Quantification of radiation dose savings in cardiac computed tomography using prospectively triggered mode and ECG pulsing: a phantom study

verfasst von: Lukas Lehmkuhl, Dieter Gosch, H. D. Nagel, Patrick Stumpp, Thomas Kahn, Matthias Gutberlet

Erschienen in: European Radiology | Ausgabe 9/2010

Einloggen, um Zugang zu erhalten

Abstract

Objective

To quantify radiation dose reduction in cardiac computed tomography (CT) using a prospectively triggered mode compared with a retrospective ECG-gated helical mode.

Methods

Absorbed organ doses in cardiac 64-row multidetector CT were quantified using an anthropomorphic male Alderson phantom with 74 thermoluminescence dosimeters. Three different imaging protocols were applied: retrospective ECG-gating, retrospective ECG-gating with additional ECG-pulsing, and a prospectively triggered mode. The measured organ doses were compared with dose estimation by a mathematical phantom.

Results

Compared with the retrospective ECG-gating mode, the mean relative organ doses were reduced by 44% using ECG pulsing and by 76% using the prospectively triggered mode. The range of dose savings varied from 34% to 49% using ECG pulsing and from 65% to 87% using the prospectively triggered mode. The effective dose was 16.5 mSv using retrospective gating, 9.2 mSv using retrospective gating with ECG pulsing and 4.0 mSv using the prospectively triggered mode.

Conclusions

Our measurements confirm the high dose-saving potential of the prospectively triggered technique in cardiac CT. The reduction in the organ doses measured corresponds to estimates determined by the mathematical phantom. The effective dose calculated by the mathematical phantom was, in some cases, significantly lower than that calculated using the anthropomorphic phantom.
Literatur
1.
Zurück zum Zitat Rosamond W, Flegal K, Furie K, Go A, Greenlund K, Haase N et al (2008) Heart disease and stroke statistics–2008 update: a report from the American Heart Association Statistics Committee and Stroke Statistics Subcommittee. Circulation 117:e25–146CrossRefPubMed Rosamond W, Flegal K, Furie K, Go A, Greenlund K, Haase N et al (2008) Heart disease and stroke statistics–2008 update: a report from the American Heart Association Statistics Committee and Stroke Statistics Subcommittee. Circulation 117:e25–146CrossRefPubMed
2.
Zurück zum Zitat Budoff MJ, Dowe D, Jollis JG, Gitter M, Sutherland J, Halamert E et al (2008) Diagnostic performance of 64-multidetector row coronary computed tomographic angiography for evaluation of coronary artery stenosis in individuals without known coronary artery disease: results from the prospective multicenter ACCURACY (Assessment by Coronary Computed Tomographic Angiography of Individuals Undergoing Invasive Coronary Angiography) trial. J Am Coll Cardiol 52:1724–1732CrossRefPubMed Budoff MJ, Dowe D, Jollis JG, Gitter M, Sutherland J, Halamert E et al (2008) Diagnostic performance of 64-multidetector row coronary computed tomographic angiography for evaluation of coronary artery stenosis in individuals without known coronary artery disease: results from the prospective multicenter ACCURACY (Assessment by Coronary Computed Tomographic Angiography of Individuals Undergoing Invasive Coronary Angiography) trial. J Am Coll Cardiol 52:1724–1732CrossRefPubMed
3.
Zurück zum Zitat Meijboom WB, Meijs MF, Schuijf JD, Cramer MJ, Mollet NR, van Mieghem CA et al (2008) Diagnostic accuracy of 64-slice computed tomography coronary angiography: a prospective, multicenter, multivendor study. J Am Coll Cardiol 52:2135–2144CrossRefPubMed Meijboom WB, Meijs MF, Schuijf JD, Cramer MJ, Mollet NR, van Mieghem CA et al (2008) Diagnostic accuracy of 64-slice computed tomography coronary angiography: a prospective, multicenter, multivendor study. J Am Coll Cardiol 52:2135–2144CrossRefPubMed
4.
Zurück zum Zitat Miller JM, Rochitte CE, Dewey M, Arbab-Zadeh A, Niinuma H, Gottlieb I et al (2008) Diagnostic performance of coronary angiography by 64-row CT. N Engl J Med 359:2324–2336CrossRefPubMed Miller JM, Rochitte CE, Dewey M, Arbab-Zadeh A, Niinuma H, Gottlieb I et al (2008) Diagnostic performance of coronary angiography by 64-row CT. N Engl J Med 359:2324–2336CrossRefPubMed
5.
Zurück zum Zitat Coles DR, Smail MA, Negus IS, Wilde P, Oberhoff M, Karsch KR et al (2006) Comparison of radiation doses from multislice computed tomography coronary angiography and conventional diagnostic angiography. J Am Coll Cardiol 47:1840–1845CrossRefPubMed Coles DR, Smail MA, Negus IS, Wilde P, Oberhoff M, Karsch KR et al (2006) Comparison of radiation doses from multislice computed tomography coronary angiography and conventional diagnostic angiography. J Am Coll Cardiol 47:1840–1845CrossRefPubMed
6.
Zurück zum Zitat Hausleiter J, Meyer T, Hadamitzky M, Huber E, Zankl M, Martinoff S et al (2006) Radiation dose estimates from cardiac multislice computed tomography in daily practice: impact of different scanning protocols on effective dose estimates. Circulation 113:1305–1310CrossRefPubMed Hausleiter J, Meyer T, Hadamitzky M, Huber E, Zankl M, Martinoff S et al (2006) Radiation dose estimates from cardiac multislice computed tomography in daily practice: impact of different scanning protocols on effective dose estimates. Circulation 113:1305–1310CrossRefPubMed
7.
Zurück zum Zitat Hurwitz LM, Reiman RE, Yoshizumi TT, Goodman PC, Toncheva G, Nguyen G et al (2007) Radiation dose from contemporary cardiothoracic multidetector CT protocols with an anthropomorphic female phantom: implications for cancer induction. Radiology 245:742–750CrossRefPubMed Hurwitz LM, Reiman RE, Yoshizumi TT, Goodman PC, Toncheva G, Nguyen G et al (2007) Radiation dose from contemporary cardiothoracic multidetector CT protocols with an anthropomorphic female phantom: implications for cancer induction. Radiology 245:742–750CrossRefPubMed
8.
Zurück zum Zitat Nickoloff EL, Alderson PO (2007) A comparative study of thoracic radiation doses from 64-slice cardiac CT. Br J Radiol 80:537–544CrossRefPubMed Nickoloff EL, Alderson PO (2007) A comparative study of thoracic radiation doses from 64-slice cardiac CT. Br J Radiol 80:537–544CrossRefPubMed
9.
Zurück zum Zitat Gopal A, Mao SS, Karlsberg D, Young E, Waggoner J, Ahmadi N et al (2008) Radiation reduction with prospective ECG-triggering acquisition using 64-multidetector computed tomographic angiography. Int J Cardiovasc Imaging 25:405–416CrossRefPubMed Gopal A, Mao SS, Karlsberg D, Young E, Waggoner J, Ahmadi N et al (2008) Radiation reduction with prospective ECG-triggering acquisition using 64-multidetector computed tomographic angiography. Int J Cardiovasc Imaging 25:405–416CrossRefPubMed
10.
Zurück zum Zitat Klass O, Jeltsch M, Feuerlein S, Brunner H, Nagel HD, Walker MJ et al (2008) Prospectively gated axial CT coronary angiography: preliminary experiences with a novel low-dose technique. Eur Radiol 19:829–836CrossRefPubMed Klass O, Jeltsch M, Feuerlein S, Brunner H, Nagel HD, Walker MJ et al (2008) Prospectively gated axial CT coronary angiography: preliminary experiences with a novel low-dose technique. Eur Radiol 19:829–836CrossRefPubMed
11.
Zurück zum Zitat Shuman WP, Branch KR, May JM, Mitsumori LM, Lockhart DW, Dubinsky TJ et al (2008) Prospective versus retrospective ECG gating for 64-detector CT of the coronary arteries: comparison of image quality and patient radiation dose. Radiology 248:431–437CrossRefPubMed Shuman WP, Branch KR, May JM, Mitsumori LM, Lockhart DW, Dubinsky TJ et al (2008) Prospective versus retrospective ECG gating for 64-detector CT of the coronary arteries: comparison of image quality and patient radiation dose. Radiology 248:431–437CrossRefPubMed
12.
Zurück zum Zitat Geleijns J, Van Unnik JG, Zoetelief J, Zweers D, Broerse JJ (1994) Comparison of two methods for assessing patient dose from computed tomography. Br J Radiol 67:360–365CrossRefPubMed Geleijns J, Van Unnik JG, Zoetelief J, Zweers D, Broerse JJ (1994) Comparison of two methods for assessing patient dose from computed tomography. Br J Radiol 67:360–365CrossRefPubMed
13.
Zurück zum Zitat Shrimpton PC (2004) Assessment of patient dose in CT. NRPB-PE/1/2004 Shrimpton PC (2004) Assessment of patient dose in CT. NRPB-PE/1/2004
14.
Zurück zum Zitat Hsieh J, Londt J, Vass M, Li J, Tang X, Okerlund D (2006) Step-and-shoot data acquisition and reconstruction for cardiac x-ray computed tomography. Med Phys 33:4236–4248CrossRefPubMed Hsieh J, Londt J, Vass M, Li J, Tang X, Okerlund D (2006) Step-and-shoot data acquisition and reconstruction for cardiac x-ray computed tomography. Med Phys 33:4236–4248CrossRefPubMed
15.
Zurück zum Zitat Huda W, Sandison GA (1984) Estimation of mean organ doses in diagnostic radiology from Rando phantom measurements. Health Phys 47:463–467PubMed Huda W, Sandison GA (1984) Estimation of mean organ doses in diagnostic radiology from Rando phantom measurements. Health Phys 47:463–467PubMed
16.
Zurück zum Zitat Golikov VY, Nikitin VV (1989) Estimation of the mean organ doses and the effective dose equivalent from Rando phantom measurements. Health Phys 56:111–115PubMed Golikov VY, Nikitin VV (1989) Estimation of the mean organ doses and the effective dose equivalent from Rando phantom measurements. Health Phys 56:111–115PubMed
17.
Zurück zum Zitat Kramer R, Zankl M, Williams G, Drexler G (1982) The calculation of dose from external photon exposures using reference human phantoms and Monte Carlo methods. Part I. The male (Adam) and female (Eva) adult mathematical phantoms. GSF-Report S-885. Neuherberg Kramer R, Zankl M, Williams G, Drexler G (1982) The calculation of dose from external photon exposures using reference human phantoms and Monte Carlo methods. Part I. The male (Adam) and female (Eva) adult mathematical phantoms. GSF-Report S-885. Neuherberg
18.
Zurück zum Zitat International Commission on Radiological Protection (ICRP) 60 (1991) Recommendations of the International Commission on Radiological Protection. Pergamon Press, Oxford International Commission on Radiological Protection (ICRP) 60 (1991) Recommendations of the International Commission on Radiological Protection. Pergamon Press, Oxford
19.
Zurück zum Zitat Stamm G, Nagel HD (2002) CT-expo–a novel program for dose evaluation in CT. Rofo 174:1570–1576PubMed Stamm G, Nagel HD (2002) CT-expo–a novel program for dose evaluation in CT. Rofo 174:1570–1576PubMed
20.
Zurück zum Zitat Zankl M, Panzer W, Drexler G (1991) The calculation of dose from external photon exposures using reference human phantoms and Monte Carlo methods. Part VI. Organ doses from computed tomographic examinations. GSF-Report 30/91. Institute for Radiation Biology, Neuherberg Zankl M, Panzer W, Drexler G (1991) The calculation of dose from external photon exposures using reference human phantoms and Monte Carlo methods. Part VI. Organ doses from computed tomographic examinations. GSF-Report 30/91. Institute for Radiation Biology, Neuherberg
21.
Zurück zum Zitat Shleien B (1973) Review of determinations of radiation dose to the active bone marrow from diagnostic X-ray examinations. DHEW publication—no. (FDA) 74–8007, Rockville Shleien B (1973) Review of determinations of radiation dose to the active bone marrow from diagnostic X-ray examinations. DHEW publication—no. (FDA) 74–8007, Rockville
22.
Zurück zum Zitat DeMarco JJ, Cagnon CH, Cody DD, Stevens DM, McCollough CH, Zankl M et al (2007) Estimating radiation doses from multidetector CT using Monte Carlo simulations: effects of different size voxelized patient models on magnitudes of organ and effective dose. Phys Med Biol 52:2583–2597CrossRefPubMed DeMarco JJ, Cagnon CH, Cody DD, Stevens DM, McCollough CH, Zankl M et al (2007) Estimating radiation doses from multidetector CT using Monte Carlo simulations: effects of different size voxelized patient models on magnitudes of organ and effective dose. Phys Med Biol 52:2583–2597CrossRefPubMed
23.
Zurück zum Zitat Hidajat N, Vogl T, Schroder RJ, Felix R (1996) Calculated organ doses and effective dosage for computerized tomography examination of the thorax and abdomen: are these doses realistic? Rofo 164:382–387PubMed Hidajat N, Vogl T, Schroder RJ, Felix R (1996) Calculated organ doses and effective dosage for computerized tomography examination of the thorax and abdomen: are these doses realistic? Rofo 164:382–387PubMed
24.
Zurück zum Zitat Veit R, Panzer W, Zankl M, Scheurer C (1992) Vergleich berechneter und gemessener Dosen an einem anthropomorphen Phantom. Z Med Phys 2:123–126 Veit R, Panzer W, Zankl M, Scheurer C (1992) Vergleich berechneter und gemessener Dosen an einem anthropomorphen Phantom. Z Med Phys 2:123–126
25.
Zurück zum Zitat Stolzmann P, Scheffel H, Schertler T, Frauenfelder T, Leschka S, Husmann L et al (2008) Radiation dose estimates in dual-source computed tomography coronary angiography. Eur Radiol 18:592–599CrossRefPubMed Stolzmann P, Scheffel H, Schertler T, Frauenfelder T, Leschka S, Husmann L et al (2008) Radiation dose estimates in dual-source computed tomography coronary angiography. Eur Radiol 18:592–599CrossRefPubMed
26.
Zurück zum Zitat Wang M, Qi HT, Wang XM, Wang T, Chen JH, Liu C (2008) Dose performance and image quality: Dual source CT versus single source CT in cardiac CT angiography. Eur J Radiol 72:396–400CrossRefPubMed Wang M, Qi HT, Wang XM, Wang T, Chen JH, Liu C (2008) Dose performance and image quality: Dual source CT versus single source CT in cardiac CT angiography. Eur J Radiol 72:396–400CrossRefPubMed
27.
Zurück zum Zitat Hausleiter J, Meyer T, Hermann F, Hadamitzky M, Krebs M, Gerber TC et al (2009) Estimated radiation dose associated with cardiac CT angiography. JAMA 301:500–507CrossRefPubMed Hausleiter J, Meyer T, Hermann F, Hadamitzky M, Krebs M, Gerber TC et al (2009) Estimated radiation dose associated with cardiac CT angiography. JAMA 301:500–507CrossRefPubMed
28.
Zurück zum Zitat Hermann F, Martinoff S, Meyer T, Hadamitzky M, Jiang C, Hendrich E et al (2008) Reduction of radiation dose estimates in cardiac 64-slice CT angiography in patients after coronary artery bypass graft surgery. Invest Radiol 43:253–260CrossRefPubMed Hermann F, Martinoff S, Meyer T, Hadamitzky M, Jiang C, Hendrich E et al (2008) Reduction of radiation dose estimates in cardiac 64-slice CT angiography in patients after coronary artery bypass graft surgery. Invest Radiol 43:253–260CrossRefPubMed
29.
Zurück zum Zitat Trabold T, Buchgeister M, Kuttner A, Heuschmid M, Kopp AF, Schroder S et al (2003) Estimation of radiation exposure in 16-detector row computed tomography of the heart with retrospective ECG-gating. Rofo 175:1051–1055PubMed Trabold T, Buchgeister M, Kuttner A, Heuschmid M, Kopp AF, Schroder S et al (2003) Estimation of radiation exposure in 16-detector row computed tomography of the heart with retrospective ECG-gating. Rofo 175:1051–1055PubMed
30.
Zurück zum Zitat Husmann L, Valenta I, Gaemperli O, Adda O, Treyer V, Wyss CA et al (2008) Feasibility of low-dose coronary CT angiography: first experience with prospective ECG-gating. Eur Heart J 29:191–197CrossRefPubMed Husmann L, Valenta I, Gaemperli O, Adda O, Treyer V, Wyss CA et al (2008) Feasibility of low-dose coronary CT angiography: first experience with prospective ECG-gating. Eur Heart J 29:191–197CrossRefPubMed
31.
Zurück zum Zitat Blankstein R, Shah A, Pale R, Abbara S, Bezerra H, Bolen M et al (2009) Radiation dose and image quality of prospective triggering with dual-source cardiac computed tomography. Am J Cardiol 103:1168–1173CrossRefPubMed Blankstein R, Shah A, Pale R, Abbara S, Bezerra H, Bolen M et al (2009) Radiation dose and image quality of prospective triggering with dual-source cardiac computed tomography. Am J Cardiol 103:1168–1173CrossRefPubMed
32.
Zurück zum Zitat Hirai N, Horiguchi J, Fujioka C, Kiguchi M, Yamamoto H, Matsuura N et al (2008) Prospective versus retrospective ECG-gated 64-detector coronary CT angiography: assessment of image quality, stenosis, and radiation dose. Radiology 248:424–430CrossRefPubMed Hirai N, Horiguchi J, Fujioka C, Kiguchi M, Yamamoto H, Matsuura N et al (2008) Prospective versus retrospective ECG-gated 64-detector coronary CT angiography: assessment of image quality, stenosis, and radiation dose. Radiology 248:424–430CrossRefPubMed
33.
Zurück zum Zitat Shuman WP, Branch KR, May JM, Mitsumori LM, Strote JN, Warren BH et al (2009) Whole-chest 64-MDCT of emergency department patients with nonspecific chest pain: radiation dose and coronary artery image quality with prospective ECG triggering versus retrospective ECG gating. AJR Am J Roentgenol 192:1662–1667CrossRefPubMed Shuman WP, Branch KR, May JM, Mitsumori LM, Strote JN, Warren BH et al (2009) Whole-chest 64-MDCT of emergency department patients with nonspecific chest pain: radiation dose and coronary artery image quality with prospective ECG triggering versus retrospective ECG gating. AJR Am J Roentgenol 192:1662–1667CrossRefPubMed
34.
Zurück zum Zitat Einstein AJ, Henzlova MJ, Rajagopalan S (2007) Estimating risk of cancer associated with radiation exposure from 64-slice computed tomography coronary angiography. JAMA 298:317–323CrossRefPubMed Einstein AJ, Henzlova MJ, Rajagopalan S (2007) Estimating risk of cancer associated with radiation exposure from 64-slice computed tomography coronary angiography. JAMA 298:317–323CrossRefPubMed
35.
Zurück zum Zitat The 2007 Recommendations of the International Commission on Radiological Protection (2007) ICRP publication 103, Ann ICRP 37:1–332 The 2007 Recommendations of the International Commission on Radiological Protection (2007) ICRP publication 103, Ann ICRP 37:1–332
36.
Zurück zum Zitat Hart D, Jones DG, Wang M (1994) Normalised organ doses for medical X-ray examinations calculated using Monte-Carlo techniques (NRPB-SR262). National Radiological Protection Board, Chilton Hart D, Jones DG, Wang M (1994) Normalised organ doses for medical X-ray examinations calculated using Monte-Carlo techniques (NRPB-SR262). National Radiological Protection Board, Chilton
37.
Zurück zum Zitat Schindera ST, Nelson RC, Lee ER, DeLong DM, Ngyen G, Toncheva G et al (2007) Abdominal multislice CT for obese patients: effect on image quality and radiation dose in a phantom study. Acad Radiol 14:486–494CrossRefPubMed Schindera ST, Nelson RC, Lee ER, DeLong DM, Ngyen G, Toncheva G et al (2007) Abdominal multislice CT for obese patients: effect on image quality and radiation dose in a phantom study. Acad Radiol 14:486–494CrossRefPubMed
Metadaten
Titel
Quantification of radiation dose savings in cardiac computed tomography using prospectively triggered mode and ECG pulsing: a phantom study
verfasst von
Lukas Lehmkuhl
Dieter Gosch
H. D. Nagel
Patrick Stumpp
Thomas Kahn
Matthias Gutberlet
Publikationsdatum
01.09.2010
Verlag
Springer-Verlag
Erschienen in
European Radiology / Ausgabe 9/2010
Print ISSN: 0938-7994
Elektronische ISSN: 1432-1084
DOI
https://doi.org/10.1007/s00330-010-1767-x

Weitere Artikel der Ausgabe 9/2010

European Radiology 9/2010 Zur Ausgabe

Mammakarzinom: Brustdichte beeinflusst rezidivfreies Überleben

26.05.2024 Mammakarzinom Nachrichten

Frauen, die zum Zeitpunkt der Brustkrebsdiagnose eine hohe mammografische Brustdichte aufweisen, haben ein erhöhtes Risiko für ein baldiges Rezidiv, legen neue Daten nahe.

„Übersichtlicher Wegweiser“: Lauterbachs umstrittener Klinik-Atlas ist online

17.05.2024 Klinik aktuell Nachrichten

Sie sei „ethisch geboten“, meint Gesundheitsminister Karl Lauterbach: mehr Transparenz über die Qualität von Klinikbehandlungen. Um sie abzubilden, lässt er gegen den Widerstand vieler Länder einen virtuellen Klinik-Atlas freischalten.

Klinikreform soll zehntausende Menschenleben retten

15.05.2024 Klinik aktuell Nachrichten

Gesundheitsminister Lauterbach hat die vom Bundeskabinett beschlossene Klinikreform verteidigt. Kritik an den Plänen kommt vom Marburger Bund. Und in den Ländern wird über den Gang zum Vermittlungsausschuss spekuliert.

Darf man die Behandlung eines Neonazis ablehnen?

08.05.2024 Gesellschaft Nachrichten

In einer Leseranfrage in der Zeitschrift Journal of the American Academy of Dermatology möchte ein anonymer Dermatologe bzw. eine anonyme Dermatologin wissen, ob er oder sie einen Patienten behandeln muss, der eine rassistische Tätowierung trägt.

Update Radiologie

Bestellen Sie unseren Fach-Newsletter und bleiben Sie gut informiert.