and assessment of tumour size. Eur Radiol 24:256–264PubMedCrossRef
33.
Zurück zum Zitat Ali-Mucheru M, Pockaj B, Patel B et al (2016) Contrast-enhanced digital mammography in the surgical management of breast cancer. Ann Surg Oncol 23:649–655PubMedCrossRef Ali-Mucheru M, Pockaj B, Patel B et al (2016) Contrast-enhanced digital mammography in the surgical management of breast cancer. Ann Surg Oncol 23:649–655PubMedCrossRef
34.
Zurück zum Zitat Patel BK, Ranjbar S, Wu T et al (2018) Computer-aided diagnosis of contrast-enhanced spectral mammography: a feasibility study. Eur J Radiol 98:207–213PubMedCrossRef Patel BK, Ranjbar S, Wu T et al (2018) Computer-aided diagnosis of contrast-enhanced spectral mammography: a feasibility study. Eur J Radiol 98:207–213PubMedCrossRef
35.
Zurück zum Zitat Luczyńska E, Heinze-Paluchowska S, Dyczek S, Blecharz P, Rys J, Reinfuss M (2014) Contrast-enhanced spectral mammography: comparison with conventional mammography and histopathology in 152 women. Korean J Radiol 15:689PubMedPubMedCentralCrossRef Luczyńska E, Heinze-Paluchowska S, Dyczek S, Blecharz P, Rys J, Reinfuss M (2014) Contrast-enhanced spectral mammography: comparison with conventional mammography and histopathology in 152 women. Korean J Radiol 15:689PubMedPubMedCentralCrossRef
36.
Zurück zum Zitat Luczynska E, Niemiec J, Heinze S et al (2018) Intensity and pattern of enhancement on CESM: prognostic significance and its relation to expression of podoplanin in tumor stroma - a preliminary report. Anticancer Res 38:1085–1095PubMed Luczynska E, Niemiec J, Heinze S et al (2018) Intensity and pattern of enhancement on CESM: prognostic significance and its relation to expression of podoplanin in tumor stroma - a preliminary report. Anticancer Res 38:1085–1095PubMed
37.
Zurück zum Zitat Iotti V, Ravaioli S, Vacondio R et al (2017) Contrast-enhanced spectral mammography in neoadjuvant chemotherapy monitoring: a comparison with breast magnetic resonance imaging. Breast Cancer Res 19:106PubMedPubMedCentralCrossRef Iotti V, Ravaioli S, Vacondio R et al (2017) Contrast-enhanced spectral mammography in neoadjuvant chemotherapy monitoring: a comparison with breast magnetic resonance imaging. Breast Cancer Res 19:106PubMedPubMedCentralCrossRef
38.
Zurück zum Zitat Patel BK, Hilal T, Covington M et al (2018) Contrast-enhanced spectral mammography is comparable to MRI in the assessment of residual breast cancer following neoadjuvant systemic therapy. Ann Surg Oncol 25:1350–1356PubMedCrossRef Patel BK, Hilal T, Covington M et al (2018) Contrast-enhanced spectral mammography is comparable to MRI in the assessment of residual breast cancer following neoadjuvant systemic therapy. Ann Surg Oncol 25:1350–1356PubMedCrossRef
39.
Zurück zum Zitat Barra FR, de Souza FF, Camelo REFA, Ribeiro ACO, Farage L (2017) Accuracy of contrast-enhanced spectral mammography for estimating residual tumor size after neoadjuvant chemotherapy in patients with breast cancer: a feasibility study. Radiol Bras 50:224–230PubMedPubMedCentralCrossRef Barra FR, de Souza FF, Camelo REFA, Ribeiro ACO, Farage L (2017) Accuracy of contrast-enhanced spectral mammography for estimating residual tumor size after neoadjuvant chemotherapy in patients with breast cancer: a feasibility study. Radiol Bras 50:224–230PubMedPubMedCentralCrossRef
40.
Zurück zum Zitat Barra FR, Sobrinho AB, Barra RR et al (2018) Contrast-enhanced mammography (CEM) for detecting residual disease after neoadjuvant chemotherapy: a comparison with breast magnetic resonance imaging (MRI). Biomed Res Int 2018:1–9CrossRef Barra FR, Sobrinho AB, Barra RR et al (2018) Contrast-enhanced mammography (CEM) for detecting residual disease after neoadjuvant chemotherapy: a comparison with breast magnetic resonance imaging (MRI). Biomed Res Int 2018:1–9CrossRef
41.
Zurück zum Zitat Tagliafico AS, Bignotti B, Rossi F et al (2016) Diagnostic performance of contrast-enhanced spectral mammography: systematic review and meta-analysis. Breast 28:13–19PubMedCrossRef Tagliafico AS, Bignotti B, Rossi F et al (2016) Diagnostic performance of contrast-enhanced spectral mammography: systematic review and meta-analysis. Breast 28:13–19PubMedCrossRef
42.
Zurück zum Zitat Zhu X, Huang J-M, Zhang K et al (2018) Diagnostic value of contrast-enhanced spectral mammography for screening breast cancer: systematic review and meta-analysis. Clin Breast Cancer 18:e985–e995PubMedCrossRef Zhu X, Huang J-M, Zhang K et al (2018) Diagnostic value of contrast-enhanced spectral mammography for screening breast cancer: systematic review and meta-analysis. Clin Breast Cancer 18:e985–e995PubMedCrossRef
43.
Zurück zum Zitat Sardanelli F, Boetes C, Borisch B et al (2010) Magnetic resonance imaging of the breast: recommendations from the EUSOMA working group. Eur J Cancer 46:1296–1316CrossRefPubMed Sardanelli F, Boetes C, Borisch B et al (2010) Magnetic resonance imaging of the breast: recommendations from the EUSOMA working group. Eur J Cancer 46:1296–1316CrossRefPubMed
44.
48.
Zurück zum Zitat Moher D, Liberati A, Tetzlaff J, Altman DG (2009) Preferred Reporting Items for Systematic Reviews and Meta-Analyses: the PRISMA statement. PLoS Med 6:e1000097PubMedPubMedCentralCrossRef Moher D, Liberati A, Tetzlaff J, Altman DG (2009) Preferred Reporting Items for Systematic Reviews and Meta-Analyses: the PRISMA statement. PLoS Med 6:e1000097PubMedPubMedCentralCrossRef
50.
Zurück zum Zitat Luczyńska E, Heinze S, Adamczyk A, Rys J, Mitus JW, Hendrick E (2016) Comparison of the mammography, contrast-enhanced spectral mammography and ultrasonography in a group of 116 patients. Anticancer Res 36:4359–4366PubMed Luczyńska E, Heinze S, Adamczyk A, Rys J, Mitus JW, Hendrick E (2016) Comparison of the mammography, contrast-enhanced spectral mammography and ultrasonography in a group of 116 patients. Anticancer Res 36:4359–4366PubMed
51.
Zurück zum Zitat Tohamey YM, Youssry SW, Abd El Aziz AI (2018) Interpretation of patterns of enhancement on contrast-enhanced spectral mammography: an approach to a standardized scheme. Egypt J Radiol Nucl Med 49:854–868CrossRef Tohamey YM, Youssry SW, Abd El Aziz AI (2018) Interpretation of patterns of enhancement on contrast-enhanced spectral mammography: an approach to a standardized scheme. Egypt J Radiol Nucl Med 49:854–868CrossRef
52.
Zurück zum Zitat Dromain C, Thibault F, Diekmann F et al (2012) Dual-energy contrast-enhanced digital mammography: initial clinical results of a multireader, multicase study. Breast Cancer Res 14:R94PubMedPubMedCentralCrossRef Dromain C, Thibault F, Diekmann F et al (2012) Dual-energy contrast-enhanced digital mammography: initial clinical results of a multireader, multicase study. Breast Cancer Res 14:R94PubMedPubMedCentralCrossRef
53.
Zurück zum Zitat Łuczyńska E, Heinze-Paluchowska S, Hendrick E et al (2015) Comparison between breast MRI and contrast-enhanced spectral mammography. Med Sci Monit 21:1358–1367PubMedPubMedCentralCrossRef Łuczyńska E, Heinze-Paluchowska S, Hendrick E et al (2015) Comparison between breast MRI and contrast-enhanced spectral mammography. Med Sci Monit 21:1358–1367PubMedPubMedCentralCrossRef
54.
Zurück zum Zitat Jochelson MS, Pinker K, Dershaw DD et al (2017) Comparison of screening CEDM and MRI for women at increased risk for breast cancer: a pilot study. Eur J Radiol 97:37–43PubMedCrossRef Jochelson MS, Pinker K, Dershaw DD et al (2017) Comparison of screening CEDM and MRI for women at increased risk for breast cancer: a pilot study. Eur J Radiol 97:37–43PubMedCrossRef
55.
Zurück zum Zitat Jochelson MS, Dershaw DD, Sung JS et al (2013) Bilateral contrast-enhanced dual-energy digital mammography: feasibility and comparison with conventional digital mammography and MR imaging in women with known breast carcinoma. Radiology 266:743–751PubMedCrossRef Jochelson MS, Dershaw DD, Sung JS et al (2013) Bilateral contrast-enhanced dual-energy digital mammography: feasibility and comparison with conventional digital mammography and MR imaging in women with known breast carcinoma. Radiology 266:743–751PubMedCrossRef
56.
Zurück zum Zitat James JR, Pavlicek W, Hanson JA, Boltz TF, Patel BK (2017) Breast radiation dose with CESM compared with 2D FFDM and 3D tomosynthesis mammography. AJR Am J Roentgenol 208:362–372PubMedCrossRef James JR, Pavlicek W, Hanson JA, Boltz TF, Patel BK (2017) Breast radiation dose with CESM compared with 2D FFDM and 3D tomosynthesis mammography. AJR Am J Roentgenol 208:362–372PubMedCrossRef
57.
Zurück zum Zitat Mokhtar O, Mahmoud S (2014) Can contrast enhanced mammography solve the problem of dense breast lesions? Egypt J Radiol Nucl Med 45:1043–1052CrossRef Mokhtar O, Mahmoud S (2014) Can contrast enhanced mammography solve the problem of dense breast lesions? Egypt J Radiol Nucl Med 45:1043–1052CrossRef
58.
Zurück zum Zitat Houben IPL, Van de Voorde P, Jeukens CRLPN et al (2017) Contrast-enhanced spectral mammography as work-up tool in patients recalled from breast cancer screening has low risks and might hold clinical benefits. Eur J Radiol 94:31–37PubMedCrossRef Houben IPL, Van de Voorde P, Jeukens CRLPN et al (2017) Contrast-enhanced spectral mammography as work-up tool in patients recalled from breast cancer screening has low risks and might hold clinical benefits. Eur J Radiol 94:31–37PubMedCrossRef
59.
Zurück zum Zitat Dromain C, Thibault F, Muller S et al (2011) Dual-energy contrast-enhanced digital mammography: initial clinical results. Eur Radiol 21:565–574PubMedCrossRef Dromain C, Thibault F, Muller S et al (2011) Dual-energy contrast-enhanced digital mammography: initial clinical results. Eur Radiol 21:565–574PubMedCrossRef
60.
Zurück zum Zitat Jeukens CRLPN, Lalji UC, Meijer E et al (2014) Radiation exposure of contrast-enhanced spectral mammography compared with full-field digital mammography. Invest Radiol 49:659–665PubMedCrossRef Jeukens CRLPN, Lalji UC, Meijer E et al (2014) Radiation exposure of contrast-enhanced spectral mammography compared with full-field digital mammography. Invest Radiol 49:659–665PubMedCrossRef
61.
Zurück zum Zitat Diekmann F, Freyer M, Diekmann S et al (2011) Evaluation of contrast-enhanced digital mammography. Eur J Radiol 78:112–121PubMedCrossRef Diekmann F, Freyer M, Diekmann S et al (2011) Evaluation of contrast-enhanced digital mammography. Eur J Radiol 78:112–121PubMedCrossRef
62.
Zurück zum Zitat Richter V, Hatterman V, Preibsch H et al (2018) Contrast-enhanced spectral mammography in patients with MRI contraindications. Acta Radiol 59:798–805PubMedCrossRef Richter V, Hatterman V, Preibsch H et al (2018) Contrast-enhanced spectral mammography in patients with MRI contraindications. Acta Radiol 59:798–805PubMedCrossRef
63.
Zurück zum Zitat Gluskin J, Click M, Fleischman R, Dromain C, Morris EA, Jochelson MS (2017) Contamination artifact that mimics in-situ carcinoma on contrast-enhanced digital mammography. Eur J Radiol 95:147–154PubMedCrossRefPubMedCentral Gluskin J, Click M, Fleischman R, Dromain C, Morris EA, Jochelson MS (2017) Contamination artifact that mimics in-situ carcinoma on contrast-enhanced digital mammography. Eur J Radiol 95:147–154PubMedCrossRefPubMedCentral
64.
Zurück zum Zitat Travieso Aja MM, Rodríguez Rodríguez M, Alayón Hernández S, Vega Benítez V, Luzardo OP (2014) Dual-energy contrast-enhanced mammography. Radiologia 56:390–399CrossRef Travieso Aja MM, Rodríguez Rodríguez M, Alayón Hernández S, Vega Benítez V, Luzardo OP (2014) Dual-energy contrast-enhanced mammography. Radiologia 56:390–399CrossRef
65.
Zurück zum Zitat Patel BK, Naylor ME, Kosiorek HE et al (2017) Clinical utility of contrast-enhanced spectral mammography as an adjunct for tomosynthesis-detected architectural distortion. Clin Imaging 46:44–52PubMedCrossRef Patel BK, Naylor ME, Kosiorek HE et al (2017) Clinical utility of contrast-enhanced spectral mammography as an adjunct for tomosynthesis-detected architectural distortion. Clin Imaging 46:44–52PubMedCrossRef
66.
Zurück zum Zitat Francescone MA, Jochelson MS, Dershaw DD et al (2014) Low energy mammogram obtained in contrast-enhanced digital mammography (CEDM) is comparable to routine full-field digital mammography (FFDM). Eur J Radiol 83:1350–1355PubMedCrossRef Francescone MA, Jochelson MS, Dershaw DD et al (2014) Low energy mammogram obtained in contrast-enhanced digital mammography (CEDM) is comparable to routine full-field digital mammography (FFDM). Eur J Radiol 83:1350–1355PubMedCrossRef
67.
Zurück zum Zitat Łuczyńska E, Niemiec J, Hendrick E et al (2016) Degree of enhancement on contrast enhanced spectral mammography (CESM) and lesion type on mammography (MG): comparison based on histological results. Med Sci Monit 22:3886–3893PubMedPubMedCentralCrossRef Łuczyńska E, Niemiec J, Hendrick E et al (2016) Degree of enhancement on contrast enhanced spectral mammography (CESM) and lesion type on mammography (MG): comparison based on histological results. Med Sci Monit 22:3886–3893PubMedPubMedCentralCrossRef
68.
Zurück zum Zitat Kariyappa KD, Gnanaprakasam F, Anand S, Krishnaswami M, Ramachandran M (2016) Contrast enhanced dual energy spectral mammogram, an emerging addendum in breast imaging. Br J Radiol 89:20150609PubMedPubMedCentralCrossRef Kariyappa KD, Gnanaprakasam F, Anand S, Krishnaswami M, Ramachandran M (2016) Contrast enhanced dual energy spectral mammogram, an emerging addendum in breast imaging. Br J Radiol 89:20150609PubMedPubMedCentralCrossRef
69.
Zurück zum Zitat Danala G, Patel B, Aghaei F et al (2018) Classification of breast masses using a computer-aided diagnosis scheme of contrast enhanced digital mammograms. Ann Biomed Eng 46:1419–1431PubMedCrossRefPubMedCentral Danala G, Patel B, Aghaei F et al (2018) Classification of breast masses using a computer-aided diagnosis scheme of contrast enhanced digital mammograms. Ann Biomed Eng 46:1419–1431PubMedCrossRefPubMedCentral
70.
Zurück zum Zitat Phillips J, Miller MM, Mehta TS et al (2017) Contrast-enhanced spectral mammography (CESM) versus MRI in the high-risk screening setting: patient preferences and attitudes. Clin Imaging 42:193–197PubMedCrossRef Phillips J, Miller MM, Mehta TS et al (2017) Contrast-enhanced spectral mammography (CESM) versus MRI in the high-risk screening setting: patient preferences and attitudes. Clin Imaging 42:193–197PubMedCrossRef
71.
Zurück zum Zitat Yagil Y, Shalmon A, Rundstein A et al (2016) Challenges in contrast-enhanced spectral mammography interpretation: artefacts lexicon. Clin Radiol 71:450–457PubMedCrossRef Yagil Y, Shalmon A, Rundstein A et al (2016) Challenges in contrast-enhanced spectral mammography interpretation: artefacts lexicon. Clin Radiol 71:450–457PubMedCrossRef
72.
Zurück zum Zitat Cheung Y-C, Tsai H-P, Lo Y-F, Ueng S-H, Huang P-C, Chen S-C (2016) Clinical utility of dual-energy contrast-enhanced spectral mammography for breast microcalcifications without associated mass: a preliminary analysis. Eur Radiol 26:1082–1089PubMedCrossRef Cheung Y-C, Tsai H-P, Lo Y-F, Ueng S-H, Huang P-C, Chen S-C (2016) Clinical utility of dual-energy contrast-enhanced spectral mammography for breast microcalcifications without associated mass: a preliminary analysis. Eur Radiol 26:1082–1089PubMedCrossRef
73.
Zurück zum Zitat ElSaid NAE, Farouk S, Shetat OMM, Khalifa NM, Nada OM (2015) Contrast enhanced digital mammography: is it useful in detecting lesions in edematous breast? Egypt J Radiol Nucl Med 46:811–819CrossRef ElSaid NAE, Farouk S, Shetat OMM, Khalifa NM, Nada OM (2015) Contrast enhanced digital mammography: is it useful in detecting lesions in edematous breast? Egypt J Radiol Nucl Med 46:811–819CrossRef
74.
Zurück zum Zitat Mori M, Akashi-Tanaka S, Suzuki S et al (2017) Diagnostic accuracy of contrast-enhanced spectral mammography in comparison to conventional full-field digital mammography in a population of women with dense breasts. Breast Cancer 24:104–110PubMedCrossRef Mori M, Akashi-Tanaka S, Suzuki S et al (2017) Diagnostic accuracy of contrast-enhanced spectral mammography in comparison to conventional full-field digital mammography in a population of women with dense breasts. Breast Cancer 24:104–110PubMedCrossRef
75.
Zurück zum Zitat Savaridas SL, Taylor DB, Gunawardana D, Phillips M (2017) Could parenchymal enhancement on contrast-enhanced spectral mammography (CESM) represent a new breast cancer risk factor? Correlation with known radiology risk factors. Clin Radiol 72:1085.e1–1085.e9CrossRef Savaridas SL, Taylor DB, Gunawardana D, Phillips M (2017) Could parenchymal enhancement on contrast-enhanced spectral mammography (CESM) represent a new breast cancer risk factor? Correlation with known radiology risk factors. Clin Radiol 72:1085.e1–1085.e9CrossRef
76.
Zurück zum Zitat Fallenberg EM, Schmitzberger FF, Amer H et al (2017) Contrast-enhanced spectral mammography vs. mammography and MRI – clinical performance in a multi-reader evaluation. Eur Radiol 27:2752–2764PubMedCrossRef Fallenberg EM, Schmitzberger FF, Amer H et al (2017) Contrast-enhanced spectral mammography vs. mammography and MRI – clinical performance in a multi-reader evaluation. Eur Radiol 27:2752–2764PubMedCrossRef
77.
Zurück zum Zitat Brandan M-E, Cruz-Bastida JP, Rosado-Méndez IM et al (2016) Clinical study of contrast-enhanced digital mammography and the evaluation of blood and lymphatic microvessel density. Br J Radiol 89:20160232PubMedPubMedCentralCrossRef Brandan M-E, Cruz-Bastida JP, Rosado-Méndez IM et al (2016) Clinical study of contrast-enhanced digital mammography and the evaluation of blood and lymphatic microvessel density. Br J Radiol 89:20160232PubMedPubMedCentralCrossRef
78.
Zurück zum Zitat Deng C-Y, Juan Y-H, Cheung Y-C et al (2018) Quantitative analysis of enhanced malignant and benign lesions on contrast-enhanced spectral mammography. Br J Radiol 91:20170605 Deng C-Y, Juan Y-H, Cheung Y-C et al (2018) Quantitative analysis of enhanced malignant and benign lesions on contrast-enhanced spectral mammography. Br J Radiol 91:20170605
79.
Zurück zum Zitat Mohamed Kamal R, Hussien Helal M, Wessam R, Mahmoud Mansour S, Godda I, Alieldin N (2015) Contrast-enhanced spectral mammography: impact of the qualitative morphology descriptors on the diagnosis of breast lesions. Eur J Radiol 84:1049–1055PubMedCrossRef Mohamed Kamal R, Hussien Helal M, Wessam R, Mahmoud Mansour S, Godda I, Alieldin N (2015) Contrast-enhanced spectral mammography: impact of the qualitative morphology descriptors on the diagnosis of breast lesions. Eur J Radiol 84:1049–1055PubMedCrossRef
80.
Zurück zum Zitat Sogani J, Morris EA, Kaplan JB et al (2017) Comparison of background parenchymal enhancement at contrast-enhanced spectral mammography and breast MR imaging. Radiology 282:63–73PubMedCrossRef Sogani J, Morris EA, Kaplan JB et al (2017) Comparison of background parenchymal enhancement at contrast-enhanced spectral mammography and breast MR imaging. Radiology 282:63–73PubMedCrossRef
81.
Zurück zum Zitat Hobbs MM, Taylor DB, Buzynski S, Peake RE (2015) Contrast-enhanced spectral mammography (CESM) and contrast enhanced MRI (CEMRI): patient preferences and tolerance. J Med Imaging Radiat Oncol 59:300–305PubMedCrossRef Hobbs MM, Taylor DB, Buzynski S, Peake RE (2015) Contrast-enhanced spectral mammography (CESM) and contrast enhanced MRI (CEMRI): patient preferences and tolerance. J Med Imaging Radiat Oncol 59:300–305PubMedCrossRef
82.
Zurück zum Zitat Phillips J, Mihai G, Hassonjee SE et al (2018) Comparative dose of contrast-enhanced spectral mammography (CESM), digital mammography, and digital breast tomosynthesis. AJR Am J Roentgenol 211:839–846PubMedCrossRef Phillips J, Mihai G, Hassonjee SE et al (2018) Comparative dose of contrast-enhanced spectral mammography (CESM), digital mammography, and digital breast tomosynthesis. AJR Am J Roentgenol 211:839–846PubMedCrossRef
83.
Zurück zum Zitat Knogler T, Homolka P, Hörnig M et al (2016) Contrast-enhanced dual energy mammography with a novel anode/filter combination and artifact reduction: a feasibility study. Eur Radiol 26:1575–1581PubMedCrossRef Knogler T, Homolka P, Hörnig M et al (2016) Contrast-enhanced dual energy mammography with a novel anode/filter combination and artifact reduction: a feasibility study. Eur Radiol 26:1575–1581PubMedCrossRef
84.
Zurück zum Zitat Xing D, Lv Y, Sun B et al (2018) Diagnostic value of contrast-enhanced spectral mammography in comparison to magnetic resonance imaging in breast lesions. J Comput Assist Tomogr 43:245–251PubMedCrossRef Xing D, Lv Y, Sun B et al (2018) Diagnostic value of contrast-enhanced spectral mammography in comparison to magnetic resonance imaging in breast lesions. J Comput Assist Tomogr 43:245–251PubMedCrossRef
85.
Zurück zum Zitat Fallenberg EM, Dromain C, Diekmann F et al (2014) Contrast-enhanced spectral mammography: does mammography provide additional clinical benefits or can some radiation exposure be avoided? Breast Cancer Res Treat 146:371–381PubMedCrossRef Fallenberg EM, Dromain C, Diekmann F et al (2014) Contrast-enhanced spectral mammography: does mammography provide additional clinical benefits or can some radiation exposure be avoided? Breast Cancer Res Treat 146:371–381PubMedCrossRef
86.
Zurück zum Zitat Cheung Y-C, Lin Y-C, Wan Y-L et al (2014) Diagnostic performance of dual-energy contrast-enhanced subtracted mammography in dense breasts compared to mammography alone: interobserver blind-reading analysis. Eur Radiol 24:2394–2403PubMedCrossRef Cheung Y-C, Lin Y-C, Wan Y-L et al (2014) Diagnostic performance of dual-energy contrast-enhanced subtracted mammography in dense breasts compared to mammography alone: interobserver blind-reading analysis. Eur Radiol 24:2394–2403PubMedCrossRef
87.
Zurück zum Zitat Kim EY, Youn I, Lee KH et al (2018) Diagnostic value of contrast-enhanced digital mammography versus contrast-enhanced magnetic resonance imaging for the preoperative evaluation of breast cancer. J Breast Cancer 21:453PubMedPubMedCentralCrossRef Kim EY, Youn I, Lee KH et al (2018) Diagnostic value of contrast-enhanced digital mammography versus contrast-enhanced magnetic resonance imaging for the preoperative evaluation of breast cancer. J Breast Cancer 21:453PubMedPubMedCentralCrossRef
88.
Zurück zum Zitat Klang E, Krosser A, Amitai MM et al (2018) Utility of routine use of breast ultrasound following contrast-enhanced spectral mammography. Clin Radiol 73:908.e11–908.e16PubMedCrossRef Klang E, Krosser A, Amitai MM et al (2018) Utility of routine use of breast ultrasound following contrast-enhanced spectral mammography. Clin Radiol 73:908.e11–908.e16PubMedCrossRef
89.
Zurück zum Zitat Tsigginou A, Gkali C, Chalazonitis A et al (2016) Adding the power of iodinated contrast media to the credibility of mammography in breast cancer diagnosis. Br J Radiol 89:20160397PubMedPubMedCentralCrossRef Tsigginou A, Gkali C, Chalazonitis A et al (2016) Adding the power of iodinated contrast media to the credibility of mammography in breast cancer diagnosis. Br J Radiol 89:20160397PubMedPubMedCentralCrossRef
90.
Zurück zum Zitat Navarro ME, Razmilic D, Araos I, Rodrigo A, Andia ME (2018) Contrast-enhanced spectral mammography. Experience in 465 examinations. Rev Med Chil 146:141–149 Navarro ME, Razmilic D, Araos I, Rodrigo A, Andia ME (2018) Contrast-enhanced spectral mammography. Experience in 465 examinations. Rev Med Chil 146:141–149
91.
Zurück zum Zitat Luczynska E, Niemiec J, Ambicka A et al (2015) Correlation between blood and lymphatic vessel density and results of contrast-enhanced spectral mammography. Pol J Pathol 3:310–322CrossRef Luczynska E, Niemiec J, Ambicka A et al (2015) Correlation between blood and lymphatic vessel density and results of contrast-enhanced spectral mammography. Pol J Pathol 3:310–322CrossRef
92.
Zurück zum Zitat Sorin V, Yagil Y, Yosepovich A et al (2018) Contrast-enhanced spectral mammography in women with intermediate breast cancer risk and dense breasts. AJR Am J Roentgenol 211:W267–W274PubMedCrossRef Sorin V, Yagil Y, Yosepovich A et al (2018) Contrast-enhanced spectral mammography in women with intermediate breast cancer risk and dense breasts. AJR Am J Roentgenol 211:W267–W274PubMedCrossRef
93.
Zurück zum Zitat Bicchierai G, Nori J, De Benedetto D et al (2018) Role of contrast-enhanced spectral mammography in the post biopsy management of B3 lesions: preliminary results. Tumori J. [Epub ahead of print] Bicchierai G, Nori J, De Benedetto D et al (2018) Role of contrast-enhanced spectral mammography in the post biopsy management of B3 lesions: preliminary results. Tumori J. [Epub ahead of print]
94.
Zurück zum Zitat Knogler T, Homolka P, Hoernig M et al (2017) Application of BI-RADS descriptors in contrast-enhanced dual-energy mammography: comparison with MRI. Breast Care (Basel) 12:212–216PubMedPubMedCentralCrossRef Knogler T, Homolka P, Hoernig M et al (2017) Application of BI-RADS descriptors in contrast-enhanced dual-energy mammography: comparison with MRI. Breast Care (Basel) 12:212–216PubMedPubMedCentralCrossRef
95.
Zurück zum Zitat Kamal RM, Helal MH, Mansour SM et al (2016) Can we apply the MRI BI-RADS lexicon morphology descriptors on contrast-enhanced spectral mammography? Br J Radiol 89:20160157PubMedPubMedCentralCrossRef Kamal RM, Helal MH, Mansour SM et al (2016) Can we apply the MRI BI-RADS lexicon morphology descriptors on contrast-enhanced spectral mammography? Br J Radiol 89:20160157PubMedPubMedCentralCrossRef
96.
Zurück zum Zitat Lewin JM, Isaacs PK, Vance V, Larke FJ (2003) Dual-energy contrast-enhanced digital subtraction mammography: feasibility. Radiology 229:261–268PubMedCrossRef Lewin JM, Isaacs PK, Vance V, Larke FJ (2003) Dual-energy contrast-enhanced digital subtraction mammography: feasibility. Radiology 229:261–268PubMedCrossRef
97.
Zurück zum Zitat Wang Q, Li K, Wang L, Zhang J, Zhou Z, Feng Y (2016) Preclinical study of diagnostic performances of contrast-enhanced spectral mammography versus MRI for breast diseases in China. Springerplus 5:763PubMedPubMedCentralCrossRef Wang Q, Li K, Wang L, Zhang J, Zhou Z, Feng Y (2016) Preclinical study of diagnostic performances of contrast-enhanced spectral mammography versus MRI for breast diseases in China. Springerplus 5:763PubMedPubMedCentralCrossRef
98.
Zurück zum Zitat Cheung Y-C, Juan Y-H, Lin Y-C et al (2016) Dual-energy contrast-enhanced spectral mammography: enhancement analysis on BI-RADS 4 non-mass microcalcifications in screened women. PLoS One 11:e0162740PubMedPubMedCentralCrossRef Cheung Y-C, Juan Y-H, Lin Y-C et al (2016) Dual-energy contrast-enhanced spectral mammography: enhancement analysis on BI-RADS 4 non-mass microcalcifications in screened women. PLoS One 11:e0162740PubMedPubMedCentralCrossRef
99.
Zurück zum Zitat Helal M, Abu Samra MF, Ibraheem MA, Salama A, Hassan EE, Hassan NE-H (2017) Accuracy of CESM versus conventional mammography and ultrasound in evaluation of BI-RADS 3 and 4 breast lesions with pathological correlation. Egypt J Radiol Nucl Med 48:741–750CrossRef Helal M, Abu Samra MF, Ibraheem MA, Salama A, Hassan EE, Hassan NE-H (2017) Accuracy of CESM versus conventional mammography and ultrasound in evaluation of BI-RADS 3 and 4 breast lesions with pathological correlation. Egypt J Radiol Nucl Med 48:741–750CrossRef
100.
Zurück zum Zitat Li L, Roth R, Germaine P et al (2017) Contrast-enhanced spectral mammography (CESM) versus breast magnetic resonance imaging (MRI): a retrospective comparison in 66 breast lesions. Diagn Interv Imaging 98:113–123PubMedCrossRef Li L, Roth R, Germaine P et al (2017) Contrast-enhanced spectral mammography (CESM) versus breast magnetic resonance imaging (MRI): a retrospective comparison in 66 breast lesions. Diagn Interv Imaging 98:113–123PubMedCrossRef
101.
Zurück zum Zitat Badr S, Laurent N, Régis C, Boulanger L, Lemaille S, Poncelet E (2014) Dual-energy contrast-enhanced digital mammography in routine clinical practice in 2013. Diagn Interv Imaging 95:245–258PubMedCrossRef Badr S, Laurent N, Régis C, Boulanger L, Lemaille S, Poncelet E (2014) Dual-energy contrast-enhanced digital mammography in routine clinical practice in 2013. Diagn Interv Imaging 95:245–258PubMedCrossRef
102.
Zurück zum Zitat Covington MF, Pizzitola VJ, Lorans R et al (2018) The future of contrast-enhanced mammography. AJR Am J Roentgenol 210:292–300PubMedCrossRef Covington MF, Pizzitola VJ, Lorans R et al (2018) The future of contrast-enhanced mammography. AJR Am J Roentgenol 210:292–300PubMedCrossRef
103.
Zurück zum Zitat Lancaster RB, Gulla S, De Los Santos J, Umphrey HR (2018) Contrast-enhanced spectral mammography in breast imaging. Semin Roentgenol 53:294–300PubMedCrossRef Lancaster RB, Gulla S, De Los Santos J, Umphrey HR (2018) Contrast-enhanced spectral mammography in breast imaging. Semin Roentgenol 53:294–300PubMedCrossRef
104.
Zurück zum Zitat James JJ, Tennant SL (2018) Contrast-enhanced spectral mammography (CESM). Clin Radiol 73:715–723PubMedCrossRef James JJ, Tennant SL (2018) Contrast-enhanced spectral mammography (CESM). Clin Radiol 73:715–723PubMedCrossRef
105.
Zurück zum Zitat Patel BK, Gray RJ, Pockaj BA (2017) Potential cost Savings of Contrast-Enhanced Digital Mammography. AJR Am J Roentgenol 208:W231–W237PubMedCrossRef Patel BK, Gray RJ, Pockaj BA (2017) Potential cost Savings of Contrast-Enhanced Digital Mammography. AJR Am J Roentgenol 208:W231–W237PubMedCrossRef
106.
Zurück zum Zitat Lewin J (2018) Comparison of contrast-enhanced mammography and contrast-enhanced breast MR imaging. Magn Reson Imaging Clin N Am 26:259–263PubMedCrossRef Lewin J (2018) Comparison of contrast-enhanced mammography and contrast-enhanced breast MR imaging. Magn Reson Imaging Clin N Am 26:259–263PubMedCrossRef
107.
Zurück zum Zitat Minsinger KD, Kassis HM, Block CA, Sidhu M, Brown JR (2014) Meta-analysis of the effect of automated contrast injection devices versus manual injection and contrast volume on risk of contrast-induced nephropathy. Am J Cardiol 113:49–53PubMedCrossRef Minsinger KD, Kassis HM, Block CA, Sidhu M, Brown JR (2014) Meta-analysis of the effect of automated contrast injection devices versus manual injection and contrast volume on risk of contrast-induced nephropathy. Am J Cardiol 113:49–53PubMedCrossRef
108.
Zurück zum Zitat Endrikat J, Barbati R, Scarpa M, Jost G, Ned Uber AE 3rd (2018) Accuracy and repeatability of automated injector versus manual administration of an MRI contrast agent—results of a laboratory study. Invest Radiol 53:1–5PubMedCrossRef Endrikat J, Barbati R, Scarpa M, Jost G, Ned Uber AE 3rd (2018) Accuracy and repeatability of automated injector versus manual administration of an MRI contrast agent—results of a laboratory study. Invest Radiol 53:1–5PubMedCrossRef
109.
Zurück zum Zitat Jost G, Endrikat J, Pietsch H (2017) The impact of injector-based contrast agent administration on bolus shape and magnetic resonance angiography image quality. Magn Reson Insights 10:1178623X1770589CrossRef Jost G, Endrikat J, Pietsch H (2017) The impact of injector-based contrast agent administration on bolus shape and magnetic resonance angiography image quality. Magn Reson Insights 10:1178623X1770589CrossRef
110.
Zurück zum Zitat Auler MA, Heagy T, Aganovic L, Brothers R, Costello P, Schoepf UJ (2006) Saline chasing technique with dual-syringe injector systems for multi-detector row computed tomographic angiography: rationale, indications, and protocols. Curr Probl Diagn Radiol 35:1–11PubMedCrossRef Auler MA, Heagy T, Aganovic L, Brothers R, Costello P, Schoepf UJ (2006) Saline chasing technique with dual-syringe injector systems for multi-detector row computed tomographic angiography: rationale, indications, and protocols. Curr Probl Diagn Radiol 35:1–11PubMedCrossRef
111.
Zurück zum Zitat Kidoh M, Nakaura T, Awai K et al (2013) Novel connecting tube for saline chaser in contrast-enhanced CT: the effect of spiral flow of saline on contrast enhancement. Eur Radiol 23:3213–3218PubMedCrossRef Kidoh M, Nakaura T, Awai K et al (2013) Novel connecting tube for saline chaser in contrast-enhanced CT: the effect of spiral flow of saline on contrast enhancement. Eur Radiol 23:3213–3218PubMedCrossRef
112.
Zurück zum Zitat Perry N, Broeders M, de Wolf C, Tornberg S, Holland R, von Karsa L (2007) European guidelines for quality assurance in breast cancer screening and diagnosis. Fourth edition--summary document. Ann Oncol 19:614–622PubMedCrossRef Perry N, Broeders M, de Wolf C, Tornberg S, Holland R, von Karsa L (2007) European guidelines for quality assurance in breast cancer screening and diagnosis. Fourth edition--summary document. Ann Oncol 19:614–622PubMedCrossRef
113.
Zurück zum Zitat Wang CL, Cohan RH, Ellis JH, Caoili EM, Wang G, Francis IR (2008) Frequency, outcome, and appropriateness of treatment of nonionic iodinated contrast media reactions. AJR Am J Roentgenol 191:409–415PubMedCrossRef Wang CL, Cohan RH, Ellis JH, Caoili EM, Wang G, Francis IR (2008) Frequency, outcome, and appropriateness of treatment of nonionic iodinated contrast media reactions. AJR Am J Roentgenol 191:409–415PubMedCrossRef
114.
Zurück zum Zitat Mortelé KJ, Oliva M-R, Ondategui S, Ros PR, Silverman SG (2005) Universal use of nonionic iodinated contrast medium for CT: evaluation of safety in a large urban teaching hospital. AJR Am J Roentgenol 184:31–34PubMedCrossRef Mortelé KJ, Oliva M-R, Ondategui S, Ros PR, Silverman SG (2005) Universal use of nonionic iodinated contrast medium for CT: evaluation of safety in a large urban teaching hospital. AJR Am J Roentgenol 184:31–34PubMedCrossRef
115.
Zurück zum Zitat Huston P, Moher D (1996) Redundancy, disaggregation, and the integrity of medical research. Lancet 347:1024–1026CrossRef Huston P, Moher D (1996) Redundancy, disaggregation, and the integrity of medical research. Lancet 347:1024–1026CrossRef
116.
Zurück zum Zitat Murphy L, Wyllie A (2009) Duplicate patient data in a meta-analysis: a threat to validity. J Crit Care 24:466–467PubMedCrossRef Murphy L, Wyllie A (2009) Duplicate patient data in a meta-analysis: a threat to validity. J Crit Care 24:466–467PubMedCrossRef
117.
Zurück zum Zitat Sardanelli F, Alì M, Hunink MG, Houssami N, Sconfienza LM, Di Leo G (2018) To share or not to share? Expected pros and cons of data sharing in radiological research. Eur Radiol 28:2328–2335PubMedCrossRef Sardanelli F, Alì M, Hunink MG, Houssami N, Sconfienza LM, Di Leo G (2018) To share or not to share? Expected pros and cons of data sharing in radiological research. Eur Radiol 28:2328–2335PubMedCrossRef
Metadaten
Titel
Technique, protocols and adverse reactions for contrast-enhanced spectral mammography (CESM): a systematic review
verfasst von
Moreno Zanardo
Andrea Cozzi
Rubina Manuela Trimboli
Olgerta Labaj
Caterina Beatrice Monti
Simone Schiaffino
Luca Alessandro Carbonaro
Francesco Sardanelli
Publikationsdatum
01.12.2019
Verlag
Springer Berlin Heidelberg
Erschienen in
Insights into Imaging / Ausgabe 1/2019
Elektronische ISSN: 1869-4101
DOI
https://doi.org/10.1186/s13244-019-0756-0

Weitere Artikel der Ausgabe 1/2019

Insights into Imaging 1/2019 Zur Ausgabe

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Ein Drittel der jungen Ärztinnen und Ärzte erwägt abzuwandern

07.05.2024 Klinik aktuell Nachrichten

Extreme Arbeitsverdichtung und kaum Supervision: Dr. Andrea Martini, Sprecherin des Bündnisses Junge Ärztinnen und Ärzte (BJÄ) über den Frust des ärztlichen Nachwuchses und die Vorteile des Rucksack-Modells.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.