Effect of catheter movement and contact during application of radiofrequency energy on ablation lesion characteristics

verfasst von: Matthew D. Olson, Nicholas Phreaner, Joseph L. Schuller, Duy T. Nguyen, David F. Katz, Ryan G. Aleong, Wendy S. Tzou, Raphael Sung, Paul D. Varosy, William H. Sauer

Erschienen in: Journal of Interventional Cardiac Electrophysiology | Ausgabe 2/2013

Einloggen, um Zugang zu erhalten

Abstract

Background

The efficient delivery of radiofrequency (RF) energy through an endocardial ablation catheter is affected by variable tissue contact due to cardiac motion with myocardial contraction and respiration. In addition, many operators intentionally move an ablation catheter during the delivery of radiofrequency energy when targeting specific arrhythmias that require lines of conduction block such as atrial flutter and atrial fibrillation. We sought to characterize and quantify any effects of catheter movement and intermittent ablation catheter contact on lesion characteristics.

Methods

An ex vivo model consisting of recently excised viable bovine myocardium, a circulating saline bath at 37 °C, a submersible load cell, and a deflectable sheath with an ablation catheter was assembled. A stepper motor attached to an ablation catheter apparatus was programmed to simulate linear drag lesions and series of point lesions with variable contact using constant force. Lesion volumes were analyzed using a digital micrometer by measuring depth, max width, depth at max width, and surface width and compared.

Results

The drag lesion was significantly larger than a point-by-point linear lesion using a constant force of 15 g (2,088 ± 122 mm3 vs. 1,595 ± 121.6; p = 0.01) when controlling for RF time and power. For single point lesion assessment, constant contact lesions were significantly larger than lesions created with intermittent contact using the same duration of RF (194 ± 68 mm3 vs. 112.5 ± 53; p = 0.03). There was no significant difference in lesion size between the constant contact at 60 s and 90-s intermittent contact lesions (194 ± 68 mm3 vs.186 ± 69).

Conclusions

In our ex vivo model, externally irrigated radiofrequency catheters produced drag lesion volumes equal to or larger than those created by a point-by-point method. We also found decreased lesion size due to intermittent contact can be overcome by increasing duration of ablation time.
Literatur
1.
Zurück zum Zitat Nakagawa, H., Yamanashi, W. S., Pitha, J. V., Arruda, M., Wang, X., Ohtomo, K., et al. (1995). Comparison of in vivo tissue temperature profile and lesion geometry for radiofrequency ablation with a saline-irrigated electrode versus temperature control in a canine thigh muscle preparation. Circulation, 91, 2264–2273.PubMedCrossRef Nakagawa, H., Yamanashi, W. S., Pitha, J. V., Arruda, M., Wang, X., Ohtomo, K., et al. (1995). Comparison of in vivo tissue temperature profile and lesion geometry for radiofrequency ablation with a saline-irrigated electrode versus temperature control in a canine thigh muscle preparation. Circulation, 91, 2264–2273.PubMedCrossRef
2.
Zurück zum Zitat Wood, M. A., Goldberg, S. M., Parvez, B., Pathak, V., Holland, K., Ellenbogen, A. L., et al. (2009). Effect of electrode orientation on lesion sizes produced by irrigated radiofrequency ablation catheters. Journal of Cardiovascular Electrophysiology, 20, 1262–1268.PubMedCrossRef Wood, M. A., Goldberg, S. M., Parvez, B., Pathak, V., Holland, K., Ellenbogen, A. L., et al. (2009). Effect of electrode orientation on lesion sizes produced by irrigated radiofrequency ablation catheters. Journal of Cardiovascular Electrophysiology, 20, 1262–1268.PubMedCrossRef
3.
Zurück zum Zitat Thiagalingam, A., D’avila, A., Foley, L., Guerrero, J. L., Lambert, H., Leo, G., et al. (2010). Importance of catheter contact force during irrigated radiofrequency ablation: evaluation in a porcine ex vivo model using a force-sensing catheter. Journal of Cardiovascular Electrophysiology, 21, 806–811.PubMed Thiagalingam, A., D’avila, A., Foley, L., Guerrero, J. L., Lambert, H., Leo, G., et al. (2010). Importance of catheter contact force during irrigated radiofrequency ablation: evaluation in a porcine ex vivo model using a force-sensing catheter. Journal of Cardiovascular Electrophysiology, 21, 806–811.PubMed
4.
Zurück zum Zitat Shah, D. C., Lambert, H., Nakagawa, H., Langenkamp, A., Aeby, N., & Leo, G. (2010). Area under the real-time contact force curve (force–time integral) predicts radiofrequency lesion size in an in vitro contractile model. Journal of Cardiovascular Electrophysiology, 21, 1038–1043.PubMedCrossRef Shah, D. C., Lambert, H., Nakagawa, H., Langenkamp, A., Aeby, N., & Leo, G. (2010). Area under the real-time contact force curve (force–time integral) predicts radiofrequency lesion size in an in vitro contractile model. Journal of Cardiovascular Electrophysiology, 21, 1038–1043.PubMedCrossRef
5.
Zurück zum Zitat Demazumder, D., Mirotznik, M. S., & Schwartzman, D. (2001). Comparison of irrigated electrode designs for radiofrequency ablation of myocardium. Journal of Interventional Cardiac Electrophysiology, 5, 391–400.PubMedCrossRef Demazumder, D., Mirotznik, M. S., & Schwartzman, D. (2001). Comparison of irrigated electrode designs for radiofrequency ablation of myocardium. Journal of Interventional Cardiac Electrophysiology, 5, 391–400.PubMedCrossRef
6.
Zurück zum Zitat Nakagawa, H., Wittkampf, F. H. M., Yamanashi, W. S., Pitha, J. V., Imai, S., Campbell, B., et al. (1998). Inverse relationship between electrode size and lesion size during radiofrequency ablation with active electrode cooling. Circulation, 98, 458–465.PubMedCrossRef Nakagawa, H., Wittkampf, F. H. M., Yamanashi, W. S., Pitha, J. V., Imai, S., Campbell, B., et al. (1998). Inverse relationship between electrode size and lesion size during radiofrequency ablation with active electrode cooling. Circulation, 98, 458–465.PubMedCrossRef
7.
Zurück zum Zitat Jaïs, P., Hocini, M., Hsu, L.-F., Sanders, P., Scavee, C., Weerasooriya, R., et al. (2004). Technique and results of linear ablation at the mitral isthmus. Circulation, 110, 2996–3002.PubMedCrossRef Jaïs, P., Hocini, M., Hsu, L.-F., Sanders, P., Scavee, C., Weerasooriya, R., et al. (2004). Technique and results of linear ablation at the mitral isthmus. Circulation, 110, 2996–3002.PubMedCrossRef
8.
Zurück zum Zitat Miyazaki, S., Takahashi, A., Kuwahara, T., Kobori, A., Yokoyama, Y., Nozato, T., et al. (2007). Randomized comparison of the continuous vs point-by-point radiofrequency ablation of the cavotricuspid isthmus for atrial flutter. Circulation Journal, 71, 1922–1926.PubMedCrossRef Miyazaki, S., Takahashi, A., Kuwahara, T., Kobori, A., Yokoyama, Y., Nozato, T., et al. (2007). Randomized comparison of the continuous vs point-by-point radiofrequency ablation of the cavotricuspid isthmus for atrial flutter. Circulation Journal, 71, 1922–1926.PubMedCrossRef
9.
Zurück zum Zitat Okumura, Y., Johnson, S. B., Bunch, T. J., Henz, B. D., O’brien, C. J., & Packer, D. L. (2008). A systematical analysis of in vivo contact forces on virtual catheter tip/tissue surface contact during cardiac mapping and intervention. Journal of Cardiovascular Electrophysiology, 19, 632–640.PubMedCrossRef Okumura, Y., Johnson, S. B., Bunch, T. J., Henz, B. D., O’brien, C. J., & Packer, D. L. (2008). A systematical analysis of in vivo contact forces on virtual catheter tip/tissue surface contact during cardiac mapping and intervention. Journal of Cardiovascular Electrophysiology, 19, 632–640.PubMedCrossRef
10.
Zurück zum Zitat Yokokawa, M., Bhandari, A. K., Tada, H., Suzuki, A., Kawamura, M., Ho, I., et al. (2011). Comparison of the point-by-point versus catheter dragging technique for curative radiofrequency ablation of atrial fibrillation. Pacing and Clinical Electrophysiology, 34, 15–22.PubMedCrossRef Yokokawa, M., Bhandari, A. K., Tada, H., Suzuki, A., Kawamura, M., Ho, I., et al. (2011). Comparison of the point-by-point versus catheter dragging technique for curative radiofrequency ablation of atrial fibrillation. Pacing and Clinical Electrophysiology, 34, 15–22.PubMedCrossRef
11.
Zurück zum Zitat Ernst, S., Ouyang, F., Löber, F., Antz, M., & Kuck, K.-H. (2003). Catheter-induced linear lesions in the left atrium in patients with atrial fibrillation: an electroanatomic study. Journal of the American College of Cardiology, 42, 1271–1282.PubMedCrossRef Ernst, S., Ouyang, F., Löber, F., Antz, M., & Kuck, K.-H. (2003). Catheter-induced linear lesions in the left atrium in patients with atrial fibrillation: an electroanatomic study. Journal of the American College of Cardiology, 42, 1271–1282.PubMedCrossRef
12.
Zurück zum Zitat van Rensburg, H., Willems, R., Holemans, P., Anné, W., Ector, H., & Heidbüchel, H. (2002). Simultaneous creation and evaluation of linear radiofrequency lesions. Journal of Interventional Cardiac Electrophysiology, 6, 215–224.PubMedCrossRef van Rensburg, H., Willems, R., Holemans, P., Anné, W., Ector, H., & Heidbüchel, H. (2002). Simultaneous creation and evaluation of linear radiofrequency lesions. Journal of Interventional Cardiac Electrophysiology, 6, 215–224.PubMedCrossRef
13.
Zurück zum Zitat Everett, T. H., IV, Lee, K. W., Wilson, E. E., Guerra, J. M., Varosy, P. D., & Olgin, J. E. (2009). Safety profiles and lesion size of different radiofrequency ablation technologies: a comparison of large tip, open and closed irrigation catheters. Journal of Cardiovascular Electrophysiology, 20, 325–335.PubMedCrossRef Everett, T. H., IV, Lee, K. W., Wilson, E. E., Guerra, J. M., Varosy, P. D., & Olgin, J. E. (2009). Safety profiles and lesion size of different radiofrequency ablation technologies: a comparison of large tip, open and closed irrigation catheters. Journal of Cardiovascular Electrophysiology, 20, 325–335.PubMedCrossRef
14.
Zurück zum Zitat Demazumder, D., Mirotznik, M. S., & Schwartzman, D. (2001). Biophysics of radiofrequency ablation using an irrigated electrode. Journal of Interventional Cardiac Electrophysiology, 5, 377–389.PubMedCrossRef Demazumder, D., Mirotznik, M. S., & Schwartzman, D. (2001). Biophysics of radiofrequency ablation using an irrigated electrode. Journal of Interventional Cardiac Electrophysiology, 5, 377–389.PubMedCrossRef
15.
Zurück zum Zitat Su, W. W., Johnson, S. B., Jain, M. K., Hall, J., & Packer, D. L. (2005). Creating continuous linear lesions in the atria: a comparison of the multipolar ablation technique versus the conventional drag-and-burn. Journal of Cardiovascular Electrophysiology, 16, 905–911.PubMedCrossRef Su, W. W., Johnson, S. B., Jain, M. K., Hall, J., & Packer, D. L. (2005). Creating continuous linear lesions in the atria: a comparison of the multipolar ablation technique versus the conventional drag-and-burn. Journal of Cardiovascular Electrophysiology, 16, 905–911.PubMedCrossRef
16.
Zurück zum Zitat McRury, I. D., Panescu, D., Mitchell, M. A., & Haines, D. E. (1997). Nonuniform heating during radiofrequency catheter ablation with long electrodes monitoring the edge effect. Circulation, 96, 4057–4064.PubMedCrossRef McRury, I. D., Panescu, D., Mitchell, M. A., & Haines, D. E. (1997). Nonuniform heating during radiofrequency catheter ablation with long electrodes monitoring the edge effect. Circulation, 96, 4057–4064.PubMedCrossRef
17.
Zurück zum Zitat Yokoyama, K., Nakagawa, H., Shah, D. C., Lambert, H., Leo, G., Aeby, N., et al. (2008). Novel contact force sensor incorporated in irrigated radiofrequency ablation catheter predicts lesion size and incidence of steam pop and thrombus. Circulation. Arrhythmia and Electrophysiology, 1, 354–362.PubMedCrossRef Yokoyama, K., Nakagawa, H., Shah, D. C., Lambert, H., Leo, G., Aeby, N., et al. (2008). Novel contact force sensor incorporated in irrigated radiofrequency ablation catheter predicts lesion size and incidence of steam pop and thrombus. Circulation. Arrhythmia and Electrophysiology, 1, 354–362.PubMedCrossRef
Metadaten
Titel
Effect of catheter movement and contact during application of radiofrequency energy on ablation lesion characteristics
verfasst von
Matthew D. Olson
Nicholas Phreaner
Joseph L. Schuller
Duy T. Nguyen
David F. Katz
Ryan G. Aleong
Wendy S. Tzou
Raphael Sung
Paul D. Varosy
William H. Sauer
Publikationsdatum
01.11.2013
Verlag
Springer US
Erschienen in
Journal of Interventional Cardiac Electrophysiology / Ausgabe 2/2013
Print ISSN: 1383-875X
Elektronische ISSN: 1572-8595
DOI
https://doi.org/10.1007/s10840-013-9824-4

Weitere Artikel der Ausgabe 2/2013

Journal of Interventional Cardiac Electrophysiology 2/2013 Zur Ausgabe

Nach Herzinfarkt mit Typ-1-Diabetes schlechtere Karten als mit Typ 2?

29.05.2024 Herzinfarkt Nachrichten

Bei Menschen mit Typ-2-Diabetes sind die Chancen, einen Myokardinfarkt zu überleben, in den letzten 15 Jahren deutlich gestiegen – nicht jedoch bei Betroffenen mit Typ 1.

Erhöhtes Risiko fürs Herz unter Checkpointhemmer-Therapie

28.05.2024 Nebenwirkungen der Krebstherapie Nachrichten

Kardiotoxische Nebenwirkungen einer Therapie mit Immuncheckpointhemmern mögen selten sein – wenn sie aber auftreten, wird es für Patienten oft lebensgefährlich. Voruntersuchung und Monitoring sind daher obligat.

GLP-1-Agonisten können Fortschreiten diabetischer Retinopathie begünstigen

24.05.2024 Diabetische Retinopathie Nachrichten

Möglicherweise hängt es von der Art der Diabetesmedikamente ab, wie hoch das Risiko der Betroffenen ist, dass sich sehkraftgefährdende Komplikationen verschlimmern.

TAVI versus Klappenchirurgie: Neue Vergleichsstudie sorgt für Erstaunen

21.05.2024 TAVI Nachrichten

Bei schwerer Aortenstenose und obstruktiver KHK empfehlen die Leitlinien derzeit eine chirurgische Kombi-Behandlung aus Klappenersatz plus Bypass-OP. Diese Empfehlung wird allerdings jetzt durch eine aktuelle Studie infrage gestellt – mit überraschender Deutlichkeit.

Update Kardiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.