Open Access 01.11.2014 | Echocardiography (T Buck, Section Editor)

Quantitative Analysis of Aortic Valve Stenosis and Aortic Root Dimensions by Three-Dimensional Echocardiography in Patients Scheduled for Transcutaneous Aortic Valve Implantation

verfasst von: Rolf Alexander Jánosi, Björn Plicht, Philipp Kahlert, Mareike Eißmann, Daniel Wendt, Heinz Jakob, Raimund Erbel, Thomas Buck

Erschienen in: Current Cardiovascular Imaging Reports | Ausgabe 11/2014

Abstract

Accurate assessment of the aortic valve area (AVA) and evaluation of the aortic root are important for clinical decision-making in patients being considered for transcatheter aortic valve implantation (TAVI). Real-time three-dimensional transesophageal echocardiography (RT3D-TEE) provides accurate and reliable quantitative assessment of aortic valve stenosis and the aortic root. We performed two-dimensional transthoracic echocardiography (2D-TTE), real-time 2D transesophageal echocardiography (RT2D-TEE) and RT3D-TEE in 71 consecutive patients referred for TAVI. RT3D-TEE multiplanar reconstruction was used to measure aortic root parameters, including left ventricular outflow tract (LVOT) diameter and area, aortic annulus diameter, aortic annulus area, and AVA. RT3D-TEE methods for planimetry and the LVOT-derived continuity equation for the estimation of AVA showed a good correlation. As iatrogenic coronary ostium occlusion is a potentially life-threatening complication, we evaluated the distances from the aortic annulus to the coronary ostia using RT3D-TEE. Based on our findings, we conclude that the geometry of the aortic root and aortic valve can be reliably and feasibly evaluated using RT3D-TEE, which is important for protecting against potential complications of TAVI, such as underestimation of the size of the aortic annulus that can result in aortic regurgitation and dislocation of the valve, or overestimation can lead to annulus rupture.
Hinweise
This article is part of the Topical Collection on Echocardiography

Introduction

Aortic valve (AV) disease has emerged as a frequent clinical problem in the rapidly growing elderly population, with an increasing proportion of patients being at high risk for traditional AV surgery [1]. Transcatheter AV implantation (TAVI) has accordingly grown in popularity as an alternative to surgery for such high-risk patients [2]. However, since it is an interventional procedure without the benefit of a direct view of the AV, detailed anatomical information prior to the procedure is mandatory to avoid devastating complications. Assessment of the AV area (AVA) is only the first step in the evaluation and selection of suitable patients. For prosthesis selection from a range of defined sizes and a variety of prosthesis types, the accurate measurement of aortic root geometry is essential [3, 4]. Moreover, a rare but extremely critical complication is the intraoperative occlusion of the coronary ostium, which underlines the importance of prior evaluation of the distance from the aortic annulus to the coronary ostia [5]. Finally, paravalvular regurgitation—which is known to be associated with higher mortality in patients undergoing TAVI—can be avoided by correct prosthesis sizing and positioning when the aortic root dimensions are known [6].
The major reasons for the regular use of echocardiography in the clinical setting and for follow-up assessment are its availability, lack of relevant procedure-related risks, and low cost as compared to invasive techniques and cardiovascular magnetic resonance imaging (MRI). The indications for surgery are based on symptoms and the severity of AV stenosis (AS); in current guidelines, this is defined as an AVA of less than 1 cm2 (<0.6 cm2/m2) or a mean AV gradient ≥40 mmHg [1, 79]. Two-dimensional transthoracic echocardiography (2D-TTE) is now typically used as the first step in quantification of AS, based on the determination of flow-dependent variables using the continuity equation (CE) and the effective AVA [10]. However, in patients with subvalvular obstruction, reduced left ventricular function, or aortic regurgitation, the accurate assessment of AS may be difficult [11]. Multiplane transesophageal echocardiography (TEE) is known to provide more exact information by allowing direct planimetric quantification of the AVA. One of the difficulties of this procedure is obtaining the correct cross-sectional 2D short-axis view at the tip of the AV cusps due to the motion of the aortic annulus and valve cusps, which may result in overestimation of the AVA [12, 13]. The use of real-time 3D-TEE (RT3D-TEE) may lead to better results than standard 2D-TEE because of the use of a 3D volume dataset of the AV, where the true orthogonal imaging plane of the aortic cusps can be measured directly or analyzed later by an experienced echocardiographer [14]. Moreover, RT3D-TEE is less time-consuming and does not have limitations, such as those faced when using other 3D techniques such as computed tomography (CT) and MRI [15].
For the calculation of AVA by echocardiography using the CE, the determination of the left ventricular outflow tract (LVOT) cross-sectional area (LVOTA) is necessary. AVA is usually estimated using CE, including the LVOTA calculated from a measured LVOT diameter (LVOTD) with the assumption that the LVOT is circular [7]. Measurements using CE have shown good agreement with those using invasive techniques; however, consistent underestimation is noted since the commonly used formulation of the CE relies on the assumption of a circular LVOT [16].
The aortic root is the direct continuation of the LVOT, beginning from the insertion of the aortic valvular cusps to the level of the sinotubular junction [4]. Its anatomy and diameter vary in patients with AS or aortic regurgitation as well as in the healthy population [17]. The measurement of the aortic annulus is the key measurement for the preparation of patients undergoing TAVI. Undersizing may lead to prosthesis migration or paravalvular regurgitation [18]. Even oversizing may cause problems such as complications related to vascular access, due to the delivery system or insufficient expansion of the prosthesis, reducing the valve’s durability [4]. We have previously described a method for the rapid, detailed 3D measurement of the aortic annulus using the ability of 3D measurements to identify the accurate diameter [19]. The aortic annulus should typically be measured in systole between the insertion of the AV cusps, not including the calcification of the aortic cusps (Fig. 1) [19, 20••]. Particularly when the annulus diameter is in the borderline range, the use of RT3D-TEE with analysis of volumetric data is helpful.
Piazza et al. [4] defined four parts of the aortic root: a virtual annulus formed by the joining basal attachments of the AV leaflets, the anatomic annulus formed by the ventriculoarterial junction, the sinotubular junction, and a crown-like ring formed by the insertion of the leaflets. Thus the structure of the aortic annulus as measured by echocardiographic means is the virtual ring or the hinge point of the AV cusps (Fig. 2).
In this study, we tested the hypothesis that the LVOT is circular, based on a formula for eccentricity, and its impact on the calculation of AVA using the CE in comparison to direct planimetry. Further, in this context, we studied the geometry of the LVOT and AV in patients undergoing TAVI at our institution.

Subjects and Methods

2D-TTE, 2D-TEE, and RT3D-TEE echocardiography were performed using a commercial ultrasound imaging system (iE33; Philips Medical Systems, Andover, MA) with 2D matrix S5-1 and a 3D matrix-array transesophageal transducer (X7-2 t). The examination protocol was as previously described [19] and is further documented in Fig. 1.
The study group comprised 71 patients undergoing 2D-TTE, 2D-TEE, and RT3D-TEE for evaluation as candidates for TAVI. The 2D-TTE protocol included a complete transthoracic echo-Doppler study with color-pulsed and continuous-wave Doppler measurements of the AV [21]. The LV end-systolic and LV end-diastolic volume indices were calculated using the Simpson biplane method of discs [21].
Routine 2D-TTE was followed by 2D-TEE and 3D-TEE full-volume acquisitions. All images were digitally stored for offline analysis (QLAB; Philips Medical Systems). For the acquisition of the volumetric 3D datasets, the live 3D zoom function was used to gain a magnified zoomed image of the aortic root in the 35° short-axis or 120° long-axis view.
AVA was calculated by 2D-TTE-derived flow methods (AVA was calculated using the CE and the time–velocity integral of AV and LVOT spectral Doppler curves): in TEE using the standard 2D-TEE planimetry method, and in 3D the planimetric AVA was extracted from volumetric 3D datasets using RT3D-TEE. In both methods, planimetry was repeated three times and values were averaged [22]. 2D measurement of the LVOTD was obtained 0.5–1.0 cm below the AV when the aortic cusps were maximally separated, and the approximated cross-sectional area of the LVOT was calculated as π × r 2 [7]. For the CE-derived calculation of LVOT, the systolic LVOTD was introduced into the equation: LVOTAc = π × (d/2)2. In contrast, in 3D using RT3D-TEE, direct planimetry for LVOT was performed again 0.5–1.0 cm below the AV (LVOTAp).
An eccentricity index (EI) was calculated using the shortest and longest LVOTDs (EI = 1 − LVOTDshort/LVOTDlong) [23]. Similarly, an EI was used for the aortic annular area. For CE, the assumed perfect circle would be represent by an EI of zero, and an ellipsoid shape by a higher EI.

Real-time 3D Transesophageal Echocardiography

Loops were viewed after acquisition to ensure that the live 3D zoom data included LVOT and AV measurements. These full-volume DICOM loops were analyzed using QLAB. First, the orientation of an imaging plane within the 3D full volume was manipulated to gain an LVOT plane aligned parallel and about 0.5–1.0 cm below the AV. Planimetry was then performed by tracing the outline of the LVOT. Second, in multiplanar reconstruction mode, two orthogonal long-axis views of the AV and ascending aorta were selected; then a third plane perpendicular to the long-axis view was moved to the smallest AV orifice area at the moment of maximal aortic valve opening. Subsequently, the annulus was then measureed by shifting the short-axis to show the aortic annulus, where the minimum and maximum diameters and annulus area were measured. The correct level of the aortic annulus was defined as the lowest plane of the valve hinge-point (inferior virtual basal ring) [4]. In addition, as a third step, in an end-diastolic frame using two different imaging planes, the distances from the ostia of the left and right coronary arteries (LCA and RCA) to the annulus were measured. The measurement procedures are illustrated in Fig. 1.
Written informed consent was obtained from all patients at the time of consent for the clinical TEE procedure.

Statistical Analysis

Continuous data are expressed as means ± SD. Categorical data are presented as percentages. Pearson’s correlation coefficient and variability were calculated. Variables were compared using a paired t-test. Values of P <0.05 were considered statistically significant.

Results

The baseline characteristics of the study population are shown in Table 1.
Table 1
Baseline clinical and 2D echocardiographic characteristics of the 71 study patients
Variable
Value
Age (years)
75 ± 13
Body surface area (m2)
1.83 ± 0.21
Body mass index (kg/m2)
30.6
Left ventricular ejection fraction (%)
47 ± 13
Aortic valve area (m2)
0.71 ± 0.26
Mean transaortic pressure gradient (mmHg)
39.9 ± 21.5
Peak transaortic pressure gradient (mmHg)
67.3 ± 34.5
LV end-diastolic diameter (mm)
52.2 ± 9.7

LVOT Eccentricity and LVOTA

LVOTDlong was significantly greater than LVOTDshort by a mean of 0.32 ± 0.21 cm (P < 0.001). LVOTAp was significantly greater than LVOTAc (3.51 ± 0.82 cm2 vs. 3.24 ± 1.07 cm2; P < 0.001). Using the equation of an ellipse (π × a × b), the LVOTA 3Dellip [=π × (LVOTDshort)/2 × (LVOTDlong)/2] was comparable to LVOTAp (3.25 ± 1.17 cm2), resulting in improved agreement with LVOTAp (r = 0.97). The mean EI was 0.14 ± 0.10, with most patients showing some degree of eccentricity.

Measurement of AVA

The mean AVA calculated by 2D-TTE was 0.69 ± 0.28 cm2 with peak and mean pressure gradients across the AV of 67.3 ± 34.5 and 39.9 ± 21.5 mmHg, respectively (Table 1). A high correlation was noted between the planimetry of AVA by 2D-TEE and RT3D-TEE (r = 0.81, P < 0.1); however, 2D-TEE planimetry showed a significantly larger AVA than RT3D-TEE (0.71 ± 0.26 cm2 vs. 0.61 ± 0.23 cm2, P < 0.05); Fig. 2). Figure 3 shows different imaging planes for the same AV, demonstrating the reason for the overestimation of AV size. Further, considering the RT3D-derived LVOTAp in the CE yielded a higher calculated AVA (0.8 ± 0.58 cm2) (Fig. 4).

Aortic Annulus Diameter and Area

The mean minimal and maximal diameters of the aortic annulus measured by RT3D-TEE were comparable to those published in previous reports [24] (Table 2). The aortic annulus diameter was measured by 2D-TTE, 2D-TEE, and RT3D-TEE, and the annulus area was measured by 2D-TEE and RT3D-TEE successfully in all patients. The aortic annulus area measured by 2D-TEE was significantly smaller than that calculated by RT3D-TEE (3.63 ± 0.72 cm2 vs. 4.05 ± 0.71 cm2; P < 0.05). There was a high correlation between the planimetry estimated by 2D-TEE and RT3D-TEE (r = 0.81). The RT3D-TEE data underlined again that the aortic annulus has an ellipsoid shape and is not circular.
Table 2
Aortic valve area calculated by planimetry and the continuity equation, and comparison of three imaging methods for measuring the aortic annulus and area
Variable
2D-TTE
2D-TEE
RT3D-TEE
Aortic valve area (cm2)
 Planimetry
0.82 ± 0.32*
0.71 ± 0.26*
0.61 ± 0.23*
 Continuity equation
0.69 ± 0.28
0.75 ± 0.39
0.8 ± 0.58
Aortic annulus
21.3 ± 3.0
22.3 ± 2.9
22.0 ± 3.0
 Maximum diameter (mm)
  
24.1 ± 3.9
 Minimum diameter (mm)
  
21.3 ± 4.4
 EI
  
0.15 ± 0.14
 Area (cm2)
 
3.63 ± 0.72
4.04 ± 0.71
LVOT
 Maximum diameter (mm)
19.6 ± 2.3
 
21.6 ± 3.6
 Minimum diameter (mm)
  
18.5 ± 3.5
 EI
  
0.14 ± 0.10
 Area (cm2)
2.94 ± 0.44
 
3.24 ± 1.07
Distance to coronary artery ostia (mm)
 LCA
  
12.1 ± 3.7
 RCA
  
15.3 ± 3.7
*P < 0.05 versus continuity equation value in each method, paired t-test

Distances from the Coronary Ostia to the Aortic Annulus

The distances from the ostia of the LCA and RCA to the aortic annulus could be measured in 92 % and 86 % of the patients, respectively. The mean distance from the LCA ostium to the aortic annulus was 12.1 ± 3.7 mm and from the RCA ostium to the aortic annulus was 15.3 ± 3.7 mm, with large ranges of 7 – 21 mm and 7.9 – 24 mm, respectively.

Discussion

In the current study, we demonstrated the feasibility of direct evaluation using RT3D-TEE for measuring the AV and aortic root and demonstrated that the aortic annulus and LVOT are oval in shape.
Our findings are consistent with those of previous studies showing that LVOTA is significantly underestimated by 2D minor axis measurements as compared with 3D data and that the geometry of the LVOTA is best described as elliptical and is rarely circular [23].
The accuracy in estimating AVA using the CE depends on the exact measurement of LVOTD and the correct placement of the pulsed Doppler flow immediately below the AV. Errors in the measurement of the diameter are squared in the process of calculating AVA [25], leading to an underestimation of up to 13 % for LVOTA and 26.3 % for AVA due to the multiplicative relationship between these two terms [26]. These findings are in agreement with those of other studies using CT [23, 26] and MRI [27]. This can lead to a reclassification of up to 25 % of patients having moderate AS when using AVA calculated by RT3D-TEE [26]. Thus, knowledge of the LVOT geometry is crucial for potential TAVI patients, affecting the success of the procedure [28]. The accurate measurement of LVOTA is moreover important in the assessment of shunt ratios for hypertrophic cardiomyopathy and in the evaluation of subaortic membranes [29].
Nakai et al. showed the dynamic movements of the aortic annulus during the cardiac cycle in speckle tracking analysis, with a more cranial position in early systole and a more caudal position during isovolumetric relaxation [12]. Different approaches for the exact estimation of the AVA include: (1) planimetric measurement from selected slices extracted from a volumetric 3D-TEE dataset, and (2) flow-derived calculation of AVA using the CE after 3D evaluation of the LVOTA. This may lead to a more accurate estimation than 2D-TEE since this method is limited by difficulties in selecting the correct cross-sectional view of the tip of the aortic cusps [12, 13, 30]. RT3D-TEE enables rapid data acquisition even of huge volumetric datasets for later accurate evaluation [31]. We favor using the live 3D zoom mode during the collection of the volumetric dataset, which helps reduce artifacts.
3D echocardiography may facilitate the accurate measurement of the aortic annulus and significantly reduce interobserver and intraobserver variability in comparison to 2D techniques [19]. This is because the 2D long-axis view (sagittal view) of the aortic annulus is difficult to align in the center of the annulus using only 2D techniques, as recently described by Piazza et al. [4]. Ng et al. demonstrated that aortic annular areas are more significantly underestimated by 2D-TEE than by 3D-TEE in comparison to the “gold standard” multislice CT (MSCT), and that such underestimation decreases after TAVI when the annular areas change to a more circular shape [26].
Previously, cardiologists and cardiac surgeons have hesitated employing TAVI in patients with bicuspid AVs. However, newer patient registries have shown that TAVI can be performed in such patients with satisfactory results despite a higher rate of relevant aortic regurgitation. 3D echocardiography can provide important information in such patients, as recent reports have shown that patients with bicuspid AS have larger transverse areas and longitudinal distances [32], helping to reduce unnecessary paravalvular leakage in the future.

Distances from the Coronary Ostia to the Aortic Annulus

In 2D-TEE examinations, the distance from the RCA ostium to the annulus can be evaluated but it is rarely possible to measure the distance from the LCA ostium to the annulus as the LCA ostium usually lies in a different coronal plane is not accessible by 2D imaging. 3D-TEE may be an alternative for the evaluation of these distances instead of the typically used multidetector CT (MDCT), especially in patients with renal dysfunction, as earlier studies have shown good correlations for these techniques [24, 33]. In previous studies, MDCT measurements yielded an average distance from the aortic annulus to the coronary ostia from 12.8 ± 3.5 to 13.4 ± 3.2 mm for the LCA and from 13.2 ± 1.8 to 16 ± 2.8 mm for the RCA [3436]. Others have evaluated only the distance to the LCA ostium and reported a distance of 13.5 ± 2.2 mm [37•] with correlations of r = 0.83 (P < 0.001) for RT3DTEE vs. MDCT. We have also evaluated the distance from the aortic annulus to the RCA ostium, which although possible more rarely than for the LCA, is possible in 86 % of patients, but only few data have been published regarding the RCA ostium [20••]. However, as the distance from the aortic annulus to the RCA ostium is usually greater, its occlusion is likely to be more uncommon.
Especially in patients with extensive calcification of the AV, which impairs exact measurement of the annulus–coronary ostium distance due to artifacts, RT3D-TEE poses certain disadvantages in comparison to other imaging modalities. An annulus–coronary ostium distance of less than 10 mm increases the risk of coronary occlusion during TAVI [4, 38, 39]. In addition, several studies have shown that the distance from the aortic annulus to the coronary artery ostia is reduced in patients with AS [17, 24, 34]. Measurement of the annulus–coronary ostium distance is also included in the expert consensus document of the European Association of Echocardiography and American Society of Echocardiography [20••].
In our study, accurate measurement of the distance from the aortic annulus to the LCA ostium was possible in 92 % of the patients, while others have found a feasibility up to 97.5 % [37•]. Tamborini et al. also demonstrated the use of RT3D-TEE for immediate postoperative evaluation of the distance from the LCA ostium to the aortic prosthesis, which was feasible in about 90 % of patients [37•]. Recently it was demonstrated that CT-based measurements differ from the corresponding intraoperative measurements [36]. But undoubtedly, knowing that the distance from the aortic annulus and coronary ostia is small is highly valuable for the surgeon, who can take additional care in positioning the prosthesis.
Recently increasing attention has been paid to the evaluation of the lengths of the AV cusps in relation to the distance to the ostia [17, 37•]. If they are greater than the distance from the aortic annulus to the coronary ostia, the patient is at risk of ostial coronary occlusion by the cusp when the valve is implanted.

Conclusions

The shape of the LVOT and the annulus is more frequently elliptical and not circular, which may lead to significant underestimation of the true LVOT and aortic annulus areas. RT3D-TEE provides accurate estimation of the AVA and quantification of the aortic root, with important clinical implications for the exact estimation of the AVA and annulus area for optimal selection of prosthetic valves. An advantage of RT3D-TEE is that it can be performed at the bedside or directly before or even after prosthetic valve implantation for the evaluation of the implanted prosthesis, e.g., for identifying paravalvular regurgitation. Moreover, it is a good alternative to MDCT, particularly when it is essential to decrease the burden of contrast agent administration as in patients with compromised renal function—thereby potentially reducing the incidence of acute kidney failure [40] —as well as in patients with cardiac arrhythmia. RT3D-TEE may be performed immediately before and after procedures in the hybrid operating room [37•, 41].

Compliance with Ethics Guidelines

Conflict of Interest

Björn Plicht received lecture honoraria from Abbott Vascular, Wetzlar, Germany, and travel expenses and lecture honoraria from Philips Medical.
Philipp Kahlert received personal fees from Edwards Lifesciences Inc.
Rolf Alexander Jánosi, Mareike Eißmann, Daniel Wendt, Heinz Jakob, Raimund Erbel, and Thomas Buck declare that they have no conflict of interest.

Human and Animal Rights and Informed Consent

This article does not contain any studies with human or animal subjects performed by any of the authors.
Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made.
The images or other third party material in this article are included in the article’s Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article’s Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder.

Unsere Produktempfehlungen

e.Med Interdisziplinär

Kombi-Abonnement

Jetzt e.Med zum Sonderpreis bestellen!

Für Ihren Erfolg in Klinik und Praxis - Die beste Hilfe in Ihrem Arbeitsalltag

Mit e.Med Interdisziplinär erhalten Sie Zugang zu allen CME-Fortbildungen und Fachzeitschriften auf SpringerMedizin.de.

Jetzt bestellen und 100 € sparen!

e.Med Innere Medizin

Kombi-Abonnement

Mit e.Med Innere Medizin erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Innere Medizin, den Premium-Inhalten der internistischen Fachzeitschriften, inklusive einer gedruckten internistischen Zeitschrift Ihrer Wahl.

Jetzt bestellen und 100 € sparen!

e.Med Radiologie

Kombi-Abonnement

Mit e.Med Radiologie erhalten Sie Zugang zu CME-Fortbildungen des Fachgebietes Radiologie, den Premium-Inhalten der radiologischen Fachzeitschriften, inklusive einer gedruckten Radiologie-Zeitschrift Ihrer Wahl.

Literatur
1.
Zurück zum Zitat Camm AJ, Lip GY, De Caterina R, Savelieva I, Atar D, Hohnloser G, et al. 2012 focused update of the ESC Guidelines for the management of atrial fibrillation: an update of the 2010 ESC Guidelines for the management of atrial fibrillation. Developed with the special contribution of the European Heart Rhythm Association. Eur Heart J. 2012;33(21):2719–47.PubMedCrossRef Camm AJ, Lip GY, De Caterina R, Savelieva I, Atar D, Hohnloser G, et al. 2012 focused update of the ESC Guidelines for the management of atrial fibrillation: an update of the 2010 ESC Guidelines for the management of atrial fibrillation. Developed with the special contribution of the European Heart Rhythm Association. Eur Heart J. 2012;33(21):2719–47.PubMedCrossRef
2.
Zurück zum Zitat Grube E, Sinning JM, Vahanian A. The Year in Cardiology 2013: valvular heart disease (focus on catheter-based interventions). Eur Heart J. 2014;35(8):490–5.PubMedCrossRef Grube E, Sinning JM, Vahanian A. The Year in Cardiology 2013: valvular heart disease (focus on catheter-based interventions). Eur Heart J. 2014;35(8):490–5.PubMedCrossRef
3.
Zurück zum Zitat Kasel AM, Cassese S, Bleiziffer S, Amaki M, Hahn RT, Kastrati A, et al. Standardized imaging for aortic annular sizing: implications for transcatheter valve selection. JACC Cardiovasc Imaging. 2013;6(2):249–62.PubMedCrossRef Kasel AM, Cassese S, Bleiziffer S, Amaki M, Hahn RT, Kastrati A, et al. Standardized imaging for aortic annular sizing: implications for transcatheter valve selection. JACC Cardiovasc Imaging. 2013;6(2):249–62.PubMedCrossRef
4.
Zurück zum Zitat Piazza N, de Jaegere P, Schultz C, Becker AE, Serruys PW, Anderson RH. Anatomy of the aortic valvar complex and its implications for transcatheter implantation of the aortic valve. Circ Cardiovasc Interv. 2008;1(1):74–81.PubMedCrossRef Piazza N, de Jaegere P, Schultz C, Becker AE, Serruys PW, Anderson RH. Anatomy of the aortic valvar complex and its implications for transcatheter implantation of the aortic valve. Circ Cardiovasc Interv. 2008;1(1):74–81.PubMedCrossRef
5.
Zurück zum Zitat Webb JG, Pasupati S, Humphries K, Thompson C, Altwegg L, Moss R, et al. Percutaneous transarterial aortic valve replacement in selected high-risk patients with aortic stenosis. Circulation. 2007;116(7):755–63.PubMedCrossRef Webb JG, Pasupati S, Humphries K, Thompson C, Altwegg L, Moss R, et al. Percutaneous transarterial aortic valve replacement in selected high-risk patients with aortic stenosis. Circulation. 2007;116(7):755–63.PubMedCrossRef
6.
Zurück zum Zitat Patsalis PC, Konorza TFM, Al-Rashid F, Plicht B, Riebisch M, Wendt D, et al. Incidence, outcome and correlates of residual paravalvular aortic regurgitation after transcatheter aortic valve implantation and importance of haemodynamic assessment. EuroIntervention. 2013;8(12):1398–406.PubMedCrossRef Patsalis PC, Konorza TFM, Al-Rashid F, Plicht B, Riebisch M, Wendt D, et al. Incidence, outcome and correlates of residual paravalvular aortic regurgitation after transcatheter aortic valve implantation and importance of haemodynamic assessment. EuroIntervention. 2013;8(12):1398–406.PubMedCrossRef
7.
Zurück zum Zitat Baumgartner H, Hung J, Bermejo J, Chambers JB, Evangelista A, Griffin BP, et al. Echocardiographic assessment of valve stenosis: EAE/ASE recommendations for clinical practice. Eur J Echocardiogr. 2009;10(1):1–25.PubMedCrossRef Baumgartner H, Hung J, Bermejo J, Chambers JB, Evangelista A, Griffin BP, et al. Echocardiographic assessment of valve stenosis: EAE/ASE recommendations for clinical practice. Eur J Echocardiogr. 2009;10(1):1–25.PubMedCrossRef
8.
Zurück zum Zitat Vahanian A, Baumgartner H, Bax J, Butchart E, Dion R, Filippatos G, et al. Guidelines on the management of valvular heart disease: The Task Force on the Management of Valvular Heart Disease of the European Society of Cardiology. Eur Heart J. 2006;28(2):230–68.CrossRef Vahanian A, Baumgartner H, Bax J, Butchart E, Dion R, Filippatos G, et al. Guidelines on the management of valvular heart disease: The Task Force on the Management of Valvular Heart Disease of the European Society of Cardiology. Eur Heart J. 2006;28(2):230–68.CrossRef
9.
Zurück zum Zitat Gersh BJ, Maron BJ, Bonow RO, Dearani JA, Fifer MA, Link MS, et al. 2011 ACCF/AHA Guideline for the Diagnosis and Treatment of Hypertrophic Cardiomyopathy: A Report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines. Circulation. 2011;124(24):e783–831.PubMedCrossRef Gersh BJ, Maron BJ, Bonow RO, Dearani JA, Fifer MA, Link MS, et al. 2011 ACCF/AHA Guideline for the Diagnosis and Treatment of Hypertrophic Cardiomyopathy: A Report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines. Circulation. 2011;124(24):e783–831.PubMedCrossRef
10.
Zurück zum Zitat Zoghbi WA, Farmer KL, Soto JG, Nelson JG, Quinones MA. Accurate noninvasive quantification of stenotic aortic valve area by Doppler echocardiography. Circulation. 1986;73(3):452–9.PubMedCrossRef Zoghbi WA, Farmer KL, Soto JG, Nelson JG, Quinones MA. Accurate noninvasive quantification of stenotic aortic valve area by Doppler echocardiography. Circulation. 1986;73(3):452–9.PubMedCrossRef
11.
Zurück zum Zitat Grayburn PA. Assessment of low-gradient aortic stenosis with dobutamine. Circulation. 2006;113(5):604–6.PubMedCrossRef Grayburn PA. Assessment of low-gradient aortic stenosis with dobutamine. Circulation. 2006;113(5):604–6.PubMedCrossRef
12.
Zurück zum Zitat Nakai H, Takeuchi M, Yoshitani H, Kaku K, Haruki N, Otsuji Y. Pitfalls of anatomical aortic valve area measurements using two-dimensional transoesophageal echocardiography and the potential of three-dimensional transoesophageal echocardiography. Eur J Echocardiogr. 2010;11(4):369–76.PubMedCrossRef Nakai H, Takeuchi M, Yoshitani H, Kaku K, Haruki N, Otsuji Y. Pitfalls of anatomical aortic valve area measurements using two-dimensional transoesophageal echocardiography and the potential of three-dimensional transoesophageal echocardiography. Eur J Echocardiogr. 2010;11(4):369–76.PubMedCrossRef
13.
Zurück zum Zitat Bernard Y, Meneveau N, Vuillemenot A, Magnin D, Anguenot T, Schiele F, et al. Planimetry of aortic valve area using multiplane transoesophageal echocardiography is not a reliable method for assessing severity of aortic stenosis. Heart. 1997;78(1):68–73.PubMedPubMedCentral Bernard Y, Meneveau N, Vuillemenot A, Magnin D, Anguenot T, Schiele F, et al. Planimetry of aortic valve area using multiplane transoesophageal echocardiography is not a reliable method for assessing severity of aortic stenosis. Heart. 1997;78(1):68–73.PubMedPubMedCentral
14.
Zurück zum Zitat Plicht B, Janosi RA, Kahlert P, Buck T, Erbel R. Aortic valve with ‘calcified valvular stenosis’: precise clarification of morphology and severity using transesophageal three dimensional echocardiography. Herz. 2008;33(4):315–6.PubMedCrossRef Plicht B, Janosi RA, Kahlert P, Buck T, Erbel R. Aortic valve with ‘calcified valvular stenosis’: precise clarification of morphology and severity using transesophageal three dimensional echocardiography. Herz. 2008;33(4):315–6.PubMedCrossRef
15.
Zurück zum Zitat La Morena DG, Saura D, Oliva MJ, Soria F, Gonzalez J, Garcia M, et al. Real-time three-dimensional transoesophageal echocardiography in the assessment of aortic valve stenosis. Eur J Echocardiogr. 2010;11(1):9–13.PubMedCrossRef La Morena DG, Saura D, Oliva MJ, Soria F, Gonzalez J, Garcia M, et al. Real-time three-dimensional transoesophageal echocardiography in the assessment of aortic valve stenosis. Eur J Echocardiogr. 2010;11(1):9–13.PubMedCrossRef
16.
Zurück zum Zitat Garcia D, Dumesnil JG, Durand L-G, Kadem L, Pibarot P. Discrepancies between catheter and Doppler estimates of valve effective orifice area can be predicted from the pressure recovery phenomenon. J Am Coll Cardiol. 2003;41(3):435–42.PubMedCrossRef Garcia D, Dumesnil JG, Durand L-G, Kadem L, Pibarot P. Discrepancies between catheter and Doppler estimates of valve effective orifice area can be predicted from the pressure recovery phenomenon. J Am Coll Cardiol. 2003;41(3):435–42.PubMedCrossRef
17.
Zurück zum Zitat Calleja A, Thavendiranathan P, Ionasec RI, Houle H, Liu S, Voigt I, et al. Automated quantitative 3-dimensional modeling of the aortic valve and root by 3-dimensional transesophageal echocardiography in normals, aortic regurgitation, and aortic stenosis: comparison to computed tomography in normals and clinical implications. Circ Cardiovasc Imaging. 2013;6(1):99–108.PubMedCrossRef Calleja A, Thavendiranathan P, Ionasec RI, Houle H, Liu S, Voigt I, et al. Automated quantitative 3-dimensional modeling of the aortic valve and root by 3-dimensional transesophageal echocardiography in normals, aortic regurgitation, and aortic stenosis: comparison to computed tomography in normals and clinical implications. Circ Cardiovasc Imaging. 2013;6(1):99–108.PubMedCrossRef
18.
Zurück zum Zitat Detaint D, Lepage L, Himbert D, Brochet E, Messika-Zeitoun D, Iung B, et al. Determinants of significant paravalvular regurgitation after transcatheter aortic valve implantation. JACC Cardiovasc Interv. 2009;2(9):821–7.PubMedCrossRef Detaint D, Lepage L, Himbert D, Brochet E, Messika-Zeitoun D, Iung B, et al. Determinants of significant paravalvular regurgitation after transcatheter aortic valve implantation. JACC Cardiovasc Interv. 2009;2(9):821–7.PubMedCrossRef
19.
Zurück zum Zitat Janosi RA, Kahlert P, Plicht B, Wendt D, Eggebrecht H, Erbel R, et al. Measurement of the aortic annulus size by real-time three-dimensional transesophageal echocardiography. Minim Invasive Ther Allied Technol. 2011;20(2):85–94.PubMedCrossRef Janosi RA, Kahlert P, Plicht B, Wendt D, Eggebrecht H, Erbel R, et al. Measurement of the aortic annulus size by real-time three-dimensional transesophageal echocardiography. Minim Invasive Ther Allied Technol. 2011;20(2):85–94.PubMedCrossRef
20.••
Zurück zum Zitat Zamorano JL, Badano LP, Bruce C, Chan KL, Goncalves A, Hahn RT, et al. EAE/ASE recommendations for the use of echocardiography in new transcatheter interventions for valvular heart disease. Eur J Echocardiogr. 2011;12(8):557–84. The current recommendations for the use of echocardiography in transcatheter interventions for valvular heart disease developed by the European Association of Echocardiography (EAE) and the American Society of Echocardiography (ASE).PubMedCrossRef Zamorano JL, Badano LP, Bruce C, Chan KL, Goncalves A, Hahn RT, et al. EAE/ASE recommendations for the use of echocardiography in new transcatheter interventions for valvular heart disease. Eur J Echocardiogr. 2011;12(8):557–84. The current recommendations for the use of echocardiography in transcatheter interventions for valvular heart disease developed by the European Association of Echocardiography (EAE) and the American Society of Echocardiography (ASE).PubMedCrossRef
21.
Zurück zum Zitat Buck T, Breithardt O-A, Faber L, Fehske W, Flachskampf FA, Franke A, et al. Manual zur Indikation und Durchführung der Echokardiographie. Clin Res Cardiol Suppl. 2009;4 Suppl 1:3–51.CrossRef Buck T, Breithardt O-A, Faber L, Fehske W, Flachskampf FA, Franke A, et al. Manual zur Indikation und Durchführung der Echokardiographie. Clin Res Cardiol Suppl. 2009;4 Suppl 1:3–51.CrossRef
22.
Zurück zum Zitat Khaw AV, Bardeleben RS, Strasser C, Mohr-Kahaly S, Blankenberg S, Espinola-Klein C, et al. Direct measurement of left ventricular outflow tract by transthoracic real-time 3D-echocardiography increases accuracy in assessment of aortic valve stenosis. Int J Cardiol. 2009;136(1):64–71.PubMedCrossRef Khaw AV, Bardeleben RS, Strasser C, Mohr-Kahaly S, Blankenberg S, Espinola-Klein C, et al. Direct measurement of left ventricular outflow tract by transthoracic real-time 3D-echocardiography increases accuracy in assessment of aortic valve stenosis. Int J Cardiol. 2009;136(1):64–71.PubMedCrossRef
23.
Zurück zum Zitat Doddamani S, Grushko MJ, Makaryus AN, Jain VR, Bello R, Friedman MA, et al. Demonstration of left ventricular outflow tract eccentricity by 64-slice multi-detector CT. Int J Cardiovasc Imaging. 2009;25(2):175–81.PubMedCrossRef Doddamani S, Grushko MJ, Makaryus AN, Jain VR, Bello R, Friedman MA, et al. Demonstration of left ventricular outflow tract eccentricity by 64-slice multi-detector CT. Int J Cardiovasc Imaging. 2009;25(2):175–81.PubMedCrossRef
24.
Zurück zum Zitat Otani K, Takeuchi M, Kaku K, Sugeng L, Yoshitani H, Haruki N, et al. Assessment of the aortic root using real-time 3D transesophageal echocardiography. Circ J. 2010;74(12):2649–57.PubMedCrossRef Otani K, Takeuchi M, Kaku K, Sugeng L, Yoshitani H, Haruki N, et al. Assessment of the aortic root using real-time 3D transesophageal echocardiography. Circ J. 2010;74(12):2649–57.PubMedCrossRef
25.
Zurück zum Zitat Bednarz JE, Krauss D, Lang RM. An echocardiographic approach to the assessment of aortic stenosis. J Am Soc Echocardiogr. 1996;9(3):286–94.PubMedCrossRef Bednarz JE, Krauss D, Lang RM. An echocardiographic approach to the assessment of aortic stenosis. J Am Soc Echocardiogr. 1996;9(3):286–94.PubMedCrossRef
26.
Zurück zum Zitat Ng AC, Delgado V, van der Kley F, Shanks M, van de Veire NRL, Bertini M, et al. Comparison of aortic root dimensions and geometries before and after transcatheter aortic valve implantation by 2- and 3-dimensional transesophageal echocardiography and multislice computed tomography. Circ Cardiovasc Imaging. 2010;3(1):94–102.PubMedCrossRef Ng AC, Delgado V, van der Kley F, Shanks M, van de Veire NRL, Bertini M, et al. Comparison of aortic root dimensions and geometries before and after transcatheter aortic valve implantation by 2- and 3-dimensional transesophageal echocardiography and multislice computed tomography. Circ Cardiovasc Imaging. 2010;3(1):94–102.PubMedCrossRef
27.
Zurück zum Zitat Tanaka K, Makaryus AN, Wolff SD. Correlation of aortic valve area obtained by the velocity-encoded phase contrast continuity method to direct planimetry using cardiovascular magnetic resonance. J Cardiovasc Magn Reson. 2007;9(5):799–805.PubMedCrossRef Tanaka K, Makaryus AN, Wolff SD. Correlation of aortic valve area obtained by the velocity-encoded phase contrast continuity method to direct planimetry using cardiovascular magnetic resonance. J Cardiovasc Magn Reson. 2007;9(5):799–805.PubMedCrossRef
28.
Zurück zum Zitat Saitoh T, Shiota M, Izumo M, Gurudevan SV, Tolstrup K, Siegel RJ, et al. Comparison of left ventricular outflow geometry and aortic valve area in patients with aortic stenosis by 2-dimensional versus 3-dimensional echocardiography. Am J Cardiol. 2012;109(11):1626–31.PubMedCrossRef Saitoh T, Shiota M, Izumo M, Gurudevan SV, Tolstrup K, Siegel RJ, et al. Comparison of left ventricular outflow geometry and aortic valve area in patients with aortic stenosis by 2-dimensional versus 3-dimensional echocardiography. Am J Cardiol. 2012;109(11):1626–31.PubMedCrossRef
29.
Zurück zum Zitat Pérez De Isla L, Zamorano J, Malangatana G, Martínez Quesada M, Almeria C, Rodrigo JL. Morphological determinants of subaortic stenosis in hypertrophic cardiomyopathy: insights from real-time 3-dimensional echocardiography. J Am Soc Echocardiogr. 2005;18(8):802–4.PubMedCrossRef Pérez De Isla L, Zamorano J, Malangatana G, Martínez Quesada M, Almeria C, Rodrigo JL. Morphological determinants of subaortic stenosis in hypertrophic cardiomyopathy: insights from real-time 3-dimensional echocardiography. J Am Soc Echocardiogr. 2005;18(8):802–4.PubMedCrossRef
30.
Zurück zum Zitat Ge S, Warner JG, Abraham TP, Kon ND, Brooker RF, Nomeir AM, et al. Three-dimensional surface area of the aortic valve orifice by three-dimensional echocardiography: clinical validation of a novel index for assessment of aortic stenosis. Am Heart J. 1998;136(6):1042–50.PubMedCrossRef Ge S, Warner JG, Abraham TP, Kon ND, Brooker RF, Nomeir AM, et al. Three-dimensional surface area of the aortic valve orifice by three-dimensional echocardiography: clinical validation of a novel index for assessment of aortic stenosis. Am Heart J. 1998;136(6):1042–50.PubMedCrossRef
31.
Zurück zum Zitat Sugeng L, Shernan SK, Salgo IS, Weinert L, Shook D, Raman J, et al. Live 3-dimensional transesophageal echocardiography initial experience using the fully-sampled matrix array probe. J Am Coll Cardiol. 2008;52(6):446–9.PubMedCrossRef Sugeng L, Shernan SK, Salgo IS, Weinert L, Shook D, Raman J, et al. Live 3-dimensional transesophageal echocardiography initial experience using the fully-sampled matrix array probe. J Am Coll Cardiol. 2008;52(6):446–9.PubMedCrossRef
32.
Zurück zum Zitat Wu VC, Kaku K, Takeuchi M, Otani K, Yoshitani H, Tamura M, et al. Aortic root geometry in patients with aortic stenosis assessed by real-time three-dimensional transesophageal echocardiography. J Am Soc Echocardiogr. 2014;27(1):32–41.PubMedCrossRef Wu VC, Kaku K, Takeuchi M, Otani K, Yoshitani H, Tamura M, et al. Aortic root geometry in patients with aortic stenosis assessed by real-time three-dimensional transesophageal echocardiography. J Am Soc Echocardiogr. 2014;27(1):32–41.PubMedCrossRef
33.
Zurück zum Zitat Mukherjee C, Hein F, Holzhey D, Lukas L, Mende M, Kaisers UX, et al. Is real time 3D transesophageal echocardiography a feasible approach to detect coronary ostium during transapical aortic valve implantation? J Cardiothorac Vasc Anesth. 2013;27(4):654–9.PubMedCrossRef Mukherjee C, Hein F, Holzhey D, Lukas L, Mende M, Kaisers UX, et al. Is real time 3D transesophageal echocardiography a feasible approach to detect coronary ostium during transapical aortic valve implantation? J Cardiothorac Vasc Anesth. 2013;27(4):654–9.PubMedCrossRef
34.
Zurück zum Zitat Akhtar M, Tuzcu EM, Kapadia SR, Svensson LG, Greenberg RK, Roselli EE, et al. Aortic root morphology in patients undergoing percutaneous aortic valve replacement: evidence of aortic root remodeling. J Thorac Cardiovasc Surg. 2009;137(4):950–6.PubMedCrossRef Akhtar M, Tuzcu EM, Kapadia SR, Svensson LG, Greenberg RK, Roselli EE, et al. Aortic root morphology in patients undergoing percutaneous aortic valve replacement: evidence of aortic root remodeling. J Thorac Cardiovasc Surg. 2009;137(4):950–6.PubMedCrossRef
35.
Zurück zum Zitat Stolzmann P, Knight J, Desbiolles L, Maier W, Scheffel H, Plass A, et al. Remodelling of the aortic root in severe tricuspid aortic stenosis: implications for transcatheter aortic valve implantation. Eur Radiol. 2009;19(6):1316–23.PubMedCrossRef Stolzmann P, Knight J, Desbiolles L, Maier W, Scheffel H, Plass A, et al. Remodelling of the aortic root in severe tricuspid aortic stenosis: implications for transcatheter aortic valve implantation. Eur Radiol. 2009;19(6):1316–23.PubMedCrossRef
36.
Zurück zum Zitat Wendt D, Thielmann M, Price V, Kahlert P, Kühl H, Kamler M, et al. Coronary ostium topography: an implication for transcatheter aortic valve implantation? Minim Invasive Ther Allied Technol. 2013;22(2):65–72.PubMedCrossRef Wendt D, Thielmann M, Price V, Kahlert P, Kühl H, Kamler M, et al. Coronary ostium topography: an implication for transcatheter aortic valve implantation? Minim Invasive Ther Allied Technol. 2013;22(2):65–72.PubMedCrossRef
37.•
Zurück zum Zitat Tamborini G, Fusini L, Gripari P, Muratori M, Cefalu C, Maffessanti F, et al. Feasibility and accuracy of 3DTEE versus CT for the evaluation of aortic valve annulus to left main ostium distance before transcatheter aortic valve implantation. JACC Cardiovasc Imaging. 2012;5(6):579–88. In a large series of patients undergoing TAVI the authors compared the feasibility and accuracy of 3D-TEE with that of CT in the measurement of left main coronary ostium to the aortic annulus distance and to the aortic prosthesis postoperatively.PubMedCrossRef Tamborini G, Fusini L, Gripari P, Muratori M, Cefalu C, Maffessanti F, et al. Feasibility and accuracy of 3DTEE versus CT for the evaluation of aortic valve annulus to left main ostium distance before transcatheter aortic valve implantation. JACC Cardiovasc Imaging. 2012;5(6):579–88. In a large series of patients undergoing TAVI the authors compared the feasibility and accuracy of 3D-TEE with that of CT in the measurement of left main coronary ostium to the aortic annulus distance and to the aortic prosthesis postoperatively.PubMedCrossRef
38.
Zurück zum Zitat Masson J-B, Lee M, Boone RH, Al Ali A, Al Bugami S, Hamburger J, et al. Impact of coronary artery disease on outcomes after transcatheter aortic valve implantation. Catheter Cardiovasc Interv. 2010;76(2):165–73.PubMedCrossRef Masson J-B, Lee M, Boone RH, Al Ali A, Al Bugami S, Hamburger J, et al. Impact of coronary artery disease on outcomes after transcatheter aortic valve implantation. Catheter Cardiovasc Interv. 2010;76(2):165–73.PubMedCrossRef
39.
Zurück zum Zitat Gautier M, Pepin M, Himbert D, Ducrocq G, Iung B, Dilly M-P, et al. Impact of coronary artery disease on indications for transcatheter aortic valve implantation and on procedural outcomes. EuroIntervention. 2011;7(5):549–55.PubMedCrossRef Gautier M, Pepin M, Himbert D, Ducrocq G, Iung B, Dilly M-P, et al. Impact of coronary artery disease on indications for transcatheter aortic valve implantation and on procedural outcomes. EuroIntervention. 2011;7(5):549–55.PubMedCrossRef
40.
Zurück zum Zitat Madershahian N, Scherner M, Rudolph T, Slottosch I, Grundmann F, Kuhn E, et al. Impact of sex on the risk to develop contrast-induced nephropathy after transcatheter aortic valve implantation. Innovations (Phila). 2014;9(2):87–92. discussion 92.CrossRef Madershahian N, Scherner M, Rudolph T, Slottosch I, Grundmann F, Kuhn E, et al. Impact of sex on the risk to develop contrast-induced nephropathy after transcatheter aortic valve implantation. Innovations (Phila). 2014;9(2):87–92. discussion 92.CrossRef
41.
Zurück zum Zitat Janosi RA, Kahlert P, Plicht B, Böse D, Wendt D, Thielmann M, et al. Guidance of percutaneous transcatheter aortic valve implantation by real-time three-dimensional transesophageal echocardiography – a single-center experience. Minim Invasive Ther Allied Technol. 2009;18(3):142–8.PubMedCrossRef Janosi RA, Kahlert P, Plicht B, Böse D, Wendt D, Thielmann M, et al. Guidance of percutaneous transcatheter aortic valve implantation by real-time three-dimensional transesophageal echocardiography – a single-center experience. Minim Invasive Ther Allied Technol. 2009;18(3):142–8.PubMedCrossRef
Metadaten
Titel
Quantitative Analysis of Aortic Valve Stenosis and Aortic Root Dimensions by Three-Dimensional Echocardiography in Patients Scheduled for Transcutaneous Aortic Valve Implantation
verfasst von
Rolf Alexander Jánosi
Björn Plicht
Philipp Kahlert
Mareike Eißmann
Daniel Wendt
Heinz Jakob
Raimund Erbel
Thomas Buck
Publikationsdatum
01.11.2014
Verlag
Springer US
Erschienen in
Current Cardiovascular Imaging Reports / Ausgabe 11/2014
Print ISSN: 1941-9066
Elektronische ISSN: 1941-9074
DOI
https://doi.org/10.1007/s12410-014-9296-7

Weitere Artikel der Ausgabe 11/2014

Current Cardiovascular Imaging Reports 11/2014 Zur Ausgabe

Nach Herzinfarkt mit Typ-1-Diabetes schlechtere Karten als mit Typ 2?

29.05.2024 Herzinfarkt Nachrichten

Bei Menschen mit Typ-2-Diabetes sind die Chancen, einen Myokardinfarkt zu überleben, in den letzten 15 Jahren deutlich gestiegen – nicht jedoch bei Betroffenen mit Typ 1.

Erhöhtes Risiko fürs Herz unter Checkpointhemmer-Therapie

28.05.2024 Nebenwirkungen der Krebstherapie Nachrichten

Kardiotoxische Nebenwirkungen einer Therapie mit Immuncheckpointhemmern mögen selten sein – wenn sie aber auftreten, wird es für Patienten oft lebensgefährlich. Voruntersuchung und Monitoring sind daher obligat.

GLP-1-Agonisten können Fortschreiten diabetischer Retinopathie begünstigen

24.05.2024 Diabetische Retinopathie Nachrichten

Möglicherweise hängt es von der Art der Diabetesmedikamente ab, wie hoch das Risiko der Betroffenen ist, dass sich sehkraftgefährdende Komplikationen verschlimmern.

TAVI versus Klappenchirurgie: Neue Vergleichsstudie sorgt für Erstaunen

21.05.2024 TAVI Nachrichten

Bei schwerer Aortenstenose und obstruktiver KHK empfehlen die Leitlinien derzeit eine chirurgische Kombi-Behandlung aus Klappenersatz plus Bypass-OP. Diese Empfehlung wird allerdings jetzt durch eine aktuelle Studie infrage gestellt – mit überraschender Deutlichkeit.

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.